K212391 · Mci Medical Concept Innovation, Inc. · GWO · Oct 21, 2022 · Neurology
Device Facts
Record ID
K212391
Device Name
MCI-Neuro Fixation System
Applicant
Mci Medical Concept Innovation, Inc.
Product Code
GWO · Neurology
Decision Date
Oct 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MCI - Neuro Fixation System is indicated for use in selective trauma of the cranial skeleton, cranial surgery and reconstructive procedure.
Device Story
MCI-Neuro Fixation System consists of titanium plates, meshes, and screws used for cranial bone fixation; intended for use in selective trauma, cranial surgery, and reconstructive procedures. Devices are non-sterile, single-use, and require steam sterilization by qualified surgeons before implantation. System components are available in various shapes and sizes to accommodate specific fracture sites and anatomical needs. The device provides mechanical stabilization of cranial bone segments; clinical benefit derived from restoring structural integrity to the cranium following trauma or surgery. No software or electronic components involved.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by bench testing, including metallographic analysis, cyclic polarization corrosion testing, four-point bending tests (ASTM F382), and screw mechanical testing (torsion, pull-out, driving torque per ASTM F543).
Technological Characteristics
Materials: Commercially pure titanium (ASTM F67) for plates/meshes; titanium alloy Ti-6Al-4V (ASTM F136) for screws. Surface: Color-anodized. Form factor: Various shapes (straight, mesh, burr hole covers, etc.) and dimensions (0.3-0.6mm plate thickness; 1.5-1.7mm screw diameter). Sterilization: Steam sterilization required by user. Mechanical principle: Rigid fixation via bone plates and screws.
Indications for Use
Indicated for patients requiring fixation in selective trauma of the cranial skeleton, cranial surgery, and reconstructive procedures. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 21, 2022
MCI Medical Concept Innovation Inc. % Graziela Brum Regulatory Affairs Specialist Passarini Regulatory Affairs/ PR Servicos Regulatorios Administrativos Ltda ME Rua Alice Alem Saadi, 855/2402 Ribeirao Preto, Sao Paulo 14096-570 Brazil
Re: K212391
Trade/Device Name: MCI-Neuro Fixation System Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO, GXR, HBW Dated: September 20, 2022 Received: September 21, 2022
Dear Graziela Brum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212391
Device Name MCI - Neuro Fixation System
Indications for Use (Describe)
MCI - Neuro Fixation System is indicated for use in selective trauma of the cranial skeleton, cranial surgery and reconstructive procedure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| ADMINISTRATIVE INFORMATION | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor/Manufacturer Name | MCI Medical Concept Innovation Inc.<br>4592 North Hiatus Road<br>Sunrise, Florida, USA 33351<br>Telephone +01 (954) 306-2521 | |
| Contact Person and Preparer | Graziela Brum and Ana Carolina Carvalho<br>Regulatory Affairs Specialist<br>Passarini Regulatory Affairs<br>PR Serviços Regulatórios Administrativos Ltda<br>E-Mail: graziela@passarini.com.br<br>Telephone +55 (16) 3421-8488 | |
| Date Prepared | 20/September/2022 | |
| DEVICE NAME AND CLASSIFICATION | | |
| Trade/ Proprietary Name<br>Common Name | MCI - Neuro Fixation System<br>Plate, Cranioplasty, Preformed, Alterable; Cover, Bur<br>Hole; Fastener, Plate, Cranioplasty | |
| Classification Regulation | 21 CFR 882.5320 | |
| Product Code | GWO, GXR, HBW | |
| Classification Panel<br>Reviewing Branch | Neurology<br>Neurology | |
| PREDICATE DEVICE INFORMATION | | |
| Predicate Device | K141911 - OPTIMUS NEURO SYSTEM | |
| | SUBJECT DEVICE | PREDICATE DEVICE |
| Device Name | MCI - Neuro Fixation System | Neuro Plating System |
| 510(K) # | K212391 | K141911 |
| Manufacturer | MCI Medical Concept Innovation Inc. | OPTIMUS NEURO SYSTEM |
| Indications for<br>use | MCI - Neuro Fixation System is<br>indicated for use in selective trauma<br>of the cranial skeleton, cranial<br>surgery, and reconstructive<br>procedure. | Optimus Neuro System is intended for use<br>in selective trauma of the cranial skeleton,<br>cranial surgery, and reconstructive<br>procedure. |
| Class | 2 | 2 |
| Product Code | GWO, GRX, HBW | GWO, GRX, HBW |
| Material<br>(Chemical<br>composition) | •Plate- Pure Titanium ASTM F67<br>(Anodized)<br>•Screw- Titanium Alloy ASTM F136<br>(Anodized) | •Plate- Pure Titanium ASTM F67 (Anodized)<br>•Screw- Titanium Alloy ASTM F136<br>(Anodized) |
| Device types<br>and<br>Dimension | Straight, Square, Rectangular, Y-shape,<br>X-shape, Z -shape, clover, triangular,<br>mesh, burr hole with<br>various lengths and thickness (03 to<br>06.mm). The screws range in 1.5 mm diameter to 1.7 mm and in lengths from | Straight, angle, Y-shape, X-shape, burr hole,<br>square, matrix and mesh with various<br>lengths and thickness Plate has various<br>length and thickness (0.1 to 0.6mm). The<br>screws range in diameters of 0.8 to 1.8mm and lengths from 3.0 to 6.0mm. |
Reference Devices
K971297 - CENTRE-DRIVE DRILL-FREE SCREW K161821- Stryker Universal Mesh K141452 - Leforte Neuro System Bone Plate and Screw K121624 - Biomet Microfixation Neuro Plating System K182758 - MCI -CMF System K944565- KLS-MARTIN Micro Osteosynthesis System (1.5mm)
{4}------------------------------------------------
### INDICATIONS FOR USE
MCI - Neuro Fixation System is indicated for use in selective trauma of the cranial skeleton, cranial surgery, and reconstructive procedure.
### DEVICE DESCRIPTION
The MCI – Neuro Fixation System is composed of plates (including burr hole covers), screws, and meshes. The bone plates and meshes are made from commercially pure titanium (ASTM F67) and the bone screws are manufactured from titanium alloy - Ti-6Al-4V (ASTM F136) and are available in different sizes and shapes, according to the site of the implantation and the extension of the fracture. The surface of plates, meshes and screws are colored-anodized.
MCI - Neuro Fixation System devices are for single use. The devices are provided non-sterile and must be properly sterilized before use, according to the recommendations provided in the Instructions for Use.
The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in cranial surgery.
#### Comparison of Technological Characteristics with the Predicate Device
{5}------------------------------------------------
| | 3.0mm to 6.0 mm. | |
|---------------|-----------------------------------------------|-----------------------------------------------|
| Single Use | Yes | Yes |
| Sterilization | Nonsterile, steam<br>sterilization before use | Nonsterile, steam<br>sterilization before use |
The subject device is substantially equivalent in indications and design principles to the following predicate device: K141911 - OPTIMUS NEURO SYSTEM
Any difference in the technological characteristics do not raise new issues of safety or efficacy.
## Performance Data
| Test | Test Method Summary | Results |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Metallographic<br>Test | Applied Standard ASTM F67-13.<br>Perform metallography on implant<br>through an Olympus BX41M-LED<br>microscope, with Kroll attack -<br>immersion. | The material complied with the<br>specifications set forth in the applied<br>standard. |
| Metallographic<br>Test | Applied Standard ASTM F136 -13.<br>Perform metallography on implant<br>through an Olympus BX41M-LED<br>microscope, with Kroll attack -<br>immersion. | The material complied with the<br>specifications set forth in the applied<br>standard. |
| Cyclic<br>polarization<br>tests | Applied Standard ASTM F2129-19 a.<br>A scan rate of 1,0 mV /s was used<br>increasing from the corrosion<br>potential, after 1 h of open-circuit or<br>a potential variation less than 3 mV<br>in 60 seconds. The scan was reversed<br>to cathodic direction at 800 mVECS<br>or when the current density reaches<br>two orders of magnitude greater<br>than the passivation threshold. | The devices have high resistance to<br>initiation and to the propagation of<br>localized corrosion. |
| Cycle Bend<br>Testing of<br>Metallic Bone<br>Plates | Applied Standard ASTM F382-17 Annex<br>A.1.<br><br>Testing machine: Universal Testing<br>Machine.<br><br>The test was performed in ambient<br>conditions (ambient air at room<br>temperature.<br><br>OBJECTIVE: To determine the following<br>four-point bending properties (static) of<br>metallic bone plate: | The Structural Stiffness (Ele) of a bone<br>plate, is the bone plate normalized<br>effective bending stiffness. The test<br>reports showed a Structural Stiffness of<br>0.003 Nm² for the subject and<br>reference devices.<br><br>The Bending Stiffness (K) of a bone<br>plate, is the maximum slope of the<br>linear elastic portion of the load versus<br>load-point displacement curve. The<br>test reports show that the reference<br>device plate Bending Stiffness was 70.4 |
| | - Structural Stiffness<br>- Stiffness<br>- Bending Strength<br>- Proof Load | N/mm and to the subject device plate<br>75.7 N/mm. |
| | | The Bending Strength (R) of a bone<br>plate, is the bending moment<br>necessary to produce a 0.2 % offset<br>displacement in the bone plate when<br>tested as the standard prescription.<br>The test reports showed a Bending<br>Strength of 0.04 Nm for the subject<br>and reference devices. |
| | | The proof load (P) is the applied load at<br>the intersection point of line. BC with<br>the load versus load-point<br>displacement curve. The test reports<br>show a reference device plate Proof<br>Load of<br>16.08 N and to the subject device plate<br>of 15.74 N. |
| | | The above results allow the conclusion<br>that the subject device worst case plate<br>proved to be slightly stiffer and<br>mechanically resistant than the<br>reference device K944565 and<br>therefore, can be considered<br>equivalents for the performance<br>standpoint. |
| Screw Torsion<br>yield Test | Applied Standard ASTM F543-17 Annex<br>A.1. | The table A5.5 of the standard ASTM<br>F543-17 - Standard Specification and<br>Test Methods for Metallic Medical Bone<br>Screws establish that the Screw type 1.5<br>should presents minimum Torsional<br>Strength of 0.2 Nm. The worst case of<br>the device in question and the reference<br>device K944565 met the requirement of<br>the applied standard. |
| | Testing machine: Torsion Equipment.<br>The test was performed in ambient<br>conditions (ambient air at room<br>temperature. | |
| | OBJECTIVE: To determine the following<br>torsional properties of bone screw:<br>Torsional Yield Strength<br>Maximum Torque | The Torsion test reports show the<br>reference device screws presented a<br>maximum torque 0.28 Nm, while the<br>subject screws presented a maximum<br>torque of 0.29 Nm. Both tested screws<br>(subject and reference device)<br>accomplish the standard requirements<br>and therefore can be considered<br>equivalents. |
| | According to the ASTM F543-17<br>Standard, the specimen needs to be<br>placed in the holding device so that five<br>threads, below the head of the screw,<br>are exposed outside the holding device.<br>The screws that were evaluated do not<br>attend this condition because they are<br>too small. So, the "Breaking Angle" | |
| | | |
| | request is not applicable in this case. | |
| Pull out testing | Applied Standard ASTM F543-17 Annex A.3.<br><br>OBJECTIVE: Determine axial pullout strength of bone screw | The test reports show for the reference device screw a maximum force of 46.71 N, and for the subject device screw of 45.54 N. A maximum force variation of less than 2.5% was observed between reference and subject device and therefore they were considered equivalent. |
| Driving torque testing | Applied Standard ASTM F543-17 Annex A.2.<br><br>OBJECTIVE: To determine the driving torque of bone screw<br><br>Testing machine: Torsion Equipment.<br><br>The test was performed in ambient conditions (ambient air at room temperature. | The driving torque testing shows for the subject device an insertion torque of 0.05 Nm and a removal torque of 0.02 Nm. The reference device has presented an insertion and removal torque of 0.09 Nm and 0.03 Nm, respectively. For both cases, subject and reference, the insertion and removal torque are below than the Torsional Yield Strength. Therefore, considering the overall results of mechanical performance, the behavior of the subject and reference device can be deemed equivalent. |
{6}------------------------------------------------
{7}------------------------------------------------
Biocompatibility of the subject devices are supported by the reference device K182758.
No clinical data were included in this submission.
The subject devices are provided non-sterile and have no expiration date defined.
### CONCLUSION
The documentation submitted in this premarket notification demonstrates that the subject devices have equivalent features and performance and, therefore, are substantially equivalent to the identified predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.