The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).
Device Story
Stryker Universal Mesh consists of titanium mesh implants (1.2mm, 1.5mm, 1.7mm) and associated manual instrumentation (cutters, benders, clippers). Implants are malleable for 3D anatomical adaptation and modified via trimming/bending. Used in cranial surgery to provide structural support and stabilization; fixated to bone using previously marketed Stryker CMF screws. System includes storage containers. Operated by surgeons in clinical/OR settings. Output is a physical implant providing rigid fixation; aids in bone healing and structural integrity post-cranial procedures. Non-sterile; single-use implants; reusable instruments.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993), cleaning/sterilization validation (ISO 17664, 14937, 17665), mechanical performance (static/dynamic load), corrosion testing, and MR compatibility (per FDA guidance).
Technological Characteristics
Material: Commercially pure titanium (ASTM F67). Form factor: 2D mesh (0.3mm-0.8mm thickness, 40x40mm to 200x200mm). Fixation: 1.2mm-1.7mm screws. Sterilization: End-user sterilized. Connectivity: None. Software: None.
Indications for Use
Indicated for reconstruction, stabilization, and rigid fixation following craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12+).
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, with three human profiles forming the staff and a cloth draped around the base.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2016
Stryker Craniomaxillofacial Mr. Jonathan Schell Sr. Regulatory Affairs Specialist 750 Trade Centre Way, Suite 200 Portage, MI 49002
Re: K161821
Trade/Device Name: Stryker Universal Mesh Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: November 4, 2016 Received: November 7, 2016
Dear Mr. Schell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161821
Device Name Stryker Universal Mesh
Indications for Use (Describe)
The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
- SUBMITTER [§807.92(a)(1)] I.
| 510(k) Owner: | Stryker Leibinger GmbH & Co. KG |
|---------------|---------------------------------|
| | Boetzinger Strasse 41 |
| | 79111 Freiburg, Germany |
- Submitter/ Contact Jonathan Schell Person: Sr. Regulatory Affairs Specialist Stryker Craniomaxillofacial 750 Trade Centre Way Portage, MI 49002 Phone: 269-389-5596 Fax: 877-648-7114
- Date prepared: November 4, 2016
#### II. DEVICE [§807.92(a)(2)]
| Trade Name: | Stryker Universal Mesh |
|--------------------------|---------------------------------------------------------|
| Common or Usual<br>name: | Bone Plate |
| Classification<br>name: | Preformed Alterable Cranioplasty Plate; 21 CFR 882.5320 |
| Regulatory Class: | Class II |
| Product Code: | GWO |
| Classification<br>Panel: | Neurology |
#### PREDICATE DEVICE [§807.92(a)(3)] III.
Universal CMF System – K022185
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION [§807.92(a)(4)]
The Stryker Universal Mesh consists of mesh implants in different shapes, thicknesses and perforation patterns which include the 1.2mm and 1.5/ 1.7mm Dynamic Mesh implants and Hybrid Mesh implants. The meshes can be modified with trimming and bending instruments for an anatomical fit and are malleable for three-dimensional adaptations. The necessary instrumentation for mesh modification, Mesh Cutter (coarse), Mesh Step Bender, and the Mesh Clipper are included with the system. The Stryker Universal Mesh implants are provided nonsterile and are for single use only. The instrumentation and storage system is reusable.
Also, the mesh can be fixated to the bone with screws available in previously marketed Stryker CMF implant systems. The previously marketed screws are not specifically packaged with the subject device, but the implant tray of the Universal Mesh Storage system contains space to store the screw disc if that option is chosen.
#### INDICATIONS FOR USE [§807.92(a)(5 )] V.
| | Subject Device | Predicate - K022185 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Universal Mesh is indicated for<br>reconstruction, stabilization and/or<br>rigid fixation subsequent to<br>craniotomy, craniectomy,<br>cranioplasty, and cranial fractures in<br>adults and adolescents (age 12 and<br>higher). | The Stryker Leibinger Universal<br>CMF System is a Cranio-<br>maxillofacial (CMF) plate and screw<br>system intended for osteotomy,<br>stabilization and rigid fixation of<br>CMF fractures and reconstruction. |
## TABLE 1: COMPARISON OF INDICATIONS FOR USE
The predicate device is indicated for use in a range of anatomic areas (cranial, craniofacial skeleton, mid-face, maxilla, mandible) for a variety of surgical applications (craniofacial surgery, reconstructive procedures). The Indications for Use statement of the Stryker Universal Mesh falls within the scope of the broader Indications statement of the predicate device. Additionally, the Subject device is intended to be used in adolescent patients 12 years of age and older. The differences in the Indications statement for the proposed device in comparison to the predicate does not constitute a new intended use.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE [§807.92(a)(6)]
{5}------------------------------------------------
The Stryker Universal Mesh is compared to its predicate devices for substantial equivalence based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
## A. Principle of Operation
The Stryker Universal Mesh implants are designed for the reconstruction, stabilization and rigid fixation of the bony areas of the cranial skeleton. The necessary instrumentation and product storage are included with the system.
## B. Technological and Operational Characteristics
-Substantial Equivalence of the Subject devices and Predicate devices are based on the following technological elements:
| Feature | Stryker Universal Mesh | Universal CMF System<br>K022185 | Explanation of Differences |
|--------------------------------------------------------|-----------------------------------|---------------------------------|----------------------------|
| | Subject Device | Predicate Device | |
| Sterility: End<br>User Sterilized | X | X | Same |
| Sizing | | | |
| Shape | Square, partly with rounded edges | Square | Similar |
| Sizes | 40x40mm<br>to<br>200x200mm | 40x40mm<br>to<br>200x200m | Similar |
| Implant<br>Thickness | 0.3mm -<br>0.8mm | 0.3mm -<br>0.6mm | Similar |
| Design | | | |
| 2-dimensional<br>(2D) | X | X | Same |
| Mode of<br>Fixation | | | |
| 1.2mm - 1.7mm<br>screws | X | X | Same |
| Material | | | |
| Titanium:<br>Commercially<br>pure titanium<br>ASTM F67 | X | X | Same |
Overall these changes do not alter the Subject device significantly and it does not represent a new worst case in design.
Strvk
{6}------------------------------------------------
## VII. PERFORMANCE DATA [§807.92(b)(1)]
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The biocompatibility evaluation for the Stryker Universal Mesh system was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The tests supported the biocompatibility of the device.
The implant is made from titanium which conforms to ASTM F67 for chemical composition, the same as the predicate. Cytotoxicity testing was performed using DIN EN 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity. Chemical residual analysis was conducted according to validated methods with results that met the established acceptance criteria. These tests support the biocompatibility of the device.
## Performance Testing
| Test | Test Method Summary | Results |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Life time<br>reprocessing | Verify the reprocessing life of the Subject device, instruments, and storage<br>container are according to internal standard. | Internal Acceptance<br>Criteria |
| Cleaning<br>Validation | Verify the cleaning process for Subject device, instruments and storage<br>container are according to recommendations in ISO 17664, AAMI TIR 30,<br>and AAMI TIR 12. | Method validated. |
| Sterilization<br>Validation | Validate the sterilization instructions for the Subject device, instruments,<br>and storage container are according to internal standard. Referenced<br>standards: ISO 14937, ISO 17664, ISO 17665 - 1&2, AAMI TIR 12,<br>AAMI ST 81, AAMI ST79. | Method validated. |
| Mechanical<br>performance<br>Testing | Static Load Test and Dynamic Load Test | Acceptance criterion<br>were met, tests were<br>passed. |
| Corrosion<br>Testing | Verify corrosion behavior of Subject device, instruments, and storage<br>device. | Acceptance criterion<br>was met, test was<br>passed. |
| MR<br>Compatibility | Magnetic Resonance (MR) testing was performed on the subject device<br>mesh implants in accordance with FDA draft guidance for industry<br>"Assessment of Radiofrequency - Induced Heating in the Magnetic<br>Resonance (MR) Environment for Multi-Configuration Passive Medical | Based on the results of<br>these evaluations, the<br>Stryker Universal mesh<br>will be labeled as MR |
The following performance tests were completed:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is black and appears to be a logo. A small circled letter 'c' is present to the upper right of the word.
| Devices," June 29, 2015. MR heat testing has been performed on the<br>Subject device according to FDA final guidance "Establishing Safety and<br>Compatibility of Passive Implants in the Magnetic Resonance (MR)<br>Environment," December 11, 2014. | conditional. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
The Subject device met all pre-defined acceptance criteria and standards and was found to not represent a new worst case. Overall, the results of the performance bench tests support the substantial equivalence of the Subject device.
## Animal Testing
Animal testing was not required as a basis for substantial equivalence.
## Clinical Testing [§807.92(b)(2)]
Clinical testing was not required as a basis for substantial equivalence.
## VIII. CONCLUSIONS [§807.92(b)(3)]
The results of the non-clinical data demonstrate the Stryker Universal Mesh will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.92 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.