GelPOINT Transvaginal Access Platform

K143308 · Applied Medical Resources · HEW · Oct 13, 2015 · Obstetrics/Gynecology

Device Facts

Record IDK143308
Device NameGelPOINT Transvaginal Access Platform
ApplicantApplied Medical Resources
Product CodeHEW · Obstetrics/Gynecology
Decision DateOct 13, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1640
Device ClassClass 2

Indications for Use

The GelPOINT Transvaginal Access Platform is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments while maintaining pneumoperitoneum during laparoscopic hysterectomy. The instrument is indicated for use in laparoscopic assisted vaginal hysterectomies. The instrument is also a conduit for the extraction of specimens.

Device Story

Sterile, single-use transvaginal access platform; facilitates laparoscopic assisted vaginal hysterectomy. Device comprises cannula with suture ties, flexible gel cap with two stopcocks for insufflation/smoke evacuation, and four trocars for instrument insertion. Used in OR by surgeons; provides conduit for instruments and specimen extraction while maintaining pneumoperitoneum. Introducer facilitates insertion; sutures secure device to patient. Benefits include maintained pneumoperitoneum and access for minimally invasive instrumentation.

Clinical Evidence

Bench testing only. No clinical data. Mechanical testing confirmed sealing against insufflation pressure and vaginal fixation stability. Biocompatibility testing (cytotoxicity, sensitization, irritation) leveraged from reference device (K133393) per ISO 10993-1 and FDA G95-1.

Technological Characteristics

Materials: polymers, silicone, stainless steel. Components: cannula, gel cap, trocars. Sterilization: Gamma irradiation (SAL 10^-6). Connectivity: None. Form factor: Three diameter/length configurations. Sensing/Actuation: Mechanical sealing and insufflation management via stopcocks.

Indications for Use

Indicated for patients undergoing laparoscopic assisted vaginal hysterectomies requiring transvaginal access for minimally invasive instruments and specimen extraction.

Regulatory Classification

Identification

A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles, stacked one behind the other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 13, 2015 Applied Medical Resources Frans VandenBroek Principal Regulatory Affairs Specialist 22872 Avenida Empresa Rancho Santa Margarita, CA 92688 Re: K143308 > Trade/Device Name: GelPOINT Transvaginal Access Platform, Models: C2A00, C2A01, C2A02 Regulation Number: 21 CFR 884.1640 Regulation Name: Culdoscope and accessories Regulatory Class: II Product Code: HEW Dated: September 10, 2015 Received: September 29, 2015 Dear Frans VandenBroek, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {2}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K143308 Device Name GelPOINT Transvaginal Access Platform, Models: C2A00, C2A01, C2A02. Indications for Use (Describe) The GelPOINT Transvaginal Access Platform is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments while maintaining pneumoperitoneum during laparoscopic hysterectory. The instrument is indicated for use in laparoscopic assisted vaginal hysterectomies. The instrument is also a conduit for the extraction of specimens. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Applied Medical. The logo consists of the words "Applied Medical" in black font, with "Applied" being larger than "Medical". To the right of the text is a teal-colored triangle with curved lines inside, resembling a stylized mountain or wave. # 510(k) SUMMARY #### I. SUBMITTER Applied Medical Resources Corporation 22872 Avenida Empresa Rancho Santa Margarita, CA, 92688 Phone: (949) 713-8000 FAX: (949) 713-8205 Contact Person: Frans VandenBroek Date Prepared: Oct 12, 2015 #### II. DEVICE Name: GelPOINT Transvaginal Access Platform, Models: C2A00, C2A01, C2A02 Common name: Transvaginal Access Device Classification Name: Culdoscope and Accessories, (21CFR 884.1640) Regulatory Class: II Product code: HEW. #### PREDICATE DEVICE III. McCartney Access Tube, Gynetech Pty. LTD (K051594) Reference Device: Applied Medical GelPOINT Path Transanal Surgery Device (K133393) #### IV. DEVICE DESCRIPTION The GelPOINT Transvaginal Access Platform is a sterile, single use instrument that allows access to the vagina to perform laparoscopic assisted hysterectomy. The device consists of a cannula with suture ties, and a cap made of a flexible gel-like material. The gel cap has two stopcocks that can be used for insufflation or smoke evacuation. Four trocars are included and may be placed through the gel cap. The trocars allow insertion of standard laparoscopic instruments into the vagina. Construction materials include various polymers, silicone and stainless steel. Packaging consists of a PTEG tray that is placed in a Tyvek/Mylar pouch and shipped in a carton. Sterilization is via Gamma irradiation; sterility assurance level is 10°. {5}------------------------------------------------ #### V. INDICATIONS FOR USE The GelPOINT Transvaginal Access Platform is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments while maintaining pneumoperitoneum during laparoscopic hysterectomy. The instrument is indicated for use in laparoscopic assisted vaginal hysterectomies. The instrument is also a conduit for the extraction of specimens. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE GelPOINT and predicate device have different technological characteristics as follows: - GelPOINT is delivered in three diameters and lengths; predicate is delivered in two ● diameters. - GelPOINT is shorter. - GelPOINT has (up to) four access ports. The predicate has three. - GelPOINT uses an introducer to facilitate insertion. ● - GelPOINT uses sutures for fixation to the patient. The predicate connects to the surgical drapes using towel clips. #### VII. PERFORMANCE DATA The following performance data support determination of substantial equivalence. # Biocompatibility testing The subject device is an External Communicating Device that contacts Tissue/Bone/Dentin for less than 24 hours. The portion of the device that contacts tissue is the cannula, dilator and the cap ring. Biocompatibility testing on the reference device (K133393), which is identical in materials and composition to the subject device, were leveraged to support the biocompatibility of the subject device. All patient-contacting device components have passed biocompatibility testing required by ISO 10993-1 and FDA General Program Memorandum G95-1. These tests include: - Cytotoxicity ● - Sensitization ● - Irritation ● # Mechanical testing There are no published performance standards for devices of this type. Therefore, Applied Medical created a dedicated test method designed to confirm substantial equivalence to the Gynetech McCartney Access Tube. These tests focused on: - Sealing against insufflation pressure, with and without instruments in place ● - o Fixation in the vagina #### VIII. CONCLUSIONS The Applied Medical GelPOINT Transvaginal Access Platform is substantially equivalent to the predicate Gynetech McCartney Access Tube.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...