ANATOMIC PEEK CERVICAL FUSION SYSTEM CORNERSTONE PSR CERVICAL FUSION SYSTEM PEEK PREVAIL CERVICAL INTERBODY DEVICE CAPST
K122037 · Medtronic Sofamor Danek, Inc. · MAX · Mar 22, 2013 · Orthopedic
Device Facts
Record ID
K122037
Device Name
ANATOMIC PEEK CERVICAL FUSION SYSTEM CORNERSTONE PSR CERVICAL FUSION SYSTEM PEEK PREVAIL CERVICAL INTERBODY DEVICE CAPST
Applicant
Medtronic Sofamor Danek, Inc.
Product Code
MAX · Orthopedic
Decision Date
Mar 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ANATOMIC PEEK CERVICAL FUSION SYSTEM is indicated for cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The ANATOMIC PEEK CERVICAL FUSION SYSTEM is to be used with any cleared supplemental fixation device for cervical. The ANATOMIC PEEK CERVICAL FUSION SYSTEM is also required to be used with autogenous bone graft and is to be implanted via an open, anterior approach.
Device Story
Passive interbody fusion implants (cages) for cervical and lumbar spine; manufactured from PEEK with tantalum radiopaque markers; some models include titanium alloy screws (e.g., PEEK PREVAIL, SOVEREIGN). Implants inserted between vertebral bodies to provide support/correction during fusion; packed with autogenous bone graft. Used in OR by surgeons via open, minimally invasive, lateral, or transforaminal approaches. Devices provide structural support to facilitate fusion; some include internal screw fixation to prevent expulsion. Update provides MRI safety labeling (MR Conditional) per ASTM F2503. No changes to design, materials, or clinical indications.
Clinical Evidence
Bench testing only. Testing performed per FDA guidance 'Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment' including ASTM F2052 (displacement force), ASTM F2213 (torque), ASTM F2119 (image artifacts), ASTM F2182 (RF heating), and ASTM F2503 (marking).
Technological Characteristics
Materials: Implant-grade PEEK, tantalum markers, titanium alloy screws (ZEPHIR), Nitinol locking mechanism. Form factor: Cages of various widths/heights. Connectivity: None (passive implant). Sterilization: Not specified. MR safety: MR Conditional per ASTM F2503.
Indications for Use
Indicated for skeletally mature patients with cervical disc disease (C2-T1) or lumbar degenerative disc disease (L2-S1) with/without Grade 1 spondylolisthesis/retrolisthesis. Requires 6 weeks (cervical) or 6 months (lumbar) non-operative treatment. Contraindications not explicitly listed beyond skeletal maturity requirements.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# 510(k) SUMMARY MEDTRONIC Sofamor Danek MRI Update for PEEK Interbody Fusion Devices July 2012
#### I. Medtronic Sofamor Danek, Company: USA Inc. 1800 Pyramid Place Memphis, Tennessee 38132 (901) 396-3133 II. Contact: Becky Ronner Regulatory Affairs Specialist Telephone: (901) 399-2757 Fax: (901) 346-9738
III. Proprietary Trade Name: Cervical Interbody Fusion Devices ANATOMIC PEEK Cervical Fusion System CORNERSTONE® PSR Cervical Fusion System PEEK PREVAIL® CERVICAL INTERBODY DEVICE
Lumbar Interbody Fusion Devices CAPSTONE® CONTROL Spinal System CLYDESDALE® Spinal System CRESCENT® Spinal System (PEEK) SOVEREIGN® Spinal System TELAMON® PEEK Spinal System
IV. Common &Classification Names: Cervical & Lumbar Interbody Fusion Devices
Class:
II
# MAR 2 2 2013
{1}------------------------------------------------
## Product Code:
| ODP (21 CFR<br>888.3080) | ANATOMIC PEEK Cervical Fusion<br>System<br>CORNERSTONE® PSR Cervical<br>Fusion System |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| OVE (21 CFR<br>888.3080) | PEEK PREVAIL® CERVICAL<br>INTERBODY DEVICE |
| MAX (21 CFR<br>888.3080) | CAPSTONE® CONTROL Spinal<br>System<br>CLYDESDALE® Spinal System<br>CRESCENT® Spinal System<br>(PEEK)<br>TELAMON® PEEK Spinal System |
| OVD (21 CFR<br>888.3080) | SOVEREIGN® Spinal System |
#### V. Description:
# a. ANATOMIC PEEK Cervical Fusion Implants
The ANATOMIC PEEK Cervical Fusion Implants consists of cages of various widths and heights which can be inserted between two cervical vertebral bodies to give support and correction during cervical interbody fusion surgeries. These implants are manufactured from implant grade polyetheretherketone (PEEK). The hollow geometry of the implants allows them to be packed with autogenous bone graft in cervical interbody fusion procedures. The ANATOMIC PEEK devices are intended to be implant via an open anterior approach and used with supplemental fixation and autogenous bone graft.
# b. CORNERSTONE® PSR Cervical Fusion Implants
The CORNERSTONE® PSR Cervical Fusion Implants consist of cages of various widths and heights, which can be inserted between two cervical vertebral bodies to give support and correction during cervical interbody fusion surgeries. These implants are manufactured from implant grade PEEK. The hollow geometry of the implants allows them to be packed with autogenous bone graft in cervical fusion procedures. The CORNERSTONE® PSR device is to be used with supplemental instrumentation and is to be implanted via an open, anterior approach.
{2}------------------------------------------------
# c. PEEK PREVAIL™ Cervical Interbody Device
The PEEK PREVAIL™ Cervical Interbody Device is an intervertebral body fusion device with internal screw fixation. The screws protrude through the interbody portion of the device and stabilize the vertebral body while preventing expulsion of the implant. The implant is "I-Beam" shaped with a 2 screw midline configuration. This device is intended to be radiolucent and the interior space of the product is to be used with autograft.
The PEEK PREVAIL™ Cervical Interbody device implant is manufactured from PEEK and contains tantalum radiopaque markers and a Nitinol screw locking mechanism. The screws used with this device (ZEPHIR® Anterior Cervical Screws) are manufactured from Titanium Alloy.
# d. CAPSTONE® CONTROL Spinal System Implants
The CAPSTONE CONTROL™ Spinal System Implants consist of PEEK cages of various widths and heights which include Tantalum markers. These implants can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The CAPSTONE CONTROL™ Spinal System Implants are to be used with supplemental instrumentation and are to be implanted via an open or minimally invasive, posterior or transforaminal approach.
# e. CLYDESDALE® Spinal System Implants
The CLYDESDALE® Spinal System Implants consist of PEEK cages of various widths and heights, which include Tantalum markers. These devices can be inserted via a lateral approach between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The CLYDESDALE® Spinal System Implants are to be used with supplemental instrumentation.
{3}------------------------------------------------
# f. CRESCENT® Spinal System (PEEK) Implants
The CRESCENT® Spinal System Implants consist of PEEK cages of various widths and heights which include Tantalum markers. These implants can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The implants may be implanted via a posterior, transforaminal or lateral approach and the procedure may be open or minimally invasive. The CRESCENT® Spinal System Implants are to be used with supplemental instrumentation.
# g. SOVEREIGN® Spinal System Implants
The SOVEREIGN® Spinal System Implants are an intervertebral body fusion devices with internal screw fixation. The screws protrude through the interbody portion of the device and stabilize the vertebral body while preventing expulsion of the implant. The implant is lens-shaped with three holes for placement of titanium screws. The SOVEREIGN® Spinal System contains both a fixed and a variable angle screw option. The fixed angle screw option provides an interference fit with the PEEK interbody implant. The variable angle screw option provides a slight clearance between the PEEK interbody implant and the screw which allows for a small amount of variable screw angulation. This system is intended to be radiolucent and the interior space of the product is to be used with autogenous bone graft. The accompanying cover plate is designed to resist screw backout and must be used when the variable angle screws are implanted.
The SOVEREIGN® Spinal System interbody device is manufactured from PEEK and contains tantalum radiopaque markers. The screws used with this device are manufactured from titanium alloy.
{4}------------------------------------------------
### h. TELAMON® PEEK Spinal System Implants
The TELAMON® PEEK Spinal System Implants consist of PEEK cages of various widths and heights, which include Tantalum markers. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. These implants may be implanted via an open or a minimally invasive posterior approach. The TELAMON® PEEK Spinal System Implants are to be used with supplemental instrumentation.
#### VI. Indications for Use:
# a. ANATOMIC PEEK Cervical Fusion Implants
The ANATOMIC PEEK device is indicated for cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The ANATOMIC PEEK device is to be used with supplemental fixation. The ANATOMIC PEEK device is also required to be used with autograft and is to be implanted via an open, anterior approach.
## b. CORNERSTONE® PSR Cervical Fusion Implants
The CORNERSTONE® PSR device is indicated for cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by
{5}------------------------------------------------
radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The CORNERSTONE® PSR device is to be used with supplemental fixation. The CORNERSTONE® PSR device is also required to be used with autograft.
#### PEEK PREVAIL™ Cervical Interbody Device ﻦ
The PEEK PREVAIL™ Cervical Interbody Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The PEEK PREVAIL™ Cervical Interbody Device must be used with internal screw fixation provided by ZEPHIR® Anterior Cervical Screws. The PEEK PREVAIL™ Cervical Interbody Device implants are to be used with autograft and implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
# d. CAPSTONE® CONTROL Spinal System Implants
The CAPSTONE CONTROLTM Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants
{6}------------------------------------------------
are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
#### CLYDESDALE® Spinal System Implants e.
The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with degenerative disc disease (DDD) at one or two contiguous levels from L2 to SI. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment. These implants may be implanted via a minimally invasive lateral approach.
#### f. CRESCENT® Spinal System (PEEK) Implants
The CRESCENT® Spinal System is indicated for interbody fusion with . autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
# g. SOVEREIGN® Spinal System Implants
The SOVEREIGN® Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogenic back pain
{7}------------------------------------------------
with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment. These implants may be implanted via a laparoscopic or an open anterior approach.
The SOVEREIGN® interbody system may be used as a stand-alone device or in conjunction with supplemental fixation.
When used as a stand-alone device, the SOVEREIGN® interbody device is intended to be used with the three titanium alloy fixed or variable angle screws. The accompanying cover plate MUST be used anytime the device is used with any number of variable angle screws. If the physician chooses to use less than three or none of the provided screws, then additional supplemental fixation for use in the lumbar spine must be used to augment stability.
# h. TELAMON® PEEK Spinal System Implants
The TELAMON® PEEK Spinal System is indicated for interbody fusion with autogenous bone graft in patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment.
These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
{8}------------------------------------------------
#### VII. Summary of the Technological Characteristics:
The purpose of this bundled 510(k) application is to provide appropriate MRI safety labeling for the subject devices, while also providing MRI technologists with a method of concluding whether an MRI scan can be performed on the device and specific instructions on how to perform the scan. The systems in this 510(k) submission have been determined to be MR conditional per ASTM F2503 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment.
The labeling that is being proposed has previously been cleared by FDA in submission K I 10063 for SOVEREIGN® Spinal System Implants. All the subject devices within this bundled submission have the same fundamental technology. They are intended to provide correction and stabilization during intervertebral body fusion procedures for treatment of degenerative disc disease. They are all generally manufactured from the same PEEK and tantalum materials and have been designed to be used with autogenous bone graft. There have been no changes to the design, to the material or the indications of the subject devices and are therefore identical to their predicates.
The design/specifications of the subject devices are not being addressed. No changes have occurred for the devices in the subject product families which are considered substantially equivalent to and previously cleared under previous 510(k) submissions such as:
- a. ANATOMIC PEEK Cervical Fusion System K112444 (SE 11/15/2011)
- b. CORNERSTONE® PSR Cervical Fusion System K100214 (SE 06/25/2010), K111264 (SE 10/12/2011)
- c. PEEK PREVAIL® Cervical Interbody Device K073285 (SE 05/15/2008), K094042 (SE 06/30/2010), K113252 (SE 01/17/2012), K030327(SE 02/26/2003)
- d. CAPSTONE® CONTROL Spinal System K120368 (04/09/2012)
{9}------------------------------------------------
- e. CLYDESDALE® Spinal Svstem K113528 (SE 12/20/2011), K100175 (SE 06/02/2010), K083026 (SE 12/29/2008)
- f. CRESCENT® Spinal System (PEEK) K094025 (SE 04/26/2010)
- g. SOVEREIGN® Spinal System K091813 (SE 11/17/2009), K110063 (SE 10/04/2011)
- h. TELAMON® PEEK Spinal System K110562 (SE 11/09/2011)
#### IX. Discussion of Non-Clinical Testing:
In accordance FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" testing has been completed on the worst case implants. The following testing has been completed and provided a determination that the subject devices in this 510(k) submission have been determined to be MR conditional:
- o ASTM F2052 "Standard test method for measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment"
- o ASTM F2213 "Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment"
- o ASTM F2119 "Standard test method for evaluation of MR image artifacts from passive implants"
- ASTM F2182 "Standard test method for measurement of radio frequency o induced heating on or near passive implant during magnetic resonance imaging"
- ASTM F2503 "Standard practice for marking medical devices and other o items for safety in the magnetic resonance environment"
#### X. Conclusion:
Non-clinical testing in accordance with the standards listed above was completed along with a risk analysis. Based on the test results and additional
{10}------------------------------------------------
supporting documentation provided within this pre-market notification, Medtronic believes that the subject devices demonstrate substantial equivalence to the listed predicate device and should be labeled as MR Conditional in accordance with ASTM F2503 – "Standard practice for marking medical devices and other items for safety in the magnetic resonance environment'.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a stylized caduceus symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The caduceus is represented by a triple-stranded helix.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2013
Medtronic Sofamor Danek USA, Incorporated % Ms. Becky Ronner Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K122037
Trade/Device Name: ANATOMIC PEEK CERVICAL FUSION SYSTEM Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class H Product Code: MAX, ODP, OVE, OVD Dated: February 20, 2013 Received: February 21, 2013
Dear Ms. Ronner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{12}------------------------------------------------
## Page 2 - Ms. Becky Ronner
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/12/Picture/6 description: The image shows the name "Erin I.D.Keith" in a stylized font. The letters "I.D.K" are designed with a layered, geometric pattern, giving them a unique visual texture. The rest of the letters are in a bold, sans-serif font, creating a contrast with the central part of the name.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
510(k) Number (if known): K122037
Device Name: ANATOMIC PEEK CERVICAL FUSION SYSTEM
#### Indications for Use:
The ANATOMIC PEEK CERVICAL FUSION SYSTEM is indicated for cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The ANATOMIC PEEK CERVICAL FUSION SYSTEM is to be used with any cleared supplemental fixation device for cervical. The ANATOMIC PEEK CERVICAL FUSION SYSTEM is also required to be used with autogenous bone graft and is to be implanted via an open, anterior approach."
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{14}------------------------------------------------
510(k) Number (if known): K122037
Device Name: CORNERSTONE® PSR Cervical Fusion System
#### Indications for Use:
The CORNERSTONE® PSR device is indicated for cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The CORNERSTONE® PSR device is to be used with supplemental fixation. The CORNERSTONE® PSR device is also required to be used with autograft.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{15}------------------------------------------------
510(k) Number (if known): __K122037
Device Name: PEEKPREVAIL® Cervical Interbody Device
#### Indications for Use:
The PEEK PREVAIL™ Cervical Interbody Device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The PEEK PREVAILTM Cervical Interbody Device must be used with internal screw fixation provided by ZEPHIR® Anterior Cervical Screws. The PEEK PREVAIL™ Cervical Interbody Device implants are to be used with autograft and implanted via an open, anterior approach. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{16}------------------------------------------------
510(k) Number (if known): _K122037
### Device Name: CAPSTONE® CONTROL Spinal System
### Indications for Use:
The CAPSTONE CONTROL™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
{17}------------------------------------------------
510(k) Number (if known): K122037
Device Name: CLYDESDALE® Spinal System
### Indications for Use:
The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment. These implants may be implanted via a minimally invasive lateral approach.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{18}------------------------------------------------
510(k) Number (if known): _K122037
Device Name: CRESCENT® Spinal System (PEEK)
### Indications for Use:
The CRESCENT® Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{19}------------------------------------------------
# 510(k) Number (if known): K122037
## Device Name: SOVEREIGN® Spinal System
### Indications for Use:
The SOVEREIGN® Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a laparoscopic or an open anterior approach.
The SOVEREIGN® interbody system may be used as a stand-alone device or in conjunction with supplemental fixation.
When used as a stand-alone device, the SOVEREIGN® interbody device is intended to be used with the three titanium alloy fixed or variable angle screws. The accompanying cover plate MUST be used anytime the device is used with any number of variable angle screws. If the physician chooses to use less than three or none of the provided screws, then additional supplemental fixation for use in the lumbar spine must be used to augment stability.
Prescription Use x AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{20}------------------------------------------------
510(k) Number (if known): K122037
Device Name: TELAMON® PEEK Spinal System
### Indications for Use:
The TELAMON® PEEK Spinal System is indicated for interbody fusion with autogenous bone graft in patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Prescription Use x (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.