MINI AND MICRO FRAGMENTS RECONSTRUCTION SYSTEM-NEOFACE
Applicant
Neoortho Produtos Ortopedicos S/A
Product Code
JEY · Dental
Decision Date
Dec 20, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mini and Micro Fragments Reconstruction System - NEOFACE is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
Device Story
Fixation system for maxillofacial surgery; consists of implant plates and screws. Plates made of commercially pure titanium; screws made of titanium alloy. Used by surgeons in clinical settings for trauma, reconstruction, and orthognathic procedures. Provides mechanical stabilization of bone segments to facilitate healing. Surgeon selects appropriate plate/screw configuration based on anatomical site and clinical need; implants are secured to bone to maintain alignment. Benefits include restoration of skeletal structure and function in the midface and mandible.
Clinical Evidence
Bench testing only. Mechanical performance evaluated via plate bending strength (ASTM F382) and screw torsional strength, breaking angle, and axial pullout strength (ASTM F543). No clinical data provided.
Technological Characteristics
Materials: Commercially pure titanium (plates), titanium alloy (screws). Mechanical testing per ASTM F382 (plates) and ASTM F543 (screws). Designs include locking and non-locking plates and screws; self-tapping and self-drilling screw options. Sterile, single-use implants.
Indications for Use
Indicated for patients requiring fixation in selective midface trauma, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible, and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Synthes (USA) Craniofacial Plate and Screw System (K050608)
Walter Lorenz Surgical Mandibular Fracture / Reconstruction Devices and Pre-bent Plates (K063052)
Lorenz Sterile Bone Plates and Bone Screws (K972322)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Summary
# NEOORTHO PRODUTOS ORTOPÉDICOS S/A
2010
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# MINI AND MICRO FRAGMENTS RECONSTRUCTION SYSTEM - NEOFACE
October 19, 2010
## ADMINISTRATIVE INFORMATION
- 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -
| Manufacturer Name: | NEOORTHO Produtos Ortopédicos S/A<br>Rua Ângelo Domingos Durigan, 607, Cascatinha<br>Curitiba - PR 82020-340, Brazil | | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|
| | Telephone: | +55 41 3535-1033 | |
| | Fax: | +55 41 3535-1018 | |
| Official Contact: | Luciane Yumi Suzuki de Oliveira<br>NEOORTHO Produtos Ortopédicos S/A | | |
| Representative/Consultant: | Linda K. Schulz, BSDH, RDH<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>11234 El Camino Real, Suite 200<br>San Diego, CA 92130 USA | | |
| | Telephone: | +1 (858) 792-1235 | |
| | Fax: | +1 (858) 792-1236 | |
| | Email: | lschulz@paxmed.com<br>kthomas@paxmed.com | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Mini and Micro Fragments Reconstruction System –<br>NEOFACE |
|-----------------------------|-------------------------------------------------------------|
| Common Name: | Bone plate, Intraosseous bone screw |
| Classification Regulations: | 21 CFR 872.4760, 21 CFR 872.4880<br>Class II |
| Product Code: | JEY, DZL |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
{1}------------------------------------------------
#### INTENDED USE
Mini and Micro Fragments Reconstruction System - NEOFACE is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
### DEVICE DESCRIPTION
Mini and Micro Fragments Reconstruction System – NEOFACE is a fixation system for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin. It consists of implant plates and implant screws. The plates of the subject system include mini and micro fragment reconstruction plates. Plates are made of commercially pure titanium and screws are made of titanium alloy.
### EQUIVALENCE TO MARKETED DEVICE
NEOORTHO Produtos Ortopédicos S/A demonstrated that, for the purposes of FDA's regulation of medical devices, Mini and Micro Fragments Reconstruction System - NEOFACE is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices:
Synthes (USA) Craniofacial Plate and Screw System cleared under K050608,
Synthes Orbital Plates cleared under K080331,
Synthes MatrixORTHOGNATHIC Plating System cleared under K083388,
Synthes MatrixMANDIBLE Preformed Reconstruction Plates cleared under K091144,
Biomet Mandibular Fracture / Reconstruction Devices (TraumaOne) cleared under K081067,
Walter Lorenz Surgical Mandibular Fracture / Reconstruction Devices and Pre-bent Plates cleared under K063052, and
Lorenz Sterile Bone Plates and Bone Screws cleared under K972322.
The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject and predicate implants are all made of commercially pure titanium or titanium alloys. The subject and predicate plates encompass the same range of physical dimensions, including length, thickness and shape. The subject and predicate plate designs include a variety of straight and geometric configurations that are commonly used in maxillofacial trauma and reconstructive surgery, in both locking and non-locking designs. The subject and predicate screws encompass the same range of physical dimensions and include locking, non-locking, self-tapping, and self-drilling designs. The subject and predicate devices are packaged in similar materials and are sterilized using similar methods.
{2}------------------------------------------------
Mechanical testing was performed to determine plate bending strength according to ASTM F382. Screw torsional strength, breaking angle, and axial pullout strength were determined according to ASTM F543.
Overall, Mini and Micro Fragments Reconstruction System - NEOFACE has the following similarities to the predicate devices:
- . has the same intended use,
- . uses the same operating principle,
- . incorporates the same basic design,
- incorporates the same or very similar materials, and .
- has similar packaging and is sterilized using the same processes. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NEOORTHO Produtos Ortopedicos S/A C/O Ms. Linda K. Schulz Paxmed International. LLC 11234 El Camino Real. Suite 200 San Diego, California 92130
DEC 2 0 2010
Re: K102641
Trade/Device Name: Mini and Micro Fragments Reconstruction System -NEOFACE Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: October 19, 2010 Received: October 20, 2010
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Schulz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
h. for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Mini and Micro Fragments Reconstruction System – NEOFACE
## ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
DEC 2 0 2010
510(k) Number: K102641
Mini and Micro Fragments Reconstruction System - NEOF ACE Device Name:
Indications for Use:
Mini and Micro Fragments Reconstruction System – NEOFACE is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
Prescription Use _X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | K102641 |
|----------------|---------|
|----------------|---------|
Page 1 of _
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.