CLYDESDALE(R) SPINAL SYSTEM

K113528 · Medtronic Sofamor Danek USA, Inc. · MAX · Dec 20, 2011 · Orthopedic

Device Facts

Record IDK113528
Device NameCLYDESDALE(R) SPINAL SYSTEM
ApplicantMedtronic Sofamor Danek USA, Inc.
Product CodeMAX · Orthopedic
Decision DateDec 20, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach.

Device Story

The CLYDESDALE® Spinal System is an intervertebral body fusion device consisting of PEEK™ OPTIMA™ LT-1 cages with tantalum markers. The device is implanted between lumbar or lumbosacral vertebral bodies via a minimally invasive lateral approach to provide support and correction during fusion surgery. The hollow geometry of the cage is packed with autogenous bone graft to facilitate fusion. The system requires use with supplemental fixation. It is intended for use by surgeons in a clinical setting to treat patients with discogenic back pain and disc degeneration.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing (engineering theoretical analysis) conducted per ASTM F2077-03 and ASTM F2267-04.

Technological Characteristics

Materials: PEEK™ OPTIMA™ LT-1 with tantalum markers. Form factor: Hollow cage of various widths and heights. Testing standards: ASTM F2077-03 (Test Methods for Intervertebral Body Fusion Devices) and ASTM F2267-04 (Load Induced Subsidence).

Indications for Use

Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 Spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113528 Page 1 of 2 DEC 2 0 2011 # CLYDESDALE® Spinal System 510(k) Summary November 2011 - I. COMPANY: Medtronic Sofamor Danek USA, Inc 1800 Pyramid Place Memphis. Tennessee 38132 II. CONTACT: Becky Ronner Regulatory Affairs Specialist Telephone: (901) 399-2757 Fax: (901) 346-9738 #### III. PROPOSED PROPRIETARY TRADE NAME: CLYDESDALE® Spinal System Intervertebral Body Fusion Device ### IV. CLASSIFICATION NAMES: CLASS: MAX (21 CFR 888.3080) # V. PRODUCT DESCRIPTION: PRODUCT CODE: The CLYDESDALE® Spinal System is intended to help provide support in the intervertebral body space during fusion of vertebral bodies in the lumbar spine. This system is intended to be used with supplemental fixation. II The CLYDESDALE® Spinal System consists of PEEK™ OPTIMA™ LT-1 cages of various widths and heights, which include tantalum markers. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. #### VI. INDICATIONS FOR USE: The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may {1}------------------------------------------------ # Page Zof 2 K113528 also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach. #### VII. Summary of the Technological Characteristics: The purpose of this Special 510(k) submission is to seek clearance to incorporate additional size options into the CLYDESDALE® Spinal System family. The design, material and fundamental technology are the same as the current CLYDESDALE® Spinal System with minor dimensional changes. # VIII. Identification of Legally Marketed Devices: The design features and indications for use for the subject CLYDESDALE® Spinal System are substantially equivalent to the following predicates: - CLYDESDALE® Spinal System K100175 (S.E. 06/02/2010) - · PERIMETER™ Spinal System K090353 (S.E. 09/29/2009) - · CAPSTONE® Spinal System K073291 (S.E. 04/24/08) # IX. Discussion of Non-Clinical Testing: Rationales based on Engineering Theoretical Analysis were completed using predicate testing that were performed in accordance with ASTMF 2077-03 "Test Methods for Intervertebral Body Fusion Devices" and ASTM F2267-04 "Standard Test Method for Measuring Load Induced Subsidence for the Intervertebral Body Fusion Device under Static Axial Compression". Data to support these rationales were provided to demonstrate that the subject devices are substantially equivalent to the predicate devices. # X. Conclusion: A risk analysis was completed. Based on the risk analysis and additional supporting documentation provided in this premarket notification, Medtronic believes the subject devices demonstrate substantial equivalence to listed predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains an image of an eagle. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the edge of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC 2 0 2011 Medtronic Sofamor Danek USA, Inc. % Ms. Becky Ronner 1800 Pyramid Place Memphis, Tennessee 38132 Re: K113528 Trade/Device Name: Clydesdale® Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: November 29, 2011 Received: November 30, 2011 Dear Ms. Ronner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Ms. Becky Ronner comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ely yours, for N. Mello Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 # 510(k) Number (if known): _K || 3528 Device Name: CLYDESDALE® Spinal System #### Indications for Use: The CLYDESDALE® Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE® Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach. AND/OR X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number_KIT3 5 28
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...