K210118 · Longnan Renzhong Medical Equipment Co., Ltd. · FLL · Sep 15, 2021 · General Hospital
Device Facts
Record ID
K210118
Device Name
INFRARED BODY THERMOMETER (Model: YK-001)
Applicant
Longnan Renzhong Medical Equipment Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Sep 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
INFRARED BODY THERMOMETER (Model: YK-001) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature using the center of the forehead as the measurement site. The device is indicated for use by people over one month old. It can be used by consumers in household environment and doctors in clinic as reference.
Device Story
Hand-held, battery-powered, non-contact infrared thermometer; detects infrared energy radiated from forehead (3-5cm distance). Thermopile sensor converts infrared energy to electronic signal; signal amplified and converted to digital via MCU AD module. MCU calculates temperature; displays result on LCD with 3-color background. Features: 99-memory recall, °C/°F switchable, auto-shutoff, low battery indicator. Used by consumers (home) or clinicians (clinic) for intermittent temperature monitoring. Provides rapid (approx. 2s) temperature readings to assist in clinical decision-making and patient health assessment.
Clinical Evidence
Clinical accuracy study conducted on 137 subjects across three age groups: >1 month to <1 year, 1-5 years, and >5 years. Febrile subjects comprised 30-50% of each group. Exclusions: inflammation at measurement site, temperature-altering medications. Results confirmed accuracy within ISO 80601-2-56:2017 requirements.
Technological Characteristics
ABS plastic enclosure; thermopile infrared sensor; 3.0V DC (2x AAA batteries). Dimensions: 138x95x40mm; weight: ~90g. Standards: ISO 80601-2-56:2017, ASTM E 1965-98, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11. Biocompatibility per ISO 10993-5/10. Software: IEC 62304 compliant.
Indications for Use
Indicated for intermittent human body temperature measurement via center of forehead. Population: individuals over one month old. Suitable for household and clinical use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 15, 2021
Longnan RenZhong Medical Equipment Co., Ltd. % Iris Fung Regulation Manager SGS-CSTC Standards Technical Services Co., Ltd. 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology Development District Guangzhou, Guangdong, 510000 China
Re: K210118
Trade/Device Name: Infrared Body Thermometer (Model: YK-001) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 6, 2021 Received: August 16, 2021
Dear Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210118
Device Name INFRARED BODY THERMOMETER (Model: YK-001)
Indications for Use (Describe)
INFRARED BODY THERMOMETER (Model: YK-001) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature using the center of the forehead as the measurement site. The device is indicated for use by people over one month old. It can be used by consumers in household environment and doctors in clinic as reference.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### 1. Summary Prepared Date
15 September 2021
#### Submitter Information 2
Sponsor Company Name: Longnan Renzhong Medical Equipment Co., Ltd.
- Address: Information Industry Technology City, Longnan Economic � Development Zone, Longnan, Ganzhou, Jiangxi Province, 341000, China
- Phone: +86-13715212441 �
- Contact Person (including title): Liangping Liao (Manager) �
- � E-mail: In@akkord.com.cn
Application Correspondent: SGS-CSTC Standards Technical Services Co., Ltd.
- Address: 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology � Development District, Guangzhou, Guangdong, 510000, China
- � Contact Person: Ms. Iris Fung
- Tel: +86-20-32136908 �
- Email: Iris.Fung@sgs.com �
#### Subject Device Information ನ
Trade Name: INFRARED BODY THERMOMETER (Model: YK-001) Common Name: Clinical Electronic Thermometer Requlation Name: Clinical Electronic Thermometer Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Regulation Class: II
## 4. Predicate Device Information
Sponsor: Shenzhen Jumper Medical Equipment Co., Ltd. Trade Name: Non-contact Infrared Thermometer 510(k) number: K131243 Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Regulation Class: II
{4}------------------------------------------------
## 5. Device Description
The proposeddevice, INFRARED BODY THERMOMETER, which includes model YK-001 hand-held, reusable, battery powered device, which is intended to detect body temperature from forehead. The proposed device measure temperatures of people by detecting the infrared energy radiated directly from the forehead without physical contact. The distance of the measurement is 3cm~5cm. The proposed device uses a temperature sensor, which can detect the human forehead temperature, object temperature and environment temperature; these temperatures are then transfer to electronic signal and amplified; and then it is transferred to digital signal by AD module in MCU of the proposed device. MCU will calculate the body temperature, and then transfer to screen for display. The device has the following features: About two-second measuring time, measuring Body or Ambient temperature, 99-memory recalls, ℃/ί unit switchable, over range message (Hi/Lo), low battery indication, auto shut-off when the device is idle for 15 seconds. When completes, the results will be displayed on the LCD display screen. The device will display 3 different background colors according to the result.
The powersupply of INFRARED BODY THERMOMETER (YK-001)is 3.0V DC, it is powered by two AAA batteries.
The reference body site of INFRARED BODY THERMOMETER (Model: YK-001) is armpit.
## 6. Indications for Use
INFRARED BODY THERMOMETER (Model: YK-001) is a non-contactinfrared thermometer intended f or the intermittentmeasurement of human body temperature using the center of the forehead as the measurement site. The device is indicated for use by people over one month old. It can be used by consumers in household environment and doctors in clinic as reference.
#### 7. Comparison to Predicate Device
A comparison of key technological characteristics between the subject device and predicate device was listed as below:
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | Longnan Renzhong Medical<br>Equipment Co.,Ltd. | Shenzhen Jumper Medical<br>Equipment Co., Ltd | -- |
| 510 (k) Number | K210118 | K131243 | -- |
| Product Name | INFRARED BODY<br>THERMOMETER | Non-contact<br>Infrared<br>Thermometer | -- |
| Models | YK-001 | JPD-FR100 | -- |
| Intended Use | INFRARED BODY<br>THERMOMETER (Model: YK-001) is<br>a non-contact infrared thermometer<br>intended for the intermittent<br>measurement of human body<br>temperature using the center of the<br>forehead as the measurement site. | The Non-contact<br>Forehead Thermometer<br>is an infrared<br>thermometer intended for<br>the intermittent<br>measurement of human<br>body temperature in<br>people of all ages. | Note 1 |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| | The device is indicated for use by<br>people over one month old. It can<br>be used by consumers in<br>household environment and<br>doctors in clinic as reference. | | |
| Sensor | Infrared Sensor | Infrared Sensor | Identical |
| Principle<br>operation | The proposed device measure<br>temperatures of people by detecting<br>the infrared energy radiated directly<br>from the forehead without physical<br>contact.<br>The proposed device uses a<br>temperature sensor, which can<br>detect the human forehead<br>temperature, object temperature<br>and environment temperature; these<br>temperatures are then transfer to<br>electronic signal and amplified; and<br>then it is transferred to digital signal<br>by AD module in MCU of the<br>proposed device. MCU will calculate<br>the body temperature,<br>and then transfer to screen for<br>display. | The device uses infrared<br>sensor(thermopile)to detect<br>the radiated infrared energy<br>emitted by the object, solid,<br>liquid orgas. The intensity of<br>the emitted energy depends<br>on the temperature of the<br>object and the infrared sensor<br>can recognize it to transfer to<br>the proper electronic signal.<br>The signal can be processed<br>in the subject device to convert<br>to the temperature reading. | Identical |
| Measurement<br>Mode | Forehead measurement Mode | Forehead measurement<br>Mode | Identical |
| Measurement<br>type | Non-contact type | Non-contact type | Identical |
| Measurement<br>Distance | 3-5cm | 1-6 cm | Note 2 |
| Measuring<br>Range | $32°C~42.9°C (89.6°F~109.2°F)$ | $32.2°C~43.3°C(90.0°F~109.9°F)$ | Note 3 |
| Accuracy | Forehead mode: $±0.2°C (±0.4°F)$ , for range $35.0°C ~42.0°C(95°F ~107.6°F)$ ;<br>$±0.3°C(±0.5°F)$ , outside this<br>temperature range | Forehead mode:<br>$±0.2°C (0.4°F)$ | Note 3 |
| Sensor type | Thermopile | Thermopile | Identical |
| High<br>temperature<br>prompt | Yes | Yes | Identical |
| Buzzer | Yes | Yes | Identical |
| Display type | LCD | LCD | Identical |
| Auto power-off<br>while no<br>operation | Yes | Yes | Identical |
| °C/°F<br>switchable | Yes | Yes | Identical |
| Memory | 99 sets | 20 sets | Note 4 |
| Power Supply | DC 3V, 2×AAA Batteries | DC 3V, 2 x AAA Batteries | Identical |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| Operating<br>Conditions | Temperature: +15°C~+40°C<br>(59°F<br>~104°F)<br>Humidity: 15% to 95%<br>non-condensing<br>Atmospheric pressure<br>:<br>70kPa~106kPa | 10~40°C(50°F ~104 °F) RH<br><<br>85% | Note 5 |
| Storage<br>Conditions | Temperature: -20℃~50℃ (-4°F<br>~122 °F)<br>Humidity: ≤95% non-condensing<br>Atmospheric pressure:<br>70kPa~106kPa | -25°C ~ +55°C (-13° F-+131<br>) RH<br>≤ 90% | Note 5 |
| Dimension (Lx<br>W x H) | 138x95x40mm (L* W * H) | 145X60X50 mm | Note 6 |
| Weight | About 90g(without battery) | 180 g | Note 6 |
| Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 | Identical |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 | Identical |
| Performance | Complied with ISO 80601-<br>2- 56:2017 | Complied with ISO 80601-2-56 | Identical |
| | Complied with ASTM E 1965 -98<br>(2016) | Complied with ASTM E 1965 -<br>98<br>(2003) | Identical |
| Components | It composed by a measuring<br>sensor, set buttons, battery<br>compartment, Buzzer, a LCD and<br>a ABS plastic enclosure, and<br>measuring without probe cover. | It composed by a measuring<br>sensor, set buttons, battery<br>compartment, Buzzer, a<br>LCD and a ABS plastic<br>enclosure, and measuring<br>without probe cover. | Identical |
| Biocompatibility | All the patient contacting<br>materials<br>are compliance with ISO 10993-<br>5; ISO10993-10 | All the patient contacting<br>materials are compliance<br>with ISO 10993-5;<br>ISO10993-10 | Identical |
| Materials | ABS for enclosure | ABS for enclosure | Identical |
| Labeling | OTC | OTC | Identical |
{5}------------------------------------------------
{6}------------------------------------------------
## Note 1
The patient population between the subject device and predicate is different. The clinical accuracy test was conducted in accordance with ISO 80601-2-56:2017. The test result complies with the standard. The difference does not raise new safety and effectiveness questions.
## Note 2
Measurement distance of the subject device is 3-5cm whereas the predicate device's is in the range of 1- 6cm. But the performance testresult of subject device shows the accuracy meets the requirements within the distance range. The difference does not raise new safety and effectiveness questions.
## Note 3
The subject device and predicate device have different measurement range, but they have the same accuracy. The measurement range of subject device meet the requirements of ASTM E 1965-98. The difference does not raise new safety and effectiveness questions.
{7}------------------------------------------------
## K210118 Summary
#### Note 4
The memory capacity of predicate device and subject device is different, but the memory function does not affectaccuracy of measurement and does not impact the performance of subject device.
#### Note 5
The operating condition and storage condition are different between the subject device and predicate. The operating condition and storage condition met the requirements of ISO 80601-2-56 and performance test was conducted in accordance with the standard. The difference does not impact the performance of subject device.
#### Note 6
There is minor difference on the size and weight between predicate device and subject device. But they both are portable device; such minor difference would not cause issue of safety and effectiveness.
#### 8. Test Summary
#### Non-clinical test:
INFRARED BODY THERMOMETER conforms to applicable standards that include:
- � ASTM E 1965-98 Standard Specification for Infrared Thermometers for Internittent Determination of Patient Temperature
- � IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- � ISO 10993-1:2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process
- � ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity
- � ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- � IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
- � IEC 60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 80601-2-56:2017/AMD 1:2018 Medical electrical equipment-part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
#### Clinical Data:
The clinical accuracy study for the non-contact thermometer was performed to determine the clinical accuracy. The three groups of subjects being tested were: 1) over 1 month to under 1 year old, 2) children between 1 and 5 years old, and 3) patients over 5 years old.
The study includes 137 subjects. For each age group, theratio of febrile subjects was more than 30% and less than 50%. The study excluded subjects with medical conditions such as inflammation at the
{8}------------------------------------------------
## K210118 Summary
measuring sites and subjects using medications known to affect body temperature. The clinical test results showed that the accuracy of the proposed device is within acceptable scope specified in ISO 81061-2-56.
## 9. Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. Based on the performance testing, comparison and analysis, the subject device INFRARED BODY THERMOMETER (Model: YK-001) is substantially equivalent to the Non-contact Infrared Thermometer cleared under K131243.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.