K203288 · Cala Health, Inc. · QBC · Oct 5, 2021 · Neurology
Device Facts
Record ID
K203288
Device Name
Cala Trio
Applicant
Cala Health, Inc.
Product Code
QBC · Neurology
Decision Date
Oct 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5897
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cala Trio device is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor.
Device Story
Cala Trio is a wrist-worn neuromodulation system for essential tremor symptom relief. System components: rechargeable stimulator, wrist-worn electrode band, and charging base station. Device applies Transcutaneous Afferent Patterned Stimulation (TAPS) to median and radial nerves. Input: tremor motion measured by onboard triaxial accelerometer. Processing: embedded firmware controls stimulation delivery parameters, calibration, and amplitude adjustments based on user input via three housing buttons. Output: TAPS therapy delivered through integrated electrodes; device display provides instructions and status. Used in home setting by patient. Healthcare providers use device output to monitor therapy; device does not provide diagnostic data. Benefits: temporary reduction in hand tremor severity.
Clinical Evidence
Clinical evidence includes a prospective, multi-center, single-arm, 3-month home-use study (n=263 enrolled, 205 completed). Results showed 62% (TETRAS) and 68% (BF-ADL) of moderate/severe patients improved to mild/slight. Clinician-reported improvement (CGI-I) was 68%; patient-reported improvement (PGI-I) was 60%. Additional studies (Isaacson et al., 2020; Yu et al., 2020) support persistent tremor relief post-stimulation (median 60 minutes) and efficacy of TAPS therapy.
Technological Characteristics
Wrist-worn stimulator with triaxial accelerometer; TAPS delivered via three embedded electrodes in a biocompatible nylon/plastic band. Energy source: 3.8V Li-ion battery. Connectivity: Base station with LTE for cloud data transfer (monitoring only). Software: Embedded firmware for calibration and stimulation control. Standards: IEC 60601-1 (safety/EMC), ISO 10993-1:2018 (biocompatibility), IEC 62133 (battery). Output: Biphasic, rectangular waveform, 150Hz, 650µs pulse duration.
Indications for Use
Indicated for adults with essential tremor to aid in temporary relief of hand tremors in the treated hand via stimulation.
Regulatory Classification
Identification
An external upper limb tremor stimulator is a prescription device which is placed externally on the upper limb and designed to aid in tremor symptom relief of the upper limb.
Special Controls
In combination with the general controls of the FD&C Act, the external upper limb tremor stimulator is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must assess the following:
(i) Characterization of the electrical stimulation, including the following, must be performed: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(ii) Impedance testing, current distribution across the electrode surface area, adhesive integrity, and shelf life testing of the electrodes and gels must be conducted.
(iii) Simulated use testing of sensor performance and the associated algorithms that determine the stimulation output must be conducted.
(2) Patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Physician and patient labeling must include:
(i) Summaries of electrical stimulation parameters;
(ii) Instructions on how to correctly use and maintain the device;
(iii) Instructions and explanations of all user-interface components;
(iv) Instructions on how to clean the device;
(v) A shelf life for the electrodes and gel; and
(vi) Reuse information.
{0}------------------------------------------------
October 5, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Cala Health, Inc. Danielle Boyd Director, Regulatory Affairs 875 Mahler Road, Suite 168 Burlingame, California 94010
Re: K203288
Trade/Device Name: Cala Trio Regulation Number: 21 CFR 882.5897 Regulation Name: External upper limb tremor stimulator Regulatory Class: Class II Product Code: QBC Dated: November 6, 2020 Received: November 9, 2020
Dear Danielle Boyd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182706
Device Name Cala Trio
Indications for Use (Describe)
The Cala Trio device is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cala Health. The logo is blue and features the word "CALA" in a stylized font with a curved line above it. Below the word "CALA" is the word "HEALTH" in a smaller, sans-serif font. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.
### 510(k) Summary 510(k): Device Modification [K203288]
- I. Submitter Manufacturer: Cala Health, Inc. 875 Mahler Rd, Suite 168 Burlingame, CA 94010 Phone: (415) 890-3961 Fax: None Primary Contact: Danielle McDonnell Boyd Director, Regulatory Affairs, Cala Health danielle(@calahealth.com (415) 819-2935 Date Prepared: October 4, 2021 II. Subiect Device Cala Trio™ Trade Name: Classification Name: External upper limb tremor stimulator Device Classification: Class II Regulation: 21 CFR 882.5897 Product Code: QBC III. Predicate Device Predicate Device: Cala ONE DEN170028, K182706 Prior Submissions: Indications for Use: Cala ONE is indicated to aid in the transient relief of hand tremors in the treated hand following stimulation in
adults with essential tremor.
{4}------------------------------------------------
### IV. Description of Device
Cala Trio is a small, lightweight, wrist-worn stimulator device designed to aid in essential tremor symptom relief by applying Transcutaneous Afferent Patterned Stimulation (TAPS) to the median and radial nerves of a patient's wrist. The Cala Trio system includes three main components: (1) a rechargeable stimulator, (2) a wrist-worn electrode band, and (3) a base station that charges the device.
The stimulator component contains the electronics for delivering TAPS to the patient's wrist. There are three buttons on the stimulator that are used for calibration and stimulation amplitude adjustments, among other functions. Text prompts, stimulation delivery parameters, timer, and other messages are provided on the stimulator's full color display to provide instructions and stimulation delivery information.
To deliver therapy, the stimulator is attached to the wrist band, which includes integrated electrodes placed at appropriate intervals around the inner diameter of the band to properly target the median and radial nerves. To accommodate a broad distribution of wrist sizes, the band is available in three sizes (small, medium, and large). Each band size is also available in right or left-handed versions to target the appropriate nerves of the prescribed hand.
#### V. Indications for Use
Cala Trio is indicated to aid in the temporary relief of hand tremors in the treated hand following stimulation in adults with essential tremor.
{5}------------------------------------------------
### VI. Technological Characteristics
The following table provides a summary comparison for technical characteristics between Cala ONE (predicate device) and Cala Trio (subject device).
| | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Discussion of<br>Differences |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Delivery of transcutaneous<br>afferent patterned<br>stimulation (TAPS) for<br>hand tremors in essential<br>tremor patients | Delivery of transcutaneous<br>afferent patterned<br>stimulation (TAPS) for<br>hand tremors in essential<br>tremor patients | Same |
| Indications<br>for Use | Cala ONE is indicated to<br>aid in the transient relief of<br>hand tremors in adults with<br>essential tremor. | Cala Trio is indicated to<br>aid in the temporary relief<br>of hand tremors in adults<br>with essential tremor. | Substantially Equivalent<br>The indications for use<br>is changed from<br>"transient" to<br>"temporary” but does<br>not impact safety or<br>effectiveness. Cala Trio<br>is indicated for the same<br>condition and patient<br>population.<br>Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the indications for use<br>change does not impact<br>the safety or<br>effectiveness of the<br>device. |
| | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Discussion of<br>Differences |
| Design | Stimulator with on- board<br>motion sensors<br>permanently mounted on<br>wristband | Stimulator with on- board<br>motion sensors that is<br>detachable from wristband | Substantially Equivalent<br>The difference in<br>component attachment<br>does not impact the<br>safety or effectiveness of<br>the device.<br><br>The TAPS therapy<br>parameters as<br>controlled by Cala Trio<br>are unchanged from<br>Cala ONE. |
| | Wrist-worn band with three<br>4x4 cm² electrodes<br>manually attached by the<br>user | Wrist-worn band with three<br>4.84 cm² electrodes<br>embedded in the band | Substantially Equivalent<br>The difference in<br>electrode configuration<br>does not impact the<br>safety or effectiveness of<br>the device.<br><br>Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the surface area change<br>does not impact the<br>safety or effectiveness of<br>the device.<br><br>The Cala Trio electrode<br>surface area is<br>equivalent and the<br>electrode material are<br>unchanged from Cala |
| Cala ONE<br>DEN170028<br>K182706 | Cala Trio<br>K203288 | Discussion of<br>Differences | |
| (Predicate Device) | (Subject Device) | | |
| AC-powered charger that<br>connects to the stimulator<br>via USB port | AC-powered base station<br>for recharging the<br>stimulator<br>Base Station includes an<br>LTE antenna that provides<br>connection and data<br>transfer to cloud system<br>which supports device<br>operations such as device<br>assistance and maintenance | Substantially Equivalent<br>The difference in the<br>charging component<br>does not impact the<br>safety or effectiveness of<br>the device.<br>The Cala Trio energy<br>source (battery) and<br>stimulator circuitry is<br>unchanged from Cala<br>ONE.<br>The cloud provides<br>support and maintenance<br>functions only.<br>Transferred data is used<br>for monitoring<br>purposes only and is not<br>used for diagnosis or<br>other treatment-related<br>recommendations or<br>treatment adjustments. | |
| Embedded firmware<br>control of device<br>calibration, stimulation<br>delivery, and device<br>function | Embedded firmware<br>control of device<br>calibration, stimulation<br>delivery, and device<br>function | Same | |
| | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Discussion of<br>Differences |
| | Device sensors (gyroscope<br>and triaxial accelerometer)<br>measure tremor motion | Device sensors (triaxial<br>accelerometer) measure<br>tremor motion | Substantially Equivalent<br>The difference in the<br>motion sensors does not<br>impact safety or<br>effectiveness.<br><br>Non-clinical<br>performance testing<br>demonstrated that the<br>measurements obtained<br>from the Cala Trio<br>sensor are equivalent to<br>those obtained from the<br>Cala ONE sensors. |
| Features | • Set-up calibration<br>• Stimulation intensity<br>control<br>• Therapy available on<br>demand | • Set-up calibration<br>• Stimulation intensity<br>control<br>• Therapy available on<br>demand | Same |
| User<br>Interface | Stimulator display<br>guides device setup,<br>operation, and usage<br>information.<br>Four buttons on the device<br>housing allow<br>user control of the<br>stimulation amplitude and<br>device calibration | Stimulator display<br>guides device setup,<br>operation, and usage<br>information.<br>Three buttons on the<br>device housing allow<br>user control of the<br>stimulation amplitude and<br>device calibration | Substantially Equivalent<br>The difference in the<br>user interface does not<br>impact safety or<br>effectiveness.<br>The Cala Trio buttons<br>serve the same combined<br>functions as those<br>controlled separately by<br>the buttons in Cala<br>ONE. The device setup,<br>operation, and usage<br>information are<br>unchanged. |
| Prescription<br>or OTC | Prescription | Prescription | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Materials | Biocompatible wrist band | Biocompatible wrist band | Same |
| | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Discussion of<br>Differences |
| Energy Source | Permanent 3.7V Lithium-<br>Ion rechargeable<br>battery | Permanent 3.8V Lithium-<br>Ion rechargeable<br>battery | Substantially Equivalent<br>The difference in battery<br>voltage does not impact<br>safety or effectiveness.<br>The Cala Trio and Cala<br>ONE device batteries<br>both comply with IEC<br>62133 standard. |
| Frequency<br>of Use | The device is used as<br>needed by the patient | The device is used as<br>needed by the patient | Same |
| Band Use Life | Single Use (DEN170028)<br>30 Days (K182706) | 90 Days | Substantially Equivalent<br>The difference in band<br>use-life does not impact<br>safety or effectiveness.<br>Non-clinical<br>performance testing<br>demonstrated that the<br>Cala Trio electrode<br>maintains electrode<br>integrity and delivers<br>therapy as intended for<br>the duration of the 90-<br>day use-life. |
| Principle of<br>Operations | Biphasic Waveform,<br>rectangular | Biphasic Waveform,<br>rectangular | Same |
| Output Modes | 1 channel with 2<br>alternating outputs | 2 alternating channels | Substantially Equivalent<br>The difference in the<br>number of output<br>channels does not<br>impact safety or<br>effectiveness.<br>The functionality of the<br>output channel and the<br>therapy parameters are<br>unchanged. |
| | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Discussion of<br>Differences |
| Output | Current Regulated | Current Regulated | Same |
| Housing<br>Materials | Plastic<br>Velcro Straps (Nylon) | Plastic<br>Velcro Straps (Nylon) | Same |
| Weight | 56g | Stimulator: 28g<br>Base Station: 209g<br>Band: 13.4g - 15.9g | Substantially Equivalent<br>The difference in device<br>weight does not impact<br>safety or effectiveness.<br><br>The total weight of the<br>patient-worn<br>components of Cala<br>Trio and Cala ONE<br>(band and stimulator)<br>have equivalent weights. |
| Dimensions<br>(mm) | 52 x 79 x 16 | Stimulator: 42 x 55 x 13<br>Base Station: 104 x 36 x 91<br>Band Length: 252.5 - 327.5<br>Band Width: 20.8 -54.1 | Substantially Equivalent<br>The difference in device<br>dimensions does not<br>impact safety or<br>effectiveness.<br><br>The patient-worn<br>components) of Cala<br>Trio and Cala ONE<br>(band and stimulator)<br>have equivalent<br>dimensions |
| Output<br>Specification<br>Category | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Comparison Result |
| Waveform | Biphasic | Biphasic | Same |
| Shape | Rectangular | Rectangular | Same |
| Maximum<br>Output Voltage<br>(volts) | 7.5 @ 500Ω | 7.5 @ 500Ω | Same |
| | 120 @ 10kΩ | 80 @ 10kΩ | Substantially<br>Equivalent<br>The therapy pattern<br>parameters are<br>unchanged.<br>Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the reduction in<br>maximum output<br>voltage does not<br>impact safety or<br>effectiveness of the<br>device |
| Maximum Output<br>Current (mA) | 15 @ 500Ω | 8 @ 500Ω | Substantially<br>Equivalent<br>The therapy pattern<br>parameters are<br>unchanged. |
| | 12 @ 10kΩ | 8 @ 10kΩ | Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the reduction in<br>maximum output<br>current does not<br>impact safety or<br>effectiveness of the<br>device |
| Output<br>Specification<br>Category | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Comparison Result |
| Duration of<br>primary<br>(depolarizing)<br>phase<br>(µsec) | 300 | 300 | Same |
| Pulse Duration<br>(µsec) | 650 | 650 | Same |
| Frequency (Hz) | 150 | 150 | Same |
| Symmetrical<br>phases? | Yes | Yes | Same |
| Phase Duration<br>(µS) | 300<br>each phase | 300<br>each phase | Same |
| Net Charge<br>(µC) | 0 @500Ω | 0 @500Ω | Same |
| Maximum<br>Phase Charge<br>(µC) | 4.5 @ 500Ω | 2.4 @ 500Ω | Substantially<br>Equivalent<br><br>The therapy pattern<br>parameters are<br>unchanged.<br><br>Clinical evidence<br>summarized in<br>Section VII<br>demonstrates that<br>the reduction in<br>maximum phase<br>charge does not<br>impact safety or<br>effectiveness of the<br>device |
| Output<br>Specification<br>Category | Cala ONE<br>DEN170028<br>K182706<br>(Predicate Device) | Cala Trio<br>K203288<br>(Subject Device) | Comparison Result |
| Maximum<br>Current Density<br>(mA/cm², r.m.s.) | 2.41 @ 500Ω | 1.17 @ 500Ω | Substantially<br>Equivalent<br><br>The therapy pattern<br>parameters are<br>unchanged.<br><br>Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the reduction in<br>maximum current<br>density does not<br>impact safety or<br>effectiveness of the<br>device |
| Maximum<br>Average Current<br>(mA) (average<br>absolute value) | 1.35 @ 500Ω | 0.72 @ 500Ω | Substantially<br>Equivalent<br><br>The therapy pattern<br>parameters are<br>unchanged.<br><br>Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the reduction in<br>maximum average<br>current does not<br>impact safety or<br>effectiveness of the<br>device |
| Output<br>Specification<br>Category | Cala ONE<br>DEN170028<br>K182706 | Cala Trio<br>K203288 | Comparison Result |
| | (Predicate Device) | (Subject Device) | |
| Maximum<br>Average Power<br>Density<br>(mW/cm²) | 2.30 @ 500Ω | 1.12 @ 500Ω | Substantially<br>Equivalent |
| | | | The therapy pattern<br>parameters are<br>unchanged. |
| | | | Clinical evidence<br>summarized in Section<br>VII demonstrates that<br>the reduction in<br>maximum average<br>power density does not<br>impact safety or<br>effectiveness of the<br>device |
## Technical Comparison of the Subject Device and Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
### VII. Performance and Non-Clinical Testing
The following tables provide a summary comparison for performance, and nonclinical testing of Cala ONE (predicate device) and Cala Trio (subject device).
# Performance Comparison of the Modified Device and Predicate Device
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
## Nonclinical Testing Comparison of the Modified Device and Predicate Device
| Output<br>Specification<br>Category | Cala ONE | Cala Trio | Comparison Result |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-------------------|
| | DEN170028<br>K182706<br>(Predicate Device) | K203288<br>(Subject Device) | |
| Biocompatibility | Demonstrated<br>Biocompatibility<br>in accordance with<br>ISO 10993-1:2009 | Demonstrated<br>Biocompatibility in<br>accordance with<br>ISO 10993-1:2018 | Same |
| EMC and<br>Electrical<br>Safety | Meets all requirements<br>of IEC60601-1 and<br>relevant collateral and<br>particular standards | Meets all requirements<br>of IEC60601-1 and<br>relevant collateral and<br>particular standards | Same |
| Stimulation<br>Waveform<br>Testing | Stimulation waveform<br>conforms to<br>requirements of<br>internal standard test<br>method and acceptance<br>criteria | Stimulation waveform<br>conforms to<br>requirements of internal<br>standard test method and<br>acceptance criteria | Same |
{15}------------------------------------------------
| Output<br>Specification<br>Category | Cala ONE<br>DEN170028<br>K182706 | Cala Trio<br>K203288 | Comparison Result |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| | (Predicate Device) | (Subject Device) | |
| Shelf Life<b…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.