The GelPOINT V-Path Vaginal Access System (VAS) is intended to be inserted in the vagina to allow for entry of minimally invasive instruments while maintaining insufflation for vaginal gynecological procedures.
Device Story
The GelPOINT V-Path Vaginal Access System provides a transvaginal port for gynecological surgery. It consists of an intravaginal Alexis retractor, two access channels, and a GelSeal cap. The device is inserted into the vagina to create a pathway for instruments (5mm to 12mm) and an endoscope. The GelSeal cap enables CO2 insufflation to expand the surgical space and allows for smoke evacuation. Used in clinical settings by physicians, the device facilitates visualization of gynecological organs and the extraction of specimens. It benefits patients by enabling minimally invasive access, reducing the need for more invasive surgical approaches.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including functional testing (maintenance of access, instrument facilitation, device removal/retention, and insufflation maintenance) and biocompatibility testing (ISO 10993-5, 10993-10, 10993-11, and USP <151>).
Indicated for patients undergoing vaginal gynecological procedures requiring minimally invasive instrument access and insufflation.
Regulatory Classification
Identification
A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.
Special Controls
*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".
September 16, 2022
Applied Medical Resources Corporation David Yu Regulatory Affairs Specialist 22872 Avenida Empresa Rancho Santa Margarita, CA 92688
Re: K220969
Trade/Device Name: GelPOINT® V-Path Vaginal Access System Regulation Number: 21 CFR§ 884.1630 Regulation Name: Colposcope Regulatory Class: II Product Code: MOK Dated: August 17, 2022 Received: August 18, 2022
Dear David Yu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220969
Device Name GelPOINT® V-Path Vaginal Access System
Indications for Use (Describe)
The GelPOINT V-Path Vaginal Access System (VAS) is intended to be inserted in the vagina to allow for entry of minimally invasive instruments while maintaining insufflation for vaginal gynecological procedures.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# GelPoint V-Path Vaginal Access System
# K220969
| 510(K) Submitter: | Applied Medical Resources Corp.<br>22872 Avenida Empresa<br>Rancho Santa Margarita, CA, 92688<br>(949) 713-8000 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | David Yu<br>Regulatory Affairs Specialist<br>Applied Medical Resources Corp.<br>David. Yu@appliedmedical.com<br>Tel: (949) 713-8023<br>Fax: (949) 713-8205 |
| Date of Preparation: | September 15, 2022 |
| Trade Name: | GelPOINT® V-Path Vaginal Access System |
| Common Name: | Vaginal endoscopic surgery access port |
| Classification: | Regulation: 884.1630 Colposcope<br>Device Class: Class II<br>Product Code: MOK, Vaginoscope and accessories |
| Predicate Device: | Trade Name: KARL STORZ INFANT, BALLOON<br>VAGINOSCOPE<br>510(k) #: K950639<br>Product Code: MOK |
| | Through FDA database search following predicate device has<br>been identified: KARL STORZ INFANT BALLOON<br>VAGINOSCOPE manufactured by KARL STORZ<br>ENDOSCOPY-AMERICA, INC (K950639). No design-related<br>recalls for these legally marketed predicate devices have<br>been identified. |
| Indications for use: | The GelPOINT V-Path Vaginal Access System is intended to be<br>inserted in the vagina to allow for entry of minimally invasive<br>instruments while maintaining insufflation for vaginal<br>gynecological procedures. |
{4}------------------------------------------------
# Device Description
The GelPOINT V-Path Vaginal Access System consists of an Intravaginal Alexis retractor and two access channels that are placed vaginally to create a pathway to gynecological organs. A GelSeal cap attaches to the retractor at the opening of the vagina. The cap allows for insufflation and smoke evacuation. Sleeves inserted through the cap allow for passage of 5mm to 12mm instrumentation. Visualization is achieved via introduction of an endoscope through a sleeve. The device is provided sterile.
| Manufacturer | Applied Medical Resources Corp | KARL STORZ<br>ENDOSCOPY-AMERICA,<br>INC |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | GelPOINT V-Path Vaginal<br>Access System<br>K220969 | KARL STORZ INFANT<br>BALLOON<br>VAGINOSCOPE<br>K950639 |
| General Device<br>Characteristics | | |
| Device Design | The Vaginal Access System<br>consists of similarly designed<br>intravaginal Alexis retractor and a<br>vaginal access channel<br>that acts as the anchor for the<br>intravaginal Alexis retractor.<br>Intended to be inserted<br>transvaginally, to allow passage of<br>minimally invasive instruments for<br>the purpose of visualizing<br>gynecological organs, and to allow<br>for diagnostic and operative<br>gynecological procedures including<br>the extraction of specimens | The device consists of a<br>vaginoscopes, balloon<br>vaginoscopes, vaginoscope<br>sheaths, fiber optic light<br>carrier and adjustable<br>magnifiers.<br>Designed to be used for<br>endoscopic diagnostic and<br>surgical procedures involving<br>the infant vagina and cervix. |
| Anatomical Sites | Vagina | Vagina |
| Placement | Transvaginal | Transvaginal |
# Comparison of Technological Characteristics with the Predicate Device
{5}------------------------------------------------
| Visualization<br>Capability | Visualization is achieved via<br>introduction of an endoscope<br>through an access port. | N/A |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Expansion of<br>Surgical Space | Utilizes CO2 insufflation via an<br>insufflator (not included with<br>device) connected to insufflation<br>port on the device. | N/A |
| Device<br>Removals | The intravaginal Alexis retractor<br>does not include a tether because<br>the retractor is only placed at the<br>vaginal introitus and can be<br>removed after unrolling the outer<br>ring. The access channel is also<br>placed at the introitus and is<br>removed manually by the user. | N/A |
| Materials | Polyether polyurethane,<br>thermoplastic elastomer | N/A |
| Sterilization | Ethylene Oxide, Sterility<br>Assurance Level of 10-6<br>Single use device; device is<br>discarded after use | N/A |
The subject and predicate device have the same intended use, namely:
- To be inserted vaginally. ●
- To allow passage of minimally invasive instruments for the purpose of . visualizing gynecological organs.
- To allow for diagnostic and operative gynecological procedures including the . extraction of specimens.
The differences in device design, presence of imaging capability, and material do not constitute a new intended use. Additionally, there are no different questions of safety and effectiveness that arise from the differences in technology. The technological differences between GelPOINT V-Path Vaginal Access System and the predicate device can be evaluated through performance testing and do not alter the intended use of the GelPOINT V-Path Vaginal Access System.
### Discussion of Performance Data
The subject device and predicate device share the same ability to achieve vaginal access to the vaginal canal. Therefore, bench testing for the subject device was conducted to demonstrate the clinically relevant functionality to provide access to the vaginal canal.
{6}------------------------------------------------
There are no existing performance standards to evaluate the functions of the GelPOINT V-Path Vaginal Access System. A protocol was developed by Applied Medical to evaluate the ability of the subject device to provide access to the vaginal canal and remain anchored following deployment. Maintenance of insufflation and device removal were also assessed.
## Function Testing
Functional bench testing was performed to demonstrate the adequate functionality of the Vaginal Access System, as follows:
- Maintenance of transvaginal access ●
- Facilitation of instrument access to surgical site ●
- Device removal
- Device retention ●
- Maintenance of insufflation
The studies confirm that the GelPOINT V-Path Vaginal Access System has the ability to establish and maintain a path of entry for minimally invasive instruments and can remain anchored in the patient while withstanding conditions under normal clinical use. The studies also confirm that the integrity and sealing capability of the device system are adequate to maintain insufflation as a means to provide visualization of the surgical space.
### Biocompatibility
Biocompatibility studies were performed in accordance with the 2020 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and "testing" as follows:
- Cytotoxicity: ISO 10993-5:2009 ●
- Sensitization: ISO 10993-10:2010 .
- Irritation: ISO 10993-10:2010 ●
- Acute Systemic Toxicity: ISO 10993-11:2017
- Pvrogenicity Potential: USP <151>. ISO 10993-11:2017 .
The biocompatibility test results demonstrate the GelPOINT V-Path Vaginal Access System is biocompatible. The results show the subject device has met the criteria to be classified as a non-cytotoxic, non-sensitizer, non-irritant device that shows no sign of being systemically toxic or pyrogenic. The information provided for biocompatibility testing indicate the subject device is acceptable to demonstrate substantial equivalence from a biocompatibility standpoint.
{7}------------------------------------------------
# Shelf-Life
The GelPOINT V-Path Vaginal Access System has a shelf life of 3 years when packaged in Tyvek and nylon film pouches, in accordance with the results of accelerated aged stability studies. Results from testing demonstrated that the devices could maintain their specifications over the stated shelf-life duration.
## Clinical Performance Data
Clinical data are not needed to support substantial equivalence for the subject device.
# Conclusion
The GelPOINT V-Path Vaginal Access System has the same intended use as the predicate device. The GelPOINT V-Path Vaginal Access System has different technological characteristics from the predicate device, but these differences do not raise different questions of safety and effectiveness. Performance testing, as described above, demonstrates that the GelPOINT V-Path Vaginal Access System is as safe and effective as the predicate. Therefore, the GelPOINT V-Path Vaginal Access System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.