Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)
K201772 · Hartalega NGC Sdn. Bhd. · LZA · Aug 25, 2020 · General Hospital
Device Facts
Record ID
K201772
Device Name
Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)
Applicant
Hartalega NGC Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Aug 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.
Device Story
Nitrile powder-free examination gloves; non-sterile; disposable. Worn by healthcare examiners to provide barrier protection against patient-to-examiner contamination. Tested for permeation resistance against specific chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Used in clinical settings; provides protection during handling of hazardous drugs.
Clinical Evidence
Bench testing only. Permeation testing performed per ASTM D6978-05 (Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs) for a panel of chemotherapy agents and Fentanyl Citrate. Breakthrough detection times reported for all tested substances.
Technological Characteristics
Nitrile material; powder-free; blue color. Permeation resistance tested per ASTM D6978-05. Non-sterile, disposable form factor.
Indications for Use
Indicated for use as a protective barrier on the examiner's hands to prevent contamination between patient and examiner. Suitable for use with chemotherapy drugs and Fentanyl Citrate. Available for over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K200019 — Nitrile Powder Free Examination Gloved Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue), Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Black) · Hartalega NGC Sdn. Bhd. · Apr 6, 2020
K201530 — Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Fusion Dark Grey) · Hartalega NGC Sdn. Bhd. · Aug 5, 2020
K151997 — Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs-Violet Blue (VBLU) · Hartalega NGC Sdn. Bhd. · Mar 31, 2016
K160560 — Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) · Hartalega Sdn Bhd · Aug 4, 2016
K172442 — Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) · Hartalega NGC Sdn. Bhd. · Nov 6, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 25, 2020
Hartalega NGC Sdn. Bhd. Nurul Aisyah Kong Senior Manager - Quality Assurance No. 1. Persiaran Tanjung Kawasan Perindustrian Tanjung Sepang, 43900, Malaysia 603 (3280) 388-8 [voice] 603 (3271) 013-5 [fax] wkkong@hartalega.com.my
Re: K201772
Trade/Device Name: Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO Dated: June 26, 2020 Received: June 29, 2020
Dear Nurul Aisyah Kong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known) K201772
#### Device Name
Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)
## Indications for Use (Describe)
Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.
These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
|---------------------------------------|------------------------------------------------|
| Carmustine (3.3 mg/ml) | 11.5 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Dacarbazine (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| Etoposide (20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 24.3 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
| Azacytidine (25.0 mg/ml) | >240 |
| Carboplatin (10.0 mg/ml) | >240 |
| Docetaxel (10 mg/ml) | >240 |
| Epirubicin (2.0 mg/ml) | >240 |
| Gemcitabine (38 mg/ml) | >240 |
| Ifosfamide (50 mg/ml) | >240 |
| Irinotecan (20 mg/ml) | >240 |
| Mitoxantrone (2.0 mg/ml) | >240 |
| Oncovin (1.0 mg/ml) | >240 |
| Oxaliplatin (5 mg/ml) | >240 |
| Vinorelbine (10 mg/ml) | >240 |
Please note that Carmustine and Thiotepa have extremely low permeation times of 11.5 minutes and 24.3 minutes respectively.
| Fentanyl Citrate and Concentration<br>Fentanyl Citrate Injection (100 mcg/2ml) | Minimum Breakthrough Detection Time in Minutes<br>>240 |
|--------------------------------------------------------------------------------|--------------------------------------------------------|
| Type of Use (Select one or both, as applicable) | |
| ☐ Prescription Use (Part 21 CFR 801 Subpart D) | ☑ Over-The-Counter Use (21 CFR 801 Subpart C) |
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.