The KYPHON™ VuE™ Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a cementoplasty (i.e. kyphoplasty or vertebroplasty) procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
Device Story
KYPHON™ VuE™ Bone Cement is a PMMA-based acrylic resin for vertebroplasty, kyphoplasty, and sacroplasty. It consists of two sterile components: powder (methylmethacrylate-styrene copolymer, barium sulfate, benzoyl peroxide) and liquid (methylmethacrylate monomer, hydroquinone, N,N-dimethyl-p-toluidine). Components are manually mixed by a clinician immediately before use. The device incorporates both powder and granular forms of barium sulfate to enhance radiopacity and visualization under fluoroscopy during delivery. The cement is injected into the vertebral body or sacrum to stabilize pathological fractures. By providing structural support, the device aims to reduce pain and restore vertebral integrity in patients with osteoporotic or neoplastic bone lesions.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (ISO 10993 series), sterilization validation, and mechanical/physical characterization (dough time, setting time, compression, bending, fatigue, fracture toughness, and extravasation rates). All results met established specifications and were comparable to predicate devices.
Indicated for patients with pathological vertebral body fractures due to osteoporosis, cancer (multiple myeloma, metastatic lesions from breast/lung, lymphoma), or benign lesions (hemangioma, giant cell tumor). Also indicated for fixation of pathological sacral vertebral body or ala fractures via sacroplasty/vertebroplasty. Includes symptomatic microfractures without height loss.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
April 18, 2022
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. Next to the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Tecres S.p.A. % Christine L. Brauer, PhD Regulatory Affairs Consultant Brauer Device Consultants, LLC 7 Trail House Court Rockville, Maryland 20850
Re: K220131
Trade/Device Name: KYPHONTM VuE™ Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: January 14, 2022 Received: January 18, 2022
Dear Christine Brauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For:
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220131
Device Name KYPHONTM VuE™ Bone Cement
#### Indications for Use (Describe)
The KYPHON™ VuE™ Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesing a cementoplasty (i.e. kyphoplasty or vertebroplasty) procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: underline;"></span> <span style="text-decoration: overline;"></span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
---
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
#### GENERAL INFORMATION 1
### 1.1 Submitter and Owner of the 510(k)
Tecres S.p.A. Via A. Doria 6 37066 Sommacampagna Verona – Italy
### 1.2 Official Correspondent
Christine L. Brauer, PhD Regulatory Affairs Consultant 7 Trail House Court Rockville, Maryland 20850 Telephone: (301) 545-1990 E-mail: chris.brauer@comcast.net
#### Devices Subject of this 510(k) 1.3
KYPHON™ VuE™ Bone Cement
#### 1.4 Date of Preparation
March 16, 2022
### 1.5 510(k) Number
K220131
### 2 NAME OF THE DEVICE AND CLASSIFICATION INFORMATION
#### 2.1 Trade/Proprietary Name
KYPHON™ VuE™ Bone Cement
#### 2.2 Common/Usual Name
Acrylic resin for vertebroplasty/kyphoplasty
#### 2.3 Classification Information
| Classification Name: | Polymethylmethacrylate (PMMA) bone cement |
|----------------------------|-------------------------------------------|
| Classification Regulation: | 21 CFR 888.3027 |
| Class: | II |
| Product Code(s): | NDN – Cement, Bone, Vertebroplasty |
{4}------------------------------------------------
| Panel: | Orthopaedic |
|--------|-------------|
|--------|-------------|
### 3 PREDICATE DEVICE
The predicate devices are identified as follows:
- KYPHON® Xpede™ Bone Cement which was cleared originally through 510(k) . application K102397 and subsequently cleared in K151227, K163032, K171938 and K191148.
- Spineplex™ Bone Cement by Stryker which was originally cleared through 510(k) ● notification K032945.
#### DEVICE DESCRIPTION ব
KYPHON™ VuE™ Bone Cement is a polymethylmethacrylate (PMMA) acrylic resin intended for use for vertebroplasty or kyphoplasty procedures, and sacroplasty procedures.
K YPHON™ VuE™ Bone Cement is a modification of the KYPHON® Xpede™ Bone Cement. Like the predicate, KYPHON™ VuE™ Bone Cement contains approximately 30% barium sulfate. However, KYPHON™ VuE™ Bone Cement contains barium sulfate in two forms (powder and granules) whereas the predicate has only the powder form of barium sulfate. The larger radio-opaque granules of barium sulfate increase visualization of the acrylic resin during delivery using fluoroscopy.
KYPHON™ VuE™ Bone Cement is provided sterile in two components: 20 grams of powder and 9 grams of liquid. The powder contains methylmethacrylate-styrene copolymer, barium sulfate and benzoyl peroxide. The The liquid contains methylmethacrylate (monomer), hydroquinone and N,N-dimethyl-p-toluidine. The components are manually mixed immediately prior to use.
#### ഗ INDICATIONS FOR USE
Below is the indication for use.
The KYPHON™ VuE™ Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a cementoplasty (i.e. kyphoplasty or vertebroplasty) procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
{5}------------------------------------------------
### 6 COMPARISON OF THE INTENDED USE WITH THE PREDICATE DEVICE
The KYPHON™ VuE™ Bone Cement and the predicate devices, the KYPHON® Xpede™ Bone Cement and the Spineplex™ Bone Cement, are intended to be used in the same surgical procedure (kyphoplasty, vertebroplasty or sacroplasty procedure) in the same target patient population and have the same primary function of treatment of pathological fractures of the vertebral body. Thus, the KYPHON™ VuE™ Bone Cement has the same intended use as the predicate devices.
#### 7 COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
VuE™ -KYPHON™ Bone Cement shares many of the same technological characteristics compared to the predicate KYPHON® Xpede" Bone Cement (see Table 1) and to the predicate Spineplex™ Bone Cement (see Table 2).
| Characteristics | KYPHON™ VuE™<br>Bone Cement<br>(Subject Device) | KYPHON® Xpede™ Bone<br>Cement<br>(Predicate Device,<br>K102397, K151227,<br>K163032, K171938 and<br>K191148) | Comparison |
|---------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------|
| Main Components | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate<br>(MMA) | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate<br>(MMA) | Same |
| Main Components | Barium Sulfate<br>(powder and granules) | Barium Sulfate<br>(powder only) | Different |
| Other Significant<br>Components | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Same |
| Mixing/Application | Manual | Manual | Same |
| Powder Sterilization<br>Method | Gamma-ray irradiation | Gamma-ray irradiation | Same |
| Sterility Assurance Level<br>(SAL) - Powder | 10-6 | 10-6 | Same |
| Liquid Sterilization<br>Method | Filtration | Filtration | Same |
| SAL - Liquid | 10-3 | 10-3 | Same |
Table 1: Summary of Technological Characteristics between the KYPHON VuE™ Bone Cement and the Predicate Device, the KYPHON® Xpede™ Bone Cement
{6}------------------------------------------------
| Characteristics | KYPHON™ VuE™<br>Bone Cement<br>(Subject Device) | Spineplex Bone Cement<br>(Predicate Device,<br>K032945) | Comparison |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------|
| Main Components | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate<br>(MMA)<br>Barium Sulfate<br>(powder and granules) | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate<br>(MMA)<br>Barium Sulfate<br>(powder only) | Same<br>Different |
| Other Significant<br>Components | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Same |
| Mixing/Application | Manual | Manual | Same |
| Powder Sterilization<br>Method | Gamma-ray irradiation | Irradiation | Same |
| Sterility Assurance<br>Level (SAL) – Powder | 10-6 | 10-6 | Same |
| Liquid Sterilization<br>Method | Filtration | Filtration | Same |
| SAL - Liquid | 10-3 | 10-3 | Same |
Table 2: Summary of Technological Characteristics between the KYPHON™ VuE™ Bone Cement and the Predicate Device, the Spineplex™ Bone Cement
## 8 PERFORMANCE DATA
This 510(k) notification provided performance data to establish the substantial equivalence of the KYPHON™ VuE™ Bone Cement to the predicate device. Performance testing was conducted in accordance with the "FDA Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement: Guidance for Industry and FDA " dated July 17, 2002. The following is a summary of the performance data.
Sterilization and Shelf Life: The device is sterilized using standard methods and the sterilization cycles have been validated following international standards. The shelf life of 2 years has been established in stability studies in that evaluated both device functionality and sterility. Bacterial endotoxin testing was performed using the LAL test (gel clot) method, following the current edition of the European Pharmacopoeia (EP) Standard 2.6.14. Bacterial Endotoxins and met the limits (<20 EU/device). The analytical procedures described in the official pharmacopoeial texts, European Pharmacopoeia (Ph. Eur.): 2.6.14. Bacterial Endotoxins, Japanese Pharmacopoeia (JP): 4.01 Bacterial Endotoxins Test, and United States Pharmacopeia (USP) General Chapter Bacterial Endotoxins Test, can be used as interchangeable in the ICH regions (Guidance for Industry O4B Evaluation and Recommendation of Pharmacopoeial Texts for Use in the ICH Regions Annex 14 Bacterial Endotoxins Test General Chapter, October 2013).
{7}------------------------------------------------
Biocompatibility: Biocompatibility evaluation has been performed to show the device materials are safe, biocompatible, and suitable for their intended use. FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" was taken into account to evaluate the biocompatibility of the device materials and the following biocompatibility studies were completed.
| Biocompatibility Test | Test Method | Results |
|----------------------------------|--------------------------------|---------------------------------------|
| Cytotoxicity | ISO 10993-5 | Not cytotoxic |
| Sensitization | ISO 10993-10 | Not sensitizing |
| Intracutaneous reactivity | ISO 10993-10 | No signs of erythema, eschar or edema |
| Systemic toxicity (acute) | ISO 10993-11 | Non-toxic |
| Material-mediated pyrogenicity | USP 42 <151><br>ISO 10993-11 | Absence of pyrogens |
| Genotoxicity (Ames test) | ISO 10993-3 | Non-mutagenic |
| Genotoxicity (Mouse<br>Lymphoma) | ISO 10993-3<br>ISO/TR 10993-33 | Non-mutagenic |
| Implantation 13W | ISO 10993-6 | No local and systemic effects- |
Performance Testing: Performance testing was performed to characterize the bone cement in accordance with special controls guidance document. This testing included the following:
- Mixing and application characteristics (e.g., dough time, setting time)
- Chemical composition (e.g., residuals, molecular weight and polymer structure, glass ● transition temperature)
- Thermal properties (e.g., polymerization temperature) ●
- Mechanical properties (e.g., modulus and flexural properties, static compression and . bending, fatigue testing, fracture toughness and viscoelasticity)
- . Sacroplasty testing (extravasation rate)
Results show comparable performances to the predicate devices and are in compliance with ASTM F451-16, ISO 5833:2002, ISO 527:1/2, ASTM F2118-14, ASTM D2990-17, ASTM D732-17 and ASTM E399-20.
The performance data demonstrate that the new devices are substantially equivalent to the predicate device and meet the requirements of the Special Controls Guidance document.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.