K202132 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · Nov 13, 2020 · Radiology
Device Facts
Record ID
K202132
Device Name
Venue
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
Nov 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Venue is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue is intended to be used in a hospital or medical clinic. Venue clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B. M. PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.
Device Story
Venue is a mobile, general-purpose diagnostic ultrasound system for hospital/clinic use. It acquires, processes, and displays ultrasound images using linear, convex, and phased array transducers. Input includes ultrasound signals and optional ECG traces (synchronized to images) or barcode data. The system features a high-resolution multi-touch LCD monitor and articulated arm for ergonomic positioning. Automated tools simplify workflow for common assessments, including needle guidance (cNeedle), ejection fraction calculation (Real-Time EF), and renal navigation. Output is displayed on-screen, stored via DICOM, or printed. The device assists clinicians in imaging, measurement, and interventional guidance (biopsy, fluid drainage, vascular/non-vascular access). Benefits include improved workflow efficiency and diagnostic visualization for point-of-care settings.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in compliance with standards such as AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, and ISO 10993-1.
Technological Characteristics
Mobile ultrasound system with high-resolution color LCD and multi-touch interface. Employs linear, convex, and phased array transducers. Modes: B, M, PW/CW/Color/Power Doppler, Harmonic Imaging, Coded Pulse. Connectivity: DICOM, wireless, barcode reader. Power: Battery and AC. Biocompatible patient-contact materials. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA PS 3.1-3.20.
Indications for Use
Indicated for qualified and trained healthcare professionals to perform ultrasound imaging, measurement, and analysis of the human body and fluid in hospital or clinic settings. Applications include abdominal, thoracic/pleural, ophthalmic, Fetal/OB, small organ, vascular, neonatal/adult cephalic, pediatric, musculoskeletal, cardiac, transrectal, transvaginal, transesophageal, intraoperative, and interventional guidance.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
November 13, 2020
GE Medical Systems Ultrasound and Primary Care Diagnostics % Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K202132
Trade/Device Name: Venue Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: October 8, 2020 Received: October 13, 2020
Dear Tracey Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K202132
Device Name Venue
#### Indications for Use (Describe)
The Venue is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue is intended to be used in a hospital or medical clinic. Venue clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), VascularPeripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/ PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/6 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized script inside. The circle is surrounded by a decorative swirl pattern, also in blue. The logo is simple and recognizable, representing the multinational conglomerate corporation.
# K202132
GE Healthcare 510(k) Premarket Notification Submission
## 510(k) Summary
| Date: | July 30, 2020 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)470-1003 |
| Secondary Contact Person: | Karin Shimoni<br>Regulatory Affairs Leader<br>GE Healthcare |
| Device Trade Name: | Venue |
| Common/Usual Name: | Diagnostic Ultrasound System |
| Classification Names: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90 |
| Product Code: | IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560-<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| Predicate Device: | Venue (K180599) |
| Reference Devices: | Venue Go (K183362)<br>Versana Premier (K200138)<br>Vivid iq (K181727)<br>LOGIQ e (K133533)<br>LOGIQ e (K151028)<br>Voluson S8 (K120741)<br>Vivid q (K121062)<br>Venue 50 (K152758)<br>Vscan Extend (K180995)<br>Vivid S70N (K182450) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a circular emblem with the letters 'GE' intertwined in the center. The letters are stylized and have a flowing, cursive appearance. The logo is colored in a light blue hue, and the background is white.
Device Description: The proposed Venue system is a general-purpose, Track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement and analysis of the human body and fluid that provides digital acquisition, processing and display capabilities. Venue can be used in offices, clinics and hospitals. The Venue is a mobile system with a small footprint that easily fits into tight spaces and positioned to accommodate the sometimes-awkward work settings of the point of care user. The Venue has a high resolution color LCD monitor, with a simple, multi-touch user interface that makes the system intuitive. The single surface screen provides easy cleanability. Articulated monitor arm enables flexible display positions in order to be accessible and clearly visible in both user-standing and sitting positions.
The proposed Venue has the capability for displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal. The ECG signal can be input directly from the patient leads or as an output from an ECG monitoring device. ECG is not intended for monitoring or diagnosis.
The Venue has a battery that allows for scanning without the need to plug in to an electrical outlet. The system is capable of wireless communication and a barcode reader is available to be used as an input device. System meets DICOM requirements to support users image storage and archiving needs and allows for output to printing devices. The user documentation is available via electronic media.
The Venue utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes. Some transducers are compatible with OEM biopsy kits to support needle-guidance procedures. The system includes several automated tools designed to simplify and shorten the workflow time of the healthcare professional for some common assessments.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a classic and recognizable appearance.
Intended Use: The Venue is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Venue is intended to be used in a hospital or medical clinic. Venue clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, Intraoperative (vascular) and interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B. M. PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD, B/CWD, B/Color/CWD. Technology: The Venue employs the same fundamental scientific technology as its predicate and reference devices. Determination of Comparison to Predicate Device Substantial Equivalence: The Venue system is substantially equivalent to the predicate device with regards to imaging capabilities, technological characteristics and safety and effectiveness. The following is an overview of the differences between the proposed Venue and the predicate Venue. · The proposed Venue and predicate Venue (K180599) have similar clinical indications for use, however the proposed Venue is adding the Intraoperative (vascular) application which has been cleared on reference device, Venue Go (K183362). · Vascular has been added with peripheral vascular, this clarifies vessels and does not change imaging or the product. No impact to safe or effectiveness use. · Imaging guidance of interventional procedures is changed to interventional guidance as it was in the IFU tables in the predicate. Biopsy and non-vascular access (instead of nerve block) are added as sub categories from the predicate IFU table. Fluid drainage is added as it part of intended use. There is no change to the intended use from the predicate, no change to the product and no impact to safe or effective use.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.
## GE Healthcare
510(k) Premarket Notification Submission
- · The IFU statement is updated to add "trained" to the operator qualification for clarity however it is in the proposed and the predicate product documentation. No change to the product and no impact to safe or effective use.
- · Device use settings and modes of operation are added as required in the 2019 ultrasound guidance. There was no change in the system modes or the use environment from the predicate.
Transducers and Modes:
- 0 The proposed Venue and predicate Venue (K180599) have identical imaging modes.
- . The proposed Venue and predicate Venue (K180599) systems transducers are similar, except for:
- o Addition of ML6-15-RS which was first cleared on Voluson S8, K120741. The clinical indications of ML6-15-RS are similar on the proposed Venue as they are on the reference device Vivid iq, K181727.
- o Addition of 12S-RS, which was first cleared on Vivid q, K121062. The clinical indications of 12S-RS are the same on the proposed Venue as they are on the reference device Vivid q, K121062.
- o Addition of L10-22-RS which was first cleared on LOGIQ e, K133533. The clinical indications of L10-22-RS are similar on the proposed Venue as they are on the reference devices LOGIQ e, K133533.
- L4-20t-RS is a new transducer being added to the proposed ● Venue system. The L4-20t-RS has similar clinical indications for use as the predicate L4-12t-RS transducer which was first cleared on LOGIO e K133533. L4-20t-RS is a surface, linear array transducer which operates in a wider range of frequencies compared to the predicate L4-12t-RS. The clinical indications of L4-20t-RS are the same as the L4-12t-RS transducer cleared with reference device Venue Go. K183362.
- New applications are added to transducers: ●
- ML6-15-RS transducer: added Ophthalmic, Cardiac o (Pediatric and Adult), Thoracic/Pleural, Vascular and Non-vascular access, cleared with 12L-RS in predicate Venue (K180599) which is a similar transducer.
- L10-22-RS transducer: added Pediatric, cleared with o L8-18i-RS in predicate Venue (K180599) which is a similar transducer.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has a white border and a few white accents that give the impression of movement or energy around the letters.
## GE Healthcare
510(k) Premarket Notification Submission
- L8-18i-RS transducer: added Ophthalmic, cleared with o 12L-RS in predicate Venue (K180599) which is a similar transducer.
- L8-18i-RS transducer: added intraoperative, cleared o with this probe on reference Venue Go (K183362).
- Removed Vascular Access application from 3Sc-RS, C1-5-● RS and 6S-RS. No impact to safe or effective use.
- Added Coded Pulse mode to Vascular/Peripheral Vascular, ● Vascular Access and Non-vascular on L10-22-RS.
### Features/Functionality:
- Simple Screen is a modification of Full Screen that is on ● predicate Venue, K180599.
- Follow-up feature is being added with minor modifications ● to what has been cleared on Versana Premier, K200138.
- Venue View is being added and is a different name for LOGIQ View that has been cleared on Versana Premier, K200138.
- Lung Sweep allows multiple lung scans to play . simultaneously during review. Lung scanning was available also on predicate Venue, K180599.
- cNeedle enhances the needle, and projects a trajectory line. ● Capabilities are similar to needle recognition that has been cleared on predicate Venue, K180599 and Pinpoint GT feature on reference Venue 50, K152758, respectively.
- Real-Time Ejection Fraction (EF) is being added and is similar to the Auto EF feature on reference Vivid S70N, K182450.
- Renal Diagram is similar to the eFAST Navigation Tool cleared on predicate Venue, K180599.
- Added biopsy guidzones cleared on reference Versana ● Premier, K200138.
- Added additional off-the-shelf SW.
### Accessories:
- Added compatible OEM ECG triggering leads for ● pediatrics, similarly to what is on predicate Vivid iq, K181727.
- Added compatible OEM biopsy guide accessory . compatibility for the C1-5-RS, and for 12L-RS, L12n-RS, & L4-12t-RS transducers.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance.
### Summary of Non-Clinical Tests:
The proposed Venue has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to comply with applicable medical device safety standards. The Venue complies with voluntary standards:
- . AAMI/ANSI ES60601-1, Medical Electrical Equipment -Part 1: General Requirements for Safety, 2005/ A2:2012
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: o General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2014
- IEC 60601-2-37, Medical Electrical Equipment Part 2-37: ● Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015
- ISO 10993-1, Biological Evaluation of Medical Deviceso Part 1: Evaluation and Testing- Third Edition, 2009
- IEC 62359. Ultrasonics Field characterization Test ● methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017
- ISO 14971, Application of risk management to medical devices, 2019
- NEMA PS 3.1 3.20, Digital Imaging and Communications ● in Medicine (DICOM) Set, 2016
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews ●
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification & Validation)
- Safety testing (Verification) .
Transducer material and other patient contact materials are biocompatible.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has a decorative border that resembles swirling water or a stylized floral pattern. The logo is simple, recognizable, and associated with the multinational conglomerate GE.
## Summary of Clinical Tests:
The subject of this premarket submission, Venue, did not require clinical studies to support substantial equivalence.
Conclusion: GE Healthcare considers the Venue to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.