Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
Device Story
AcuityPDR is a mobile x-ray system; integrates generator/collimator, digital radiography (DR) panels, and acquisition software. System inputs: x-ray radiation; outputs: diagnostic digital images. Used in clinical settings by physicians or technicians. Operation: user selects exposure parameters (kV, mAs) via console; generator produces x-rays; DR panel captures signal; acquisition software processes and displays images. Benefits: portable diagnostic imaging for various patient populations. System components are sourced from various manufacturers; integrated and tested by RadmediX. Healthcare providers use images for diagnostic decision-making. Battery-powered generator allows mobile use; supports 200 exposures per charge.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-54 standards, and verification of system integration.
Indicated for adult and pediatric patients requiring diagnostic x-ray imaging. Contraindicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Radmedix LLC % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
April 13, 2020
### Re: K200726
Trade/Device Name: AcuityPDR Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB, LLZ Dated: March 17, 2020 Received: March 20, 2020
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200726
Device Name AcuityPDR
Indications for Use (Describe)
Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K200726
Image /page/3/Picture/1 description: The image shows a logo with the text "RadmediX". The letters "R" and "X" are in blue, while the letters "R", "m", "e", "d", and "i" are in gray. The logo is likely for a company or organization with the name RadmediX.
RadmediX, LLC 2510 Lance Rd. Dayton, OH 45409 Tel 844 723 6334 Registration Number: 3009134655
### 1. Administrative Information
Reason for Submission: 510(k) Notification for RadmediX AcuityPDR
| Submitter: | Submission contact person: | Gabriel Issa, Director of Equipment |
|----------------------------------|----------------------------|-------------------------------------|
| | Contact telephone: | 844 723 6334 |
| | Contact e-mail: | gabe@radmedix.com |
| | Date prepared: | April 9, 2020 |
| Identification: | Classification Name: | AcuityPDR |
| | Classification Name: | Mobile X-Ray System |
| | Classification Panel: | Radiology |
| | Classification Regulation: | 21 CFR §892.1720 |
| | Device Class: | Class II |
| | Product Code: | IZL, MQB, LLZ |
| Substantially equivalent device: | Trade Name: | CMDR 2CW |
| | Manufacturer: | MinXray |
| | 510(k) #: | K191451 |
| | Classification Name: | Mobile X-Ray System |
| | Classification Panel: | Radiology |
| | Classification Regulation: | 21 CFR §892.1720 |
| | Device Class: | Class II |
| | Product Code: | IZL, MQB, LLZ |
- 2. Device description: RadmediX AcuityPDR combines components into a complete mobile x-ray system, including software, a generator/collimator combination, and digital radiography panels. Radmedix combines components from various manufacturers into a complete mobile x-ray system. The customer selects one (or more) of the following digital x-ray receptor panels: DRTech 4343A, (K192400); DRTech 4343W, (K193017); AcuityDR (K171137); AcuityDR 1417 (K162552 EVS 3643, EVS 3643G) or AcuityDR 1717 (K162555 EVS 4343, EVS 4343G). In addition, the customer selects one of three software packages: Accuvue+ (K130883), AccuVueMED, (K152172) or AccuVue (K141440). The generator can be battery operated. A single battery charge produces 200 Exposures at max KV and MAS settings. The battery charger is UL Listed and the internal lithium ion battery is overcharge and overcurrent protected. A typical acquisition computer would be a Lenovo P53S or a Dell Precision 3541.
{4}------------------------------------------------
- 3. Indications for Use: Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.
| Comparable<br>Properties | CMDR 2CW, made by MinXray<br>510(k) #: K191451 | AcuityPDR | Comparison<br>Results |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Indications<br>for use | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic x-rays. Not for<br>mammography. | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic x-rays. Not for<br>mammography. | Identical |
| X-ray tube | Toshiba D-124S | TOSHIBA D-125SB 100kV, 40mA | Equivalent<br>Functionality |
| Peak<br>generator<br>output | 2 kW | 2 kW | SAME |
| Tube current<br>range | 30 mA±20% at 60kV mA<br>25 mA±20% at 80kV mA<br>20 mA±20% at 100kV | 40-60kV: 25mA.<br>61-100kV: 20mA. | Comparable<br>ranges. |
| Tube voltage<br>adjustable<br>range | 40-100 kV, 2kV steps | 40-100kV, step value 1kV. | More flexible<br>kV<br>adjustment |
| mAs range | 2.0-150 (steps not specified) | 0.4mAs ~ 50mAs, with the range of:<br>0.40, 0.50, 0.63, 0.80, 1.00, 1.25, 1.6,<br>2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8, 10, 12.5, 16,<br>20, 25, 32, 40, 50. | Comparable<br>adjustability |
| Collimator | Built in | Built in | Equivalent<br>Functionality |
| X-ray<br>Generator | Three models, from 100 to 120 kVp<br>maximum | One model, up to 100 kVp | Equivalent<br>Functionality |
| Digital X-Ray<br>Detectors | K150929 CareView 1500Cw X-ray Flat<br>Panel Detector manufactured by<br>CareRay | DRTech 4343A, (K192400);<br>DRTech 4343W, (K193017);<br>AcuityDR (K171137);<br>AcuityDR 1417 (K162552);<br>AcuityDR 1717 (K162555) | Equivalent<br>Functionality |
| Operator<br>console | Touch Control or Touch Screen | SAME | Similar<br>Functionality |
| Comparable<br>Properties | CMDR 2CW, made by MinXray<br>510(k) #: K191451 | AcuityPDR | Comparison<br>Results |
| Acquisition<br>Software | dicomPACS® DX-R K141440 | Customer Selects:<br>AccuVueMED (K152172)<br>AccuVue (K141440) or<br>AccuVue+ (K130883) | Equivalent<br>Functionality<br>Only cleared<br>software is<br>supplied. |
| Photos | Image: CMDR 2CW | Image: AcuityPDR | Similar form<br>factors |
| Power Source | AC LINE, single or three phase<br>depending on the generator | AC Line or rechargeable batteries<br>(Generator only) | SAME. |
| Standards | 60601-1:2005; 60601-1-2:2014<br>21CFR1020 | SAME<br>Generators are TUV Tested<br>AC Line Power Supply is UL Listed<br>See list below. | SAME |
## 4. Technological characteristics: Comparison Table
{5}------------------------------------------------
5. Non clinical testing: Testing was performed successfully according to the following standards:
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Standard Designation<br>Number And Date | Title Of Standard |
|------------------------------|----------------------------------------|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-4 | IEC | 60601-1:2005/(R)2012<br>And A1:2012 | Medical Electrical Equipment Part 1: General Requirements For<br>Basic Safety And Essential Performance (IEC 60601 1:2005, MOD) |
| 19-1 | IEC | 60601-1-2:2014 | Medical Electrical Equipment Part 12: General Requirements For<br>Basic Safety And Essential Performance Collateral Standard:<br>Electromagnetic Disturbances Requirements And Tests |
{6}------------------------------------------------
| FDA<br>Recognition<br>Number | Standard<br>Developing<br>Organization | Standard Designation<br>Number And Date | Title Of Standard |
|------------------------------|----------------------------------------|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 12-269 | IEC | 60601-1-3 Edition 2.1<br>2013-04 | Medical electrical equipment - Part 1-3: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Radiation protection in diagnostic X-ray equipment |
| 12-296 | IEC | 60601-2-54 -<br>2009+AMD1 :<br>2015+AMD2:2018 | Particular Requirements For The Basic Safety And Essential<br>Performance Of X-Ray Equipment for Radiography and<br>Radioscopy |
| 12-238 | NEMA | PS 3.1 - 3.20 (2011) | NEMA Digital Imaging and Communications in Medicine (DICOM)<br>Set |
| N/A | FDA | 21CFR1020 | Electronic Products; Performance Standard for Diagnostic X-Ray<br>Systems and Their Major Components |
All of the components subject to the CDRH performance standard are certified to comply with the standard by their respective manufacturers. We do not supply any non-certified components. Applicable components carry a certification label (UL, ETL, etc.). The digital panel software employed was used unmodified from clearances obtained from FDA. Since multiple configurations are available, our service engineers fully test each new system upon installation at the customer site. In recognition of possible cybersecurity threats to the software, we consulted this guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Druq Administration Stoff October 2014. As a result, we updated our own internal standard operating procedures and added cybersecurity precautions to the software users' manuals. The generator User Manual has been updated to add Pediatric Considerations. Each system is tested for proper integration prior to shipment to the customer.
6. Clinical testing. Not required for a determination of substantial equivalence.
# 7. Substantial Equivalence Discussion.
The RadmediX AcuityPDR performs the same functions using the same technological methods to produce diagnostic x-ray images. In all material aspects, the Visaris and the RadmediX systems are substantially equivalent to each other.
# 8. Substantial Equivalence Conclusion:
After analyzing bench test results, risk analysis, and clinical evaluation, it is the conclusion of RadmediX LLC that the RadmediX AcuityPDR is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.