K142559 · Medtronic Sofamor Danek USA, Inc. · MAX · Jun 9, 2015 · Orthopedic
Device Facts
Record ID
K142559
Device Name
ELEVATE Spinal System
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jun 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ELEVATE™ Spinal System Expandable Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Device Story
ELEVATE™ Spinal System consists of expandable interbody cages made of PEEK with tantalum markers and titanium alloy; designed for lumbar or lumbosacral intervertebral body fusion. Implants feature hollow geometry for packing with autogenous bone graft; bullet-nose design facilitates insertion. Used in OR by surgeons via open or minimally invasive posterior approach. Requires supplemental fixation instrumentation cleared for lumbar spine. Provides structural support and correction to vertebral bodies during fusion process. Benefits patients by stabilizing spinal segments and promoting fusion in DDD cases.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by mechanical testing (static/fatigue compression, compression-shear, subsidence) per ASTM F2077-11 and ASTM F2267-4, and MR-safety evaluation per ASTM F2052, F2213, F2119, and F2182.
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, with or without Grade 1 Spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other, representing people. The faces are rendered in a simple, abstract style.
June 9, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medtronic Sofamor Danek USA, Incorporated Ms. Becky Ronner Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K142559
Trade/Device Name: ELEVATE™ Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: May 8, 2015 Received: May 11, 2015
Dear Ms. Ronner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Becky Ronner
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
K142559
Page 1 of 1
510(k) Number (if known) K142559
Device Name ELEVATE™ Spinal System
### Indications for Use (Describe)
The ELEVATE™ Spinal System Expandable Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## MEDTRONIC Sofamor Danek ELEVATE™ Spinal System
## May 2015
| I. | Submitter | Medtronic Sofamor Danek, USA Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901)396-3133<br>Fax: (901) 346-9738 |
|------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact: | Becky Ronner<br>Senior Regulatory Affairs Specialist<br>Direct Telephone: (901)399-2757 |
| | Date Prepared: | September 5, 2014 |
| II. | Device | |
| | Name of Device: | ELEVATE™ Spinal System |
| | Classification Names: | Intervertebral Body Fusion Device<br>(21 CFR 888.3080) |
| | Class: | II |
| | Product Code: | MAX |
| III. | Predicate Devices: | Globus CALIBER® K102293(Primary)<br>(S.E. January 5th, 2011)<br>CAPSTONE® Spinal System K073291 & K123027<br>(S.E. April 24th, 2008 & July 25th, 2013)<br>CRESCENT® Spinal System K094025<br>(S.E. April 26th, 2010)<br>Globus PATRIOT® K072970<br>(S.E. January 18th, 2008)<br>SOVEREIGN® Spinal System K122037<br>(S.E. March 22nd, 2013) |
{4}------------------------------------------------
Globus RISE® Spacer K113447 (S.E. January 26, 2012) WAVE™ D Cage K083626 & K121333 (S.E. October 17, 2009 & S.E. June 29, 2012) The predicates have not been subject to a design related recall.
#### IV. Description:
The ELEVATE™ Spinal System consists of polyetheretherketone (PEEK) with tantalum markers and titanium expandable cages of various length and heights, which can be surgically implanted between two lumbar or lumbarsacral vertebral bodies to give support and correction during lumbar intervertebral body fusion. The hollow geometry of the implants allows them to be packed with autogenous bone graft. This system also includes stainless steel and silicone instruments used to facilitate the implantation of the subject cages. Finally, the subject system includes case, trays, and lids used to package, ship and can be used to sterilize the non-sterile instruments in the subject system. The case, trays, and lids are manufactured from aluminum with either silicone or nylon coated brackets.
The ELEVATE™ Spinal System will be available in similar sizes as the predicate systems.
#### V. Indications for Use:
The ELEVATE™ Spinal System Expandable Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
{5}------------------------------------------------
#### VI. Comparison of Technological Characteristics with the Predicate Devices:
The ELEVATE™ Spinal System has the same fundamental technology, PEEK material with tantalum markers as the predicate devices. The predicate and subject devices are intended to be surgically implanted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar intervertebral body fusion. The subject and predicates implants include a bullet nose and were designed to contain graft material to facilitate a fusion between two vertebral bodies.
- Globus CALIBER® Spinal System (Primary) .
- CAPSTONE® Spinal System o
- CRESCENT® PEEK Spinal System ●
- Globus PATRIOT® Spinal System o
- SOVEREIGN® Spinal System ●
- Globus RISE® Spacer ●
- WAVE™ D Cage o
Additionally, like Globus CALIBER® Spinal System, the subject ELEVATE™ Spinal System is an expandable interbody device design and includes titanium alloy.
#### VII. Performance Data:
The following performance data were provided in support of substantial equivalence.
## Biocompatibility
The biocompatibility evaluation for the ELEVATE™ Spinal System devices was conducted in accordance with FDA's Draft Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" issued April, 23, 2013
The subject ELEVATE™ Spinal System implants are permanent implants and will be classified as permanent , >30 day body contact according to with FDA's Draft Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The subject implants are manufactured from identical materials as the predicate devices, in accordance with the following ASTM standards:
- ASTM F2026: Standard Specifications for Poletheretherketone Polymers . for Surgical Implant Applications
{6}------------------------------------------------
- ASTM F 560: Standard Specification for Unalloyed Tantalum for Surgical ● Implant Applications
- ASTM F136: Standard Specification for Wrought Ti-6Al-4V ELI Alloy . for Surgical Implant
The ELEVATE™ Spinal System instruments are external communicating devices and are classified as limited, up to 24 hours of body contact according to with FDA's Draft Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". These instruments are manufactured from the same medical grade stainless steel and silicone as the predicate devices in accordance with the following ASTM standards:
- ASTM F899 Standard Specification for Wrought Stainless Steels for . Surgical Instruments
- ASTM A564 Standard Specification for Hot Rolled and Cold Finished . Age Hardening Stainless Steel Bars and Shapes
- ASTM A693 Standard Specification for Precipitation-Hardening Stainless . and Heat Resisting Steel Plate, Sheet and Strip
The case, trays, and lids used to for shipment and sterilization of instruments are manufactured from aluminum with either silicone or nylon coated brackets and are not patient contacting and do not require biocompatiblity testing.
The PEEK, tantalum, titanium alloy, medical grade stainless steel and silicone have a long history of safe and effective use in predicate spinal implants and instruments; based on this it was determined that biocompatibility testing is not required. No biocompatibility testing was conducted on the subject devices.
### Mechanical Testing
In accordance with, Guidance for Industry and FDA Staff – Spinal System 510(k)'s", Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices.
Design verification testing was completed on the subject ELEVATE™ Spinal System in accordance with ASTM F2077-11 "Test Methods for Intervertebral Body Fusion Devices" and ASTM F2267-4 (2011) "Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression". The following testing was completed and tests met the predetermined acceptance criteria.
- . Static Compression
{7}------------------------------------------------
- Compression Fatigue ●
- Static Compression-Shear ●
- Compression-Shear Fatigue o
- Subsidence ●
Additional radiographic verification testing was completed demonstrating the substantial equalivence of the subject device to the predicates included in this submission.
In accordance with the FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" the subject ELEVATE™ Spinal System was evaluated for MR-safety in accordance with the following standards:
- ASTM F2052:2006 "Standard test method for measurement of . magnetically induced displacement force on passive implants in the magnetic resonance environment"
- ASTM F2213:2006 "Standard test method for measurement of ● magnetically induced torque on medical devices in the magnetic resonance environment"
- ASTM F2119:2007(R13) "Standard test method for evaluation of MR ● image artifacts from passive implants"
- ASTM F2182:2002a, 2011, 2011a "Standard test method for 0 measurement of radio frequency induced heating on or near passive implant during magnetic resonance imaging"
#### VIII. Conclusion:
Based on the risk analysis, test results, and additional supporting documentation provided in the pre-market notification, the subject ELEVATE™ Spinal System are as safe and effective as the following predicates:
- o Globus CALIBER® K102293 (S.E. January 5th, 2011)
- CAPSTONE® Spinal System K073291 & K123027 (S.E. April 24th, 2008 o & July 25th, 2013)
- CRESCENT® Spinal System K094025 (S.E. April 26th, 2010) O
- Globus PATRIOT® K072970 (S.E. January 18th, 2008) O
- SOVEREIGN® Spinal System K122037 (S.E March 22nd, 2013) o
- Globus RISE® Spacer K113447 (S.E. January 26, 2012) O
- WAVE™ D Cage K083626 & K121333 (S.E. October 17, 2009 & S.E. O June 29, 2012)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.