The Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim (Blue) is a non - sterile disposable I no religio intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner
Device Story
Nitrile powder-free examination glove; intended for use as a protective barrier to prevent cross-contamination between patients and examiners in medical settings. Device is a disposable, non-sterile physical barrier worn on the hand. No electronic, software, or algorithmic components.
Clinical Evidence
Bench testing only.
Technological Characteristics
Nitrile rubber material; powder-free; disposable; non-sterile; examination glove form factor.
Indications for Use
Indicated for use as a non-sterile, disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K211624 — USA Gloves Nitrile Powder Free Exam Gloves · Usa Gloves · Dec 18, 2021
K973903 — POWDERFREE NITRILE EXAMINATION GLOVES (BLUE COLOUR) · Sri Johani Sdn. Bhd. · Dec 17, 1997
K080937 — POWDER FREE NITRILE EXAMINATION GLOVE, ORANGE · Gx Corporation Sdn Bhd · May 23, 2008
K984249 — SKINSENSE NITRILE POWDER-FREE SYNTHETIC MEDICAL EXAMINATION GLOVE · London International Group, Inc. · May 4, 1999
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2017
Hartalega Ngc Sdn. Bhd. Ms. Nurul Abdullah Ouality Assurance Senior Manager Dataran Sd Pju 9 Bandar Sri Damansara, 52200 MY
Re: K162646
> Trade/Device Name: Nitrile Powder Free Examination Glove With Low Dermatitis Potential Claim (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: January 13, 2017 Received: January 31, 2017
Dear Ms. Nurul Abdullah,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Lori A. Wiggins -S
for Tina Kiang, Ph.D. Acting Division Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K162646
Device Name
Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim (Blue)
Indications for Use (Describe)
The Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim (Blue) is a non - sterile disposable I no religio intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.