K223748 · Us Medical Glove Company, LLC · LZA · Jan 13, 2023 · General Hospital
Device Facts
Record ID
K223748
Device Name
Nitrile Exam Gloves (Chemotherapy Drug Tested)
Applicant
Us Medical Glove Company, LLC
Product Code
LZA · General Hospital
Decision Date
Jan 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Attributes
3rd-Party Reviewed
Intended Use
A powder-free, nitrile examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Device Story
Disposable, powder-free nitrile examination glove; worn by healthcare personnel during patient examinations to provide a protective barrier against contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05 (2019). Used in clinical settings; provides barrier protection; reduces risk of exposure to hazardous drugs during handling.
Clinical Evidence
Bench testing only. Testing performed per ASTM D6978-05 (2019) to assess permeation breakthrough times for 12 chemotherapy drugs. Results demonstrate no breakthrough up to 240 minutes for 10 drugs; specific breakthrough times provided for Carmustine and Thiotepa.
Technological Characteristics
Material: Nitrile. Form factor: Powder-free examination glove. Testing standard: ASTM D6978-05 (2019) for chemotherapy drug permeation. Class I medical device.
Indications for Use
Indicated for use as a disposable medical glove worn by examiners to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs; contraindicated for use with Carmustine and Thiotepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K201772 — Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) · Hartalega NGC Sdn. Bhd. · Aug 25, 2020
K191753 — Medline Sterile Powder-Free Nitrile Examination Gloves - Green (Tested For Use with Chemotherapy Drugs) · Medline Industries, Inc. · Mar 17, 2020
K160560 — Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) · Hartalega Sdn Bhd · Aug 4, 2016
K170794 — Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (White) · Hartalega NGC Sdn. Bhd. · Jun 30, 2017
K151997 — Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs-Violet Blue (VBLU) · Hartalega NGC Sdn. Bhd. · Mar 31, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 13, 2023
US Medical Glove Company LLC % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K223748
Trade/Device Name: Nitrile Exam Gloves (Chemotherapy Drug Tested) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ Dated: December 13, 2022 Received: December 14, 2022
## Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K223748
Device Name Nitrile Exam Gloves (Chemotherapy Drug Tested)
#### Indications for Use (Describe)
A powder-free, nitrile examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
| Drug Name | Concentration Tested | Minimum Breakthrough Detection Time (Minutes) |
|----------------------------|-------------------------|-----------------------------------------------|
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml (20,000 ppm) | No breakthrough up to 240 |
| Doxorubicin Hydrochloride | 2.0 mg/ml (2,000 ppm) | No breakthrough up to 240 |
| Etoposide (Toposar) | 20.0 mg/ml (20,000 ppm) | No breakthrough up to 240 |
| Fluorouracil | 50.0 mg/ml (50,000 ppm) | No breakthrough up to 240 |
| Paclitaxel (Taxol) | 6.0 mg/ml (6,000 ppm) | No breakthrough up to 240 |
| Cisplatin | 1.0 mg/ml (1,000 ppm) | No breakthrough up to 240 |
| Dacarbazine (DTIC) | 10.0 mg/ml (10,000 ppm) | No breakthrough up to 240 |
| Ifosfamide | 50.0 mg/ml (50,000 ppm) | No breakthrough up to 240 |
| Mitoxantrone | 2.0 mg/ml (2,000 ppm) | No breakthrough up to 240 |
| Vincristine Sulfate | 1.0 mg/ml (1,000 ppm) | No breakthrough up to 240 |
| Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | No breakthrough up to 14.3 * |
| Thiotepa | 10.0 mg/ml (10,000 ppm) | No breakthrough up to 48.2 * |
*Please note that Carmustine and Thiotepa have low permeation times. Do not use with Carmustine and/or Thiotepa.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|-------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="font-size:20px">☒</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.