K183362 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · Jan 28, 2019 · Radiology
Device Facts
Record ID
K183362
Device Name
Venue GO
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
Jan 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified healthcare professionals for ultrasound imaging, measurement and analysis of the human body and fluid. Venue Go clinical applications include the following: Abdominal (including Gynecology and Urology), Thoracic/Pleural, Ophthalmic, Fetal/Obstetrics, Small Organ (including breast, testes, thyroid), Peripheral vascular, Adult and neonatal cephalic, Musculoskeletal (Conventional and Superficial), Cardiac (Adults and Pediatric), Transesophageal, Transrectal, Transvaginal, Intraoperative and Imaging guidance for interventional procedures (e.g. Nerve block, biopsy, vascular access).
Device Story
Venue Go is a compact, portable diagnostic ultrasound system for use by qualified healthcare professionals in clinical settings. It accepts acoustic signals via linear, convex, and phased array transducers; processes these signals to generate real-time ultrasound images and fluid flow analysis; and displays output on a high-resolution multi-touch LCD monitor. The system supports ECG synchronization for image timing. It features wireless LAN for DICOM-compliant image storage, archiving, and printing. The device is used for imaging, measurement, and interventional guidance (e.g., nerve block, biopsy, vascular access). By providing high-resolution visualization, it assists clinicians in diagnostic assessment and procedural guidance, potentially improving patient outcomes through accurate, real-time anatomical and physiological evaluation.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, thermal, electrical, electromagnetic, and mechanical safety testing, all of which comply with applicable standards (e.g., IEC 60601-1, IEC 60601-2-37, ISO 10993-1).
Technological Characteristics
Portable ultrasound system with high-resolution color LCD and multi-touch interface. Supports linear, convex, and phased array transducers. Modes include B, M, PW/CW Doppler, Color/Power Doppler, Color M, Harmonic Imaging, and Coded Pulse. Connectivity via wireless LAN and DICOM. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, and ISO 10993-1. Powered by AC or internal battery.
Indications for Use
Indicated for qualified healthcare professionals to perform diagnostic ultrasound imaging, measurement, and analysis of the human body and fluid across various clinical applications including abdominal, thoracic, ophthalmic, fetal/obstetrics, small organ, vascular, cephalic, musculoskeletal, and cardiac regions, as well as interventional guidance.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
January 28, 2019
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K183362
Trade/Device Name: Venue Go Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 3, 2018 Received: December 4, 2018
Dear Tracey Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hole 2. Nils
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K183362
Device Name Venue Go
#### Indications for Use (Describe)
Venue Go is a general-purpose diagnostic ultrasound system intended for use by qualified healthcare professionals for ultrasound imaging, measurement and analysis of the human body and fluid. Venue Go clinical applications include the following: Abdominal (including Gynecology and Urology), Thoracic/Pleural, Ophthalmic, Fetal/Obstetrics, Small Organ (including breast, testes, thyroid), Peripheral vascular, Adult and neonatal cephalic, Musculoskeletal (Conventional and Superficial), Cardiac (Adults and Pediatric), Transesophageal, Transrectal, Transvaginal, Intraoperative and Imaging guidance for interventional procedures (e.g. Nerve block, biopsy, vascular access).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CER 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CER 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. There are decorative swirls or flourishes around the perimeter of the circle, adding to the logo's distinctive design.
### Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the Venue Go. Combinations identified "P" represents those previously cleared with another GE Ultrasound system. Combinations identified as "N" are new.
The following Indication for Use forms are appended:
System: Venue Go Transducer: 3Sc-RS Transducer: 9L-RS Transducer: C1-5-RS Transducer: 8C-RS Transducer: E8C-RS Transducer: 12L-RS Transducer: 6Tc-RS Transducer: L4-12t-RS Transducer: L8-18i-RS Transducer: 6S-RS
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a blue circle. There are small, white dots placed around the inner edge of the circle, adding a decorative element to the design.
510(k) Premarket Notification Submission
### GE Venue Go Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other | |
| Ophthalmic | N | | | | N | | N | | | | | |
| Fetal / Obstetrics | N | N | N | | N | N | N | N | N | N | | |
| Abdominal [1] | N | N | N | | N | N | N | N | N | N | | |
| Pediatric | N | N | N | | N | N | N | N | N | N | | |
| Small Organ [2] | N | N | N | | N | N | N | N | N | N | | |
| Neonatal Cephalic | N | N | N | | N | N | N | N | N | N | | |
| Adult Cephalic | N | N | N | | N | N | N | N | N | N | | |
| Pediatric Cardiac | N | N | N | N | N | N | N | N | N | N | 6,8 | |
| Adult Cardiac | N | N | N | N | N | N | N | N | N | N | 6,8 | |
| Peripheral Vascular | N | N | N | | N | N | N | N | N | N | | |
| Musculo-skeletal Conventional | N | N | N | | N | N | N | N | N | N | | |
| Musculo-skeletal Superficial | N | N | N | | N | N | N | N | N | N | | |
| Thoracic/Pleural [3] | N | N | N | | N | N | N | N | N | N | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | N | N | N | N | N | N | | N | N | | 6 | |
| Transrectal | N | N | N | | N | N | N | N | N | N | | |
| Transvaginal | N | N | N | | N | N | N | N | N | N | | |
| Intraoperative [9] | N | N | N | | N | N | N | N | N | N | | |
| Interventional Guidance | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | N | N | N | | N | N | N | N | N | N | 5,7 | |
| Nonvascular [4] | N | N | N | | N | N | N | N | N | N | 5,7 | |
N = new indication; P = previously cleared by FDA K170714; P1 = previously cleared by FDA K163596; P2 = previously cleared by FDA K161706;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image guidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. The logo is simple and recognizable, representing the brand identity of General Electric.
510(k) Premarket Notification Submission
### GE Venue Go with 3Sc-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-----------------------|---------------------|------------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*1] | Harmonic<br>Imaging | Coded<br>Pulse * | Other |
| Ophthalmic | P | | | | P | | P | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | | |
| Pediatric | P | P | P | | P | P | P | P | P | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | P | P | P | | P | P | P | P | P | | |
| Pediatric Cardiac | P | P | P | P | P | P | P | P | P | | 6 |
| Adult Cardiac | P | P | P | P | P | P | P | P | P | | 6 |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | P | P | P | P | P | P | | |
| Nonvascular [4] | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image quidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized, cursive font inside. The letters are also blue, matching the color of the circle. The logo is simple and recognizable, representing the well-known multinational corporation.
### GE Venue Go with 9L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[1] | Harmonic<br>Imaging | Coded<br>Pulse• | Other | |
| Ophthalmic | P | | | | P | | P | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | P | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | | |
| Small Organ [2] | P | P | P | | P | P | P | P | P | P | | |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | | |
| Adult Cephalic | | | | | | | | | | | | |
| Pediatric Cardiac | P | P | P | | P | P | P | P | P | P | 8 | |
| Adult Cardiac | P | P | P | | P | P | P | P | P | P | 8 | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | | |
| Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | P | 5,7 | |
| Nonvascular [4] | P | P | P | | P | P | P | P | P | P | 5,7 | |
N = new indication; P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image quidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. The logo is simple and recognizable, representing the brand identity of General Electric.
### 510(k) Premarket Notification Submission
### GE Venue Go with C1-5-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[¹] | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | P | |
| Pediatric | P | P | P | | P | P | P | P | P | P | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Pediatric Cardiac | P | P | P | | P | P | P | P | P | P | 8 |
| Adult Cardiac | P | P | P | | P | P | P | P | P | P | 8 |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | |
| Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | P | 7 |
| Nonvascular [4] | P | P | P | | P | P | P | P | P | P | 7 |
N = new indication; P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image quidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The blue color is consistent throughout the logo, providing a clean and recognizable design.
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | P | |
| Pediatric | P | P | P | | P | P | P | P | P | P | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | |
| Adult Cephalic | | | | | | | | | | | |
| Pediatric Cardiac | P | P | P | | P | P | | P | P | P | 8 |
| Adult Cardiac | P | P | P | | P | P | | P | P | P | 8 |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nonvascular [4] | P | P | P | | P | P | P | P | P | P | |
#### GE Venue Go with 8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication: P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P² = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD: [7] Image guidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized, cursive font inside. The letters are also blue, matching the color of the circle. The logo is simple and recognizable, representing the well-known multinational corporation.
510(k) Premarket Notification Submission
### GE Venue Go with E8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-------------------|------------------|--------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes[1] | Harmonic Imaging | Coded Pulse* | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | P | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | P | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Pediatric Cardiac | | | | | | | | | | | |
| Adult Cardiac | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural [3] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P | P | P | | P | P | P | P | P | P | |
| Transvaginal | P | P | P | | P | P | P | P | P | P | |
| Intraoperative [9] | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nonvascular [4] | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image quidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The blue color is consistent throughout the logo, providing a clean and recognizable design.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other | |
| Ophthalmic | P | | | | P | | P | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | P | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | | |
| Small Organ [2] | P | P | P | | P | P | P | P | P | P | | |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | | |
| Adult Cephalic | | | | | | | | | | | | |
| Pediatric Cardiac | P | P | P | | P | P | P | P | P | P | 8 | |
| Adult Cardiac | P | P | P | | P | P | P | P | P | P | 8 | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | | |
| Thoracic/Pleural [3] | P | P | P | | P | P | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | P | 5,7 | |
| Nonvascular [4] | P | P | P | | P | P | P | P | P | P | 5,7 | |
#### GE Venue Go with 12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication: P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
- [3] Including detection of fluid and pleural motion/sliding;
- [4] Nonvascular includes nerve block or biopsy;
- [5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD: [7] Image guidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a blue circle. There are small, white dots placed around the inner edge of the circle, adding a decorative element to the design.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse* | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal [1] | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Pediatric Cardiac | | | | | | | | | | | | |
| Adult Cardiac | P | P | P | P | P | P | P | P | P | | 6 | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural [3] | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | P | P | P | P | P | P | P | P | P | | 6 | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nonvascular [4] | | | | | | | | | | | | |
#### GE Venue Go with 6Tc-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
- [3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image guidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized, cursive font inside. The letters are also blue, matching the color of the circle. The logo is simple and recognizable, representing the well-known multinational corporation.
### 510(k) Premarket Notification Submission
### GE Venue Go with L4-12t-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other |
|----------------------------------------------------|----|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Ophthalmic | P2 | | | | P2 | | P2 | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | N | N | N | | N | N | N | N | N | N | |
| Pediatric | P2 | P2 | P2 | | P2 | P1 | P2 | P1 | P2 | P1 | |
| Small Organ [2] | P2 | P2 | P2 | | P2 | P1 | P2 | P1 | P2 | P1 | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Pediatric Cardiac | N | N | N | | N | N | N | N | N | N | 8 |
| Adult Cardiac | N | N | N | | N | N | N | N | N | N | 8 |
| Peripheral Vascular | P2 | P2 | P2 | | P2 | P1 | P2 | P1 | P2 | P1 | |
| Musculo-skeletal Conventional | P2 | P2 | P2 | | P2 | P1 | P2 | P1 | P2 | P1 | |
| Musculo-skeletal Superficial | P2 | P2 | P2 | | P2 | P1 | P2 | P1 | P2 | P1 | |
| Thoracic/Pleural [3] | P2 | P2 | P2 | | P2 | N | P2 | N | P2 | N | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative [9] | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P2 | P2 | P2 | | P2 | N | P2 | N | P2 | P2 | 5,7 |
| Nonvascular [4] | P2 | P2 | P2 | | P2 | N | P2 | N | P2 | P2 | 5,7 |
N = new indication; P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P2 = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image quidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized, cursive font inside. The letters are also blue, matching the color of the circle. The logo is simple and recognizable, representing the well-known multinational corporation.
510(k) Premarket Notification Submission
### GE Venue Go with L8-18i-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
|----------------------------------------------------|----|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Pediatric | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Small Organ [2] | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Neonatal Cephalic | P1 | P1 | P1 | | P1 | | P1 | P1 | P1 | P1 | |
| Adult Cephalic | | | | | | | | | | | |
| Pediatric Cardiac | | | | | | | | | | | |
| Adult Cardiac | | | | | | | | | | | |
| Peripheral Vascular | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Musculo-skeletal Conventional | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Musculo-skeletal Superficial | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Thoracic/Pleural [3] | P2 | P2 | P2 | | P2 | | P2 | N | P2 | P2 | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative [9] | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | 7 |
| Nonvascular [4] | P2 | P2 | P2 | | P2 | | P2 | P1 | P2 | P2 | 7 |
N = new indication: P = previously cleared by FDA K180599; P1 = previously cleared by FDA K181783; P² = previously cleared by FDA K151028;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Including detection of fluid and pleural motion/sliding;
[4] Nonvascular includes nerve block or biopsy;
[5] Biopsy bracket available;
[6] Combined modes as defined in [*], but exclude B/Power/PWD, and include: B/CWD,
B/Color/CWD;
[7] Image quidance supports freehand needle/catheter placement;
[8] Combined modes as defined in [*], but exclude B/Power/PWD;
[9] Intraoperative includes vascular;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized, cursive font inside. The letters are also blue, matching the color of the circle. The logo is simple and recognizable, representing the well-known multinational corporation.
510(k) Premarket Notification Submission
### GE Venue Go with 6S-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal [1] | P2 | P2 | P2 | | P2 | P2 | P2 | N | P2 | | | |
| Pediatric | P2 | P2 | P2 | | P2 | P2 | P2 | P1 | P2 | | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | P2 | P2 | P2 | | P2 | P2 | P2 | P1 | P2 | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Pediatric Cardiac | P2 | P2 | P2 | P2 | P2 | P2 | P2 | P1 | P2 | | 6 | |
| Adult Cardiac | P2 | P2 | P2 | P2…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.