K242385 · Shenzhen Borria Technology Co., Ltd. · OHS · Apr 7, 2025 · General, Plastic Surgery
Device Facts
Record ID
K242385
Device Name
LED Therapy Mask (MN1, M226)
Applicant
Shenzhen Borria Technology Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Apr 7, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Red light: Treatment of full-face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of full face wrinkles.
Device Story
Wearable LED phototherapy mask for home use; treats facial wrinkles, acne, and provides topical heating. Device inputs: electrical power from rechargeable lithium battery. Principle of operation: LEDs emit narrowband light (Red 630nm, Blue 470nm, Infrared 850nm, Mixed) to target skin tissue. User operates via controller to select light mode and power on/off. Output: light energy applied to face/neck. Clinical benefit: wrinkle reduction, acne management, and muscle/joint relief via thermal effects. Used by patients in home environment.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83, IEC 62133-2), photobiological eye safety (IEC 62471), software verification/validation, and usability engineering per FDA guidance.
Technological Characteristics
Wearable LED mask; materials compliant with ISO 10993-5, -10, -23. Light source: LEDs (Red 630nm, Blue 470nm, NIR 850nm). Energy source: 3.7V 2500mAh rechargeable lithium battery. Connectivity: Type-C charging. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471, IEC 62133-2.
Indications for Use
Indicated for individuals seeking treatment for full-face wrinkles, mild to moderate inflammatory acne (models M226/MN1 only), and topical heating for arthritis, muscle spasms, stiffness, muscle relaxation, and increased local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LED light therapy mask (FM-01, FM-02, FM-03) (K223544)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Build Correspondence
Convert to PDF
April 7, 2025
Shenzhen Borria Technology Co., Ltd
Chardy Wang
Project Leader
501, Floor 5, Building 6, Industrial Zone, No.2015, Shenyan
Road, Pengwan Community, Haishan Street, Yantian District
Shenzhen,
China
Re: K242385
Trade/Device Name: LED Therapy Mask (MN1, M226)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, ILY, OLP
Dated: August 12, 2024
Received: August 12, 2024
Dear Chardy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K242385 - Chardy Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K242385 - Chardy Wang
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.04.07
17:08:19 -04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242385 | |
| Device Name LED Therapy Mask (MN1, M226) | |
| Indications for Use (Describe) Red light: Treatment of full- face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of full face wrinkles. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
# 510(k) Summary-K242385
"510(k) Summary" as required by 21 CFR Part 807.92.
April 2, 2025
## I. Submitter
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
501, Floor 5, Building 6, Industrial Zone, No.2015, Shenyan Road, Pengwan Community, Haishan Street, Yantian District, Shenzhen
Chardy Wang
Project Leader
Tel: +86 133 1290 5505
Email: 466144494@qq.com
## II. Device
Name of Device: LED Therapy Mask
Model(s): MN1, M226
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: II
Product Code: OHS, OLP, ILY
Regulation Number: 21 CFR 878.4810
## III. Predicate Device and Reference Device
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Approval Date |
| --- | --- | --- | --- |
| Guangdong Newdermo Biotech Co., Ltd | LED light therapy mask (FM-01, FM-02, FM-03) | K223544 | Feb 23, 2023 |
Page 1 of 9
{5}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
Reference device 1:
| Manufacturer | Reference Device | 510(k) Number | Approval Date |
| --- | --- | --- | --- |
| myBlend | MyLedMask | K223147 | May 14, 2023 |
Reference device 2:
| Manufacturer | Reference Device | 510(k) Number | Approval Date |
| --- | --- | --- | --- |
| La Lumiere, LLC | Pro X OTC 5 Light Therapy Device | K140471 | Jan 15, 2025 |
Reference device 3:
| Manufacturer | Reference Device | 510(k) Number | Approval Date |
| --- | --- | --- | --- |
| Trophy Skin Incorporated | Rejuvalite MD | K133896 | Nov 13, 2014 |
IV. Device Description
The subject device LED Therapy Mask is a home use wearable LED phototherapy device whose purpose is to produce an even and narrowband of light for the treatment of cosmetic indications including facial wrinkles and acnes.
The subject device consists of a mask body unit that contains light emitting diodes (LEDs), a controller, straps and Type-C charging cable. And the device is powered by built-in rechargeable lithium battery on the controller.
Page 2 of 9
{6}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
The LEDs generate 4 kinds of light which include Blue light (460nm, only suitable model M226 and MN1 LED Facial Mask), Red light (630nm), Infrared light (850nm) and Mixed light (630nm and 850nm). A controller is connect to the mask body unit to control the device, such as turn on/off the device, switch LED output. And the straps used for securing the mask unit to the body part.
## V. Indications for Use
Red light: Treatment of full-face wrinkles.
Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Mixed light: Treatment of full face wrinkles.
## VI. Comparison of Technological Characteristics With the Predicate Devices
| Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| 510(k) Number | Pending | K223544 | K223147 | K140471 | K133896 | / |
| Trade name | LED Therapy Mask (MN1, M226) | LED light therapy mask (FM-01, FM-02, FM-03) | MyLedMask | Pro X OTC 5 Light Therapy Device | Rejuvalite MD | / |
Page 3 of 9
{7}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| Manufacturer | SHENZHEN BORRIA TECHNOLOGY CO., LTD | Guangdong Newdermo Biotech Co., Ltd | myBlend | La Lumiere, LLC | Trophy Skin Incorporated | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHS, OLP, ILY | OHS, OLP, ILY | OHS | OHS | OHS | |
| Device classification | Class II | Class II | Class II | Class II | Class II | Same |
| Indication for use/ Intended use | Red light: Treatment of full-face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily | Red light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily | MyLedMask is an over-the-counter device that is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full-face wrinkles | The Pro XOTC 5is an Over-The-Counter device intended for use in the treatment of facial wrinkles. It is for people with wrinkles on their face and who have Fitzpatrick skin types I,II and/or III. | The Rejuvalite MD is an Over the Counter device that is intended for the use in the treatment of full face wrinkles. | Same |
Page 4 of 9
{8}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of full face wrinkles. | increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne. | | | | |
| Location for use | Face and Neck | Face and body | Face and Neck | Full Face | Full Face | Same |
| Prescription or OTC | OTC | OTC | OTC | OTC | OTC | Same |
| Power supply | Input: DC 5V, 1A Built-in rechargeable lithium battery: DC 3.7V 2500mAh | Input: 100-240 V~, 50/60 Hz, 0.25 A Output: DC 5 V, 500 mA | Voltage: 100 to 240 volts, AC Frequency:50-60Hz Intensity: 0.35 A | Unknown | AC to DC | Same |
| Light source | LEDs | LEDs | LEDs | Light-emitting diodes | LEDs | Same |
| Wavelength | Red: 630nm±5nm
Blue: 470 nm±5nm
Near-Infrared: 850 nm±5nm | Red: 620nm
Blue: 460nm
Infrared: 850nm
Mixed: 620nm and 850nm and 460nm | Red:630 nm (OSRAM LS T67F)
NIR: 850 nm (OSRAM SFH 4253) | Red: 620-640nm
IR: 820-880nm | Red 600,622,660nm
IR 860nm | Similar Note 1 |
Page 5 of 9
{9}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | Mixed light: 630nm and 850nm | | | | | |
| LED Intensity | Red: 0.89~2.55mW/cm²
Blue: 1.44~4.09mW/cm²
Near-Infrared: 1.83~3.05mW/cm²
Mixed light: 0.95~2.64mW/cm² | Red light: 2.0~3.0 mW/cm²
Blue light: 2.0~4.0 mW/cm²
Infrared light: 2.0~4.0 mW/cm²
Mixed light: 9.0~12.0 mW/cm² | 18.7 mW/cm² (average of 6 measurement location (including one on the neck) between 10 to 27mW/cm²) | Red light and IR light: 0.76 mW/cm² | Red light and IR light: 2.35 mW/cm² | Similar Note 2 |
| Dimensions (mm) | MN1: LED Facial Mask: 300*208.5*5.5mm
LED Neck Mask: 338.5*249.7*5.5mm
M226: LED Facial Mask: 300*208.5*5.5mm | FM-01: 207X277X43mm,
FM-02: 198X383X33.5mm,
FM-03: 237.5X108X8.1mm | LED mask: 176 x342 x 122
Storage case:365 x 225 x150 | Unknown | Unknown | Different Note 3 |
| Treatment Time | 10 minutes each time | Manual Mode: 15minutes each time,
Automatic Mode: 10minutes each time | According to skin phototype, daily - 1 fair skin: 5 min 35 (335 s) - 2 moderately dark skin: 11 min 10 (670 s) - 3 dark | 900 seconds | 3 minutes daily, 5 days per week for 8 weeks | |
Page 6 of 9
{10}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Reference Device 1 | Reference Device 2 | Reference Device 3 | Remark |
| --- | --- | --- | --- | --- | --- | --- |
| | | | skin: 13 min 55 (835 s) during 6 to 8 week | | | |
| Electrical safety | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-83
IEC 62471
IEC 62133-2 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57 | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-57 | Unknown | IEC 60601-1
IEC 60601-1-2 | Same |
| Biocompatibility feature | All body-contacting materials are complied with ISO 10993-5, ISO 10993-10 and ISO 10993-23 | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 | Unknown | Unknown | Unknown | Same |
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Evaluation
The biocompatibility evaluation for the body-contacting components of the Intense pulsed light device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including:
{11}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
- ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23:2021, Biological evaluation of medical devices – Part 23: Tests for skin irritation
## 2) Electrical Safety and EMC
Electrical safety and EMC testing was performed to, and passed, as per the following standards:
- IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: electromagnetic compatibility
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-11 Edition 2.1 2020-07 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
- IEC 62133-2 Edition 1.0 2017-02, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
## 3) Eye Safety
- IEC 62471 Photobiological safety of lamps and lamp systems
## 4) Software Verification and Validation
Page 8 of 9
{12}
SHENZHEN BORRIA TECHNOLOGY CO.,LTD
K242385 -510(k) Summary
Software documentation consistent with *moderate level* of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## 5) Usability
The product usability has been evaluated and verified according to the following FDA guidance
- Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016.
## VIII. Conclusions
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device LED Therapy Mask is as safe, as effective, and performs as well as the legally marketed predicate device and reference devices.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.