CRESCENT SPINAL SYSTEM

K094025 · Medtronic Sofamor Danek · MAX · Apr 26, 2010 · Orthopedic

Device Facts

Record IDK094025
Device NameCRESCENT SPINAL SYSTEM
ApplicantMedtronic Sofamor Danek
Product CodeMAX · Orthopedic
Decision DateApr 26, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CRESCENT™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.

Device Story

CRESCENT™ Spinal System consists of hollow, bullet-nosed, anatomically shaped intervertebral body spacers; features diamond V teeth on superior/inferior surfaces for expulsion resistance; contains axial voids for autogenous bone graft. Implanted via transforaminal or direct lateral approaches in lumbar spine (L2-S1). Used by surgeons in clinical settings; requires supplemental fixation instrumentation cleared for lumbar use. Tantalum markers included for radiographic visualization. Device provides structural support for interbody fusion; benefits patients by stabilizing disc space and facilitating bone growth.

Clinical Evidence

Bench testing only. Mechanical testing performed included static and dynamic axial compression, static compression shear, compression shear fatigue, and subsidence testing in accordance with ASTM 2077-03 and ASTM F2267-04. All acceptance criteria were met.

Technological Characteristics

Materials: Medical Grade PEEK (polyetheretherketone) with Tantalum markers. Dimensions: 7mm-15mm height, 25mm-36mm length. Design: Hollow, bullet-nosed, anatomically shaped with diamond V teeth. Mechanical testing standards: ASTM 2077-03, ASTM F2267-04.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### CRESCENTTM Spinal System 510(k) Summary April 23, 2010 - I. Company: Medtronic Sofamor Danek 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 ## APR 2 6 2010 | Contact: | Jennifer Hackney | |----------|-------------------------------| | | Regulatory Affairs Specialist | #### Proprietary Trade Name: CRESCENT™ Spinal System II. Classification Name: Intervertebral Body Fusion Device (21 CFR 888.3080) III. - IV. Product Code: MAX #### V. Product Description The subject CRESCENT™ Spinal System consists of a variety of hollow intervertebral body spacers featuring a bullet-nosed, anatomically shaped design with axial voids designed to hold autogenous bone graft. The subject devices are designed with diamond V teeth across both superior and inferior surfaces to allow the implant to grip the superior and inferior end plates, thus allowing expulsion resistance. The devices are designed to be implanted through the transforaminal and direct lateral approaches. The devices range from 7mm to 15mm in height and from 25mm to 36mm in length. The devices are manufactured from Medical Grade PEEK (polyetheretherketone). These devices also contain Tantalum markers used for imaging purposes. #### V. Indications The CRESCENT™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine. {1}------------------------------------------------ ### VI. Identification of the Legally Marketed Predicate Devices Used to Claim Substantial Equivalence Documentation was provided which demonstrated that the subject CRESCENT™ Spinal System is substantially equivalent to several interbody cages, including the CAPSTONE® Spinal System (K073291, SE 04/24/08), the LT-CAGE® PEEK Lumbar Tapered Fusion Device (P970015, Medtronic Sofamor Danek, Approved 9/10/03), the RAY® Threaded Fusion Cage (P950019, Stryker, Approved 9/4/03), and the BRANTIGEN I/F CAGE® (DePuy, P960025), as well as to the VERTE-STACK® Spinal System (K052261, SE 10/07/05). #### VII. Brief Discussion of the Non-Clinical Tests Submitted The following mechanical tests of the subject CRESCENT™ Spinal System were performed : - . static axial compression testing in accordance with ASTM 2077-03: - dynamic axial compression testing in accordance with ASTM 2077-03; . - . static compression shear testing in accordance with ASTM 2077-03; - compression shear fatigue testing in accordance with ASTM 2077-03; and . - subsidence testing in accordance with ASTM F2267-04. . #### VIII. Conclusions Drawn from the Non-Clinical Tests Results of mechanical testing indicated that all acceptance criteria were met, demonstrating substantial equivalence to the listed predicate devices. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Medtronic Sofamor Danek % Ms. Jennifer Hackney Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132 # APR 2 6 2010 Re: K094025 Trade/Device Name: CRESCENT™ Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: April 15, 2010 Received: April 16, 2010 Dear Ms. Hackney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Ms. Jennifer Hackney CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you don't oper.its aov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regardions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Barbara Buehrn Mark N. Melke Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): K094025 Device Name: CRESCENT™ Spinal System_ ### Indications for Use: The CRESCENT™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine. × Prescription Use (Part 21 CFR 801 Subpart D) ﺮ Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) K (Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K094025
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...