K222237 · Cala Health, Inc. · QBC · Nov 22, 2022 · Neurology
Device Facts
Record ID
K222237
Device Name
Cala kIQ
Applicant
Cala Health, Inc.
Product Code
QBC · Neurology
Decision Date
Nov 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5897
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cala kIQ is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor. Cala kIQ is indicated to aid in the temporary relief of postural and kinetic hand tremor symptoms that impact some activities of daily living in the treated hand following stimulation in adults with Parkinson's disease.
Device Story
Cala kIQ is a wrist-worn stimulator for hand tremor relief via Transcutaneous Afferent Patterned Stimulation (TAPS) to median and radial nerves. Device includes stimulator, conductive electrode band, and charging base station. During setup, patient performs postural hold; onboard sensors measure tremor frequency to calibrate patient-specific stimulation pattern. Patient uses buttons to initiate therapy, adjust amplitude, and complete calibration. Used in-home by patients after initial calibration in clinic or home. Output is biphasic, rectangular electrical stimulation. Therapy sessions last 40 minutes. Device provides temporary tremor relief, potentially improving performance of daily living activities (e.g., writing, pouring, using utensils).
Clinical Evidence
Two clinical studies provided. Study 1 (n=40, Parkinson's): Single-arm, non-significant risk study. TAPS therapy showed improvement in MDS-UPDRS tremor tasks and Bain and Findley ADL tasks. Responder rates (≥0.5-point improvement) ranged from 28-70% across tasks. Study 2 (n=19, crossover): Prospective, two-arm study comparing Cala kIQ to Cala Trio. Results showed equivalent safety and effectiveness between devices. No device-related serious adverse events reported in either study.
Technological Characteristics
Wrist-worn electrical stimulator; TAPS therapy; biphasic rectangular waveform; 150 Hz frequency; 650 µs pulse duration; 300 µs phase duration. Materials: elastic, microfiber, conductive electrodes. Connectivity: standalone with base station charging. Complies with IEC 60601 (electrical/mechanical/thermal safety) and ISO 10993-1:2018 (biocompatibility).
Indications for Use
Indicated for adults with essential tremor or Parkinson's disease experiencing hand tremors or postural/kinetic tremor symptoms impacting activities of daily living.
Regulatory Classification
Identification
An external upper limb tremor stimulator is a prescription device which is placed externally on the upper limb and designed to aid in tremor symptom relief of the upper limb.
Special Controls
In combination with the general controls of the FD&C Act, the external upper limb tremor stimulator is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must assess the following:
(i) Characterization of the electrical stimulation, including the following, must be performed: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(ii) Impedance testing, current distribution across the electrode surface area, adhesive integrity, and shelf life testing of the electrodes and gels must be conducted.
(iii) Simulated use testing of sensor performance and the associated algorithms that determine the stimulation output must be conducted.
(2) Patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Physician and patient labeling must include:
(i) Summaries of electrical stimulation parameters;
(ii) Instructions on how to correctly use and maintain the device;
(iii) Instructions and explanations of all user-interface components;
(iv) Instructions on how to clean the device;
(v) A shelf life for the electrodes and gel; and
(vi) Reuse information.
{0}------------------------------------------------
Build Correspondence
Image /page/0/Picture/2 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
11/22/2022
Cala Health, Inc. Danielle McDonnell Boyd Senior Director, Regulatory Affairs 1800 Gateway Dr. Suite 300 San Mateo, California 94404
Re: K222237
Trade/Device Name: Cala kIQTM Regulation Number: 21 CFR 882.5897 Regulation Name: External upper limb tremor stimulator Regulatory Class: Class II Product Code: QBC Dated: October 21, 2022 Received: October 21, 2022
Dear Danielle Boyd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Date: 2022.11.22 Xiaorui Tang 13:34:34 -05'00'
For:
CDR Jitendra Virani Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K22237
Device Name Cala kIQ
Indications for Use (Describe)
Cala kIQ is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor.
Cala kIQ is indicated to aid in the temporary relief of postural and tremor symptoms that impact some activities of daily living in the treated hand following stimulation in adults with Parkinson's disease.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cala Health. The logo consists of the word "CALA" in a stylized, sans-serif font, with a curved line extending from the bottom of the "C" to the top of the second "A". Below the word "CALA" is the word "HEALTH" in a smaller, sans-serif font. The logo is a light blue color.
## 510(k) Summary
#### I. SUBMITTER
| Manufacturer: | Cala Health, Inc.<br>1800 Gateway Drive, Suite 300 San Mateo, CA 94404<br>Phone: (415) 890-3961<br>Fax: None |
|---------------|--------------------------------------------------------------------------------------------------------------|
|---------------|--------------------------------------------------------------------------------------------------------------|
- Primary Contact: Danielle McDonnell Boyd Senior Director, Regulatory Affairs & Quality Assurance Cala Health danielle@calahealth.com (415) 819-2935
July 22, 2022 Date Prepared:
#### II. SUBJECT DEVICE
| Trade Name: | Cala kIQTM |
|------------------------|---------------------------------------|
| Classification Name: | External upper limb tremor stimulator |
| Device Classification: | Class II |
| Regulation: | 21 CFR 882.5897 |
| Product Code: | QBC |
#### III. PREDICATE DEVICE
| Predicate Device: | Cala Trio™ |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Prior Submissions: | DEN170028, K182706, K203288 |
| Indications for Use: | Cala Trio is indicated to aid in the temporary relief of hand tremors in<br>the treated hand following stimulation in adults with essential tremor. |
#### DEVICE DESCRIPTION IV.
Cala kIQ is a small, lightweight, wrist-worn stimulator device designed to aid in the relief of hand tremors by applying a calibrated Transcutaneous Afferent Patterned Stimulation (TAPS) to the median and radial nerves of a patient's wrist.
{4}------------------------------------------------
The TAPS output pattern is calibrated to the individual patient's tremor frequency. During device setup, the patient performs a postural hold, during which the onboard sensors measure the tremor frequency which is then used to customize the TAPS output pattern to deliver therapy to the median and radial nerves at the appropriate, patient-specific, optimal stimulation frequency.
Cala kIQ is comprised of the following components:
- Stimulator: Contains sensors, electronics, and firmware for delivering TAPS therapy . and provides user interface
- Band: Contains multi-use, conductive electrodes spaced at intervals to align with . targeted nerves and attachment points for Cala kIQ Stimulator
- . Base Station: Provides charging function and contains an indicator light to alert the user to operating conditions.
The Cala kIQ Stimulator contains the electronics, sensors, and firmware for calibrating and delivering TAPS therapy. The patient uses the Cala kIQ Stimulator buttons to complete device calibration, adjust the stimulation amplitude, and initiate therapy sessions. The Cala kIQ Stimulator is attached to the Cala kIQ band, which includes embedded electrodes placed at appropriate intervals to properly target the median and radial nerves. To accommodate a broad distribution of wrist sizes, the band is available in three sizes (small, medium, and large).
#### V. INDICATIONS FOR USE
Cala kIQ is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor.
Cala kIQ is indicated to aid in the temporary relief of postural and kinetic hand tremor symptoms that impact some activities of daily living in the treated hand following stimulation in adults with Parkinson's Disease.
#### VI. SUBSTANTIAL EQUIVALENCE
Table 1 below provides a summary comparison between the Subject Device, Cala kIQ and the Predicate Device, Cala Trio, with respect to intended use, technological characteristics, and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
{5}------------------------------------------------
## Table 1. Cala kIQ Substantial Equivalence
| | Predicate Device | Subject Device | Substantially<br>Equivalent? |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Cala Trio | Cala kIQ | |
| | K203288 | K222237 | |
| 510(k) Number | K203288 | K222237 | |
| Manufacturer | Cala Health | Cala Health | |
| Intended Use | Delivery of transcutaneous afferent patterned<br>stimulation ("TAPS") for treatment of hand<br>tremors | Delivery of transcutaneous afferent patterned<br>stimulation ("TAPS") for treatment of hand<br>tremors | Yes<br>Subject Device is identica<br>to Predicate Device |
| Indications for Use | Cala Trio is indicated to aid in the temporary<br>relief of hand tremors in the treated hand<br>following stimulation in adults with essential<br>tremor | Cala kIQ is indicated to aid in the temporary<br>relief of hand tremors in the treated hand<br>following stimulation in adults with essential<br>tremor.<br>Cala kIQ is indicated to aid in the temporary<br>relief of postural and kinetic hand tremor<br>symptoms that impact some activities of daily<br>living in the treated hand following<br>stimulation in adults with Parkinson's<br>disease. | Yes<br>Substantial equivalence<br>demonstrated through<br>clinical evidence |
| Target Population | Adults with essential tremor | Adults with essential tremor or Parkinson's<br>Disease | Yes<br>Substantial equivalence<br>demonstrated through<br>clinical evidence |
| Anatomical site | Median and radial nerves | Median and radial nerves | Yes<br>Subject Device is identica<br>to Predicate Device |
| Intended Users | Patient | Patient | Yes<br>Subject Device is identica<br>to Predicate Device |
| | Predicate Device<br>Cala Trio<br>K203288 | Subject Device<br>Cala kIQ<br>K222237 | Substantially<br>Equivalent? |
| Clinical Setting | In-home use after an initial calibration in the<br>patient's home or in the physician's office. | In-home use after an initial calibration in the<br>patient's home or in the physician's office. | Yes<br>Subject Device is identical<br>to Predicate Device |
| Rx or OTC use | Prescription Use only | Prescription Use only | Yes<br>Subject Device is identical<br>to Predicate Device |
| | | Design | |
| Technology | Calibrated Transcutaneous Afferent Patterned<br>Stimulation (TAPS) delivered through<br>electrodes embedded on wearable band | Calibrated Transcutaneous Afferent Patterned<br>Stimulation (TAPS) delivered through<br>electrodes embedded on wearable band | Yes<br>Subject Device is identical<br>to Predicate Device |
| Energy used or<br>delivered | Electrical stimulation | Electrical stimulation | Yes<br>Subject Device is identical<br>to Predicate Device |
| Human Factors | Wrist-worn electrical stimulator with<br>detachable band containing multi-use<br>electrodes. Base station provides charging<br>function and contains indicator lights for<br>operating status | Wrist-worn electrical stimulator with<br>detachable band containing multi-use<br>electrodes. Base station provides charging<br>function and contains indicator lights for<br>operating status | Yes<br>Subject Device is identical<br>to Predicate Device |
| Patient Contacting<br>Materials | The following components of the Cala Trio<br>Band have Surface, Intact Skin Long-Term<br>contact (>30 days):<br>• Elastic<br>• Electrodes<br>• Microfiber<br>• Elastic<br>• Band thread | The following components of the Cala kIQ<br>Band have Surface, Intact Skin Long-Term<br>contact (>30 days):<br>• Elastic<br>• Electrodes<br>• Microfiber<br>• Elastic<br>• Band thread | Yes<br>Subject Device is identical<br>to Predicate Device |
| | Predicate Device<br>Cala Trio<br>K203288 | Subject Device<br>Cala kIQ<br>K222237 | Substantially<br>Equivalent? |
| Operating Principle | Transcutaneous Afferent Patterned<br>Stimulation (TAPS) to the median and<br>radial nerves of a patient's wrist | Transcutaneous Afferent Patterned<br>Stimulation (TAPS) to the median and radial<br>nerves of a patient's wrist | Yes<br>Subject Device is identical<br>to Predicate Device |
| Electrodes | 1 common, 2 active electrodes<br><br>Individual Electrode Surface area<br>$22mm x 22mm = 4.84 cm^2$ | 2 common electrodes paired with 2 active<br>electrodes<br><br>Individual Electrode Surface area<br>$22mm x 6mm = 1.32 cm^2$ | Yes<br>Substantial equivalence<br>demonstrated through<br>non- clinical performance<br>testing and clinical<br>evidence |
| Battery and Base<br>Station | Rechargeable Lithium-ion battery and AC-<br>powered charger. | Rechargeable Lithium-ion battery and AC-<br>powered charger. | Yes<br>Subject Device is identical<br>to Predicate Device |
| | Performance | | |
| User Workflow | 1. Calibration<br>2. Set stimulation intensity<br>3. Therapy available on demand | 1. Calibration<br>2. Set stimulation intensity<br>3. Therapy available on demand | Yes<br>Subject Device is identical<br>to Predicate Device |
| Treatment Time | 40 minutes | 40 minutes | Yes<br>Subject Device is identical<br>to Predicate Device |
| Wristband life | 90 days | 90 days | Yes<br>Subject Device is identical<br>to Predicate Device. |
| | Predicate Device<br>Cala Trio<br>K203288 | Subject Device<br>Cala kIQ<br>K222237 | Substantially<br>Equivalent? |
| Shelf-life | Cala Trio Band: 24 months | Cala kIQ Band: 3 months | Yes<br>Shelf-life testing confirms<br>Cala kIQ Band meets<br>product specification at<br>labeled shelf-life.<br>Shelf-life difference from<br>predicate due to<br>availability of test data,<br>not performance<br>characteristics |
| Output Specifications | | | |
| Waveform (e.g.,<br>pulsed monophasic,<br>biphasic) | Biphasic | Biphasic | Yes<br>Subject Device is identical<br>to Predicate Device |
| Shape (e.g.,<br>rectangular, spike,<br>rectified sinusoidal) | Rectangular | Rectangular | Yes<br>Subject Device is identical<br>to Predicate Device |
| Maximum Output<br>Voltage<br>(volts) | 4 @ 500Ω | 4 @ 500Ω | Yes<br>Subject Device is identical<br>to Predicate Device |
| | 80 @ 10kΩ | 80 @ 10kΩ | |
| Maximum Output<br>Current<br>(mA) | 8 @ 500Ω | 8 @ 500Ω | Yes<br>Subject Device is identical<br>to Predicate Device |
| | 8 @ 10kΩ | 8 @ 10kΩ | |
| Duration of primary<br>(depolarizing) phase<br>(μsec) | 300 | 300 | Yes<br>Subject Device is identical<br>to Predicate Device |
| | Predicate Device<br>Cala Trio<br>K203288 | Subject Device<br>Cala kIQ<br>K222237 | Substantially Equivalent? |
| Pulse Duration<br>(μsec) | 650 | 650 | Yes<br>Subject Device is identical<br>to Predicate Device |
| Frequency<br>(Hz) | 150 | 150 | Yes<br>Subject Device is identical<br>to Predicate Device |
| Symmetrical phases? | Yes | Yes | Yes<br>Subject Device is identical<br>to Predicate Device |
| Phase Duration<br>(μS) | 300 each phase | 300 each phase | Yes<br>Subject Device is identical<br>to Predicate Device |
| Net Charge<br>(μC) | 0 @500Ω | 0 @500Ω | Yes<br>Subject Device is identical<br>to Predicate Device |
| Maximum Phase<br>Charge<br>(μC) | 2.4 @ 500Ω | 2.4 @ 500Ω | Yes<br>Subject Device is identical<br>to Predicate Device |
| Maximum Current<br>Density<br>(mA/cm², r.m.s.) | 0.50 @ 500Ω | 1.29 @ 500Ω | Yes<br>Substantial equivalence<br>demonstrated through non-<br>clinical performance testing<br>and clinical evidence |
| Maximum Average<br>Current<br>(mA) (average<br>absolute value) | 0.72 @ 500Ω | 0.72 @ 500Ω | Yes<br>Subject Device is identical<br>to Predicate Device |
| | Predicate Device<br>Cala Trio<br>K203288 | Subject Device<br>Cala kIQ<br>K222237 | Substantially<br>Equivalent? |
| Maximum Average<br>Power Density<br>(mW/cm²) | 0.59 @ 500Ω<br>(.0006 W/cm²) | 2.18 @ 500Ω<br>(0.0022 W/cm²) | Yes<br>Substantial equivalence<br>demonstrated through non-<br>clinical performance testing<br>and clinical evidence |
| Safety | | | |
| Electrical safety | Conforms to IEC 60601 Electrical Safety | Conforms to IEC 60601 Electrical Safety | Yes<br>Subject Device is identical<br>to Predicate Device |
| Compatibility<br>with intended<br>environments | Conforms to EMC requirements | Conforms to EMC requirements | Yes<br>Subject Device is identical<br>to Predicate Device |
| Mechanical safety | Conforms to IEC 60601 Electrical Safety | Conforms to IEC 60601 Electrical Safety | Yes<br>Subject Device is identical<br>to Predicate Device |
| Chemical safety | Not applicable | Not applicable | Yes<br>Subject Device is identical<br>to Predicate Device |
| Thermal safety | Conforms to IEC 60601 Electrical Safety | Conforms to IEC 60601 Electrical Safety | Yes<br>Subject Device is identical<br>to Predicate Device |
| Radiation safety | Not applicable | Not applicable | Yes<br>Subject Device is identical<br>to Predicate Device |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
#### VII. PERFORMANCE DATA
Non-clinical performance testing confirmed that Cala kIQ complies with performance standards and functions as intended.
Performance testing included bench testing to confirm electrode function, physical characteristics, mechanical function, shelf life, and system performance. Testing confirmed that the Cala kIQ meets the product requirements.
EMC and Electrical Safety testing demonstrated that Cala kIO is compliant with the applicable IEC 60601 clauses. Biocompatibility evaluations and assessments per FDA Guidance documents demonstrate that Cala kIQ meets the ISO 129993-1:2018 standard and is acceptable for use.
### VIII. CLINICAL DATA
Clinical data to support substantial equivalence for the differences in indicated patient populations and technological characteristics includes evidence from two clinical studies.
The first clinical study was a single-arm. non-significant risk study that evaluated calibrated TAPS therapy in patients with Parkinson's Disease who also had postural hand tremor.
A total of forty patients were enrolled in the study. There were no reports of device-related serious adverse events, and all device-related adverse events were resolved with minimal intervention.
Across visits, TAPS therapy improved 0.5±0.5 points on an unvalidated subset of MDS-UPDRS dominant hand tasks listed below:
- Rest Tremor Amplitude ●
- Postural Tremor ●
- Kinetic Tremor ●
- Pronation-Supination ●
- Finger Tapping ●
- Hand Movements ●
Also, across visits, TAPS therapy improved the ability to perform some activities of daily living by 0.4±0.5 points based on an unvalidated subset of BF-ADL tasks listed below:
- Use a spoon to drink soup .
- Hold a cup of tea ●
- Pour milk from a bottle or carton ●
- Dial a telephone
- Pick up your change in a shop
- Insert an electric plug into a socket ●
- . Unlock your front door with a key
- Write a letter ●
{12}------------------------------------------------
These results were descriptive in nature, different patients showed improvement in different tasks tested. None of the patients showed improvement in all of the tasks tested and some of the patients showed no improvement in any of the tasks tested.
Pre-stimulation to post-stimulation change was evaluated for each of the MDS-UPDRS Part III dominant hand tasks and the Bain and Findley ADL tasks at both Visits 2 and 3. The MDS-UPDRS and BF-ADL scales are assessed at 1-point increments and improvement was calculated for the average scores as well as for each individual task. The responder rate, defined as the percent of subjects that had a ≥ 0.5-point average per-task improvement, was calculated for the average scores as well as for each individual task. The study design included two visits with study clinicians and a 2-week home use period of during which subjects were instructed to complete TAPS therapy sessions twice daily. Assessment performed during study visits were completed while subjects were off medication, home-use sessions completed with subjects on medication per their standard of care.
Table 2 below provides study data for all enrolled subjects and includes responder rates for assessment averages and individual tasks. Of note, the CGI-I were used as secondary endpoints in the study:
Table 3 below provides the responder rates for study subjects with a score of > 2 for a specific task at the applicable visit. The data in Table 3 are a subset of the results summarized in Table 2.
{13}------------------------------------------------
| | Visit 2 | | | | Visit 3 | | | | | |
|--------------------------------------------------------------|---------|------------------|-------------------|---------------------|-----------------------|---------|------------------|-------------------|---------------------|-----------------------|
| | N | Pre | Post | Change | Responder<br>Rate | N | Pre | Post | Change | Responder<br>Rate |
| | | Mean ± SD | Mean ± SD | Mean ± SD | (%) | | Mean ± SD | Mean ± SD | Mean ± SD | (%) |
| | | | | | MDS-UPDRS | | | | | |
| Average<br>across 6<br>tasks* | 40 | 1.63 ± 0.55 | 1.20 ± 0.69 | -0.44 ± 0.44 | 40% | 36 | 1.51 ± 0.50 | 1.11 ± 0.57 | -0.40 ± 0.37 | 44% |
| Rest Tremor | 40 | 1.85 ± 1.10 | 1.33 ± 1.07 | -0.53 ± 0.82 | 45% | 36 | 1.94 ± 0.95 | 1.44 ± 1.05 | -0.50 ± 0.94 | 39% |
| Postural<br>Tremor | 40 | 2.30 ± 0.88 | 1.35 ± 1.08 | -0.95 ± 0.88 | 70% | 36 | 2.14 ± 0.96 | 1.31 ± 1.06 | -0.83 ± 0.88 | 61% |
| Kinetic<br>Tremor | 40 | 1.08 ± 0.92 | 0.75 ± 0.81 | -0.33 ± 0.53 | 35% | 36 | 1.06 ± 0.92 | 0.83 ± 0.74 | -0.22 ± 0.42 | 22% |
| Pronation-<br>Supination | 40 | 1.50 ± 0.91 | 1.23 ± 0.95 | -0.28 ± 0.72 | 33% | 36 | 1.31 ± 0.95 | 1.17 ± 1.00 | -0.14 ± 0.76 | 28% |
| Finger<br>Tapping | 40 | 1.75 ± 0.87 | 1.45 ± 0.96 | -0.30 ± 0.76 | 28% | 36 | 1.47 ± 0.84 | 1.03 ± 0.77 | -0.44 ± 0.56 | 42% |
| Hand<br>Movements | 40 | 1.30 ± 0.82 | 1.10 ± 0.87 | -0.20 ± 0.79 | 35% | 36 | 1.14 ± 0.80 | 0.89 ± 0.67 | -0.25 ± 0.69 | 28% |
| | Visit 2 | | | | Visit 3 | | | | | |
| | N | Pre<br>Mean ± SD | Post<br>Mean ± SD | Change<br>Mean ± SD | Responder<br>Rate (%) | N | Pre<br>Mean ± SD | Post<br>Mean ± SD | Change<br>Mean ± SD | Responder<br>Rate (%) |
| BF-ADLs | | | | | | | | | | |
| Average<br>ADL Rating<br>(8 Tasks,<br>8-32)* | 40 | 1.93 ± 0.50 | 1.48 ± 0.43 | -0.44 ± 0.43 | 43% | 36 | 1.76 ± 0.48 | 1.44 ± 0.37 | -0.32 ± 0.39 | 25% |
| Use a spoon<br>to drink soup | 40 | 2.23 ± 0.97 | 1.60 ± 0.74 | -0.63 ± 0.74 | 55% | 36 | 2.19 ± 0.92 | 1.78 ± 0.76 | -0.42 ± 0.77 | 33% |
| Hold a cup of<br>tea | 40 | 2.13 ± 0.94 | 1.75 ± 0.84 | -0.38 ± 0.84 | 40% | 36 | 2.03 ± 1.00 | 1.69 ± 0.79 | -0.33 ± 0.68 | 25% |
| Pour milk<br>from a bottle<br>or carton | 40 | 2.00 ± 0.72 | 1.48 ± 0.68 | -0.53 ± 0.78 | 45% | 36 | 1.89 ± 0.67 | 1.56 ± 0.73 | -0.33 ± 0.68 | 42% |
| Dial a<br>telephone | 40 | 1.75 ± 0.74 | 1.23 ± 0.58 | -0.53 ± 0.72 | 45% | 36 | 1.53 ± 0.61 | 1.17 ± 0.38 | -0.36 ± 0.54 | 33% |
| Pick up your<br>change in a<br>shop | 40 | 1.60 ± 0.71 | 1.33 ± 0.53 | -0.28 ± 0.64 | 23% | 36 | 1.42 ± 0.60 | 1.31 ± 0.52 | -0.11 ± 0.52 | 19% |
| Insert an<br>electric plug<br>into a socket | 40 | 1.65 ± 0.70 | 1.30 ± 0.52 | -0.35 ± 0.70 | 38% | 36 | 1.42 ± 0.65 | 1.14 ± 0.42 | -0.28 ± 0.61 | 25% |
| Unlock your<br>front door<br>with a key | 40 | 1.59 ± 0.72 | 1.33 ± 0.53 | -0.26 ± 0.59 | 23% | 36 | 1.44 ± 0.61 | 1.17 ± 0.38 | -0.28 ± 0.51 | 25% |
| | Visit 2 | | | | | Visit 3 | | | | |
| | N | Pre | Post | Change | Responder<br>Rate | N | Pre | Post | Change | Responder<br>Rate |
| | | Mean ± SD | Mean ± SD | Mean ± SD | (%) | | Mean ± SD | Mean ± SD | Mean ± SD | (%) |
| Write a letter | 40 | 2.48 ± 0.78 | 1.85 ± 0.74 | -0.63 ± 0.70 | 53% | 36 | 2.19 ± 0.86 | 1.72 ± 0.81 | -0.47 ± 0.61 | 42% |
| Global Impression - Improvement | | | | | | | | | | |
| Clinical<br>Global<br>Impression –<br>Improvement<br>(CGI-I) | 40 | N/A | | | 78%<br>(31/40) | 36 | N/A | | | 83%<br>(30/36) |
| Patient<br>Global<br>Impression –<br>Improvement<br>(PGI-I) | 40 | N/A | | | 75%<br>(30/40) | 36 | N/A | | | 81%<br>(29/36) |
# Table 2. PD-02 Responder Rates for All Enrolled Subjects
{14}------------------------------------------------
{15}------------------------------------------------
*For averaged scores, a Responder was defined as subjects that had a ≥ 0.5-point average per-task improvement
{16}------------------------------------------------
| Visit 2 | | | Visit 3 | | | |
|---------------------------------------|---------------------|---------------------|--------------------|---------------------|---------------------|--------------------|
| N | Change<br>Mean ± SD | Responder Rate (%) | N | Change<br>Mean ± SD | Responder Rate (%) | |
| MDS-UPDRS | | | | | | |
| Action Tremor (Postural and Kinetic)* | 18 | -0.61 ± 0.58 | 67% | 16 | -0.59 ± 0.61 | 63% |
| Rest Tremor Amplitude | 28 | -0.79 ± 0.79 | 61% | 28 | -0.57 ± 1.00 | 39% |
| Postural Tremor | 33 | -1.00 ± 0.94 | 70% | 28 | -0.93 ± 0.94 | 64% |
| Kinetic Tremor | 13 | -0.54 ± 0.52 | 54% | 12 | -0.58 ± 0.51 | 58% |
| Pronation-Supination | 17 | -0.59 ± 0.80 | 53% | 13 | -0.23 ± 1.01 | 38% |
| Finger Tapping | 21 | -0.57 ± 0.81 | 38% | 17 | -0.65 ± 0.61 | 59% |
| Hand Movements | 15 | -0.47 ± 0.64 | 40% | 10 | -0.90 ± 0.74 | 70% |
| BF-ADLs | | | | | | |
| Use a spoon to drink soup | 30 | -0.87 ± 0.68 | 73% | 27 | -0.56 ± 0.85 | 44% |
| Hold a cup of tea | 29 | -0.59 ± 0.87 | 55% | 23 | -0.52 ± 0.79 | 39% |
| Pour milk from a bottle or carton | 30 | -0.73 ± 0.78 | 60% | 26 | -0.46 ± 0.76 | 58% |
| Dial a telephone | 23 | -0.91 ± 0.73 | 78% | 17 | -0.76 ± 0.56 | 71% |
| Visit 2 | | | Visit 3 | | | |
| | N | Change<br>Mean ± SD | Responder Rate (%) | N | Change<br>Mean ± SD | Responder Rate (%) |
| Pick up your change in a shop | 19 | $-0.63 \pm 0.76$ | 47% | 13 | $-0.54 \pm 0.52$ | 54% |
| Insert an electric plug into a socket | 21 | $-0.81 \pm 0.60$ | 71% | 12 | $-0.83 \pm 0.83$ | 75% |
| Unlock your front door with a key | 18 | $-0.61 \pm 0.70$ | 50% | 14 | $-0.71 \pm 0.61$ | 64% |
| Write a letter | 36 | $-0.69 \pm 0.71$ | 58% | 27 | $-0.63 \pm 0.63$ | 56% |
# Table 3. PD-02 Responder Rates for Subjects with Score ≥ 2 per Task
{17}------------------------------------------------
*For averaged scores, a Responder was defined as subjects that had a ≥ 0.5-point average per-task improvement
{18}------------------------------------------------
Considering existing data for versions of this device, reviewed and cleared under previous submissions for patients with similar symptoms, this study provides sufficient information to demonstrate that Cala kIQ may be effective to aid in the temporary relief of postural and kinetic hand tremor symptoms that impact some activities of daily living in the treated hand following stimulation in adults with Parkinson's disease.
The second study was a prospective, single-center, two-arm, non-significant risk, crossover study demonstrating that Cala kIO provides TAPS therapy with equivalent safety and effectiveness as compared to Cala Trio.
A total of 19 subjects completed the study, 6 of which had Parkinson's Disease. There were no reports of device-related serious adverse events, and all were resolved with minimal intervention. Clinician-rated tremor assessments (TETRAS and MDS-UPDRS), and subject-rated assessments (BF- ADLs) results were similar for Cala kIQ and Cala Trio.
#### SUBSTANTIAL EQUIVALENCE IX.
Substantial equivalence of the Subject Device, Cala kIQ, to the Predicate Device, Cala Trio, is demonstrated based on the comparison of labeling, technical characteristics, performance, and both clinical and non-clinical testing.
The intended use is unchanged and the difference in the indications for use has been supported by clinical data as described above.
Performance testing demonstrated that Cala kIQ complies with the same special controls and the same consensus and performance standards as Cala Trio.
#### CONCLUSION X.
The Subject Device, Cala kIQ, has the same intended use as the Predicate Device, Cala Trio. The differences between the subject and predicate device indications for use do not alter the intended use of the device. Clinical data support the temporary relief of postural and kinetic hand tremor symptoms of some activities of daily living in Parkinson's Disease and demonstrate that Cala kIQ delivers TAPS therapy with equivalent safety and effectiveness as Cala Trio.
The subject and predicate device have similar technological characteristics and the differences do not raise new questions of safety or effectiveness. Performance testing demonstrates that Cala kIQ is as safe, as effective, and performs as well as Cala Trio.
Cala kIQ and Cala Trio are substantially equivalent. Cala Health has provided evidence as described above to demonstrate that any differences in the indications for use and technological characteristics do not impact safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.