The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The HemoSphere Advanced Monitoring Platform can be used with the Viewfinder Remote mobile application for supplemental near real-time remote display of tissue oximetry and non-invasive hemodynamic parameter data monitored with a FORE-SIGHT ELITE oximeter module and HemoSphere ClearSight Module, respectively. The HemoSphere Advanced Monitoring Platform is intended for use with the Edwards Swan-Ganz and Oximetry Catheters and FloTrac, Acumen IQ, TruWave DPT sensors and ClearSight cuffs.
Device Story
HemoSphere Advanced Monitoring Platform processes hemodynamic and volumetric data from connected modules (Swan-Ganz, Oximetry, Pressure, Tissue Oximetry, ClearSight). System transforms sensor inputs into real-time hemodynamic parameters, waveforms, and physiological alarms. Viewfinder Remote mobile application enables supplemental near-real-time remote display of non-invasive data and alarms via Viewfinder Hub and Cloud. Used in hospital settings (OR, ICU, ER) by clinicians; provides supportive visual aid for patient status communication. Output allows clinicians to monitor multiple patients simultaneously; assists in hemodynamic optimization and goal-directed therapy protocols. Benefits include improved situational awareness and remote access to patient status.
Clinical Evidence
No clinical data was required for this submission. Performance was established through bench testing, system verification, electrical safety/EMC testing (IEC 60601 series), wireless coexistence testing (ANSI C63.27), and usability studies.
Technological Characteristics
Modular hemodynamic monitoring platform. Components: HemoSphere Advanced Monitor, Swan-Ganz/Oximetry/Pressure/Tissue Oximetry/ClearSight modules. Connectivity: Wireless streaming to mobile application via Viewfinder Hub/Cloud. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366, IEC 60601-2-34, IEC 60601-2-49. Software: Major level of concern for monitor, moderate for mobile app.
Indications for Use
Indicated for adult and pediatric critical care patients requiring monitoring of cardiac output, hemodynamic parameters, venous oxygen saturation, or regional hemoglobin oxygen saturation. Used in hospital environments (OR, ICU, ER) for continuous assessment of cardiac function, fluid status, vascular resistance, and pressure. Acumen HPI feature indicated for surgical or non-surgical patients receiving advanced hemodynamic monitoring to provide insight into likelihood of future hypotensive events (MAP < 65 mmHg). Fore-Sight Elite module indicated for individuals at risk for reduced-flow or no-flow ischemic states.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 8, 2021
Edwards Lfesciences Chirag Shah Manager, Regulatory Affairs One Edwards Way Irvine, California 92614
Re: K211465
Trade/Device Name: HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, HemoSphere Tissue Oximetry Module, HemoSphere ForeSight Module, HemoSphere ClearSight Module, Acumen Hypotension Prediction Index Feature, Viewfinder Remote Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DQE, QAQ, MUD, DXN, DSB Dated: May 10, 2021 Received: May 11, 2021
Dear Chirag Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K211465
#### Device Name
HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, HemoSphere Tissue Oximetry Module, HemoSphere ClearSight Module, HemoSphere ClearSight Module, Acumen Hypotension Prediction Index, Viewfinder Remote
#### Indications for Use (Describe)
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cable and Edwards is indicated for use in adult and pediatric critical care patients requiring of venous oxygen saturation (SvO2 and SevO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function. fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
{3}------------------------------------------------
HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module and FORE-SIGHT ELITE Tissue Oximeter Module
The non-invasive Fore-Sight Elite tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no-flow ischemic states. The Fore-Sight Elite tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
· When used with large sensors, the Fore-Sight Elite tissue oximeter module is indicated for use on adults and transitional adolescents ≥40 kg.
· When used with medium sensors, the Fore-Sight Elite tissue oximeter module is indicated for use on pediatric subjects >3 kg.
· When used with small sensors, the Fore-Sight Elite tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere ClearSight Module
The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the non-invasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters. Refer to the ClearSight finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number: | 2015691 |
| Contact Person: | Chirag Shah<br>Manager, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 250-1580<br>Fax: (949) 809-2972 |
| Date: | July 1, 2021 |
| Platform Name | HemoSphere Advanced Monitoring Platform |
| Trade Name: | HemoSphere Advanced Monitor<br>HemoSphere Swan-Ganz Module<br>HemoSphere Oximetry Cable<br>HemoSphere Pressure Cable<br>HemoSphere Tissue Oximetry Module<br>HemoSphere ForeSight Module<br>HemoSphere ClearSight Module<br>Acumen Hypotension Prediction Index feature<br>Viewfinder Remote |
| Common Name: | Cardiac Output/Oximetry/Ejection Fraction Computer |
| Classification<br>Name: | Programmable Diagnostic Computer<br>21 CFR 870.1425<br>Fiberoptic Oximeter Catheter<br>21 CFR 870.1230<br>Adjunctive Predictive Cardiovascular Indicator<br>21 CFR 870.2210<br>Oximeter<br>21 CFR 870.2700<br>Oximeter, Tissue Saturation<br>21 CFR 870.1130<br>System, Measurement, Blood-Pressure, Non-Invasive<br>21 CFR 870.2770<br>Plethysmograph, Impedance |
# 510(k) Summary – HemoSphere Advanced Monitoring Platform
{5}------------------------------------------------
| Product Code: | DQK, Class II<br>DQE, Class II<br>QAQ, Class II<br>MUD, Class II<br>DSB, Class II<br>DXN, Class II |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary<br>Predicate<br>Device: | HemoSphere Advanced Monitoring Platform manufactured by Edwards<br>Lifesciences, K201446, cleared October 1, 2020. |
| Device<br>Description: | The HemoSphere Advanced Monitoring platform was designed to simplify<br>the customer experience by providing one platform with modular solutions<br>for their hemodynamic monitoring needs. The user can choose from the<br>available optional sub-system modules or use multiple sub-system modules<br>at the same time. This modular approach provides the customer with the<br>choice of purchasing and/or using specific monitoring applications based on<br>their needs. Users are not required to have all of the modules installed at the<br>same time for the platform to function.<br><br>HemoSphere Advanced Monitoring Platform consists of the HemoSphere<br>Advanced Monitor that provides a means to interact with and visualize<br>hemodynamic and volumetric data on a screen and five (5) optional external<br>modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April<br>14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14,<br>2017), HemoSphere Pressure Cable (K180881 Cleared, November 16,<br>2018), HemoSphere Tissue Oximetry Module (K190205 August 29, 2019),<br>HemoSphere ForeSight Module (K180003, May 10, 2018), and the<br>HemoSphere ClearSight Module (K201446 cleared October 1, 2020).<br><br>The HemoSphere Advanced Monitor (K201446 most recently cleared<br>October 1, 2020), subject of this submission, is being modified to enable<br>connectivity to a Viewfinder Remote mobile application via a software-<br>based Viewfinder Hub and Viewfinder Cloud.<br><br>The Viewfinder Remote mobile application provides clinicians with a<br>supplemental near-real time display of the patient hemodynamic data from<br>the connected HemoSphere Advanced Monitoring Platform. The<br>Viewfinder Remote application is part of the Edwards Viewfinder network,<br>which includes Viewfinder Hub and Viewfinder Cloud. The Viewfinder<br>Remote mobile application functions as a supportive visual aid for patient<br>status communication between clinicians and allows them to view multiple<br>patient monitoring sessions at once from their mobile device. The near-real<br>time updates to patient monitoring sessions includes non-invasive<br>hemodynamic parameter data and the associated physiological alarm<br>notifications, historical trend data and parameter waveform data. |
{6}------------------------------------------------
#### Indications for HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module The HemoSphere Advanced Monitor when used with the HemoSphere Use: Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. When used in conjunction with a HemoSphere Pressure Cable, CO20s, CI20s, SV20s and SVI20s can be monitored. Use of HemoSphere Swan-Ganz Module in conjunction with HemoSphere Pressure Cable is restricted to adult patients only. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment.
Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
HemoSphere Advanced Monitor with HemoSphere Oximetry Cable The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Pressure Cable The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the
balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac. Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions
{7}------------------------------------------------
should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere ForeSight Module and HemoSphere Tissue Oximetry Module
The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
- . When used with large sensor is indicated for use on adults and transitional adolescents >40 kg.
- When used with Medium Sensors, the FORE-SIGHT ELITE tissue ● oximeter module is indicated for use on pediatric subjects >3 kg.
- . When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere ClearSight Module The HemoSphere Advanced Monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
Refer to the ClearSight finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
Intended Use: The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a
{8}------------------------------------------------
hospital setting. The HemoSphere Advanced Monitoring Platform can be used with the Viewfinder Remote mobile application for supplemental near real-time remote display of tissue oximetry and non-invasive hemodynamic parameter data monitored with a FORE-SIGHT ELITE oximeter module and HemoSphere ClearSight Module, respectively.
The HemoSphere Advanced Monitoring Platform is intended for use with the Edwards Swan-Ganz and Oximetry Catheters and FloTrac, Acumen IQ, TruWave DPT sensors and ClearSight cuffs.
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Swan-Ganz Module are listed in the table below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------|------------------------------|------------------------|--------------------------------------------------------------------|
| CO | continuous cardiac output | | | |
| sCO | STAT cardiac output | | | |
| CI | continuous cardiac index | | | |
| sCI | STAT cardiac index | | | |
| EDV | right ventricular end diastolic<br>volume | | | |
| sEDV | STAT right ventricular end<br>diastolic volume | | | |
| EDVI | right ventricular end diastolic<br>volume index | | | |
| sEDVI | STAT right ventricular end<br>diastolic volume index | | | |
| HRavg | averaged heart rate | | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| LVSWI | left ventricular stroke work index | | | |
| PVR | pulmonary vascular resistance | HemoSphere | | |
| PVRI | pulmonary vascular resistance<br>index | Swan-Ganz<br>Module | | |
| RVEF | right ventricular ejection fraction | | | |
| sRVEF | STAT right ventricular ejection<br>fraction | | | |
| RVSWI | right ventricular stroke work index | | | |
| SV | stroke volume | | | |
| SVI | stroke volume index | | | |
| SVR | systemic vascular resistance | | | |
| SVRI | systemic vascular resistance index | | | |
| iCO | intermittent cardiac output | | Adult and<br>Pediatric | |
| iCI | intermittent cardiac index | | | |
| iSVR | intermittent systemic vascular<br>resistance | | | |
| iSVRI | intermittent systemic vascular<br>resistance index | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere oximetry cable are as listed below:
{9}------------------------------------------------
| Parameter | Description | Sub-System Module Used | Patient Population | Hospital Environment |
|-----------|----------------------------------|---------------------------|---------------------|-----------------------------------------------------|
| SvO2 | Mixed Venous Oxygen Saturation | HemoSphere Oximetry Cable | Adult and Pediatric | Operating Room, Intensive Care Unit, Emergency Room |
| ScvO2 | Central Venous Oxygen Saturation | | | |
A comprehensive list of additional parameters that are available for adult and pediatric patient populations on the HemoSphere Advanced Monitor and a connected HemoSphere Swan-Ganz Module and a connected HemoSphere Oximetry Cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------|--------------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | | | |
| VO2 | Oxygen Consumption | HemoSphere<br>Swan-Ganz<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult and<br>Pediatric | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|---------------------------------|-----------------------|---------------------------------|
| CO/<br>CI | Continuous Cardiac Output¹/<br>Continuous Cardiac Index¹ | | | |
| CVP | Central Venous Pressure | | | |
| DIA | Systemic arterial diastolic blood<br>pressure | | | |
| DIAPAP | pulmonary artery diastolic blood<br>pressure | | | |
| dP/dt | Systolic slope² | | | |
| Eadyn | Dynamic Arterial Elastance² | | | Operating<br>Room,<br>Intensive |
| MAP | Mean Arterial Pressure | | | |
| MPAP | Mean Pulmonary Arterial¹ Pressure | HemoSphere<br>Pressure<br>Cable | Adult only | |
| PPV | pulse pressure variation¹ | | | Care Unit,<br>Emergency<br>Room |
| PRART | Pulse rate | | | |
| SV/<br>SVI | Stroke Volume¹/<br>Stroke Volume Index¹ | | | |
| SVR/<br>SVRI | Systemic Vascular Resistance¹/<br>Systemic Vascular Resistance¹<br>Index | | | |
| SVV | Stroke Volume Variation¹ | | | |
| SYS | Systolic Blood Pressure | | | |
| HPI | Acumen Hypotension Prediction<br>Index | | | |
| ¹FloTrac parameters are available when using a FloTrac/Acumen IQ sensor and if the FloTrac<br>feature is enabled. | | | | |
| ²HPI parameters are available when using an Acumen IQ sensor and if the HPI feature is activated. | | | | |
{10}------------------------------------------------
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable and a connected HemoSphere oximetry cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|--------------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSphere<br>Swan-Ganz<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz Module and pressure cable are listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|-------------------------------|------------------------------|-----------------------|--------------------------------------------------------------------|
| CO20s | 20-second cardiac output | HemoSphere | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| CI20s | 20-second cardiac index | Swan-Ganz | | |
| SV20s | 20-second stroke volume | Module and<br>HemoSphere | | |
| SVI20s | 20-second stroke volume index | Oximetry<br>Cable | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Tissue Oximetry Module, and HemoSphere ForeSight Module are listed below.
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|---------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------|--------------------------------------------------------------------|
| StO2 | Absolute regional hemoglobin<br>oxygen saturation of blood under<br>the sensors | HemoSphere<br>ForeSight<br>Module and<br>HemoSphere<br>Tissue<br>Oximetry<br>Module | Adult and<br>Pediatric | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere ClearSight module are listed below.
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|--------------------------------------------------------|------------------------------------|-----------------------|-------------------------|
| CO/CI | Continuous Cardiac Output/<br>Continuous Cardiac Index | | | Operating<br>Room, |
| DIA | noninvasive arterial diastolic blood<br>pressure | HemoSphere<br>ClearSight<br>Module | Adult<br>only | Intensive Care<br>Unit, |
| MAP | Noninvasive Mean Arterial Pressure | | | Emergency<br>Room |
| PPV | pulse pressure variation | | | |
{11}------------------------------------------------
| PR | Noninvasive Pulse rate |
|------|------------------------------------|
| SV/ | Stroke Volume/ |
| SVI | Stroke Volume Index |
| SVR/ | Systemic Vascular Resistance |
| SVRI | Systemic Vascular Resistance Index |
| SVV | Stroke Volume Variation |
| SYS | Systolic Blood Pressure |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere ClearSight module and oximetry cable are listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------|----------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSphere<br>ClearSight<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
#### Intended use - Viewfinder Remote:
Viewfinder Remote is a mobile application which provides supplemental remote near real-time display of non-invasive hemodynamic parameters measured by a connected HemoSphere advanced monitoring platform. Viewfinder Remote allows clinicians to view continuous monitoring data and alarms/alerts remotely for multiple patients. All displayed data is generated by connected HemoSphere advanced monitoring platforms, and not by Viewfinder Remote. Viewfinder Remote is intended for use by clinicians as a supportive visual aid, and not as a replacement for in-person patient monitoring with connected HemoSphere advanced monitoring platforms.
| Comparison to<br>Predicate<br>Device: | The existing HemoSphere Advanced Monitoring Platform (K201446,<br>cleared October 1, 2020) consists of: |
|---------------------------------------|---------------------------------------------------------------------------------------------------------|
| • | HemoSphere Advanced Monitor |
| • | HemoSphere Swan-Ganz Module |
| • | HemoSphere Oximetry Cable |
| • | HemoSphere Pressure Cable |
| • | HemoSphere Tissue Oximetry Module |
- HemoSphere ForeSight Module ● ● HemoSphere ClearSight Module
- Acumen Hypotension Prediction Index feature (when used with Acumen IQ sensor only)
{12}------------------------------------------------
As part of its connectivity development program, Edwards has modified the HemoSphere Advanced Monitor (subject of this 510(k)) to allow it to wirelessly stream continuously monitored non-invasive hemodynamic parameter data to the Viewfinder Remote (subject of this 510(k)), a mobile device-based application, for remote viewing of the information.
As a connected device, cybersecurity modifications have been implemented on the HemoSphere Advanced Monitor. Additionally, workflow-based updates to support the functionality of remote viewing, including the ability to end a session, provide additional metadata support, as well as multiple setup wizards for initial configuration have been made. The following predicate is used to establish substantial equivalence:
- Primary Predicate: HemoSphere Advanced Monitor (K201446, cleared October 1, 2020) utilized for substantial equivalence to the HemoSphere Advanced Monitor in terms of the graphical user interface (GUI) used, wireless module incorporated, device modularity and basic device functionality. Modifications are being made to the HemoSphere Advanced Monitor to add the capability for it to remotely connect to the Viewfinder Remote mobile application.
Per FDA's guidance, "Medical Device Accessories – Describing Accessories and Classification Pathways" the remote view application (Viewfinder Remote) is considered an accessory to the HemoSphere Advanced Monitor since it augments the performance of the HemoSphere Advanced Monitor by allowing convenience of use (remote display of noninvasive hemodynamic parameters, alarms and alerts) of the device. Hence, the primary predicate for this application is the HemoSphere Advanced Monitor (K201446, cleared October 1, 2020).
Verification and validation testing were performed to compare the performance and functionality of the HemoSphere Advanced Monitoring Platform to its predicate device. Testing included a side-by-side comparison of the output parameters using a bench test.
| Performance<br>Data (Bench<br>and/or Clinical): | The following verification activities were performed in support of a<br>substantial equivalence determination for the modifications being made as<br>part of this submission. |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## System Verification
End-to-end system verification was performed to ensure data integrity and accuracy from the Monitor to the Remote View mobile application. All tests passed.
{13}------------------------------------------------
## Electrical Safety and Electromagnetic Compatibility (EMC)
The HemoSphere Advanced Monitor and all its connected modules/cables were tested to the following standards: IEC 60601-1. IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366, IEC 60601-2-34 and IEC 60601-2-49. All tests passed.
## Wireless Coexistence Testing
ANSI C63.27/D1.0, bench and simulated environment testing were performed on the entire HemoSphere Advanced Monitoring Platform. including all sub-system modules and its interfacing analog inputs and outputs and the wirelessly connected Viewfinder Remote mobile application running on a mobile device. All tests passed.
## Software Verification
The HemoSphere Advanced Monitor is considered as software of Major Level of Concern. The Viewfinder Remote mobile application is considered as software of Moderate Level of Concern.
Software verification was performed per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". All tests passed.
## Usability Study
A usability study was performed for the HemoSphere Advanced Monitoring Platform and Viewfinder Remote in accordance with FDA's guidance, "Applying Human Factors and Usability Engineering to Medical Devices". All tests Passed.
## Clinical Performance
Clinical data was not required for this device or the modifications being made to the device.
#### Non-Clinical Performance Conclusions:
Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications. Verification activities performed confirmed that the differences in the design and materials used did not adversely affect the safety and effectiveness of the subject device.
#### Conclusions Overall Conclusion:
The nonclinical and clinical tests demonstrate that the HemoSphere Advanced Monitor and the Viewfinder Remote mobile application are substantially equivalent to their legally marketed predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.