NON-CONTACT INFRARED THERMOMETER

K131243 · Shenzhen Jumper Medical Equipment Co., Ltd. · FLL · Nov 12, 2013 · General Hospital

Device Facts

Record IDK131243
Device NameNON-CONTACT INFRARED THERMOMETER
ApplicantShenzhen Jumper Medical Equipment Co., Ltd.
Product CodeFLL · General Hospital
Decision DateNov 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The non-contact infrared thermometer, model JPD-FR100, can measure body temperature for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference.

Device Story

Non-contact infrared thermometer (JPD-FR100) measures human body temperature via infrared thermopile sensor; detects radiated energy from patient's forehead at 1-6cm distance. Device processes electronic signals from sensor to calculate temperature; converts readings to oral temperature equivalent. Operated by consumers or clinicians in home or clinical settings. Provides digital temperature display; includes memory for 20 readings. Benefits include non-invasive, rapid temperature assessment without physical contact, reducing cross-contamination risk.

Clinical Evidence

Clinical performance testing conducted in accordance with ASTM E1965 requirements. Results demonstrate the device meets clinical accuracy standards for patient temperature measurement. No specific sample size or statistical metrics provided in summary.

Technological Characteristics

Infrared thermopile sensor; 1-6cm measurement distance; LCD display; powered by two 1.5V AAA batteries. Dimensions: 145x60x50mm; Weight: 180g. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home healthcare), and ASTM E1965. Biocompatibility per ISO 10993-1, -5, -10.

Indications for Use

Indicated for infants and adults for non-contact body temperature measurement. Suitable for use by consumers in home environments and by clinicians in clinical settings.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Version:A/l # VOL 05 510(k) Summary This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR \$807.92. The assigned 510(k) number is: K131243 **NOV 1 2 2013** ## 1.0 Information of Submitter and Correspondent ### Submitter's Information: Shenzhen Jumper Medical Equipment Co., Ltd. Address: 5th Floor, Building No.34, Baoyuan Industrial Zone, Xixiang Street, Baoan District, Shenzhen 518102, Guangdong Province, P.R. China Phone: +86-755-26692192 Fax: Contact Person: Mr. Yi Hui E-mail: info@jumper-medical.com ### Submission correspondent's information: Shenzhen ZYTC Consulting Co., Ltd. Address: 4th Floor, Jinhui Building, Nanhai BLVD, Nanshan District, Shenzhen, Guangdong, China. Contact person: Mr. Field.Fu E-mail: cefda13485@163.com ## 2.0 Device Information | Type of 510(k) submission: | Traditional | |----------------------------|-----------------------------------| | Trade Name: | Non-contact Infrared Thermometer | | Model: | JPD-FR100 | | Classification name: | thermometer, electronic, clinical | | Review Panel: | General Hospital | | Product Code: | FLL | | Device Class: | II | | Regulation Number: | 880.2910 | ## 3.0 Predicate Device Information {2}------------------------------------------------ Sponsor: K-jump Health Co., Ltd. Non-contact Infrared Thermometer, model KI-8280 Device: 510(K) Number: K102947 ## 4.0 Device Description The Non-contact Infrared Thermometer, Model JPD-FR100, uses infrared sensor (thermopile) to detect the radiated infrared energy emitted by the object, solid, liquid or gas. The intensity of the emitted energy depends on the temperature of the object and the infrared sensor can recognize it to transfer to the proper electronic signal. The electronic signal can be processed in the subject device to convert to the temperature reading. Therefore, the subject device is able to measure the temperature of a person by the energy the person emits. The predicated device KI-8280, use the same detection principle to measure the patient's temperature. The subject device also use the focusing design to collect the infrared emitted from nearby area of patient's forehead. This mechanism make the subject device have the ability to detect the forehead temperature in the distance of 1-6cm. The subject device intended to detect the temperature of patients. The temperature is converted to the oral temperature. The temperature reading shown in the device display after measuring is the patient's oral temperature. The compact, small and light-weight design, the Shenzhen Jumper Medical Equipment Co., Ltd. Non-contact Infrared Thermometer, model JPD-FR100, enables to provide safe and reliable result and offers a very good clinical accuracy for human body temperature measurements. # 5.0 Intended Use The non-contact infrared thermometer, model JPD-FR100, can measure body temperature for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference. # 6.0 Performance Summary VOL 05 Page 2 of 5 {3}------------------------------------------------ Version:A/l Non-contact Infrared Thermometer, model JPD-FR100 was conforms to the following standards: - . IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988+A1:1991+A2:1995 - IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic . safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, 2007 - . IEC 60601-1-11:2010, Medical electrical equipment - General requirements for basic safety and essential performance - Part 1-11: Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - . ASTM E1965-98 (2009), Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. - IEC 62304:2006, Medical device software Software life cycle processes. ◆ - ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing � within a risk management process. 2009 - ◆ ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity. 2009 - ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for � imitation and delayed-type hypersensitivity - ISO 14971:2007, Medical devices Application of risk management to medical devices. � # 7.0 Comparison to predicate device and conclusion The subject device is substantially equivalent to predicate devices, K102947, Kl-8280. The substantial equivalence chart is provided as follows: | Characteristics | Subject Device | Predicate Device | |--------------------|------------------------------------------------|-------------------------------------| | Device name | Non-contact Infrared<br>Thermometer | Non-contact Infrared<br>Thermometer | | Model | JPD-FR100 | KI-8280 | | K number | Pending | K102947 | | Manufacturer | Shenzhen Jumper Medical<br>Equipment Co., Ltd. | K-jump Health Co., Ltd. | | Measurement method | Infrared radiation detection | Infrared radiation detection | {4}------------------------------------------------ Shenzhen Jumper Medical Equipment Co., Ltd. VOL_005:001_510K Summary Version:A/1 | Measurement mode | Forehead measure mode | Forehead and surface<br>measure mode | |------------------------|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------| | Measuring range | Forehead temperature mode:<br>$32.2°C-43.3°C (90.0°F-109.9°F)$ ; | Forehead temperature mode:<br>$32.0°C-42.2°C (89.6°F-107.9°F)$ ;<br>Surface temperature mode:<br>$0.0°C-100°C (32°F-212°F)$ ; | | Display resolution | $0.1°C (0.1°F)$ | $0.1°C (0.1°F)$ | | C/F switchable | Yes | Yes | | Measuring accuracy | Forehead temperature mode:<br>$±0.2°C(0.4°F)$ ; | Forehead temperature mode:<br>$±0.2°C(0.4°F)$ ;<br>Surface temperature mode:<br>$±1°C(1.8°F)$ | | Display | LCD display | LCD display | | Measurement distance | 1-6cm | 3-8cm | | Key | Two button (Forehead<br>measure, temperature unit<br>shift) | Two button (On/off, Memory) | | Memory | 20 sets | 10 sets | | Power source | Two 1.5V AAA batteries | Two 1.5V AAA batteries | | Low battery indication | Yes | Yes | | Waterproof | No | No | | Dimension | 145×60×50mm | 138×90×45mm | | Weight | 180g | 125g(including batteries) | | Operating condition | $10°C-40°C (50°F-104°F)$ , <<br>95%RH no-condensing | $10°C-40°C (50°F-104°F)$ , <<br>95%RH no-condensing | # 8.0 Clinical Tests Performed The clinical performance test protocol and data analysis followed the requirements of ASTM E 1965. The test report showed the clinical performance of subject device complied with the requirements of ASTM E1965. It is acceptable to measure patient's temperature. # 9.0 Conclusions Non-contact Infrared Thermometer, Model JPD-FR100, has the same intended use and similar characteristics as the predicate device. Moreover, the subject device demonstrates product safety by successful completion of testing to the IEC 60601-1 standard and {5}------------------------------------------------ Version:A/1 electromagnetic standard IEC 60601-1-2. The performance test demonstrates the JPD-FR100 meets the ASTM E 1965 and concludes that any differences in their characteristics do not raise any safety and effectiveness issues. From the above information we conclude the subject device, JPD-FR100, is substantially equivalent to the predicate devices, KI-8280. # 10.0 Summary prepared date November 11, 2013 {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or branches extending upwards, representing health and well-being. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 12, 2013 Shenzhen Jumper Medical Equipment Company, Limited C/O Mr. Field Fu Shenzhen ZYTC Consulting Company, Limited 40 Floor, Jinhui Building, Nanhai Boulevard Nanshen District, Shenzhen, Guandong CHINA 518052 Re: K131243 Trade/Device Name: Non-contact Infrared Thermometer. Model JPD-FR100 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: October 10, 2013 Received: November 5, 2013 Dear Mr. Fu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Page 2 - Mr. Fu Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.hun for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Image /page/6/Picture/7 description: The image shows a signature followed by the text 'Sincerely yours'. Below the signature, the text 'Tejashri Purohit-Sheth, M.D.' is visible, followed by 'Clinical Deputy Director' and 'DAGRID'. The word 'FOR' is at the bottom right of the image. Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K131243 ### Device Name Non-contact Infrared Thermometer, Model JPD-FR100 ### Indications for Use (Describe) Non-conact Infrared Themometer, model JPD-FR 100, can measure for infants and adults without contact to human body. It can be used by consumers in household environment and doctor in clinic as reference. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. THE ATTACT COLLECT FOR FOR FOR FOR FOR FOR THE SEARNERS AT TOTAL CONTINUE TO THE FORT IS A CATE ・・・ Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/7/Picture/12 description: The image shows the FDA logo. The logo is made up of the letters F, D, and A. The letters are all capitalized and are in a bold, sans-serif font. The letters are arranged in a horizontal line, with the F on the left, the D in the middle, and the A on the right. The logo is black and white. Richard C. Chapman 2013.11.12 15:08:21 -05'00' Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...