K182758 · Mci Medical Concept Innovation, Inc. · JEY · Dec 5, 2019 · Dental
Device Facts
Record ID
K182758
Device Name
MCI - CMF System
Applicant
Mci Medical Concept Innovation, Inc.
Product Code
JEY · Dental
Decision Date
Dec 5, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MCI - CMF System is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
Device Story
MCI - CMF System consists of bone plates and fixation screws for maxillofacial surgery. Plates are commercially pure titanium (ASTM F67); screws are titanium alloy (Ti-6Al-4V, ASTM F136). Both are color-anodized. Device is non-sterile; requires cleaning and steam sterilization by the user. Used by qualified maxillofacial surgeons in clinical settings for trauma, reconstruction, and orthognathic procedures. Provides rigid fixation to stabilize bone segments during healing. Benefits include structural support for bony defects and fractures, facilitating functional and anatomical restoration.
Clinical Evidence
Bench testing only. No clinical data included. Performance supported by mechanical testing (4-point bending fatigue for plates; pullout, torsion, and driving torque for screws) per ASTM F382 and ASTM F543, and biocompatibility testing per ISO 10993 standards.
Technological Characteristics
Materials: CP Titanium (ASTM F67) for plates, Ti-6Al-4V alloy (ASTM F136) for screws. Surface: Color-anodized. Form factor: Various plate geometries (straight, L, T, Y, mesh, etc.) and screw types (cortical, emergency, locking). Energy: None (mechanical fixation). Sterilization: Steam sterilization (ISO 17665-1/2). Connectivity: None.
Indications for Use
Indicated for patients requiring fixation of maxillofacial and oral fractures, orthognathic reconstructions, mandibular reconstruction, and osteotomy surgery in the midface, maxilla, mandible, and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
CranioMaxillofacial Fixation (CMF) System - CMF Visionare (K180204)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.
December 5, 2019
MCI Medical Concept Innovation Inc. % Janine Treter Regulatory Affairs Specialist Passarini Regulatory Affairs of America LLC 201 Biscayne Blvd, Suite 1200 Miami, Florida 33131
Re: K182758
Trade/Device Name: MCI - CMF System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY, DZL Dated: November 4, 2019 Received: November 6, 2019
# Dear Janine Treter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar, Ph.D. for Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182758
Device Name
MCI - CMF System
Indications for Use (Describe)
MCI - CMF System is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table style="border:none"><tr><td><div> <span> <svg height="16" width="16"> <polygon points="1 0, 16 0, 16 15, 0 15, 0 0, 1 0" style="fill:lightgrey;stroke:black;stroke-width:1"></polygon> <polyline points="2 2,8 12,14 2" style="fill:none;stroke:black;stroke-width:2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div></td><td><div> <span> <svg height="16" width="16"> <polygon points="0 0, 15 0, 15 15, 0 15, 0 0" style="fill:white;stroke:black;stroke-width:1"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div></td></tr></table> | <div> <span> <svg height="16" width="16"> <polygon points="1 0, 16 0, 16 15, 0 15, 0 0, 1 0" style="fill:lightgrey;stroke:black;stroke-width:1"></polygon> <polyline points="2 2,8 12,14 2" style="fill:none;stroke:black;stroke-width:2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="16" width="16"> <polygon points="0 0, 15 0, 15 15, 0 15, 0 0" style="fill:white;stroke:black;stroke-width:1"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| <div> <span> <svg height="16" width="16"> <polygon points="1 0, 16 0, 16 15, 0 15, 0 0, 1 0" style="fill:lightgrey;stroke:black;stroke-width:1"></polygon> <polyline points="2 2,8 12,14 2" style="fill:none;stroke:black;stroke-width:2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="16" width="16"> <polygon points="0 0, 15 0, 15 15, 0 15, 0 0" style="fill:white;stroke:black;stroke-width:1"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
# 510(k) Summary
# ADMINISTRATIVE INFORMATION
| Manufacturer Name | MCI Medical Concept Innovation Inc.<br>4592 North Hiatus Road<br>Sunrise, Florida, USA 33351 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Carlos Lacerda, Mr.<br>Manager Director<br>MCI Medical Concept Innovation Inc.<br>E-Mail: carlos.lacerda@mci-medical.com<br>Telephone +01 (954) 306-2521 |
| Date Prepared | 05/Dec/2019 |
| Preparer / Alternate Contact | Janine Treter<br>Regulatory Affairs Specialist<br>Passarini Regulatory Affairs<br>janine@rapassarini.com.br |
| DEVICE NAME AND CLASSIFICATION | |
| Trade/ Proprietary Name | MCI - CMF System |
| Common Names | Bone Plate; Fixation Screws |
| Primary Classification Name | Bone Plate |
| Primary Classification Regulation | 21 CFR 872.4760, Class II |
| Primary Product Code | JEY |
| Subsequent Classification Name | Intraosseous Fixation Screw or Wire |
| Subsequent Classification Regulation | 21 CFR 872.4880, Class II |
| Subsequent Product Code | DZL |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
| PREDICATE DEVICE INFORMATION | |
| Primary Predicate Device | K180204 - CranioMaxillofacial Fixation (CMF) System - CMF Visionare<br>- Visionare LLC |
| Reference Devices | K944565 -KLS-MARTIN MICRO OSTEOSYTHESIS SYSTEM(1.5MM) - KLS-MARTIN L.P.<br>K943347 - KLS Mini Osteosynthesis System - KLS-MARTIN L.P.<br>K083388 - Synthes MatrixORTHOGNATHIC Plate System - Synthes (USA)<br>K091233 - SYNTHES MATRIXMANDIBLE SUBCONDYLAR PLATES - SYNTHES (USA) |
{4}------------------------------------------------
K022185 - Universal CMF System - Stryker Leibinger K182609 - Delphos Implants Rigid Fixation System - Delphos Implants - Ind. Com. Importação e Exportação K140037 & K160363 - Optimus CMF System - Osteonic Co., Ltd. K050934 - MODUS® Titanium Osteosynthesis System - Medartis Inc. K091679 & K103778 - LeForte Neuro System Bone Plate - Jeil Medical Corporation
### INDICATIONS FOR USE
MCI - CMF System is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
## SUBJECT DEVICE DESCRIPTION
The bone plates are made from commercially pure titanium (ASTM F67) and the bone screws are manufactured from titanium alloy - Ti-6AI-4V (ASTM F136) and are available in different sizes and shapes, according the site of the implantation and the extension of the fracture. The surface of plates and screws are colored-anodized.
MCI - CMF System implant devices are for single use. The devices are provided non-sterile and must be properly cleaned, disinfected and sterilized before use, according the recommendations provided in the Instructions for Use.
The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in maxillofacial surgery.
# TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE
The subject and the predicate devices have a range of plates and screws with the same indication for reconstructive surgery, for fixation of maxillofacial and oral fractures, orthognathic reconstructions, mandibular reconstruction and any osteotomy surgery or trauma in maxillofacial.
The subject device is substantially equivalent to the primary predicate device K180204 in intended use, Indications for Use, designs, materials, and function.
The indication for use statement of the subject devices, primary predicate and reference devices is shown in Table 5.1. The indication for use statement of subject and primary predicate device is identical. A comparison between the design and features of the subject devices, primary predicate and reference devices is shown in tabular format for plates (Tables 5.2 and 5.3) and screws (Table 5.4).
The subject device plates are substantially equivalent to the primary predicate device K180204, or reference devices K944565, K943347, K08388, K091233, K022185, K182609, K140037 & K160363, K050934, K091679 & K103778, in designs and range of dimensions. Differences in the plates design features and dimensions between the subject device and the primary predicate device K180204 are addressed by comparison to the reference devices.
{5}------------------------------------------------
For System 1.5, the reference device K944565 is for substantial equivalence of part of the Straight Micro Plates, Straight Micro Plates Bridge, L Micro Plates Bridge, and Y Micro Plates Bridge. It is for substantial equivalence of the Orbital Micro Plate L Oblique Bridge, H Nasal Micro Plate, I Nasal Micro Plate, Orbital Floor Micro Mesh and 3D Micro Mesh of 83x83, 122x 122 and 200x200 mm. For System 2.0, the reference device K944565 is for substantial equivalence of part of the Mini Plate Double Pillar for Ment.
The reference device K943347 for System 1.5 is for substantial equivalence of T Micro Plate Bridge 4 holes in design only. For System 2.0, the reference device K943347 is for substantial equivalence of L Mini Plates and L Mini Plates Bridge of 6 holes. Both are reversible plates and are covered by two plates of the reference device in the same design (right/left). It is also for substantial equivalence of T Mini Plate Bridge of 4 holes, Straight Mini Plates Bridge Tab, Chin Mini Plates Bridge Tab and L Mini Plates Bridge Double Tab.
The reference device K083388 for System 1.5 is for substantial equivalence of part of the Micro Plate Bridge. It is for the substantial equivalence of the Micro Plates L Orbital Oblique Bridge. For system 2.0, this reference device is for substantial equivalence of the Straight Mini Plate of 20 holes, L Mini Plate Oblique Bridge, Sagittal Mini Plates Cut Bridge and Sagittal Mini Plates Cut Oblong Bridge.
The reference device K091233 for System 2.0 is for substantial equivalence of Lambda Mini Plates, Support Mini Plates, Trapezoidal Mini Plates and Trapezoidal Parallel Mini Plates.
The reference device K022185 for System 1.5 is for substantial equivalence of part of the T Micro Plate Bridge of 5 holes. It is for substantial equivalence of the Orbital Floor Plate. For System 2.0, K022185 is for substantial equivalence of T Mini Plate of 6 holes, T Orbital Mini Plate Bridge, and for the design features not emcopassed by K944595 for the Mini Plate Double Pillar for Ment.
The reference device K182609 for System 1.5 is for substantial equivalence of the Straight Micro Plate design features not encompassed by K944565, part of the 4 and 6 holes Straight Micro Plates Bridge and for the design features of the T Micro Plates Bridge 5 holes not completely emcopassed by K022185. The K182609 is also for the substantial equivalence of X Micro Plate Bridge and Support Zygomatic Micro Plates. For System 2.0, the reference device K182609 is for substantial equivalence of the Straight Mini Plate of 6 holes, Straight Mini Plate Bridge of 4 holes and for part of the Straight Mini Plate Bridge of 6 holes. It is for the substantial equivalence of the Orbital Mini Plates of 8 and 10 holes, Y Mini Plate, Mini Plates Double Line Bridge and for substantial equivalence of the Chin Mini Plates not encompassed by the primary predicate device.
The reference devices K140037 & K160363 for System 1.5 are for substantial equivalence of the Lefort Micro Plates. For System 2.0 are for substantial equivalence of Straight Mini Plates not emcopassed by K083388 and K182609 and for the Orbital Mini Plate not emcopassed by K182609. The reference devices K140037 & K160363 are also for the substantial equivalence of BSSO Dual Angled Mini Plates Bridge.
For System 1.5 the reference device K050934 is for substantial equivalence of the Straight Micro Plate Bridge features not emcopassed by K944565 and K182609, and of the L Micro Plates Bridge features not encompassed by K944565 and K083388. It is also for substantial equivalence of Y Micro Plate Bridge not emcompassed by K944565 and 3D Micro Mesh of 50x50 mm. For System 2.0 is for substantial equivalence of Straight Mini Plates Bridge not emcopassed by K182609 and L Mini Plates
{6}------------------------------------------------
Bridge not encompassed by K943347. It is also for the substantial equivalence of the Mini Plate Sagittal.
The reference devices K091679 & K103778 for System 1.5 are for substantial equivalence of Angled Micro Plates for Piriformis.
The subject device screws are substantially equivalent to the primary predicate device K180204, or reference devices K944565 and K022185, in designs and range of dimensions for all types of screws: Cortical Screw, Emergency Cortical Screw, Cortical Screw AP and Locking Screw AP. Differences in the screw design features and dimensions between the subject device and the primary predicate device K180204 are addressed by comparison to the reference devices.
For screws, the reference device K944565 is for substantial equivalence on the results of comparative performance testing. The reference device K022185 is for substantial equivalence of the length 20 mm for Cortical Screws of 2.0 mm diameter and Emergency Cortical Screw of 2.3 mm. The reference device K022185 is also for substantial equivalence of the Cortical Screws AP of 2.0 mm in diameter for the length of 4 mm.
{7}------------------------------------------------
| Table 5.1: Comparison on indication for use statement | | | |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | KNUMBER/ MANUFACTURER | INDICATION FOR USE STATEMENT | SUBSTANTIAL EQUIVALENCE<br>DISCUSSION |
| SUBJECT<br>DEVICE | K182758 – MCI - CMF System<br><br>MCI Medical Concept Innovation Inc. | MCI - CMF System is intended for use in selective trauma of the midface,<br>maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery<br>of the maxilla, mandible and chin. | Equivalent<br><br>The indication for use of subject and<br>primary predicate device is identical. |
| PRIMARY<br>PREDICATE<br>DEVICE | K180204 - CranioMaxillofacial Fixation (CMF) System<br>- CMF Visionare<br><br>Visionare LLC | CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in<br>selective trauma of the midface, maxillofacial surgery, reconstructive procedures,<br>and selective orthognathic surgery of the maxilla, mandible and chin. | The subject device indication for use is<br>within the scope of indications of the<br>reference devices. |
| REFERENCE<br>DEVICES | K944565 -KLS-MARTIN MICRO OSTEOSYTHESIS<br>SYSTEM(1.5MM)<br><br>KLS Martin LP | No 510(k) summary available | Despite of the reference devices<br>"K091679 & K103778 - LeForte Neuro<br>System Bone Plate" citing the word<br>Neuro in their proprietary name, by the<br>indication for use itself is possible to<br>understand they remain in the scope<br>since are related to midface and<br>craniomaxillofacial applications.<br>Therefore, K091679 & K103778 are<br>suitable as reference predicates. |
| | K943347 - KLS Mini Osteosynthesis System<br><br>KLS-MARTIN L.P. | No 510(k) summary available | |
| | K083388- Synthes MatrixORTHOGNATHIC Plating<br>System | The Synthes MatrixORTHOGNATHIC Plating System is intended for use in selective<br>trauma of the midface and craniofacial skeleton; craniofacial surgery; | |
| KNUMBER/ MANUFACTURER | INDICATION FOR USE STATEMENT | SUBSTANTIAL EQUIVALENCE<br>DISCUSSION | |
| Synthes (USA) | reconstructive procedures; and selective orthognathic surgery of the maxilla,<br>mandible and chin in adolescents (greater than 12 to 21 years of age) and adults. | | |
| | Specific Indications for Use: | | |
| | • Fractures ofthe midface and craniofacial skeleton | | |
| | • LeFort I osteotomies, sagittal split osteotomies and genioplasties | | |
| | • Orthognathic surgery including reconstructive procedures | | |
| K091233 - SYNTHES MATRIXMANDIBLE<br>SUBCONDYLAR PLATES | The Synthes MatrixMANDIBLE Subcondylar Plates are intended for oral,<br>maxillofacial surgery; trauma and reconstructive surgery, specifically for fractures of<br>the subcondylar region of the mandible and fractures of the condylar basis region of<br>the mandible. | | |
| SYNTHES (USA) | | | |
| K022185 - Universal CMF System | The Stryker® Leibinger Universal CMF System is a Cranio-maxillofacial (CMF) plate<br>and screw system intended for osteotomy, stabilization and rigid fixation of CMF<br>fractures and reconstruction. | | |
| Stryker Leibinger | | | |
| REFERENCE<br>DEVICES | K182609 - Delphos Implants Rigid Fixation System<br>Delphos Implants - Ind. Com. Importação e<br>Exportação | The Delpos Implants Rigid Fixation System is intended for fracture fixation in<br>maxillofacial trauma reconstruction, mandibular reconstruction and orthognatic<br>reconstruction. | |
| KNUMBER/ MANUFACTURER | INDICATION FOR USE STATEMENT | SUBSTANTIAL EQUIVALENCE<br>DISCUSSION | |
| | The Delpos Implants Rigid Fixation System implants are intended for single use only. | | |
| K140037 & K160363 - Optimus CMF System<br>Osteonic Co., Ltd. | Optimus CMF System is implantable bone plates and bone screws for maxillofacial<br>and mandible surgery procedures including:<br>1. Fractures<br>2. Osteotomies<br>3. Reconstructive procedures<br>4. Revision procedures where other treatments or devices have failed. | | |
| K050934 - MODUS® Titanium Osteosynthesis System<br>Medartis, Inc. | The MODUS® Titanium Osteosynthesis System is intended for osteotomies and<br>fractures involving any part ofthe craniofacial skeleton and requiring positional and<br>functional stability. Indications include fixation in the nasoethmoidal, intraorbital,<br>and frontal sinus areas; fixation of comminuted fractures of maxillo-facial and<br>craniofacial areas; tumor surgery for defect bridging; reconstruction of bony<br>structures by means of mesh materials; coverings for burr holes in the skull; trauma<br>of nasal bones; surgical correction of dento facial deformations; and reconstruction<br>after tumor surgery. | | |
| K091679 & K103778 - LeForte Neuro System Bone<br>Plate<br>Jeil Medical Corporation | This device is intended for use in selective trauma of the mid-face and craniofacial<br>skeleton; craniofacial surgery; reconstructive procedures; and selective orthognatic<br>surgery of the maxilla and chin. | | |
MCI Medical Concept Innovation Inc.
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
### Table 5.2: Comparison between the subject, primary and reference K944565. K943347. K083388. K091233 devices for plates to be optived
| COMPARISON | SUBJECT DEVICE | | | PRIMARY PREDICATE DEVICE | | | REFERENCE DEVICES | | | | | | | | | | | |
|-------------------|------------------------------------------|--------------------------------------|-------------------|----------------------------------|-------------------------|-------------------|------------------------------------------|-----------------------------|-------------------|------------------------------|-------------|-------------------|---------------------------|---------|-------------------|-------------|---------|-------------------|
| | K182758 | | | K180204 | | | K944565 | | | K943347 | | | K083388 | | | K091233 | | |
| Product Code | JEY, DZL | | | JEY, DZL | | | JEY | | | JEY | | | JEY, DZL | | | JEY | | |
| Raw Material | CP Titanium | | | CP Titanium | | | CP Titanium | | | CP Titanium | | | CP Titanium | | | CP Titanium | | |
| Surface treatment | Anodized | | | Anodized | | | Anodized | | | Anodized | | | Anodized | | | Anodized | | |
| | System 1.5 | | | | | | | | | | | | | | | | | |
| Design/Features | Description | Hole nº | Thickness (mm) | Description | Hole nº | Thickness (mm) | Description | Holes | Thickness (mm) | Description | Holes | Thickness (mm) | Description | Holes | Thickness (mm) | Description | Holes | Thickness (mm) |
| | Straight Micro Plate | 4, 6, 8, 10, 16, 20 | 0.6 | Straight Mini Plate Regular | 4, 6, 8, 16, 20 | 0.7 | Micro Plate | 4, 6, 8, 16, 20 | 0.6 | Mini Plate | 4, 6, 8, 16 | 0.6, 1.0 | | | | | | |
| | Straight Micro Plate Bridge | 4, 6 | 0.6 | Straight Mini Plate Medium | 4, 6 | 0.7 | Micro Plate | 4, 6 | 0.6 | | | | | | | | | |
| | Orbital Micro Plate | 8, 10 | 0.6 | Orbital Mini Plate Regular | 8, 10 | 0.7 | Micro Plate | 8, 10 | 0.6 | Champy Plate | 8, 10 | 0.6 | | | | | | |
| | L Micro Plate Bridge | 4 | 0.6 | L Mini Plate Medium & Extra Long | 4 | 0.7 | Micro Plate | 4 | 0.6 | Mini Plate | 4 | 0.6, 1.0 | 90° L Plate Medium & Long | 4 | 0.5 | | | |
| | L Micro Plate Oblique Bridge | 4, 5 | 0.6 | | | | Micro Plate | 4, 5 | 0.6 | | | | | | | | | |
| | L Micro Plate Orbital | 6 | 0.6, 0.7, 0.8 | | | | | | | | | | Anatomical L Plate | 6 | 0.5, 0.7, 0.8 | | | |
| | T Micro Plate Bridge | 4, 5 | 0.6 | T Mini Plate Medium & Extra Long | 5 | 0.7 | | | | Champy Plate | 4 | 1.0 | | | | | | |
| | X Micro Plate Bridge | 6 | 0.6 | T Mini Plate Medium & Extra Long | 6 | 0.7 | Micro Plate | 6 | 0.6 | | | | | | | | | |
| | Y Micro Plate Bridge | 5 | 0.6 | | | | Micro Plate | 5 | 0.6 | | | | | | | | | |
| | Lefort Micro Plate | 11 | 0.6 | | | | Lindorf Micro Le Fort I Plate | 11 | 1.0 | | | | | | | | | |
| | Angled Micro Plate Piriforms | 10 | 0.8 | Le Fort Mini Plate | 14 | 0.8 | | | | | | | | | | | | |
| | Support Zygomatic Micro Plate | 7 | 0.8 | | | | | | | | | | | | | | | |
| | H Nasal Micro Plate | 12 | 0.6 | | | | Micro Plate | 12 | 0.6 | Mini Plate | 12 | 0.6 | | | | | | |
| | I Nasal Micro Plate | 11, 12 | 0.6 | | | | Micro Plate | 11, 12 | 0.6 | | | | | | | | | |
| COMPARISON | SUBJECT DEVICE | | | PRIMARY PREDICATE DEVICE | | | REFERENCE DEVICES | | | | | | | | | | | |
| | | K182758 | | | K180204 | | | K944565 | | | K943347 | | | K083388 | | | K091233 | |
| Design/Features | Orbital Floor<br>Micro Mesh | | | | | | Micro Orbital<br>Mesh | | | | | | | | | | | |
| | | --- | 0.3, 0.5 | | | | | --- | 0.3, 0.5 | | | | | | | | | |
| | Orbital Floor | | | | | | | | | | | | | | | | | |
| | Plate | --- | 0.3 | | | | | | | | | | | | | | | |
| | 3D Micro Mesh | | | | | | 3D-Mesh | | | | | | | | | | | |
| | | 50x50<br>83x83<br>122x122<br>200x200 | 0.6 | | | | | 80x80<br>120x120<br>200x200 | 0.6 | | | | | | | | | |
| | System 2.0 | | | | | | | | | | | | | | | | | |
| Design/Features | Description | Hole n° | Thickness<br>(mm) | Description | Hole n° | Thickness<br>(mm) | Description | Holes | Thickness<br>(mm) | Description | Holes | Thickness<br>(mm) | Description | Holes | Thickness<br>(mm) | Description | Holes | Thickness<br>(mm) |
| | Straight Mini<br>Plate | | | Straight Mini<br>Plate Regular | | | | | | | | | Adaptation<br>Plate | | | | | |
| | | 4, 6, 12, 16,<br>20 | 0.9 | | 4, 6, 12,<br>16, 20 | 1.0 | | | | | | | | 20 | 0.8 | | | |
| | Straight Mini<br>Plate Bridge | | | Straight Mini<br>Plate Bridge | | | | | | Mini Plate<br>Bridge | | | | | | | | |
| | | 4, 6 | 0.9 | | 4, 6 | 1.0 | | | | | 4 | 1.0 | | | | | | |
| | Orbital Mini<br>Plate | | | Orbital Mini Plate | | | | | | Champy Plate<br>Bridge | | | | | | | | |
| | | 6, 8, 10 | 0.9 | | 6, 8, 10 | 1.0 | | | | | 8, 10 | 0.6 | | | | | | |
| | | L Mini Plate | | | L Mini Plate<br>Regular | | | | | | Mini Plate | | | | | | | |
| | | 4, 6 | 1.0 | | 4 | 1.0 | | | | | 4, 5 | 1.0 | | | | | | |
| | L Mini Plate<br>Bridge | | | L Mini Plate Long | | | | | | Mini Plate | | | | | | | | |
| | | 4, 6 | 1.0 | | 4 | 1.0 | | | | | 6 | 1.0 | | | | | | |
| | L Mini Plate<br>Oblique Bridge | | | | | | | | | | | | Oblique L Plate | | | | | |
| | | 6 | 1.0 | | | | | | | | | | | 6 | 0.8 | | | |
| | T Mini Plate | | | | | | | | | Mini Plate | | | | | | | | |
| | | 4 | 1.0 | | | | | | | | 4, 6 | 1.0 | | | | | | |
| | | 6 | 0.9 | | | | | | | | | | | | | | | |
| | T Orbital Mini<br>Plate | | | | | | | | | | | | | | | | | |
| | | 8 | 0.9 | | | | | | | | | | | | | | | |
| | Y Mini Plate | | | |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.