Infrared thermometer (Models: YRK-002A, T8, T9, T10, T11) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead at a distance of 2-5cm for people of all ages. The device is reusable for home use and clinical use.
Device Story
Hand-held, battery-powered, non-contact infrared thermometer; measures infrared energy emitted from forehead. Uses thermopile sensor for target reading and thermistor for ambient temperature compensation. User operates device by pressing trigger at 2-5cm distance; measurement completes in 1 second with audible beep. Results displayed on LCD; device includes memory for 30 readings and auto-off feature. Used in home or clinical settings by patients or clinicians. Provides temperature data to assist in clinical decision-making regarding patient health status.
Clinical Evidence
Clinical evaluation performed per ASTM E1965-98 (2016) on 240 subjects across four age groups: 0-3 months, 3 months-1 year, 1-5 years, and >5 years (including neonates 0-28 days). Results demonstrated compliance with ASTM E1965-98 performance requirements.
Technological Characteristics
ABS plastic enclosure; thermopile sensor with integrated thermistor; DC 3V (2x AAA) power. Dimensions: 136x87x40mm; Weight: 88g. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98. Biocompatibility: ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for intermittent measurement of human body temperature from the forehead at a distance of 2-5cm for all ages, including neonates (0-28 days). Suitable for home and clinical use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
Yibin Junxin Electronics Technology Co., Ltd. % Cassie Lee Manager Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Guangzhou, Guangdong 510000 China
Re: K203662
Trade/Device Name: Infrared thermometer, Models: YRK-002A, T8, T9, T10, T11 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 23, 2021 Received: August 27, 2021
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203662
Device Name Infrared thermometer, Models: YRK-002A, T8, T9, T10, T11
Indications for Use (Describe)
Infrared thermometer (Models: YRK-002A, T8, T9, T10, T11) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead at a distance of 2-5cm for people of all ages. The device is reusable for home use and clinical use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary for K203662
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
510(k) Owner's Name: Yibin Junxin Electronics Technology Co., Ltd. Establishment Registration Number: 3017015806 Address: Building B2, No. 136, West Section Of Xinggang Road, Lingang Economic And Technological Development Zone, Yibin Sichuan, China Tel: +86-831-3602202 Phone: +86-13530731921 Contact Person: Chaoze Guan Email: peter@tablet-china.com
## Application Correspondent:
Contact Person: Cassie Lee Share Info (Guangzhou) Medical Consultant Ltd. Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Tel: +86 20 8266 2446 Email: regulatory@share-info.com
#### Date of the summary prepared: September 23, 2021
#### 2. Subject Device Information
Trade Name: Infrared thermometer Models: YRK-002A, T8, T9, T10, T11 Common Name: Clinical Electronic Thermometer Regulation Name: Clinical electronic thermometer Review Panel: General Hospital Product Code: FLL Regulation Number: 21 CFR 880.2910 Regulatory Class: II
#### 3. Predicate Device Information
Trade/Device Name: Infrared Forehead Thermometer Model: JZK-601, JZK-602, JZK-603 510(K) Number: K203707
#### 4. Device Description
Infrared thermometer (Models: YRK-002A, T8, T9, T10, T11) is a hand-held, battery powered, infrared Thermometer that coverts a user's forehead temperature, using the infrared energy emitted in the area around the user's forehead temperature when measure from 2-5 cm of the subject's forehead with no contact.
{4}------------------------------------------------
It uses a thermopile sensor with integrated thermistor for the target reading and a thermistor mounted in the head of the thermometer for ambient temperature readings.
lt composed by a measuring sensor, PCB, 4 buttons, an LCD and an enclosure. The functions of the five models are same. Press the measurement key to turn on the device. The screen of the device lights up. The LCD screen has the power status, measurement mode, temperature unit and measurement temperature. Press the "Setting" key to open the F1: The function of "Temperature Unit" can be set. When measuring body temperature, users need to measure in body mode from 2-5 cm from their forehead. Press the trigger, after 1 second with the sound of "beep", the measurement is completed and the temperature is displayed on the LCD screen. Without any operation, it will close automatically in 17 sec. The difference between these models is only the model's name.
#### 5. Intended Use / Indications for Use
Infrared thermometer (Models: YRK-002A, T8, T9, T10, T11) is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead at a distance of 2-5cm for people of all ages. The device is reusable for home use and clinical use.
### 6. Comparison of indications for use and technological characteristics
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Thermometer is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of Comparison | Subject Device | Predicate Device | Verdict |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Company | Yibin Junxin Electronics Technology Co., Ltd. | ShenZhen ZhengKang Technology Co., Ltd. | -- |
| Trade Name | Infrared thermometer | Infrared Forehead Thermometer | -- |
| Classification Name | Thermometer, Electronic, Clinical | Thermometer, Electronic, Clinical | Same |
| 510(k) Number | K203662 | K203707 | -- |
| Product Code | FLL | FLL | Same |
| Thermometer Type | Infrared Forehead | Infrared Forehead | Same |
| Indications for Use | Infrared thermometer (Models: YRK-002A, T8, T9, T10, T11) is a non-contact infrared | The Infrared forehead thermometer is a non-contact infrared thermometer (Models JZK-601, JZK-602, JZK-603) intended for | Similar Note 1 |
| Elements of Comparison | Subject Device | Predicate Device | Verdict |
| | thermometer intended<br>for the intermittent<br>measurement of human<br>body temperature from<br>forehead at a distance<br>of 2-5cm for people of<br>all ages. The device is<br>reusable for home use<br>and clinical use. | the intermittent<br>measurement of human<br>body temperature from<br>forehead at a distance<br>of 1-5cm for people of<br>all ages. The device is<br>reusable for home use<br>and clinical use. | |
| Prescription/over-the-counter<br>use | OTC | OTC | Same |
| Intended use environment | Home, Clinical | Home, Clinical | Same |
| Measurement<br>method | Infrared radiation<br>detection | Infrared radiation<br>detection | Same |
| Measurement place | Forehead | Forehead | Same |
| Measurement<br>mode | Adjusted | Not public | Different<br>Note 2 |
| Reference Body Site | Oral | Not public | Different<br>Note 2 |
| Measuring range | $32.0°C ~ 42.9°C(89.6°F to 109.2°F)$ | $32.0°C ~ 42.9°C (89.6°Fto 109.2°F)$ | Same |
| Display<br>resolution | 0.1°C/0.1°F | 0.1°C/0.1°F | Same |
| C/F switchable | YES | YES | Same |
| Measuring<br>accuracy | $32°C ~ 39°C: ±0.2°C(89.6°F ~ 102.2°F:±0.4°F)39°C ~ 42.9°C: ±0.3°C(102.2°F ~ 109.2°F:±0.5°F)$ | $32°C ~ 34.9°C: ±0.3°C(±0.5°F)35°C ~ 42°C: ±0.2°C(±0.4°F)42.1°C ~ 42.9°C: ±0.3°C(±0.5°F)$ | Similar<br>Note 2 |
| Measurement<br>distance | 2-5cm | 1-5cm | Similar<br>Note 2 |
| Power Supply | DC3V (2 pieces 1.5V<br>AAA) | JZK-601 and JZK-603:<br>DC 3V (2x AA 1.5V<br>Alkaline batteries)<br>JZK-602: DC 3V (2x<br>AAA 1.5V Alkaline<br>batteries) | Same |
| Dimensions | 136mm(L) x 87mm(W)<br>x 40mm(H) | JZK-601 and JZK-603:<br>154 x 96 x 42mm<br>JZK-602: 143 x 81 x<br>36mm | Similar<br>Note 3 |
| Weight | 88g | JZK-601 and JZK-603: | Similar |
| Elements of Comparison | Subject Device | Predicate Device | Verdict |
| | | 93g<br>JZK-602: 86g | Note 3 |
| Auto Power off | Yes | Yes | Same |
| Memory | 30 sets | 32 sets | Similar<br>Note 3 |
| Display screen | LCD | LCD | Same |
| Operational environmental<br>conditions | Temperature: 10°C ~<br>40°C (50 ~ 104°F)<br>Humidity: ≤85% | Temperature: 10 ~ 40°C<br>(50 ~ 104°F)<br>Humidity: ≤95% | Similar<br>Note 2 |
| Storage environmental<br>conditions | Temperature: -25°C ~<br>70°C (-13 ~ 219.2°F)<br>Humidity: ≤95% | Temperature: -20°C ~<br>60°C (-4 ~ 140°F)<br>Humidity: ≤95% | Similar<br>Note 2 |
| Electric Safety<br>and EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-56<br>ASTM E1965-98 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-56<br>ASTM E1965-98 | Same |
| Materials of construction | ABS plastic | ABS | Same |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same |
{5}------------------------------------------------
{6}------------------------------------------------
#### Discussion:
#### Note 1:
Although there are a few differences in the Indications for Use, the intended user population is the same as the predicate device (K203707), and the measurement distance is within the measurement distance range of the predicate device (K203707), and the measurement positions of the subject device and the predicate device is the same, and they meet the requirement of safety and essential performance standards. The difference between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
#### Note 2:
The "Reference Body Site", "Reference Body Site", "Measuring accuracy", "Measurement distance", "Operational environmental conditions" and "Storage environmental conditions" of subject device is similar to the predicate device, all of them meet the requirement of safety and essential performance standard ISO 80601-2-56 and ASTM E1965-98. The differences between the predicate devices and subject device will not affect the safety and effectiveness of the subject device.
#### Note 3:
The "Dimensions", "Weight" and "Memory" of subject device is similar with predicate devices, both of them meet the requirement of safety and essential performance standard IEC 60601-1,
{7}------------------------------------------------
IEC 60601-1-2 and IEC 60601-1-11. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
## 7. Summary of Non-Clinical Testing
7.1 Non-clinical testing was conducted to verify that the subject devices meet all design specifications, demonstrated safety based on current industry standards, and to demonstrate substantial equivalence to the predicate. The following tests were performed:
A. Electromagnetic Compatibility, Electrical Safety, and Battery Safety:
The subject devices were tested for compliance with the following:
- � ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- � IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- � IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- � ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
B. Software Verification: Software documentation was provided in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.
C. Performance Testing:
- � ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- � ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
## 7.2 Discussion of Clinical Tests Performed
The clinical performance test protocol and data analysis is conducted as a requirement of ASTM E1965-98 (2016). The test report showed the clinical performance of the subject device complied with ASTM E1965-98 (2016).
{8}------------------------------------------------
Clinical tests were conducted on the subject device Model YRK-002A, T8, T9, T10, T11. The clinical tests evaluated 240 of subjects. The proposed thermometer was evaluated in four groups: A1 - 0 up to 3 months to 1 year; B1 -older than 1 year and younger than 5 years; and C - older than 5 years old. Test subjects included patients from the neonatal subgroup patient population (0 – 28 days). The clinical performance test protocol and data analysis were conducted in accordance with the ASTM E1965-98 (2016). The test report showed the clinical performance of the subject device complied with ASTM E1965-98 (2016).
### 8. Final Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness.The subject device Infrared thermometer (Models: YRK-002A, T8, T9, T10, T11) is substantially equivalent to the Infrared Forehead Thermometer, Model: JZK-601, JZK-602, JZK-603 cleared under K203707 with respect to the indications for use, target populations, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.