Medtronic HV-R™ Bone Cement, Kyphon™ Xpede™ Bone Cement, CD Horizon™ Fenestrated Screw Set
K191148 · Medtronic Sofamor Danek USA, Inc. · PML · Sep 12, 2019 · Orthopedic
Device Facts
Record ID
K191148
Device Name
Medtronic HV-R™ Bone Cement, Kyphon™ Xpede™ Bone Cement, CD Horizon™ Fenestrated Screw Set
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
PML · Orthopedic
Decision Date
Sep 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used without cement, CD Horizon™ Fenestrated Screws (with or without Sextant™ or Longitude™ instrumentation) are intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis), pseudarthrosis, and/or failed previous fusion. When used in conjunction with Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Xpede™ Bone Cement, CD Horizon™ Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CD Horizon™ Fenestrated Screws augmented with Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System comprises cannulated fenestrated screws and PMMA bone cements (Medtronic HV-R™ and Kyphon™ Xpede™). Screws feature fenestrations for cement injection into pedicles; used for spinal fixation or augmentation. Cements are highly viscous PMMA formulations. Used in clinical/surgical settings by surgeons. Screws provide mechanical stabilization; cement augments fixation in compromised bone or restores spinal integrity in tumor patients. Output is a stabilized spinal construct. Benefits include structural support in patients where fusion is not feasible or bone quality is compromised.
Clinical Evidence
No clinical data. Bench testing only; reliance on previously cleared predicate data.
Technological Characteristics
Implants: Medical grade titanium, titanium alloy, or cobalt-chromium-molybdenum alloy. Cements: PMMA with ~30% barium sulfate; powder/liquid components (methylmethacrylate-styrene copolymer, benzoyl peroxide, hydroquinone, N,N-dimethyl-p-toluidine). Delivery: Highly viscous state. Sterilization: Sterile.
Indications for Use
Indicated for posterior, non-cervical spinal fixation in patients with DDD, spondylolisthesis, tumor, trauma, spinal stenosis, scoliosis, kyphosis, pseudarthrosis, or failed fusion. Also indicated for spinal column integrity restoration in patients with advanced stage thoracic, lumbar, or sacral tumors with limited life expectancy. Bone cements indicated for pathological vertebral fractures due to osteoporosis, cancer, or benign lesions (hemangioma, giant cell tumor) via cementoplasty, and sacral fracture fixation.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a clear, sans-serif font. The word "FDA" is in a blue box.
September 12, 2019
Medtronic Sofamor Danek, USA Inc. Ms. Shweta Sharma Principal Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K191148
Trade/Device Name: Medtronic HV-R™ Bone Cement, Kyphon™ Xpede™ Bone Cement, CD Horizon™ Fenestrated Screw Set. Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: PML, NKB, NDN Dated: April 29, 2019 Received: April 30, 2019
Dear Ms. Sharma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K191148
Device Name CD Horizon™ Fenestrated Screw Set
### Indications for Use (Describe)
When used without cement, CD Horizon™ Fenestrated Screws (with or without Sextant™ or Longitude™ instrumentation) are intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis), pseudarthrosis, and/or failed previous fusion.
When used in conjunction with Medironic HV-R™ Fenestrated Screw Cement or Kyphon™ Xpede™ Bone Coment, CD Horizon™ Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion™ Fenestrated Screws augmented with Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|--|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191148
Device Name Kyphon™ Xpede™ Bone Cement
### Indications for Use (Describe)
Kyphon™ Xpede™ Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions a cementoplasty (i.e. kyphoplasty or vertebroplasty) procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arrsing from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
When used in conjunction with CD Horizon™ Fenestrated Screws, Kyphon™ Xpede™ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CD Horizon™ Fenestrated Screws augmented with Kyphon™ Xpede™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span> <img alt="Checked" height="20" src="checked.png" width="20"/> </span> |
|----------------------------------------------|------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K191148
Device Name Medtronic HV-RTM Fenestrated Screw Cement
### Indications for Use (Describe)
When used in conjunction with CD Horizon™ Fenestrated Screws, Medtronic HV-R™ Fenestrated Screw Cement is intended to restore the integrity of the spinal column, even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. Medronic HV-R™ Fenestrated Screw Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) SUMMARY
## August 21, 2019
| I. | Company: | Medtronic Sofamor Danek, USA Inc.<br>1800 Pyramid Place<br>Memphis, TN 38132<br>Telephone Number: (901) 396-3133 | |
|----|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Contact: | Shweta Sharma<br>Principal Regulatory Affairs Specialist<br>Telephone number: (901) 396-3133<br>Email: shweta.s.sharma@medtronic.com | |
| | Proprietary Trade Name: | Kyphon™ Xpede™ Bone Cement<br>Medtronic HV-RTM Fenestrated Screw Cement<br>CD Horizon™ Fenestrated Screw Set | |
| | Common Name: | Bone Screw, Pedicle Screw, Cement | |
| | Classification Name/<br>Regulation Numbers/<br>Classification/<br>Classification Product<br>Code/<br>Subsequent Product<br>Codes (if any) | PML, NDN<br>Kyphon™ Xpede™ Bone Cement and Medtronic HV-RTM<br>Fenestrated Screw Cement<br>Polymethylmethacrylate (PMMA) bone cement<br>888.3027<br>Class II<br>NKB<br>CD Horizon™ Fenestrated Screw Set<br>Thoracolumbosacral pedicle screw system<br>888.3070<br>Class II | |
#### II. Predicate Devices:
Primary Predicate: Medtronic HV-R™ Bone Cement (K152604, S.E. 01/06/2016)
{6}------------------------------------------------
Additional Predicates: CD Horizon™ Spinal System (K170679, S.E. 05/11/2017)
CD Horizon™ Fenestrated Screw Set (K171938, S.E. 10/23/2017) CD Horizon™ Fenestrated Screw Set (K170347, S.E. 04/04/2017) Kyphon™ HV-R™ Bone Cement (K180700, S.E. 05/18/2018) Kyphon™ Xpede™ Bone Cement (K163032, S.E. 02/27/2017)
These predicate devices have not been subject to a design-related recall.
#### III. Device Description:
## CD Horizon™ Fenestrated Screw Set
The CD Horizon™ Fenestrated Screw Set consists of a variety of cannulated screws. These screws contain a series of fenestrations which allows polymethylmethacrylate (PMMA) bone cement (Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Xpede™ Bone Cement) to be injected into the treated site. This cement is used to augment screw fixation into the pedicle in patients whose life expectancy is of insufficient duration to permit achievement of fusion. These implants may also serve as traditional pedicle screws when used without bone cement in patients with non-compromised bone quality. CD Horizon™ Fenestrated Screws are specifically designed to connect to appropriate rods and associated connecting components contained within the CD Horizon™ Spinal System. Refer to the CD Horizon™ Spinal System package insert for information regarding those implants. Care should be taken so the correct components are used in the spinal construct. CD Horizon™ Fenestrated Screw Set implant components are fabricated from medical grade titanium and/or medical grade titanium alloy and/or medical grade cobalt-chromiummolybdenum allov. Never use stainless steel and titanium implant components in the same construct.
To achieve best results, do not use CD Horizon™ Fenestrated Screw implants with components from any system other than the CD Horizon™ Spinal System. As with all orthopedic and neurosurgical implants, CD Horizon™ Fenestrated Screw implants should never be reused under any circumstances.
{7}------------------------------------------------
## Kyphon™ Xpede™ Bone Cement
Kyphon™ Xpede™ Bone Cement is a polymethylmethacrylate (PMMA) that contains approximately 30% barium sulfate. It is designed for delivery in a highly viscous state. Kyphon™ Xpede™ Bone Cement is provided sterile in two components: 20 grams of powder and nine grams of liquid. The powder contains methylmethacrylate-styrene copolymer, barium sulfate and benzoyl peroxide. The liquid contains methylmethacrylate (monomer), hydroquinone and N, N dimethyl-p-toluidine.
## Medtronic HV-RTM Fenestrated Screw Cement
Medtronic HV-R™ Fenestrated Screw Cement is a polymethylmethacrylate (PMMA) that contains approximately 30% barium sulfate. It is designed for delivery in a highly viscous state. Medtronic HV-RTM Fenestrated Screw Cement is provided sterile in two components: 20 grams of powder and nine grams of liquid. The powder contains methylmethacrylate-styrene co-polymer, barium sulfate, and benzovl peroxide. The liguid contains methylmethacrylate (monomer), hydroquinone and N. N dimethyl-p-toluidine.
#### Indications for Use: IV.
# CD Horizon™ Fenestrated Screw Set
When used without cement, CD Horizon™ Fenestrated Screws (with or without Sextant™ or Longitute™ instrumentation) are intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, tumor and/or trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), pseudarthrosis, and/or failed previous fusion. When used in conjunction with Medtronic HV-R™ Fenestrated Screw Cement or Kyphon™ Xpede™ Bone Cement, CD Horizon™ Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, lumbar, and sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CD Horizon™ Fenestrated Screws augmented with Medtronic HV-RTM Fenestrated
{8}------------------------------------------------
Screw Cement or Kyphon™ Xpede™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
## Kyphon™ Xpede™ Bone Cement
Kyphon™ Xpede™ Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using a cementoplasty (i.e. kyphoplasty or vertebroplasty) procedure. It is also indicated for the fixation of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and giant cell tumor. Pathological fracture may include a symptomatic microfracture (as documented by appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
When used in conjunction with CD Horizon™ Fenestrated Screws, Kyphon™ XpedeTM Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic, and lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CD Horizon™ Fenestrated Screws augmented with Kyphon™ Xpede™ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
## Medtronic HV-RTM Fenestrated Screw Cement
When used in conjunction with CD Horizon™ Fenestrated Screws, Medtronic HV-R™ Fenestrated Screw Cement is intended to restore the integrity of the spinal column, even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic, and lumbar, or sacral spine in whom life expectancy is of insufficient duration to permit achievement of fusion. Medtronic HV-R™ Fenestrated Screw Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
{9}------------------------------------------------
#### V. Comparison of the Technological Characteristics with the Predicate Device:
The primary predicate for the subject devices is CD Horizon™ Fenestrated Screw Set (K171938, S.E. 10/23/2017). The subject CD Horizon™ fenestrated screws have the same or similar indications, intended use, same or similar overall design, packaging, sterilization and materials as the following FDA primary predicate cleared in K171938 (S.E. 10/23/2017) and K170679 (S.E. 05/11/2017). The predicate and subject new fenestrated screws have the identical function and scientific fundamental technology.
There are no technology changes to the existing CD Horizon™ Fenestrated Screw Set devices or the associated existing bone cements. However, the CD Horizon™ Fenestrated Screw Set system, Medtronic HV-RTM Fenestrated Screw Cement and Kyphon™ XpedeTM Bone Cement labeling has been updated to reflect the same indications for use as the predicates Kyphon™ Xpede™ Bone Cement (K163032, S.E. 02/27/2017), Kyphon™ HV-RTM Bone Cement (K180700, S.E. 05/18/2018), Medtronic HV-R™ Bone Cement (K152604, S.E. 01/06/2016), CD Horizon™ Fenestrated Screw Set (K171938, S.E. 10/23/2017) and CD Horizon™ Fenestrated Screw Set (K170347, S.E. 04/04/2017).
#### Performance Data: VI.
In accordance with the Guidance for Industry and FDA Staff - Spinal System 510(k)'s, Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices. Medtronic believes that testing is not warranted for the subject devices as they do not present a new worst case when compared to the predicate devices. All existing predicate data previously provided in the predicate 510(k)s are still applicable.
#### VII. Conclusions
The subject CD Horizon™ Fenestrated Screw Set devices, Kyphon™ Xpede™ Bone Cement and Medtronic HV-R™ Fenestrated Screw Cement have shown through supporting information provided in this premarket notification to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.