ELAZ Nitrile Powder Free Examination Glove

K213604 · Protect Gloves Co., Ltd. · LZA · Apr 3, 2022 · General Hospital

Device Facts

Record IDK213604
Device NameELAZ Nitrile Powder Free Examination Glove
ApplicantProtect Gloves Co., Ltd.
Product CodeLZA · General Hospital
Decision DateApr 3, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Intended Use

ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Disposable, non-sterile, powder-free nitrile examination glove; made of acrylonitrile-butadiene rubber. Worn on examiner's hands during medical procedures to serve as a barrier against contamination. Device is a physical barrier; no electronic or algorithmic components. Used in clinical settings by healthcare professionals. Benefits include protection against cross-contamination between patient and examiner.

Clinical Evidence

Bench testing only. Testing included physical dimensions (ASTM D6319-19), freedom from holes (ASTM D5151-19), residual powder (ASTM D6124-17), and physical properties (ASTM D412/D573). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity). Results met all acceptance criteria.

Technological Characteristics

Material: Acrylonitrile-butadiene rubber (nitrile). Form factor: Disposable, powder-free, non-sterile glove. Sizes: S, M, L, XL. Compliance: ASTM D6319-19, ASTM D5151-19, ASTM D6124-17. Biocompatibility: ISO 10993-5, ISO 10993-10, ISO 10993-11.

Indications for Use

Indicated for use by medical examiners to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. April 3, 2022 Protect Gloves Co., Ltd. % Paweena U-Thainual CEO MDR Solutions Co., Ltd. 1435 Kanchanapisek Rd., Bang Khae Nuea Bangkok, Bang Khae 10160 Thailand Re: K213604 Trade/Device Name: ELAZ Nitrile Powder Free Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: March 21, 2022 Received: March 24, 2022 Dear Paweena U-Thainual: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213604 ### Device Name ELAZ Nitrile Powder Free Examination Glove Indications for Use (Describe) ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K213604 # 510(k) Summary | 1. | General Information | | |----|----------------------|----------------------------------------| | | Applicant/Submitter: | Protect Gloves Company Limited | | | Address: | 60/138 Moo 3 Mabyangporn Pluakdang | | | | Rayong, 21140 THAILAND | | | | Tel: +66-2-384-3049 | | | | Email: nathaphon@me.com | | | Contact Person: | Paweena U-Thainual, PhD | | | Address: | MDR Solutions, Co., Ltd. | | | | 1435 Kanchanapisek Rd., Bang Khae Nuea | | | | Bang Khae, Bangkok 10160 THAILAND | | | | Email: paweena@mdrsolutions.co.th | | | Preparation Date: | November 8, 2021 | | | | | #### 2. Device Name and Code | Device Trade Name: | ELAZ Nitrile Powder Free Examination Glove | |----------------------|--------------------------------------------| | Common Name: | Nitrile Patient Examination Glove | | Classification Name: | Non-Powdered Patient Examination Glove | | Product Code: | LZA | | Regulation Number: | 21 CFR 882.6250 | | Classification: | I | | Review Panel: | General Hospital | #### Predicate Device 3. Provided below is the legally marketed predicate device. Table 1 Primary Predicate device | Applicant | Device Name | 510(k) Number | |--------------------------------------------------|------------------------------------------------------------------------------|---------------| | Tangshan Lanhai<br>Medical Supplies Co.,<br>Ltd. | Disposable Nitrile<br>Examination Gloves (Powder<br>free, Purple-Blue, Blue) | K210898 | {4}------------------------------------------------ #### 4. Device Description ELAZ Nitrile Powder Free Examination Glove is non-sterile, single use only, disposable, powder free examination glove. The glove is made of Acrylonitrile-butadiene rubber. The glove is designed to meets the specification of ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. #### Indications / Intended Use 5. ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. #### Technological Characteristics Comparison: 6. ELAZ Nitrile Powder Free Examination Glove is compared to the legally marketed predicate device. | | Proposed Subject Device | Predicate Device | Comparison | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Manufacturer | Protect Gloves Co., Ltd. | Tangshan Lanhai Medical Supplies Co., Ltd. | N/A | | Trade Name | ELAZ Nitrile Powder Free Examination Glove | Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) | N/A | | 510(k) Number | K213604 | K210898 | N/A | | Classification | Class I | Class I | YES | | Product Code | LZA | LZA | YES | | Regulation Number | 880.6250 | 880.6250 | YES | | Intended Use/ Indications For use | ELAZ Nitrile Powder Free Examination Glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner | The DISPOSABLE NITRILE EXAMINATION GLOVES (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | YES | | Powdered Free | Yes | Yes | Yes | Page 2 of 5 {5}------------------------------------------------ | Device Dimensions Comparison | | | | | | | | | | | | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|---------|-------------------------|-------------------------------------------------------|--------------|-------------------|-------------|--------|------|-----| | Size | S, M, L, XL | | | | | | XS, S, M, L, XL | | | YES* | | | Color | Blue | | | | | | Purple-Blue, Blue | | | YES | | | Sterility | Non-Sterile | | | | Non-Sterile | | | | | YES | | | | S | M | L | XL | XS | S | M | L | XL | YES | | | Length (mm) | ≥ 220 | ≥ 230 | ≥ 230 | ≥ 230 | ≥ 220 | ≥ 220 | ≥ 230 | ≥ 230 | ≥ 230 | | | | Width (mm) | S | M | L | XL | XS | S | M | L | XL | YES | | | | 80±10 | 95±10 | 110±10 | 120±10 | 70±10 | 80±10 | 95±10 | 110±10 | 120±10 | | | | Finger<br>Thickness<br>(mm) | ≥ 0.05 | | | | | ≥ 0.05 | | | | | YES | | Palm<br>Thickness<br>(mm) | ≥ 0.05 | | | | | ≥ 0.05 | | | | | YES | | Performance and Physical Comparison | | | | | | | | | | | | | | | Before Aging | | After Aging | | Before Aging | After Aging | | | | | | Tensile<br>Strength<br>(MPa) | | ≥ 14 | | ≥ 14 | | ≥ 14 | | ≥ 14 | | YES | | | Ultimate<br>Elongation<br>(%) | | Before Aging | | After Aging | | Before Aging | | After Aging | | YES | | | | | ≥ 500 | | ≥ 400 | | ≥ 500 | | ≥ 400 | | | | | Single Use | Yes | | | | Yes | | | | | YES | | | | Be free from holes when tested in | | | | Be free from holes when tested in | | | | | | | | Freedom<br>from hole | accordance with ASTM D5151<br>AQL 2.5 | | | | accordance with ASTM D5151<br>AQL 2.5 | | | | | YES | | | Powder<br>Content | Powder residue ≤ 2.0 mg | | | Powder residue ≤ 2.0 mg | | | | | YES | | | | Comply with<br>ASTM D6319 | Yes | | | Yes | | | | | YES | | | | Material<br>Used | Nitrile | | Nitrile | | | | | YES | | | | | Shelf life | | 3 years | | | N/A | | | | N/A | | | | Safety Comparison - Biocompatibility Test | | | | | | | | | | | | | Irritation | Under the conditions of the study,<br>not an irritant | | | | Under the conditions of the study, not<br>an irritant | | | | | YES | | | Sensitization | Under conditions of the study, not a<br>sensitizer | | | | Under conditions of the study, not a<br>sensitizer | | | | | YES | | | Cytotoxicity | Under conditions of the study, it<br>was considered as "non-cytotoxic"<br>at 25%, 12.5%, and 6.25% and<br>demonstrate cytotoxicity at the 50%<br>and 100% of the test item extract. | | | | N/A | | | | | N/A | | | Acute<br>Systemic<br>Toxicity | Under the conditions of the study,<br>there was no evidence of systemic<br>toxicity from the extract | | | | | | | | | YES | | {6}------------------------------------------------ * When compared to the same size, the dimensions, and physical design are identical. #### Performance Data 7. Non-clinical tests: Bench tests were conducted according to the FDA-recognized consensus standard, to verify that the proposed device met all design specifications. The proposed device complies with the following standards: - ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical । Application. - ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. - - -ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves. Biocompatibility tests were conducted according to the FDA-recognized consensus standard, to verify the safety of the device. - ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for In vitro cytotoxicity. - ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization. - -ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity. | Test<br>Methodology | Purpose | Acceptance Criteria | Results | |---------------------|-------------------------------------|-----------------------|-----------------------| | ASTM D6319-19 | Physical<br>Dimensions<br>Test (mm) | Length (mm) | Length (mm) | | | | S ≥ 220 | S ≥ 220 | | | | M ≥ 230 | M ≥ 230 | | | | L ≥ 230 | L ≥ 230 | | | | XL ≥ 230 | XL ≥ 230 | | | | Width (mm) | Width (mm) | | | | S: 80±10 | S: 87-89 | | | | M: 95±10 | M: 99-101 | | | | L: 110±10 | L: 110-113 | | | | XL: 120±10 | XL: 112-115 | | | | Finger Thickness (mm) | Finger Thickness (mm) | | | | ≥0.05 | S ≥ 0.16 | | | | | M ≥ 0.15 | | | | | L ≥ 0.15 | | | | | XL ≥ 0.13 | | | | Palm Thickness (mm) | Palm Thickness (mm) | | | | >0.05 | S > 0.11 | {7}------------------------------------------------ | | | | M ≥ 0.11<br>L ≥ 0.11<br>XL ≥ 0.10 | |------------------------|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ASTM D5151-19 | Testing for Freedom<br>from holes | AQL 2.5 | Gloves pass AQL 2.5 | | ASTM D6124-17 | Determine the<br>powder residue for<br>powder free gloves | < 2 mg/glove | Average 0.43 mg/glove | | ASTM D412<br>ASTM D573 | Testing for Physical<br>property<br>characteristics | Before Aging<br>Tensile strength ≥14 MPa<br>Ultimate Elongation<br>≥500%<br><br>After Aging<br>Tensile strength ≥14 MPa<br>Ultimate Elongation<br>≥400% | Before Aging<br>Tensile strength ≥21 MPa<br>Ultimate Elongation<br>≥502%<br><br>After Aging<br>Tensile strength ≥18 MPa<br>Ultimate Elongation<br>≥465% | | ISO 10993-5 | Tests for In vitro<br>cytotoxicity | Under the conditions of<br>the study non-cytotoxic. | Under conditions of the<br>study, it was considered<br>"non-cytotoxic" at 25%,<br>12.5%, and 6.25% and<br>demonstrate cytotoxicity<br>at the 50% and 100% of<br>the test item extract. | | ISO 10993-10 | Evaluate the<br>endpoint of<br>irritant for<br>biocompatibility | Under the conditions of<br>the study, not an irritant. | Under the conditions of<br>the study, not an irritant. | | ISO 10993-10 | Evaluate the<br>endpoint of<br>sensitization for<br>biocompatibility | Under the conditions of<br>the study, not a sensitizer. | Under the conditions of<br>the study, not a sensitizer. | | ISO 10993-11 | Tests for systemic<br>toxicity | Under the conditions of<br>the study, the device<br>extracts do not pose a<br>systemic toxicity<br>concern | Under the conditions of<br>the study, the device<br>extracts do not pose a<br>systemic toxicity<br>concern | #### Conclusions 8. The conclusions drawn from the non-clinical tests demonstrate that the subject device, ELAZ Nitrile Powder Free Examination Glove, is as safe, as effective, and performs as well as or better than the legally marketed device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...