K210425 · Medtronic Sofamor Danek USA, Inc. · MAX · May 24, 2021 · Orthopedic
Device Facts
Record ID
K210425
Device Name
Catalyft PL Expandable Interbody System
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
MAX · Orthopedic
Decision Date
May 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Catalyft™ PL Expandable Interbody System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD - defined by discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). These DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. Additionally, the Catalyft™ PL Expandable Interbody System can be used with spinal deformities as an adjunct to fusion. These patients should be skeletally mature and have undergone 6 months of non-operative treatment prior to surgery. Implants are used to facilitate fusion in the lumbar spine using autogenous bone graft and/or allograft bone graft comprised of cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate. These implants are intended for use with supplemental internal fixation systems.
Device Story
System comprises expandable titanium alloy (Ti-6Al-4V ELI) interbody fusion implants (inline 'PL' or hockey-stick 'PL40' shapes) and navigated surgical instruments; used in lumbar spine surgery. Implants feature hollow centers for bone graft packing; expandable design allows restoration of lordosis and anatomical sizing. Navigated instruments integrate with StealthStation™ system to assist surgeons in precise anatomical localization during open or minimally invasive procedures via CT/MR-based models, fluoroscopy, or digitized landmarks. Surgeon uses navigated instruments for disc preparation and implant placement; output visualized on StealthStation™ display. Device facilitates spinal fusion; benefits include improved anatomical restoration and surgical precision.
Clinical Evidence
Bench testing only. Implants tested per ASTM F2077 (compression/shear fatigue, static compression), ASTM F2267 (subsidence), and ASTM Draft F-04.25.02.02 (expulsion). Navigated instruments underwent verification and validation for compatibility with the StealthStation™ system.
Indicated for skeletally mature patients with lumbar (L2-S1) degenerative disc disease (DDD) or spinal deformities requiring fusion. Patients may have up to Grade 1 spondylolisthesis or retrolisthesis. Requires 6 months prior non-operative treatment. Used with autogenous/allograft bone graft and supplemental internal fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 24, 2021
Medtronic Sofamor Danek, USA Inc. Alex Underberg Regulatory Affairs Specialist 1800 Pvramid Place Memphis, Tennessee 38132
Re: K210425
Trade/Device Name: Catalyft™ PL Expandable Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OLO Dated: April 23, 2021 Received: April 26, 2021
Dear Alex Underberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210425
### Device Name
Catalyft(TM) PL Expandable Interbody System
#### Indications for Use (Describe)
The Catalyfi™ PL Expandable Interbody System is indicated for use as an intervertebral body fusion device in sketetally mature patients with degenerative disc disease (DDD - defined by discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). These DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. Additionally, the Catalyft™ PL Expandable Interbody System can be used with spinal deformities as an adjunct to fusion. These patients should be skeletally mature and have undergone 6 months of non-operative treatment prior to surgery. Implants are used to facilitate fusion in the lumbar spine using autogenous bone graft and/or allograft bone graft comprised of cancellous and/or corticocancellous bone, and/or demineralized allograft bone marrow aspirate. These implants are intended for use with supplemental internal fixation systems.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K210425
#### Device Name
Navigated Instruments Compatible with StealthStation™ System
Indications for Use (Describe)
Medtronic Navigated Surgical Instruments are intended to be used during preparation and placement of Medtronic implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. Medtronic Navigated Reusable Instruments can be compatible with various Medtronic spinal implant systems.
Navigated surgical instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. An instrument may incorporate a measuring function, which has uses as described on the label and the instruments.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
February 10th, 2021
| Submitter | Medtronic Sofamor Danek, USA Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133 |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Alex Underberg<br>Regulatory Affairs Specialist<br>Telephone: (901) 344-1443 (Direct) |
| Date Prepared | February 10th, 2021 |
| Name of Device | Catalyft™ PL Expandable Interbody System |
| Common Name | Interbody Cages; Navigated Instruments |
| Classification Name | Intervertebral fusion Device with Bone Graft,<br>Lumbar (21 CFR 888.3080);<br>Stereotaxic Instruments (21 CFR 882.4560) |
| Regulatory Classification | Class II |
| Product Codes | MAX, OLO |
| Predicate Devices | Predicate 1 (Primary Predicate): Artic-L™ 3D Ti<br>Spinal System and Artic-XL™ 3D Ti Spinal System<br>(K171689, S.E. 10/05/2017; K190959, S.E.<br>07/03/2019)<br><br>Predicate 2: Elevate™ Spinal System (K142559,<br>S.E. 06/09/2015)<br><br>Predicate 3: Crescent™ Spinal System (K094025,<br>S.E. 04/26/2010; K133216, S.E. 11/22/2013;<br>K171031, S.E. 07/06/2017)<br><br>Predicate 4: Clydesdale™ Spinal System<br>(K100175, S.E. 06/02/2010; K132897, S.E.<br>12/11/2013; K133577, 09/26/2014)<br><br>Medtronic Navigated Manual Reusable<br>Instruments for Use with StealthStation™<br>System<br><br>Navigated Elevate™ Inserter (K163581, S.E.<br>04/14/2017) |
| Description of Devices | Navigated Capstone™ Trials (K131425, S.E. 08/14/2013; K150231, S.E. 06/16/2015) |
| | Navigated Disc Prep Instruments (K150231, S.E. 06/16/2015, K163581, S.E. 04/14/2017; K201267, S.E. 08/26/2020) |
| | The Catalyft™ PL Expandable Interbody System consists of implants, instruments, and cases, trays, and lids. |
| | The subject implants are expandable lordotic Titanium Alloy (Ti-6Al-4V ELI) interbody fusion implants that are provided in either an inline straight tip, known as the “PL” implant, or a hockey stick-shaped tip, which is known as the “PL40” implant. The implants are expandable and have varying footprints, heights, and lengths that provide surgeons the ability to have better control of the restoration of lordosis in patients and allows more ability to appropriately size the interbody to match patient anatomy. |
| | The subject implants are designed with a hollow center region to house autogenous bone graft and/or allograft bone graft comprised of cancellous and/or corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate |
| | In addition to implants, navigated instruments compatible with the StealthStation™ system have been developed. The subject StealthStation™ System compatible instruments are specific to the subject implants, and there are no changes to the StealthStation™ software related to the stereotaxic instruments in this submission. |
| | Previously cleared Navigated Rotating Shavers (K150231, S.E. 06/16/2015, K163581, S.E. 04/14/2017; K201267, S.E. 08/26/2020) will be used as disc prep instruments and trials for the Catalyft™ PL system. |
| Indications for Use | Cases, trays, and lids have been developed for<br>transportation of the subject instruments.<br><br><b>Catalyft™ PL Expandable Interbody System</b><br>The Catalyft™ PL Expandable Interbody System is<br>indicated for use as an intervertebral body fusion<br>device in skeletally mature patients with<br>degenerative disc disease (DDD-defined by<br>discogenic back pain with generation of the disc<br>confirmed by patient history and radiographic<br>studies) at one or two contiguous levels of the<br>lumbar spine (L2-S1). These DDD patients may<br>also have up to Grade 1 spondylolisthesis or<br>retrolisthesis at the involved levels. Additionally,<br>the Catalyft™ PL Expandable Interbody System<br>can be used with patients diagnosed with spinal<br>deformities as an adjunct to fusion. These<br>patients should be skeletally mature and have<br>undergone 6 months of non-operative treatment<br>prior to surgery. Implants are used to facilitate<br>fusion in the lumbar spine using autogenous bone<br>graft and/or allograft bone graft comprised of<br>cancellous and/or corticocancellous bone, and/or<br>demineralized allograft bone with bone marrow<br>aspirate. These implants are intended for use with<br>supplemental internal fixation systems.<br><br><b>Medtronic Navigated Manual Reusable<br/>Instruments for Use with StealthStation™<br/>System</b><br>Medtronic Navigated Surgical Instruments are<br>intended to be used during preparation and<br>placement of Medtronic implants during spinal<br>surgery to assist the surgeon in precisely locating<br>anatomical structures in either open or minimally<br>invasive procedures. Medtronic Navigated<br>Reusable Instruments can be compatible with<br>various Medtronic spinal implant systems.<br>Navigated surgical instruments are specifically<br>designed for use with the StealthStation™<br>System, which is indicated for any medical<br>condition in which the use of stereotactic<br>surgery may be appropriate, and where<br>reference to a rigid anatomical structure, such |
| | identified relative to a CT or MR-based model,<br>fluoroscopy images, or digitized landmarks of |
| | the anatomy. An instrument may incorporate a |
| | measuring function, which has uses as |
| | described on the label and the instruments. |
| Comparison of Technological<br>Characteristics with the Predicate<br>Devices | Catalyft™ PL Expandable Interbody System has<br>the same fundamental scientific technology,<br>indications for use, intended use, design, and<br>material levels of attachment as the predicate<br>devices. The predicate and subject devices are<br>intended for stabilization use and to promote<br>bone fusion during the normal healing process<br>following surgical correction of disorders of the<br>spine. |
| | The subject device has a similar expansion<br>mechanism to that of the predicate Elevate™<br>Spinal System device. However, the Catalyft™ PL<br>implant's bottom endplate moves during<br>expansion, whereas the Elevate™ implant's<br>bottom endplate remains in a fixed position. |
| | Navigated instruments are non-sterile, reusable<br>instruments. The navigated disc prep instruments<br>are existing devices that are not undergoing any<br>design changes. The navigated disc prep<br>instruments are being extended to be used as disc<br>prep instruments and trials for Catalyft™ PL<br>Expandable Interbody System. |
| Performance Data | Testing on implants was completed in accordance<br>with ASTM 2077 (Compression Fatigue,<br>Compression-Shear Fatigue, Static Compression,<br>and Static Compression-Shear), ASTM 2267<br>(Subsidence), and ASTM Draft Standard F-<br>04.25.02.02 (Expulsion). Testing was also<br>completed to ensure functionality and<br>compatibility of new/existing instruments. |
| | Navigated Instruments underwent verification<br>and validation activities per the Bench<br>Performance Testing section and are compatible<br>to be used with Catalyft™ PL Expandable<br>Interbody System and StealthStation™ System. |
| Conclusion | |
| | Based on the information contained in this<br>submission, Medtronic believes that the subject<br>devices are substantially equivalent to the<br>predicate devices. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.