LED Light Therapy Device (KFB290, KFB291, KFB265, KFB293)
K241857 · Dongguan Boyuan Intelligent Technology Co.,Ltd · OHS · Oct 11, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241857
Device Name
LED Light Therapy Device (KFB290, KFB291, KFB265, KFB293)
Applicant
Dongguan Boyuan Intelligent Technology Co.,Ltd
Product Code
OHS · General, Plastic Surgery
Decision Date
Oct 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
KFB290, KFB291: Red light: Treatment of full-face wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Red+Infrared Light: Treatment of full-face wrinkles. Amber light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Mixed light(Red+Blue +Infrared): Treatment of mild to moderate inflammatory acne. KFB265, KFB293: Red+Infrared Light: Treatment of wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Amber light: Treatment of wrinkles.
Device Story
Wearable LED phototherapy masks (KFB290, KFB291, KFB265, KFB293) for home use; device includes mask, controller, and power supply. Emits specific wavelengths (Red 635-637nm, Blue 465nm, Infrared 850-854nm, Amber 605nm) to treat wrinkles and acne or provide therapeutic heating. Operated by patient via controller; user selects light mode and duration (10-30 minutes). Output is non-ionizing light energy; intended to stimulate skin or provide heat. Benefits include wrinkle reduction, acne management, and temporary relief of muscle/joint stiffness. No clinical data; safety established via bench testing and substantial equivalence to predicate LED masks.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and efficacy supported by biocompatibility testing (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-83, -2-57), battery safety (IEC 62133-2), and photobiological safety (IEC 62471). Software verification and validation confirmed mitigation of hazards to acceptable levels.
Technological Characteristics
Wearable LED mask; light sources: LEDs (Red, Blue, Infrared, Amber). Power: 100-240V AC input, 5V/1A output, 1300mAh Li-ion battery. Materials: Body-contacting components compliant with ISO 10993-5, -10, -23. Connectivity: USB charging. Software: Moderate level of concern. Sterilization: Not applicable.
Indications for Use
Indicated for adults seeking treatment of full-face wrinkles, mild to moderate inflammatory acne, and topical heating for arthritis, muscle spasms, stiffness, and increased local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2024
Dongguan Boyuan Intelligent Technology Co., Ltd. % Jilan Luo RA Specialist Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong Province 518052 China
Re: K241857
Trade/Device Name: LED Light Therapy Device (KFB290, KFB291, KFB265, KFB293) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP, ILY Dated: June 26, 2024 Received: June 27, 2024
Dear Jilan Luo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{3}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by TANISHA TANISHA I HITHE -L. HITHE -S Date: 2024.10.11
15:27:56 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241857
#### Device Name
LED Light Therapy Device(Model(s): KFB290, KFB291, KFB265, KFB293)
Indications for Use (Describe)
#### KFB290, KFB291:
Red light: Treatment of full-face wrinkles.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Red+Infrared Light: Treatment of full-face wrinkles.
Amber light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Mixed light(Red+Blue +Infrared): Treatment of mild to moderate inflammatory acne.
#### KFB265, KFB293:
Red+Infrared Light: Treatment of wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Amber light: Treatment of wrinkles.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
# "510(k) Summary" as required by 21 CFR Part 807.92.
#### I. Submitter
Company name: Dongguan Boyuan Intelligent Technology Co.,Ltd Address: Room 801, Building 3, No.3 Kuiqiao Road, Puxin Lake, Tangxia Town, Dongguan, Guangdong, China.
Contact person: Le Li Title: Deputy general manager Tel: +86-177 2259 0830 E-mail: 978953150@gq.com Preparation date: September 24, 2024
## II. Subject Device
Name of Device: LED Light Therapy Device Model(s): KFB290, KFB291, KFB 265, KFB 293 Common or Usual Name: Light based over the counter wrinkle reduction Over-The-Counter Powered Light Based Laser For Acne Lamp, Infrared, Therapeutic Heating Regulation Name:
Laser surgical instrument for use in general and plastic surgery and in dermatology Infrared lamp. Regulatory Class: II Product Code: OHS, OLP, ILY Regulation Number: 878.4810, 890.5500
#### III. Predicate and reference Devices
#### For KFB290, KFB291:
| No. | Manufacturer | Device name | Product code | 510(k) Number |
|-----|----------------------------------------------------------|---------------------------|---------------|---------------|
| 1. | Guangdong Newdermo<br>Biotech Co.,Ltd | LED light therapy<br>mask | OHS, OLP, ILY | K223544 |
| 2. | Shenzhen Kaiyan Medical<br>Equipment Co., Ltd. | LED Facial Mask | OHS, OLP | K230351 |
| 3. | Zhongshan Bisen Plastic<br>Electronic Products Co., Ltd. | RED Light Device | OHS | K162489 |
#### For KFB 265, KFB 293:
| No. | Manufacturer | Device name | Product code | 510(k) Number |
|-----|------------------------------------------------|-----------------|--------------|---------------|
| 1. | Hunan Guangye<br>Biotechnology Co., Ltd. | Beauty LED Mask | OHS, OLP | K221151 |
| 2. | Shenzhen Kaiyan Medical<br>Equipment Co., Ltd. | LED Facial Mask | OHS, OLP | K230351 |
{6}------------------------------------------------
# IV. Device Description KFB290, KFB291:
LED Light Therapy Device(Models: KFB290, KFB291) is a home use wearable LED phototherapy device which can help reduce wrinkles and acne and provide topical heating. LED Light Therapy Device is consisting of main unit(mask), controller, Silicone eye protector, USB charging cable and straps. The LED Light Therapy Device have 4 kinds of light, which include Red light (wavelength 635nm ± 5mm), Blue light (wavelength 465 ± 5nm), Infrared light (wavelength 850mm ± 5mm), Amber light (wavelength 605 ± 5nm).
#### KFB 265, KFB 293:
LED Light Therapy Device(Models: KFB 265, KFB 293) is a home use wearable LED phototherapy device which can help reduce wrinkles and acne and provide topical heating. LED Light Therapy Device is consisting of main unit(mask), controller, USB charging cable and adapter. There are several kinds of light having medical effects: Red + Infrared (wavelength 637nm (±5nm), Blue light (wavelength 465nm), Amber light (wavelength 605±5nm).
#### V. Indications for Use
#### KFB290, KFB291:
- Red light: Treatment of full-face wrinkles. a.
- Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and b. muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
- Red+Infrared Light: Treatment of full-face wrinkles. C.
- Amber light: Treatment of full-face wrinkles. d.
- e. Blue light: Treatment of mild to moderate inflammatory acne.
- f. Mixed light(Red+Blue +Infrared): Treatment of mild to moderate inflammatory acne.
#### KFB265, KFB293:
- Red+Infrared Light: Treatment of wrinkles. a.
- b. Blue light: Treatment of mild to moderate inflammatory acne.
- Amber light: Treatment of wrinkles. C.
#### VI. Comparison of Technological Characteristics With the Predicate Device
The LED Light Therapy Device has the same intended use as the predicates. The technological characteristics, features, specifications, materials are similar to the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use.
Therefore, the LED Light Therapy Device may be found substantially equivalent to its predicate devices.
{7}------------------------------------------------
| Comparison<br>Elements | Subject Device | Predicate Device | Reference Device 1 | Reference Device 2 |
|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K223544 | K230351 | K162489 |
| Trade name | LED Light Therapy<br>Device(Models:<br>KFB290, KFB291) | LED light therapy<br>mask, Models: FM-01,<br>FM-02, FM-03 | LED Facial Mask,<br>Model(s): MZ-01,<br>NEWKEY-01, SP-<br>FM-01 | RED Light Device |
| Manufacturer | Dongguan Boyuan<br>Intelligent Technology<br>Co.,Ltd | Guangdong Newdermo<br>Biotech Co.,Ltd | Shenzhen Kaiyan<br>Medical Equipment<br>Co., Ltd. | Zhongshan Bisen<br>Plastic Electronic<br>Products Co., Ltd. |
| Regulation number | 21 CFR 878.4810<br>21 CFR 890.5500 | 21 CFR 878.4810<br>21 CFR 890.5500 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Classification Name | Light Based Over The<br>Counter Wrinkle<br>Reduction(OHS),<br>Over-The-Counter<br>Powered Light Based<br>Laser For Acne(OLP),<br>Infrared Therapeutic<br>Heating(ILY) | Light Based Over The<br>Counter Wrinkle<br>Reduction(OHS),<br>Over-The-Counter<br>Powered Light Based<br>Laser For Acne(OLP),<br>Infrared Therapeutic<br>Heating(ILY) | Light Based Over The<br>Counter Wrinkle<br>and Acne Reduction | Light Based Over The<br>Counter Wrinkle<br>Reduction |
| Product code | OHS, OLP, ILY | OHS, OLP, ILY | OHS, OLP | OHS |
| Device classification | Class II | Class II | Class II | Class II |
| Indication for use/<br>Intended use | - Red light: Treatment<br>of full-face wrinkles.<br>- Infrared light: Provide<br>topical heating for the<br>purpose of elevating<br>tissue temperature;<br>arthritis and muscle<br>spasm; relieving<br>stiffness; promoting the<br>relaxation of muscle<br>tissue; and to<br>temporarily increase<br>local blood circulation.<br>- Red+Infrared Light:<br>Treatment of full-face<br>wrinkles.<br>- Amber light:<br>Treatment of full-face<br>wrinkles.<br>- Blue light: Treatment<br>of mild to moderate<br>inflammatory acne. | Red light: Treatment of<br>full-face wrinkles.<br>Blue light: Treatment of<br>mild to moderate<br>inflammatory acne.<br>Infrared light: Provide<br>topical heating for the<br>purpose of elevating<br>tissue temperature;<br>arthritis and muscle<br>spasm; relieving<br>stiffness; promoting the<br>relaxation of muscle<br>tissue; and to<br>temporarily increase<br>local blood circulation.<br>Mixed light:Treatment<br>of mild to moderate<br>inflammatory acne. | LED Facial Mask is an<br>over the counter device<br>that is intended to use<br>LED light for the<br>treatment of wrinkles<br>and mild to moderate<br>acne. | The RED Light Device<br>is an OTC device<br>indicated to emit energy<br>in the red and IR region<br>of the spectrum for use<br>in dermatology for the<br>treatment of periorbital<br>wrinkles. |
| - Mixed light(Red+Blue<br>+Infrared): Treatment<br>of mild to moderate<br>inflammatory acne. | | | | |
| Prescription or OTC | OTC | OTC | OTC | OTC |
| Dimension | KFB290:<br>Approximately 302<br>mm x 197 mm x 22<br>mm<br>KFB291:<br>Approximately 412<br>mm x 195 mm x 22<br>mm | FM-01: 207 * 277 * 43 mm,<br>FM-02: 198 * 383 * 33.5 mm,<br>FM-03: 237.5 * 108 * 8.1 mm | / | / |
| Power supply | Input: 100 -240 V, 50 /<br>60 Hz<br>Output: 5V, 1A<br>Lithium ion battery:<br>1300mAh | Input: 100-240V,<br>50/60Hz, 0.25 A<br>Output: DC 5 V, 500 mA | An external adapter<br>Input: AC 100-240V<br>50-60Hz 0.2A<br>Output: DC 12V 0.5A | Adaptor:100~240V AC<br>50/60Hz Lithium<br>battery: 2x3.7V |
| Sterilization | Not applicable | Not applicable | Not applicable | Not applicable |
| Light source | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes |
| Location for use | Face | Face and body | Face | Face |
| Wavelength | 635nm ± 5nm visible<br>red light;<br>850nm±5nm Invisible<br>red light;<br>465 ± 5nm blue light;<br>605 ± 5nm amber light | Red: 620nm<br>Blue: 460nm<br>Infrared: 850nm<br>Mixed: 620nm and<br>850nm and 460nm | Blue: 465nm±5nm<br>Red: 625nm±5nm<br>Amber: 605nm±5nm | Red: 633nm ±5nm<br>visible red light;<br>Infrared: 830nm±5nm<br>Invisible red light; |
| Irradiance | Red: 25mW/cm²;<br>IR: 3mW/cm²;<br>Red+IR: 30mW/cm²;<br>Blue: 18mW/cm²;<br>Amber: 20mW/cm²;<br>Mixed light:<br>9mW/cm²; | Red light:<br>2.0~3.0mW/cm²<br>Blue light:<br>2.0~4.0mW/cm²<br>Infrared light: 2.0~4.0<br>mW/cm²<br>Mixed light: 9.0~12.0<br>mW/cm² | Blue: 15~63mW/cm²<br>Red: 31~75mW/cm²<br>Amber:<br>15~42mW/cm² | 125 mW/cm²<br>70 mW/cm² (633 nm)<br>55 mW/cm² (830 nm) |
| Treatment time | For red, blue and<br>red+infrared: 10, 20,<br>30 minutes<br>For infrared, amber<br>light and mixed light:<br>10, 20 minutes. | Manual Mode: 15<br>minutes each time,<br>Automatic Mode: 10<br>minutes each time. | 10 minutes/day,<br>3 times per week | 15-20 minutes |
| Compliance<br>with<br>voluntary standards | IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-1-11;<br>IEC 60601-2-83;<br>IEC 62471 | IEC 60601-1;<br>IEC 60601-1-2 | IEC60601-1-2<br>IEC60601-1<br>IEC60601-1-11<br>IEC60601-2-57<br>IEC62471 | IEC60601-1-2<br>IEC60601-1<br>IEC60601-1-11<br>IEC60601-2-57<br>IEC62471 |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23 | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 | Comply with ISO 10993-1, ISO 10993-5 and ISO 10993-10 | ISO 10993-1 |
{8}------------------------------------------------
{9}------------------------------------------------
# Table2: Substantial Equivalence Comparison for KFB265, KFB293
| Comparison Elements | Subject Device | Predicate Device | Reference Device 1 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K221151 | K230351 |
| Trade name | LED Light Therapy<br>Device(Models: KFB 265,<br>KFB 293) | Beauty LED Mask, Model:<br>KFB 265 | LED Facial Mask, Model(s):<br>MZ-01, NEWKEY-01, SP-<br>FM-01 |
| Manufacturer | Dongguan Boyuan Intelligent<br>Technology Co.,Ltd | Hunan Guangye<br>Biotechnology Co., Ltd. | Shenzhen Kaiyan Medical<br>Equipment Co., Ltd. |
| Regulation number | 21 CFR 878.4810<br>21 CFR 890.5500 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Classification Name | Light Based Over The Counter<br>Wrinkle Reduction(OHS),<br>Over-The-Counter Powered<br>Light Based Laser For<br>Acne(OLP), Infrared<br>Therapeutic Heating(ILY) | Light Based Over The<br>Counter Wrinkle and Acne<br>Reduction | Light Based Over The Counter<br>Wrinkle and Acne Reduction |
| Product code | OHS, OLP, ILY | OHS, OLP | OHS, OLP |
| Device classification | Class II | Class II | Class II |
| Indication<br>for use/<br>Intended use | - Red+Infrared Light:<br>Treatment of wrinkles.<br>- Blue light: Treatment of mild<br>to moderate inflammatory<br>acne.<br>- Amber light: Treatment of<br>wrinkles. | The device is intended to use<br>LED light for the treatment of<br>wrinkles and mild to moderate<br>acne. | LED Facial Mask is an over<br>the counter device that is<br>intended to use LED light for<br>the treatment of wrinkles and<br>mild to moderate acne. |
| Prescription or OTC | OTC | OTC | OTC |
| Dimension | KFB 265: Approximately<br>183 mm x 238 mm x 98 mm<br>KFB 293: Approximately<br>228.3 mm x 139.5 mm x 90<br>mm<br>Controller: 100mm x 50<br>mm x 21.5 mm | LED Mask: Approximately<br>183 * 238 * 98 mm<br>Controller: 100 * 50 * 21.5<br>mm | / |
| Power supply | Input: 100 -240 V, 50 / 60 Hz<br>Output: 5V, 1A<br>Lithium ion battery:<br>1300mAh | Input: 100 -240 V ~ 50 /60 Hz<br>Output: 5V 1A | An external adapter Input: AC<br>100-240V 50-60Hz 0.2A<br>Output: DC 12V 0.5A |
| Comparison Elements | Subject Device | Predicate Device | Reference Device 1 |
| 510(k) Number | Pending | K221151 | K230351 |
| Sterilization | Not applicable | Not applicable | Not applicable |
| Light source | Light Emitting Diodes | Light Emitting Diodes | Light Emitting Diodes |
| Location for use | Face | Face | Face |
| Wavelength | visible red light: 637nm ±<br>5nm;<br>invisible red light: 854nm ±<br>5nm;<br>blue light: 465±5nm;<br>amber light: 605 ± 5nm; | Red (637nm±5nm) and<br>IR (854nm±5nm);<br>Blue (465±5nm) | Blue: 465nm±5nm<br>Red: 625nm±5nm<br>Amber: 605nm±5nm |
| Irradiance | Red+IR: 25.5mW/cm²<br>Blue: 1.36mW/cm²<br>Amber: 20mW/cm² | Red+IR: 25.5mW/cm²<br>Blue: 1.36mW/cm² | Blue: 15~63mW/cm²<br>Red: 31~75mW/cm²<br>Amber: 15~42mW/cm² |
| Treatment time | 10 minutes | 10 min | 10 minutes/day,<br>3 times per week |
| Compliance<br>with<br>voluntary standards | IEC 60601-1; | IEC 60601-1; | IEC60601-1-2 |
| | IEC 60601-1-2;<br>IEC 60601-1-11;<br>IEC 60601-2-57;<br>IEC 62471 | IEC 60601-1-2;<br>IEC 60601-1-11;<br>IEC 60601-2-57;<br>IEC 62471 | IEC 60601-1<br>IEC60601-1-11<br>IEC60601-2-57<br>IEC62471 |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23 | All<br>body-contacting<br>materials are complied with<br>ISO10993-5 and ISO 10993-<br>10 | Comply with ISO 10993-1,<br>ISO 10993-5 and ISO 10993-<br>10 |
{10}------------------------------------------------
## VII.Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the LED Light Therapy Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including:
- > ISO 10993-5: 2009, Biological evaluation of medical devices -Par t 5: Tests for in vitro cytotoxicity
- > ISO 10993-10: 2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization
- > ISO 10993-23: 2021, Biological evaluation of medical devices Part 23: Tests for irritation
#### 2) Electrical Safety
- IEC 60601-1: 2005+A1:2012+A2:2020 Medical electrical equipment Part 1: General requirements A for basic safety and essential performance
{11}------------------------------------------------
- A IEC 60601-1-2:2014+A1:2020 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC 60601-1-11: 2015+A1:2020 Medical Electrical Equipment -Part 1-11: General Requirements A for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- A IEC 60601-2-83: 2019+A1:2022 - Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
- > IEC 60601-2-57: 2011 - MEDICAL ELECTRICAL EQUIPMENT- Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic,diagnostic,monitoring and cosmetic/aesthetic use
- > IEC 62133-2: 2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
### 3) Eye Safety
- > IEC 62471: 2006 Photobiological safety of lamps and lamp systems
### 4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
#### VIII. Conclusions
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.