GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device
K243152 · Applied Medical Resources Corp. · HEW · Oct 2, 2025 · Obstetrics/Gynecology
Device Facts
Record ID
K243152
Device Name
GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device
Applicant
Applied Medical Resources Corp.
Product Code
HEW · Obstetrics/Gynecology
Decision Date
Oct 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1640
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K243152 · Oct 2, 2025
GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device
Applied Medical Resources Corp.
Published clinical literature (case series and comparative studies)
The literature provides clinical evidence demonstrating the feasibility and safety of the device for retroperitoneal sentinel lymph node dissection, including complication rates and successful access rates in a real-world clinical setting.
vNOTES; sentinel lymph node dissection; endometrial cancer; clinical case series
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
vNOTES retroperitoneal sentinel lymph node dissection for endometrial cancer staging: First multicenter, prospective case series; Multicenter, prospective case series; Study Period: 2024
Patients with endometrial cancer; Sample Size: 64; Number of Sites: Multicenter
Transvaginal natural orifice transluminal endoscopic surgery (VNOTES) retroperitoneal sentinel lymph node BIOPSY compared with conventional laparoscopy in patients with endometrial cancer; Comparative study; Study Period: 2024
Patients with endometrial cancer; Sample Size: 57 (Group 1 N=19, Group 2 N=38)
Conventional laparoscopy
Feasibility of retroperitoneal sentinel lymph node biopsy
Indications for Use
The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments and maintain insufflation during laparoscopic-assisted and vaginal gynecological procedures. The device may be used for sentinel pelvic lymph node dissection occurring in the retroperitoneal space. The device is also a conduit for the extraction of specimens
Device Story
Device consists of intraperitoneal and retroperitoneal Alexis retractors and a GelSeal cap; placed transvaginally to create access to peritoneum and retroperitoneum. Cap allows insufflation and smoke evacuation; sleeves accommodate 5mm-12mm laparoscopic instruments. Endoscope introduced through sleeve for visualization. Used in clinical settings by gynecological surgeons. Provides access to sentinel pelvic lymph nodes without abdominal incisions; facilitates specimen extraction. Benefits include minimally invasive approach for gynecological procedures.
Clinical Evidence
Evidence includes one prospective case series (N=64) and two retrospective datasets (N=19, N=58). Prospective study showed 96.9% success rate in accessing sentinel pelvic lymph nodes without abdominal incision. Complications included 3.1% conversion rate to traditional approach and 4.7% cystotomy rate. Complication rates for bleeding, infection, and vessel/nerve injury were comparable to traditional approaches.
Technological Characteristics
Materials: polycarbonate, polyester, polyethylene, polyisoprene, silicone, stainless steel, nylon, polyurethane, thermoplastic elastomer. Principle: mechanical retractor with GelSeal cap for insufflation. Dimensions: accommodates 5mm-12mm instruments. Sterilization: Ethylene Oxide (SAL 10^-6). Single-use.
Indications for Use
Indicated for patients undergoing laparoscopic-assisted and vaginal gynecological procedures, including sentinel pelvic lymph node dissection in the retroperitoneal space, requiring transvaginal access for minimally invasive instruments, insufflation maintenance, and specimen extraction.
Regulatory Classification
Identification
A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 6, 2025
Applied Medical Resources Corp.
Apeksha Shanbhag
Senior Manager Regulatory Affairs
22872 Avenida Empresa
Rancho Santa Margarita, California 92688
Re: K243152
Trade/Device Name: GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device
Regulation Number: 21 CFR 884.1640
Regulation Name: Culdoscope and accessories
Regulatory Class: Class II
Product Code: HEW
Dear Apeksha Shanbhag:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 2, 2025.
Specifically, FDA is updating this substantial equivalence (SE) Letter to correct the official correspondent as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jason Roberts, OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices, Jason.Roberts@fda.hhs.gov.
Sincerely,
Jason Roberts -S
Jason Roberts
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 1, 2025
Applied Medical Resources Corp.
David Yu
Senior Specialist Regulatory Affairs
22872 Avenida Empresa
Rancho Santa Margarita, California 92688
Re: K243152
Trade/Device Name: GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device
Regulation Number: 21 CFR 884.1640
Regulation Name: Culdoscope and accessories
Regulatory Class: II
Product Code: HEW
Dated: January 9, 2025
Received: September 5, 2025
Dear David Yu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, The Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K243152 - David Yu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K243152 - David Yu
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243152 | |
| Device Name GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device | |
| Indications for Use (Describe) The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments and maintain insufflation during laparoscopic-assisted and vaginal gynecological procedures. The device may be used for sentinel pelvic lymph node dissection occurring in the retroperitoneal space. The device is also a conduit for the extraction of specimens | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
510(k) Summary
# 510(k) Summary
## GelPoint V-Path Transvaginal Access Platform with Retroperitoneal Access Device
### K243152
**510(K) Submitter:**
Applied Medical Resources Corp.
22872 Avenida Empresa
Rancho Santa Margarita, CA, 92688
(949) 713-8000
**Contact Person:**
Apeksha Shanbhag
Regulatory Affairs Senior Manager
Applied Medical Resources Corp.
apeksha.shanbhag@appliedmedical.com
Tel: (949) 713-8341
Fax: (949) 713-8205
**Date of Preparation:** October 1, 2025
**Trade Name:** GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device
**Common Name:** Transvaginal endoscopic surgery access port
**Classification:**
Regulation: 21 CFR 884.1640, Culdoscope & Accessories
Device Class: Class II
Product Code: HEW
**Predicate Device:**
Trade Name: Transvaginal Access Platform
510(k) #: K191294
Product Code: HEW
No design-related recalls for the predicate device have been identified.
1
{6}
Device Description: The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device consists of an intraperitoneal Alexis retractor and retroperitoneal Alexis retractor that are placed transvaginally to create a pathway to the peritoneum and retroperitoneum. A GelSeal cap attaches to the retractors at the opening of the vagina. The cap allows for insufflation and smoke evacuation. Sleeves inserted through the cap allow for passage of 5mm to 12mm laparoscopic instrumentation. Visualization is achieved via introduction of an endoscope through a sleeve. The device is provided sterile.
Indications for use: The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments and maintain insufflation during laparoscopic assisted and vaginal gynecological procedures. The device may be used for sentinel pelvic lymph node dissection occurring in the retroperitoneal space. The device is also a conduit for the extraction of specimens.
Comparison of Technological Characteristics with the Predicate Device
| General Device Characteristics | Predicate Device: Transvaginal Access Platform K191294 | Subject Device: GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device |
| --- | --- | --- |
| Device Design | The platform facilitates access for gynecological procedures. The device consists of a retractor that is placed transvaginally. A GelSeal cap attaches to the retractor at the opening of the vagina. The cap allows for insufflation and smoke evacuation. Sleeves inserted through the cap allow for passage of laparoscopic instrumentation. Visualization is achieved via introduction of an endoscope through a sleeve. Gel technology allows for triangulation of instrumentation. Specimens can be retrieved by removing the GelSeal cap. | Same |
510(k) Summary
{7}
| Anatomical Sites | Vagina, pelvic, and peritoneal cavity | Vagina, pelvic, peritoneal cavity, and retroperitoneal cavity |
| --- | --- | --- |
| Placement | Transvaginally | Same |
| Visualization Capability | Visualization is achieved via introduction of an endoscope through a sleeve. | Same |
| Expansion of Surgical Space | Utilizes CO2 insufflation via an insufflator (not included with device) connected to insufflation port on the device | Same |
| Device Removals | The flexible Alexis retractor features a tether attached to the inner ring of the device. The tether and tether tag are positioned outside of the patient during device placement and usage. To remove the retractor, the outer ring is first unrolled until the retractor becomes loose. The tether is then pulled to remove the inner ring from the pelvic cavity and the entire retractor from the vaginal canal. | Same |
| Materials | Polycarbonate, polyester, polyethylene, polyisoprene, silicone, stainless steel, nylon, polyurethane, thermoplastic elastomer | Same |
| Sterilization | Ethylene Oxide, Sterility Assurance Level of 10-6
Single use device; device is discarded after use | Same |
There are no different questions of safety and effectiveness that arise from the differences in technology. The technological differences between GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device and the predicate device can be evaluated through performance testing and do not alter the intended use of the GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device.
The subject and predicate device have the same intended use, namely:
- To be inserted transvaginally,
- To allow passage of minimally invasive instruments for the purpose of visualizing gynecological organs, and
- To allow for diagnostic and operative gynecological procedures including the extraction of specimens.
## Discussion of Performance Data
510(k) Summary
{8}
There are currently no performance standards related to transvaginal endoscopic ports; therefore, Applied Medical designed a protocol to evaluate and compare the predicate and subject device's ability to achieve key performance specifications, namely:
- Maintenance of transvaginal access
- Facilitation of instrument access to surgical site
- Maintenance of insufflation
- Retention
- Device removal
The testing confirms that the GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device has the ability to establish and maintain a path of entry for minimally invasive instruments and can remain anchored in the patient while withstanding conditions under normal clinical use. The testing also confirms that the integrity and sealing capability of the device system are adequate to maintain insufflation as a means to provide visualization of the surgical space. Lastly, device removal was assessed to ensure the retractor component of the subject device can safely be removed from the surgical space while remaining intact. Based on the results of the performance testing conducted, the subject device, GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is considered substantially equivalent to the predicate, Transvaginal Access Platform, in terms of performance and does not raise any additional concerns of safety and effectiveness.
## Biocompatibility
Biocompatibility studies were performed in accordance with the 2020 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and “testing” as follows:
- Cytotoxicity: ISO 10993-5:2009
- Sensitization: ISO 10993-10:2010
- Irritation: ISO 10993-10:2010
- Acute Systemic Toxicity: ISO 10993-11:2017
- Pyrogenicity Potential: USP <151>, ISO 10993-11:2017
The biocompatibility test results demonstrate the GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is biocompatible. The results show the subject device has met the criteria to be classified as a non-cytotoxic, non-sensitizer, non-irritant device that shows no sign of being systemically toxic or pyrogenic. The information provided for biocompatibility testing indicate the subject device is acceptable to demonstrate substantial equivalence from a biocompatibility standpoint.
## Shelf-Life
The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device has a shelf life of 3 years when packaged in Tyvek and nylon film pouches, in accordance
510(k) Summary
{9}
with the results of accelerated aged stability studies. Results from testing demonstrated that the devices could maintain their specifications over the stated shelf-life duration.
## Clinical Performance Data
Baekelandt, J., Jespers, A., Huber, D., Badiglian-Filho, L., Stuart, A., Chuang, L., ... & Burnett, A. (2024). vNOTES retroperitoneal sentinel lymph node dissection for endometrial cancer staging: First multicenter, prospective case series. *Acta Obstetricia et Gynecologica Scandinavica*. (N=64)
Comba, C., Karakas, S., Erdogan, S. V., Demir, O., Şimşek, E., Karasabanoglu, F., ... & Ozdemir, I. A. (2024). Transvaginal natural orifice transluminal endoscopic surgery (VNOTES) retroperitoneal sentinel lymph node BIOPSY compared with conventional laparoscopy in patients with endometrial cancer. *Surgical Oncology*, 55, 102099. (Group 1 N=19, Group 2 N=38)
Burnett, A. F., Huber, D. E., Chuang, L., & Baekelandt, J. F. (2023). 10061 vNOTES Sentinel Lymph Node Dissection for Gynecologic Malignancies. *Journal of Minimally Invasive Gynecology*, 30(11), S100-S101. (N=58)
The retrospective data sets demonstrate feasibility of the use of the subject device for the expanded indications for use. The prospective dataset of 64 patients demonstrated no device-related failures. Procedure-related complications associated with retroperitoneal access for sentinel lymph node dissection included two conversions from retroperitoneal access to a traditional approach (3.1%, 2/64) and three cystotomies creating access for device placement (4.7%, 3/64). Complication rates for bleeding, infection, and injury to vessels or nerves were similar to those observed with traditional alternative approaches. The benefit of this device is providing access to the sentinel pelvic lymph node without requiring an abdominal incision, observed in 62/64 patients (96.9%).
## Conclusion
Based on the intended use, technological characteristics, and performance testing results, the subject GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is considered substantially equivalent to the predicate device.
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.