Red+Infrared light: Treatment of full-face wrinkles. Red light: Treatment of full-face wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Blue light: Treatment of mild to moderate inflammatory acne. Red+Blue light: Treatment of mild to moderate inflammatory acne. The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in home healthcare environment.
Device Story
Wearable LED phototherapy mask for home use; utilizes red, infrared, and blue light spectrums to treat skin conditions. Device consists of mask unit, controller, and Type-C charging cable. User applies mask directly to face; device emits specific wavelengths (red 620nm, infrared 850nm) to skin. Red/Infrared modes target wrinkles; blue light targets inflammatory acne; infrared provides topical heating for muscle/joint relief. Operation involves selecting treatment mode via controller. Benefits include non-invasive wrinkle reduction, acne management, and temporary relief of muscle/joint stiffness through increased local blood circulation. Device is operated by the patient in a home healthcare environment.
Clinical Evidence
Bench testing only. Device performance verified through compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home healthcare environment), IEC 60601-2-83 (light therapy safety), and IEC 62471 (photobiological safety). No clinical data provided.
Technological Characteristics
Wearable LED mask; lithium battery powered; Type BF applied part. Wavelengths: Red (620nm), Infrared (850nm). Irradiance: 4-58.7 mW/cm2 depending on mode. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471. Connectivity: Type-C charging. Sterilization: N/A (home use device).
Indications for Use
Indicated for adults for the treatment of full-face wrinkles, mild to moderate inflammatory acne, and for topical heating to elevate tissue temperature, relieve arthritis/muscle spasm/stiffness, promote muscle relaxation, and increase local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
January 28, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a seal on the left and the FDA acronym with the full name of the agency on the right. The seal features an emblem with a stylized design, while the FDA acronym is in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Shenzhen Rainbow Technology Co., Ltd. % Jilan Luo RA Specialist Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong 518052 China
Re: K243423
Trade/Device Name: LED Light Therapy Mask (Model(s): RB-008, RB-031, RB-081, RB-008G, RB-008GB, RB-008J, RB-008JB) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS, OLP, ILY Dated: October 31, 2024 Received: November 4, 2024
Dear Jilan Luo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "YAN FU -S" in large, bold letters. To the right of the name, there is a digital signature that reads "Digitally signed by YAN FU -S" followed by the date and time "Date: 2025.01.28 16:52:13". Below the date and time, there is a timezone offset of "-05'00'".
for Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K243423
Device Name
LED Light Therapy Mask (Model(s): RB-008, RB-030, RB-081, RB-008GB, RB-008J, RB-008JB)
Indications for Use (Describe)
- Red+Infrared light: Treatment of full-face wrinkles.
- Red light: Treatment of full-face wrinkles.
- Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
- Blue light: Treatment of mild to moderate inflammatory acne.
- Red+Blue light: Treatment of mild to moderate inflammatory acne.
The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in home healthcare environment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K243423 | 510(k) Summary | | | | Prepared on: 2025-01-27 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------|--|--|--|-------------------------|
| Contact Details | | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | Shenzhen Rainbow Technology Co.,Ltd. | | | | |
| Applicant Address | 5th floor, Building C, Fenda Technology Innovation Park, Xixiang, Baoan<br>District Shenzhen Guangdong 518000 China | | | | |
| Applicant Contact Telephone | 86 13590329742 | | | | |
| Applicant Contact | Ms. Layla Li | | | | |
| Applicant Contact Email | Layla@rainbowdo.com | | | | |
| Correspondent Name | Feiying Drug & Medical Consulting Technical Service Group | | | | |
| Correspondent Address | Rm 2401 Zhenye International Business Center, No. 3101-90,<br>Qianhai Road Shenzhen Guangdong 518052 China | | | | |
| Correspondent Contact Telephone | +86 13535208704 | | | | |
| Correspondent Contact | Ms. Jilan Luo | | | | |
| Correspondent Contact Email | jilan.luo@qq.com | | | | |
| Device Name | | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | LED Light Therapy Mask (Model(s): RB-008, RB-030,<br>RB-081, RB-008G, RB-008GB, RB-008J, RB-008JB) | | | | |
| Common Name | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology | | | | |
| Classification Name | Light Based Over The Counter Wrinkle Reduction | | | | |
| Regulation Number | 21 CFR 878.4810, 21 CFR 890.5500 | | | | |
| Product Code(s) | OHS, OLP, ILY | | | | |
| Legally Marketed Predicate Devices | | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | | Product Code |
| K241857 | LED Light Therapy Device | | | | OHS, OLP, ILY |
| K230351 | LED Facial Mask | | | | OHS, OLP |
| K241102 | Luna 4 Plus | | | | OHS, NFO |
| K160691 | Acne Light Therapy Wand | | | | OLP |
| K223544 | LED Light Therapy Mask | | | | OHS, OLP, ILY |
{5}------------------------------------------------
# Device Description Summary
The LED Light Therapy Mask adopts light emitting diodes (LED) in the infrared, red and blue spectrum to irradiate on the skin. The face will absorb the light energy to achieve the effects. The LED Light Therapy Mask is applied directly to the skin to ensure consistent administration of light during each treatment. LED Light Therapy Mask is a home use wearable LED phototherapy device which can help reduce wrinkles and arovide topical heating, It is consisting of main unit(mask), controller, Type-C charging cable and so on.
The RB-081 outputs 5 kinds of treatment modes: red infrared, blue, red+blue light. The RB-030 outputs 4 kinds of treatment modes: red-infrared, red, infrared, blue light. The RB-008 outputs 3 kinds of treatment modes: red+infrared, red, blue light. The RB-008G/ RB-008GB/RB-008J/RB-008JB output 4 kinds of treatment modes: red+infrared, red, blue light.
# Intended Use/Indications for Use
�Red+Infrared light: Treatment of full-face wrinkles.
മRed light: Treatment of full-face wrinkles.
�lnfrared light: Provide topical heating tissue temperature; arthritis and muscle spasm;relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
ØBlue light: Treatment of mild to moderate inflammatory acne.
�Red+Blue light: Treatment of mild to moderate inflammatory acne.
The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne for adults in home healthcare environment.
## Indications for Use Comparison
The LED Light Therapy Mask have the same indications for use as the predicate devices.
# Technological Comparison
The technical characteristics of LED Light Therapy Mask (Models:RB-008, RB-008G, RB-0008GB, RB-008J, RB-008J) are substantially equivalent to the predicate devices in the following aspects:
1.the indications for use of LED Light Therapy Mask are within the range of all predicate devices.
2.similar power supply: the product uses lithium batteries, which were tested according to FDA guidance documents and the requirements of IEC 62133-2, the tests are all passed.
3.same wavelength: the LED Light Therapy Mask has the same wavelengths (red: 620mm, infrared: 850nm) as K223544 and they all comply with IEC 60601-2-83 and IEC 62471 requirements, so it will not raise any safety or effectiveness issue.
4.similar irradiance though the irradiance of subject device is a little different from the predicate devices, the irradiance of subject device is within the range of the minimum value of the predicate device and reference predicate devices, and the subject device complies with IEC 60601-2-83 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue. The evaluation process is as follows:
a. Red+lR light: the irradiance of subject device (45.5mW/cm2) is between the range of K241857 (30mW/cm2) and K241102 (63mW/ cm2).
b. Red light: the irradiance of subject device (53.2mW/cm2) is within the range of the red light irradiance (31~75mW/cm2) of K230351. c. Infrared light: the irradiance of subject device (4mW/cm2) is the same as the K223544.
d. Blue light the irradiance of subject device (58.7mW/cm2) is within the range of the blue light irradiance (15~63mW/cm2) of K230351.
- e. Red and blue light: the irradiance of subject device (33.6mW/cm2) is similar with the K160691(34mW/cm2).
5. the same safety features: Type BF applied part.
6. dimension and weight: though the dimension and weight are a little different from the predicate devices, this difference is insignificant and do not raise any safety or effectiveness problems.
Therefore, the LED Light Therapy Mask is substantially equivalent to its predicate device and reference predicate devices.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-clinical tests have been conducted to verify that the Light Therapy Mask meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate devices. The testing results device complies with the following standards:
1. IEC 60601-1:2020, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance 2. IEC 60601-1-2:2021, Medical electrical equipments for basic safety and essential performance -Collateral standard: Electromagnetic disturbances - Requirements and tests
3. IEC 60601-1-11:2020, Medical electrical equipments for basic safety and essential performance --
# 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
## 21 CFR 807.92(a)(4)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.