Venue

K180599 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · May 2, 2018 · Radiology

Device Facts

Record IDK180599
Device NameVenue
ApplicantGE Medical Systems Ultrasound and Primary Care Diagnostics
Product CodeIYN · Radiology
Decision DateMay 2, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesAI/ML, Pediatric

Indications for Use

The Venue is a general purpose diagnostic ultrasound system for use by qualified healthcare professionals. The clinical environments where the Venue can be used include critical care and emergency room environments, as well as point-of care areas in offices, clinical and hospital settings for diagnosis of patients. The Venue is intended for ultrasound imaging, measurement and analysis of the human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/Organ (including breast, testes, thyroid). Peripheral vascular, neonatal and adult cephalic, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transvaginal, Transesophageal, and imaging guidance of interventional procedures (e.g. Nerve block, vascular access).

Device Story

Venue is a mobile, general-purpose diagnostic ultrasound system for point-of-care use by clinicians. It acquires, processes, and displays ultrasound images using various linear, convex, and phased array transducers. The system features a high-resolution multi-touch screen and battery operation for portability. It supports standard imaging modes (B, M, Doppler, etc.) and includes automated tools to simplify clinical workflows, such as the eFAST navigation tool and review summary. An optional ECG module allows synchronization of images with ECG traces for specific timing. The device supports DICOM for image storage and archiving. It is used for diagnostic imaging and interventional guidance (e.g., needle/catheter placement). By providing real-time visualization and automated measurements, the device assists clinicians in rapid assessment and decision-making in critical care and emergency environments.

Clinical Evidence

No clinical studies were required to support substantial equivalence. The device was evaluated through non-clinical bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in compliance with standards such as AAMI/ANSI ES60601-1, IEC60601-1-2, and NEMA UD 2.

Technological Characteristics

Mobile ultrasound system with high-resolution color LCD monitor and multi-touch interface. Employs linear, convex, and phased array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, and harmonic imaging. Connectivity includes wireless communication and DICOM support. Biocompatible patient-contact materials. Electrical safety per AAMI/ANSI ES60601-1; electromagnetic compatibility per IEC60601-1-2; acoustic output per NEMA UD 2.

Indications for Use

Indicated for qualified healthcare professionals to perform diagnostic ultrasound imaging, measurement, and analysis of the human body and fluid in critical care, emergency, office, clinical, and hospital settings. Applications include abdominal, thoracic/pleural, ophthalmic, fetal/obstetric, small organ (breast, testes, thyroid), peripheral vascular, neonatal/adult cephalic, pediatric, musculoskeletal, and cardiac imaging, as well as interventional guidance (nerve block, vascular access).

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC GE Healthcare % Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226 Re: K180599 Trade/Device Name: Venue Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: March 7, 2018 Received: March 8, 2018 Dear Tracey Ortiz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. May 2, 2018 {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K180599 Device Name Venue #### Indications for Use (Describe) The Venue is a general purpose diagnostic ultrasound system for use by qualified healthcare professionals. The clinical environments where the Venue can be used include critical care and emergency room environments, as well as point-of care areas in offices, clinical and hospital settings for diagnosis of patients. The Venue is intended for ultrasound imaging, measurement and analysis of the human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/Organ (including breast, testes, thyroid). Peripheral vascular, neonatal and adult cephalic, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transvaginal, Transesophageal, and imaging guidance of interventional procedures (e.g. Nerve block, vascular access). | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are white and are set against a blue circular background. The blue background is surrounded by a white border with three stylized swirls. ### Indications for Use Forms The following forms represent indications with clinical applications and exam types along with the modes of operation for the Venue. Combinations identified "P" represents those previously cleared with another GE Ultrasound system. Combinations identified as "N" are new. The following Indication for Use forms are appended: System: Venue Transducer: 3Sc-RS Transducer: 9L-RS Transducer: C1-5-RS Transducer: 8C-RS Transducer: E8C-RS Transducer: 12L-RS Transducer: L12n-RS Transducer: 6Tc-RS {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular shape. The circle is made up of a series of wave-like lines that create a swirling effect around the letters. The logo is colored in a light blue hue. GE Healthcare 510(k) Premarket Notification Submission | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other | |----------------------------------------------------|----|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------| | Ophthalmic | P | | | | P | | P | | | | | | Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | | | Abdominal [1] | P | P | P | | P | P | P | P | P | P | | | Pediatric | P | P | P | | P | P | P | P | P | P | | | Small Organ [2] | P | P | P | | P | P | P | P | P | P | | | Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | | | Adult Cephalic | P | P | P | | P | P | P | P | P | P | | | Pediatric Cardiac | P | P | P | P | P | P | P2 | P | P | P2 | 6 | | Adult Cardiac | P | P | P | P | P | P | P2 | P | P | P2 | 6 | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | | | Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | P | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | P2 | P2 | P2 | P2 | P2 | P2 | | N | P2 | P2 | 6 | | Transrectal | P | P | P | | P | P | P | P | P | P | | | Transvaginal | P | P | P | | P | P | P | P | P | P | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | P | 5,7 | | Nonvascular [5] | P | P | P | | P | P | P | P | P | P | 5,7 | ### GE Venue Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD; [7] Image guidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [+] Coded pulse is for digitally encoded harmonics; {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a decorative border with wave-like patterns around the edge. # GE Healthcare 510(k) Premarket Notification Submission ### GE Venue with 3Sc-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other | | Ophthalmic | P | | | | P | | P | | | | | | Fetal / Obstetrics | P | P | P | | P | P | P | P | P | | | | Abdominal [1] | P | P | P | | P | P | P | P | P | | | | Pediatric | P1 | P1 | P1 | | P1 | P1 | P1 | P1 | P1 | | | | Small Organ [2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | P | P | P | | P | P | P | P | P | | | | Pediatric Cardiac | P | P | P | P | P | P | P | P | P | | 6 | | Adult Cardiac | P | P | P | P | P | P | P | P | P | | 6 | | Peripheral Vascular | P | P | P | | P | P | P | P | P | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | | 7 | | Nonvascular [5] | | | | | | | | | | | | N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image quidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [ +] Coded pulse is for digitally encoded harmonics; {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the letters, giving it a classic and recognizable appearance. ## GE Venue with 9L-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|----------------|-------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse | Other | | Ophthalmic | N | | | | N | | N | | | | | | Fetal / Obstetrics | N | N | N | | N | N | N | N | N | N | | | Abdominal [1] | P | P | P | P | P | P | P | P | P | P | | | Pediatric | P | P | P | P | P | P | P | P | P | P | | | Small Organ [2] | P | P | P | P | P | P | P | P | P | P | | | Neonatal Cephalic | P2 | P2 | P2 | P2 | P2 | N | P2 | N | P2 | P2 | | | Adult Cephalic | | | | | | | | | | | | | Pediatric Cardiac | N | N | N | N | N | N | N | N | N | N | | | Adult Cardiac | N | N | N | N | N | N | N | N | N | N | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | | | Thoracic/Pleural [3] | P | P | P | P | P | P | P | P | P | P | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | P | P | P | P | P | P | P | P | P | P | 5,7 | | Nonvascular [5] | P | P | P | P | P | P | P | P | P | P | 5,7 | N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image quidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [+] Coded pulse is for digitally encoded harmonics. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the letters, giving it a classic and recognizable appearance. ## GE Healthcare 510(k) Premarket Notification Submission ### GE Venue with C1-5-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics | P | P | P | P | P | P | P | P | P | P | | | Abdominal [1] | P | P | P | P | P | P | P | P | P | P | | | Pediatric | P | P | P | P | P | P | P | P | P | P | | | Small Organ [2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Pediatric Cardiac | N | N | N | N | N | N | N | N | N | N | | | Adult Cardiac | N | N | N | N | N | N | N | N | N | N | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | | | Thoracic/Pleural [3] | P | P | P | P | P | P | P | P | P | P | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | P | P | P | P | P | P | P | P | P | P | 7 | | Nonvascular [5] | P | P | P | P | P | P | P | P | P | P | 7 | N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image quidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [↓] Coded pulse is for digitally encoded harmonics. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the letters, giving it a classic and recognizable appearance. # GE Healthcare 510(k) Premarket Notification Submission ### GE Venue with 8C-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|------------------|-------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse * | Other | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics | | | | | | | | | | | | | | Abdominal [1] | P | P | P | P | P | P | P | P | P | P | | | | Pediatric | P | P | P | P | P | P | P | P | P | P | | | | Small Organ [2] | | | | | | | | | | | | | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | | | | Adult Cephalic | P | P | P | P | P | P | P | P | P | P | | | | Pediatric Cardiac | P2 | P2 | P2 | P2 | P2 | P2 | P2 | N | P2 | P2 | | | | Adult Cardiac | P2 | P2 | P2 | P2 | P2 | P2 | P2 | N | P2 | P2 | | | | Peripheral Vascular | P2 | P2 | P2 | P2 | P2 | P2 | P2 | N | P2 | P2 | | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | | | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | | | | Thoracic/Pleural [3] | P | P | P | P | P | P | P | P | P | P | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | | Vascular Access (IV, PICC) | | | | | | | | | | | | | | Nonvascular [5] | N | N | N | | N | N | N | N | N | N | | | N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image quidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [↓] Coded pulse is for digitally encoded harmonics. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a decorative border with wave-like patterns around the edge. #### Mode of Operation Clinical Application Color Color M PW CW Power Combined Harmonic Coded Other B M Doppler Doppler Doppler Doppler Doppler Modes[*] lmaqinq Anatomy/Region of Interest Pulse * Ophthalmic Fetal / Obstetrics P P P P P P P P P P P P P P P P P P Abdominal [1] Pediatric Small Organ [2] Neonatal Cephalic Adult Cephalic Pediatric Cardiac Adult Cardiac Peripheral Vascular Musculo-skeletal Conventional Musculo-skeletal Superficial Thoracic/Pleural [3] Exam Type, Means of Access Transesophageal b P P P P P Р b b Transrectal P P P P P P P P P Transvaginal Interventional Guidance Vascular Access (IV, PICC) Nonvascular [5] ### GE Venue with E8C-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image guidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [↓] Coded pulse is for digitally encoded harmonics. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. The logo is simple and recognizable, representing the well-known multinational corporation. GE Healthcare 510(k) Premarket Notification Submission | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other | | Ophthalmic | P | | | | P | | P | | | | | | Fetal / Obstetrics | | | | | | | | | | | | | Abdominal [1] | P | P | P | P | P | P | P | P | P | P | | | Pediatric | P | P | P | P | P | P | P | P | P | P | | | Small Organ [2] | P | P | P | P | P | P | P | P | P | P | | | Neonatal Cephalic | P2 | P2 | P2 | P2 | P2 | N | P2 | P2 | P2 | P2 | | | Adult Cephalic | | | | | | | | | | | | | Pediatric Cardiac | N | N | N | N | N | N | N | N | N | N | | | Adult Cardiac | N | N | N | N | N | N | N | N | N | N | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | | | Thoracic/Pleural [3] | P | P | P | P | P | P | P | P | P | P | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | P | P | P | P | P | P | P | P | P | P | 5,7 | | Nonvascular [5] | P | P | P | P | P | P | P | P | P | P | 5,7 | ### GE Venue with 12L-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image guidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [+] Coded pulse is for digitally encoded harmonics. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The circle has decorative, swirling elements around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the long-standing brand identity of General Electric. GE Healthcare 510(k) Premarket Notification Submission | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other | | Ophthalmic | P | | | | P | | P | | | | | | Fetal / Obstetrics | | | | | | | | | | | | | Abdominal [1] | P | P | P | P | P | P | P | P | P | P | | | Pediatric | P | P | P | P | P | P | P | P | P | P | | | Small Organ [2] | P | P | P | P | P | P | P | P | P | P | | | Neonatal Cephalic | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | | | Adult Cephalic | | | | | | | | | | | | | Pediatric Cardiac | N | N | N | N | N | N | N | N | N | N | | | Adult Cardiac | N | N | N | N | N | N | N | N | N | N | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | | | Thoracic/Pleural [3] | P | P | P | P | P | P | P | P | P | P | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | P | 5,7 | | Nonvascular [5] | P | P | P | | P | P | P | P | P | P | 5,7 | ### GE Venue with L12n-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD: [7] Image guidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [+] Coded pulse is for digitally encoded harmonics. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background of the circle. GE Healthcare 510(k) Premarket Notification Submission | Clinical Application<br>Anatomy/Region of Interest | | Mode of Operation | | | | | | | | | | |----------------------------------------------------|----|-------------------|------------|------------|---------------|-----------------|---------------|-------------------|------------------|---------------|-------| | | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes[*] | Harmonic Imaging | Coded Pulse • | Other | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics | | | | | | | | | | | | | Abdominal [1] | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ [2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Pediatric Cardiac | | | | | | | | | | | | | Adult Cardiac | P2 | P2 | P2 | P2 | P2 | P2 | P2 | N | P2 | P2 | 6 | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Thoracic/Pleural [3] | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | P2 | P2 | P2 | P2 | P2 | P2 | P2 | N | P2 | P2 | 6 | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Interventional Guidance | | | | | | | | | | | | | Vascular Access (IV, PICC) | | | | | | | | | | | | | Nonvascular [5] | | | | | | | | | | | | ### GE Venue with 6Tc-RS Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706; Notes: [1] Abdominal includes GYN and Urology (includes prostate); [2] Small Organ includes breast, testes, thyroid; [3] Including detection of fluid and pleural motion/sliding; [4] Nonvascular includes nerve block or biopsy; [5] Biopsy bracket available; [6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD, B/Color/CWD; [7] Image guidance supports freehand needle/catheter placement; [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD; [ +] Coded pulse is for digitally encoded harmonics. {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a dynamic and recognizable appearance. The logo is simple, yet iconic, representing the long-standing brand identity of General Electric. ### 510(k) Summary | In accordance with 21 CFR 807.92 the following summary of information is provided: | | |------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date:<br>Submitter: | March 6, 2018<br>GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 | | Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)470-1003<br>F:(414)918-8275 | | Secondary Contact Person: | Karin Shimoni<br>Regulatory Affairs Leader<br>GE Healthcare | | Device Trade Name:<br>Common/Usual Name:<br>Classification Names: | Venue<br>Diagnostic Ultrasound System<br>Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90- | | Product Code: | IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX | | Primary Predicate Device:<br>Secondary Predicate<br>Device(s): | Venue (K170714)<br>LOGIQ e (K151028)<br>Vivid iq (K161706)<br>LOGIQ P9 and LOGIQ P7 (K163596) | | Device Description: | The proposed Venue system is a general-purpose, Track 3,<br>diagnostic ultrasound device, intended for ultrasound imaging,<br>measurement and analysis of the human body and fluid that<br>provides digital acquisition, processing and display capabilities.<br>Venue can be used in offices, clinical areas and hospitals.<br>The Venue is a mobile system with a small footprint that easily<br>fits into tight spaces and positioned to accommodate the<br>sometimes-awkward work settings of the point of care user.<br>The Venue has a high resolution color LCD monitor, with a<br>simple, multi-touch user interface that makes the system intuitive.<br>The single surface screen can be cleaned with disinfectants.<br>Articulated monitor arm enables flexible display positions in | {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters. The logo is presented in a blue color. order to be accessible and clearly visible in both user-standing and sitting positions. The proposed Venue has the capability for displaying the patient's ECG trace synchronized to the scanned image. This allows the user to view an image from a specific time of the ECG signal. The ECG signal can be input directly from the patient or as an output from an ECG monitoring device. ECG is not intended for monitoring or diagnosis. The Venue has a battery that allows for scanning without the need to plug in to an electrical outlet. The system is capable of wireless communication and a barcode reader is available to be used as an input device. System meets DICOM requirements to support users image storage and archiving needs and allows for output to printing devices. The user documentation is available via electronic media. The Venue utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes. Some biopsy kits are available for needle-guidance procedures. The system includes several automated tools designed to simplify and shorten the workflow time of the healthcare professional for some common assessments. - The Venue is a general purpose diagnostic ultrasound system for Intended Use: use by qualified healthcare professionals. The clinical environments where the Venue can be used include critical care and emergency room environments, as well as point-of care areas in offices, clinical and hospital settings for diagnosis of patients. The Venue is intended for ultrasound imaging, measurement and analysis of the human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, Transesophageal, and imaging guidance of interventional procedures (e.g. Nerve block, vascular access). - Technology: The Venue employs the same fundamental scientific technology as its predicate devices. {15}------------------------------------------------ Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo is a blue circle with the letters 'GE' in a stylized script inside. The letters are also blue, and the circle is surrounded by a series of curved lines that resemble water droplets or waves. The logo is simple and recognizable, and it is associated with a well-known and established company. GE Healthcare 510(k) Premarket Notification Submission | Determination of<br>Substantial Equivalence: | Comparison to Predicate Devices | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The Venue system is substantially equivalent to the predicate<br>devices with regards to intended use, imaging capabilities,<br>technological characteristics and safety and effectiveness. | | | The systems are all intended for ultrasound imaging,<br>measurement and analysis of the human body and fluid<br>for multiple clinical applications. The Venue and predicate Venue (K170714) have similar<br>clinical indications for use however the proposed Venue<br>has the Transesophageal indication which has been<br>cleared on predicate Vivid iq (K161706). Proposed Venue<br>has removed Pinpoint™ GT Technology SW. The Venue and predicate Venue (K170714) have identical<br>imaging modes. The Venue and predicate Venue (K170714) systems<br>transducers are similar, except for adding 6Tc-RS, cleared<br>in Vivid iq (K161706), and adding clinical applications to<br>the 9L-RS, C1-5-RS, 8C-RS, 12L-RS and L12n-RS<br>probes. Automated features cleared on predicate Venue<br>(K170714) have improvements, to assist the user<br>workflow. Addition of the ECG module for ECG trace/image<br>synchronization that was cleared with predicate LOGIQ e<br>(K151028). Adding eFAST Navigation Tool and Review Summary<br>feature that automates the workflow for the eFAST exam<br>that can be done manually. The Venue and predicate Venue (K170714) have similar<br>capability in…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...