Med-link Pulse Oximeter is intended for spot checking in measuring and displaying functional arterial oxygen saturation (SpO2) and pulse rate of patients in hospitals, physician's office, clinical settings and home care environment. It's a reusable device intended for adults and pediatrics who are well or poorly perfused.
Device Story
Device is a reusable, non-sterile, finger-worn pulse oximeter for spot-checking SpO2 and pulse rate. Input: dual-wavelength light (660nm red, 905nm infrared) transmitted through finger tissue; photo detector measures light attenuation. Principle: photoplethysmography; calculates SpO2 based on differential absorption of light by oxygenated vs. deoxygenated hemoglobin in pulsating arterial bed. Output: SpO2, pulse rate, pulse waveform, bar graph, and battery status displayed on device screen. Used by clinicians or patients in hospitals, clinics, or home environments. Healthcare providers use output for spot-check assessment of patient oxygenation status; aids in clinical decision-making regarding patient respiratory health. Benefits: non-invasive, portable, rapid assessment of oxygen saturation.
Clinical Evidence
Clinical study conducted on 12 healthy adult volunteers (ages 19-34) with varying skin pigmentation. Subjects underwent induced hypoxia (70-100% SaO2) under stationary conditions. Device SpO2 accuracy compared against arterial blood CO-Oximetry. Results demonstrated an ARMS value of 3% over the 70-100% range. No adverse events reported.
Technological Characteristics
Materials: ABS enclosure, PC lens, silica gel finger pad. Sensing: dual-wavelength (660nm/905nm) LED emitter and photo detector. Power: 2x 1.5V AAA alkaline batteries. Form factor: finger-worn. Connectivity: standalone. Software: moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61, ISO 10993-5/10.
Indications for Use
Indicated for spot-check measurement of SpO2 and pulse rate in adult and pediatric patients, including those who are well or poorly perfused, in hospitals, clinics, physician offices, and home care settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shenzhen Med-link Electronics Tech Co., Ltd Jialing Zhang Regulatory Affairs Specialist 4th and 5th Floor, Building Two, Hualian Industrial Zone Xinshi Community, Dalang Street Shenzhen, Guangdong 518109 China
Re: K202743
Trade/Device Name: Med-Link Temp-pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: January 22, 2021 Received: January 25, 2021
Dear Jialing Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing. Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202743
Device Name Med-link Pulse Oximeter
Indications for Use (Describe)
Med-link Pulse Oximeter is intended for spot checking in measuring and displaying functional arterial oxygen saturation (SpO2) and pulse rate of patients in hospitals, physician's office, clinical settings and home care environment. It's a reusable device intended for adults and pediatrics who are well or poorly perfused.
Type of Use (Select one or both, as applicable)
| <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for MedLinket. The logo is composed of the letter 'M' in orange and the rest of the word 'MedLinket' in blue. The logo is simple and modern.
## 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
Type of submission:Traditional
The assigned 510(k) number is: K202743
### 1. Submitter information
Manufacturer Name: Shenzhen Med-link Electronics Tech Co., Ltd. Address: 4th and 5th Floor, Building Two, Hualian Industrial Zone, Xinshi Community, Dalang Street, Longhua District, 518109 Shenzhen, PEOPLE'S REPUBLIC OF CHINA Tel: 0086-755-61568825 Fax: 0086-755-6112005 5 Establishment Registration Number: 3006636961
### 2. Corres pondent
Jialing Zhang (Regulatory Affairs Specialist, Primary Contact) E-mail: user60@med-linket.com
Yi Liu E-mail: user22@med-linket.com
### 3. Data of Preparation
22nd, Jan. 2021
#### 4. Identification of the Device
Trade Name: Med-link Pulse Oximeter Common Name: pulse oximeter Classification Regulation: 21 CFR 870.2700 Product Code: DQA Class: II Review Panel: Anesthesiology
{4}------------------------------------------------
K202743
### Shenzhen Med-link Electronics Tech Co., Ltd.
### 5. Identification of the Predicate Device
### Table 1 Predicate Device Information
| No. | Device Name | Common<br>Name | Manufacturer | Classification<br>and Code | Classification<br>regulation | 510(k)<br>number |
|-----|-------------------------|--------------------------|------------------------|----------------------------|------------------------------|------------------|
| 1 | Nonin Onyx 3, Model9591 | Finger Pulse<br>Oximeter | Nonin<br>Medical, Inc. | Class II,<br>DQA | 21 CFR<br>870.2700 | K191403 |
#### Intended Use and Indications for Use of the Subject Device 6.
Med-link Pulse Oximeter is intended for spot checking in measuring functional arterial oxygen saturation (SpO2) and pulse rate of patients in hospitals, physician's office, clinical settings and home care environment. It's a reusable device intended for adults and pediatrics who are well or poorly perfused.
### 7. Device Description
Med-Link Pulse Oximeter is to spot-check oxygen saturation in blood (SpO2) and pulse oximeter is used on adults and pediatrics at hospital, clinics, and/or home. The device mainly consists of power supply module, detector and emitter LED, signal collection and processor module, user interface and button control. There is no visual or audio alarm. The device is reusable and non-sterile. The device has AM801. The device contains a dual light source (Red LED and Infrared red LED) and a photo detector. Bone, tissue, pigmentation, and venous vessels normally absorb a constant amount of light over time. The arteriolar bed normally pulsates and absorbs variable amounts of light during the pulsations. The ratio of light absorbed in an oxygen saturation measurement (SpO2). Because a measurement of SpO2 is dependent on light from the device, excessive ambient light can interfere with this measurement. The wavelength of red LED is 660mm and infrated LED is 905nm with maximum optical output power of 15mW.
### 8. Comparison to the Predicate Device
| Item | Proposed Device | Predicate Device | Verdict |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Trade name | Med-link Pulse Oximeter | Nonin Onyx 3, Model9591 Finger<br>Pulse Oximeter | / |
| 510(K) Submitter | Shenzhen Med-link Electronics | Nonin Medical, Inc. | / |
| Item | Proposed Device | Predicate Device | Verdict |
| | Tech Co., Ltd. | | |
| 510(K) Number | K202743 | K191403 | / |
| Classification<br>Regulation | 21 CFR 870.2700 | 21 CFR 870.2700 | Same |
| Classification and<br>Code | Class II,<br>DQA | Class II,<br>DQA | Same |
| Common name | Pulse Oximeter | Finger Pulse Oximeter | Same |
| Type of Use | Prescription | Prescription | Same |
| Intended use | Med-link Pulse Oximeter is intended<br>for spot checking in measuring and<br>displaying functional arterial oxygen<br>saturation (SpO2) and pulse rate of<br>patients in hospitals, physician's<br>office, clinical settings and home<br>care environment. It's a reusable<br>device intended for adults and<br>pediatrics who are well or poorly<br>perfused. | The Nonin® Model 9591 Onyx® 3<br>Finger Pulse Oximeter is a small,<br>lightweight, portable and reusable<br>spot-check device indicated for use<br>in measuring and displaying<br>functional oxygen saturation of<br>aterial hemoglobin(%SpO2) and<br>pulse rate of patients who are well<br>or poorly perfused. The Respiration<br>Rate parameter provides a<br>non-invasive measurement of<br>respiration rate, in breaths per<br>minute.<br>For %SpO2, and pulse rate, the 9591<br>is intended for use in hospitals,<br>clinics, long-term care facilities,<br>skilled nursing facilities, and home<br>healthcare services. It is intended for<br>adult and pediatric patients who are<br>well or poorly perfused, with digits<br>that are between 0.3~1.0<br>inch(0.8~2.5cm) thick, under<br>non-motion conditions.<br>For Respiration rate, the 9591 is<br>intended for use in hospitals, clinics, | Different note<br>1 |
| Item | Proposed Device | Predicate Device | Verdict |
| | | long-term care facilities, skilled<br>nursing facilities, and home<br>healthcare services. It it intended for<br>adult who are well perfused, with<br>digits that are between 0.3~1.0<br>inch (0.8~2.5cm) thick, under non<br>-motion conditions. It is not<br>intended for use in high-acuity<br>environments, such as ICU or<br>operating rooms where continuous<br>monitoring is expected. | |
| Operating<br>Principle | The measurement of PULSE<br>OXIMETER uses a multi-functional<br>oxyhemoglobinometer to transmit<br>some narrow spectrum light bands<br>through blood samples, and to<br>measure attenuation of spectrum<br>with different wavelengths<br>according to the characteristic that<br>RHb, O2Hb, Met Hb and COHb<br>absorb the light of different<br>wavelength, thereby determining<br>O2Hb saturation of different<br>fractions. O2Hb saturation is called<br>"fractional" O2Hb saturation.<br>Present SpO2 oximeter transmits<br>light of two wavelengths only, red<br>light and infrared, to differentiate<br>HbO2 from HbR. One side of the<br>sensor contains two LEDs, and the<br>other side contains a photoelectric<br>detector. SpO2 oximeter measures<br>HbO2 saturation in the blood by the<br>light plethysmograph when the pulse | Pulse oximetry is a non-invasive<br>method that passes red and infrared<br>light through perfused tissue and<br>detects the fluctuating signals<br>caused by arterial pulses.<br>Well-oxygenated blood is bright red,<br>while poorly<br>oxygenated blood is dark red. The<br>pulse oximeter determines<br>functional oxygen saturation of<br>arterial hemoglobin (SpO2) from<br>this color difference by measuring<br>the ratio of absorbed red and<br>infrared light as volume fluctuates<br>with each pulse. Additionally, the<br>pulse oximeter uses variations in the<br>pulse volume fluctuation's<br>amplitude, baseline shift, and timing<br>to determine the respiratory rate. | Different note<br>1 |
| Item | Proposed Device | Predicate Device | Verdict |
| | beats. The result is quite precise<br>when HbO2 saturation is between<br>70% to 100%. | | |
| Application site | Finger | Finger | Same |
| Usage | Reusable | Reusable | Same |
| SpO2 range | 70% to 100% | 0% to 100% SpO2 | Different note<br>2 |
| SpO2 resolution | 1% | Not provided | Different note<br>2 |
| SpO2 accuracy under<br>good perfusion | 90% to 100% range: ±2%;<br>70% to 89% range: ±3%;<br><70%: unspecified. | ±2 digits | Different note<br>2 |
| SpO2 accuracy under<br>low perfusion | 90% to 100% range: ±2%;<br>70% to 89% range: ±3%;<br><70%: unspecified. | ±2 digits | Different note<br>2 |
| Pulse rate range | 30 to 245 bpm | 18~321 BPM | Different note<br>2 |
| Pulse rate resolution | 1 bpm | Not provided | Different note<br>2 |
| Pulse rate accuracy<br>under good perfusion | ±3 bpm | 20 to 250 BPM±3 digits | Different note<br>2 |
| Pulse rate accuracy<br>underlow perfusion | ±3 bpm | 40 to 240 BPM±3 digits | Different note<br>2 |
| Applicable population | Adult and pediatric | Adult and pediatric | Same |
| Measurement<br>wavelength | Red: approximately 660nm;<br>Infrared: approximately 905nm | 660 and 910 nanometers | Different note<br>3 |
| Operation<br>Environment | Temperature:<br>41°F~104°F(5°C~40°C);<br>Humidity: ≤80%; atmospheric<br>pressure: 86kPa~106kPa | Temperature: -5°C~40°C; humidity:<br>10~95% non-condensing; | Different note<br>4 |
| Storage Environment | Temperature<br>14°F~104°F(-10°C~40°C); humidity<br>≤80%; atmospheric pressure:<br>86kPa~106kPa | Temperature: -40°C~70°C;<br>humidity: 10~95% non-condensing; | Different note<br>4 |
| Item | Proposed Device | Predicate Device | Verdict |
| Power requirements | 1.5V (AAA)alkaline batteryX2<br>(IEC Type<br>LR03) | 3V DC | Same |
| Battery life(operating) | 2400 spot checks<br>20 hours continuous | 2000 spot checks<br>25 hours continuous | Different note<br>5 |
| Battery life(storage) | 6 months, with batteries installed | 1 month, with batteries installed | Different note<br>5 |
| User interface | 8 directions for display | Not provided | Different note<br>5 |
| Display parameter | SpO2, Pulse Rate, Pulse Waveform<br>Display, Bar Graph(Pulse<br>Amplitude Indicator) and Battery<br>Indicator | SpO2, Pulse Rate and Respiration<br>Rate | Different note<br>5 |
| Electrical-type and<br>degree of protection | Type BF internally powered | Type BF internally powered | Same |
| Enclosure-degree<br>of<br>ingress<br>protection<br>against liquids | IPX2 | IP32 | Same |
| Measuring Mode | Spot-check…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.