The DePuy PULSE Thoracolumbar Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The DePuy PULSE Thoracolumbar Screw System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
Device Story
Spinal fixation systems (pedicle screws, rods, connectors, plates) provide immobilization and stabilization of thoracic, lumbar, and sacral spine segments. Systems are used by surgeons in clinical settings to treat spinal instabilities, deformities, and degenerative conditions. Components are implanted via posterior or percutaneous approaches to facilitate fusion; fenestrated screws allow bone cement injection for tumor-related stabilization. The device story focuses on updating labeling for magnetic resonance (MR) compatibility. No changes to design, materials, or intended use were made; substantial equivalence is based on existing predicate systems.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing for MR compatibility according to ASTM F2213, F2052, F2119, and F2182 standards.
Technological Characteristics
Metallic spinal fixation components (screws, rods, connectors, plates). Materials: Pure Titanium/Titanium Alloy (ASTM F-67, F-136, F-1472), Stainless Steel (ASTM F-138, F-134, F-2229), Cobalt-Nickel-Chromium-Molybdenum Alloy (ASTM F-562), Cobalt-Chromium-Molybdenum Alloy (ASTM F-1537). Mechanical fixation via pedicle/non-pedicle screws, hooks, and rod-based constructs. Fenestrated screws allow bone cement injection. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients (and pediatric patients for specific systems) requiring spinal immobilization as an adjunct to fusion for acute/chronic instabilities or deformities (scoliosis, lordosis, kyphosis, spondylolisthesis, spinal stenosis, tumor, pseudoarthrosis, failed fusion, degenerative disc disease). Fenestrated variants indicated for advanced stage tumors in thoracic/lumbar spine to restore column integrity without fusion. Contraindications include cases where structural integrity is severely compromised.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2022
Medos International SARL % Sergio M. Cordeiro Senior Regulatory Affairs Specialist Depuy Synthes Spine 325 Paramount Drive Raynham, Massachusetts 02767
### Re: K191212
Trade/Device Name: CrossOver Cross Connectors, DePuy PULSE Thoracolumbar Screw System, EXPEDIUM Fenestrated Screw System, EXPEDIUM SFX Cross Connectors, EXPEDIUM Spine System. EXPEDIUM VERSE Spine System. E-Z Link Cross Connectors, ISOLA Spine System, MONARCH Spine System, MOSS MIAMI Spine System, TiMX Low Back System, VIPER Fenestrated Screw System, VIPER PRIME System, VIPER PRIME Fenestrated Screws, VIPER SAI (Sacral-Alar-Iliac), VIPER System, VIPER 2 System, VSP Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: PGM, NKB, JDQ, KWQ, KWP, JDN
Dear Sergio M. Cordeiro:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 24, 2019. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation number, 21 CFR 888.3070.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ronald Jean, OHT6: Office of Orthopedic Devices, (301)796-5650, Ronald.Jean@fda.hhs.gov.
Sincerely,
# Ronald P. Jean -S
Ronald P. Jean, Ph.D. Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
September 24, 2019
Medos International SARL % Mr. Sergio M. Cordeiro Senior Regulatory Affairs Specialist DePuy Synthes Spine 325 Paramount Drive Raynham, Massachusetts 02767
### Re: K191212
Trade/Device Name: CrossOver Cross Connectors, DePuy PULSE Thoracolumbar Screw System, EXPEDIUM Fenestrated Screw System, EXPEDIUM SFX Cross Connectors, EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, E-Z Link Cross Connectors, ISOLA Spine System, MONARCH Spine System, MOSS MIAMI Spine System, TiMX Low Back System, VIPER Fenestrated Screw System, VIPER PRIME System, VIPER PRIME Fenestrated Screws, VIPER SAI (Sacral-Alar-Iliac), VIPER System, VIPER 2 System, VSP Spine System Regulatory Class: Unclassified
Product Code: PGM, NKB, JDQ, KWQ, KWP, JDN Dated: May 3, 2019 Received: May 6, 2019
Dear Mr. Cordeiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ronald P. Jean -S
Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K191212
Device Name DePuy PULSE Thoracolumbar Screw System
### Indications for Use (Describe)
The DePuy PULSE Thoracolumbar Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The DePuy PULSE Thoracolumbar Screw System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------|--|
| <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
### K191212
### Device Name
ISOLA® Spine System, VSP® Spine System, MOSS® MIAMI Spine System, TiMX Low Back System, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH® Spine System
### Indications for Use (Describe)
The ISOLA, VSP, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are pedicle screw systems intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
The ISOLA, VSP, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The ISOLA, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are also a hook and sacral/iliac screw fixation system of the indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
The ISOLA, VSP, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems when used with pedicle screws are indicated for degenerative discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine.
The ISOLA and MOSS MIAMI Spine Systems when used as anterior thoracic/lumbar screw fixation systems, are indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
The MONARCH Spine System Dual Rod Connectors can be used to connect MONARCH Spine System Rods to rods of other DePuy Spine 4.75mm, 5.5mm, and 6.35mm diameter rod systems.
Type of Use (Select one or both, as applicable)
|X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K191212
Device Name EXPEDIUM® SFX® Cross Connectors
### Indications for Use (Describe)
The EXPEDIUM SFX Cross Connector System is designed to transversely connect two rods used in posterior spinal instrumentation constructs. The EXPEDIUM SFX Cross Connector System devices are intended for use with components of the commercially available EXPEDIUM, VIPER, VSP, ISOLA, MONARCH, MOSS MIAMI, and TiMX Spine Systems. Please refer to the labeling of the aforementioned spinal systems indications for use.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K191212
### Device Name
EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System, VIPER® 2 System, VIPER® 2 System, VIPER SAI, and VIPER PRIME
### Indications for Use (Describe)
The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER 2 System, VIPER SAI, and VIPER PRIME is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic. Iumbar and sacral spine.
The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER 2 System, VIPER SAI, and VIPER PRIME is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER System, VIPER 2 System, VIPER PRIME is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER 2 System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER System, VIPER 2 System, and VIPER PRIME are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The EXPEDIUM Growing System is indicated for patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The EXPEDIUM Growing System may be used with any cleared traditional 4.5 and 5.5 EXPEDIUM Spine Systems. The EXPEDIUM Growing Spine System is not intended to be used with 4.0mm diameter screws.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K191212
### Device Name
EXPEDIUM® Fenestrated Screw System, VIPER® Fenestrated Screw System, and VIPER PRIME™ Fenestrated Screws
### Indications for Use (Describe)
When used in conjunction with CONFIDENCE High Viscosity Spinal Cement, the VIPER and EXPEDIUM Fenestrated Screw Systems are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER and EXPEDIUM Fenestrated Screw Systems augmented with the CONFIDENCE High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) | | | | |
|--------------------------------------------------|---------------------------------------------|--|--|--|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{10}------------------------------------------------
# 510(k) SUMMARY
| A. | Submitter Information | |
|----|--------------------------|----------------------------------------|
| | Manufacturer: | Medos International SARL |
| | | Chemin-Blanc 38 |
| | | 2400 Le Locle, Switzerland |
| | Submitter: | DePuy Synthes Spine |
| | | 325 Paramount Drive |
| | | Raynham, MA 02767 |
| | Contact Person: | Sergio M. Cordeiro |
| | | 325 Paramount Drive |
| | | Raynham, MA 02767 |
| | Telephone: | (508) 977-2640 |
| | Fax: | (508) 828-3797 |
| | Email: | scordei1(@its.jnj.com |
| B. | Date Prepared | August 22, 2019 |
| C. | Device Name | |
| | Trade/Proprietary Names: | CrossOver Cross Connectors |
| | | DePuy PULSE Thoracolumbar Screw System |
| | | EXPEDIUM® Fenestrated Screw System |
| | | EXPEDIUM® SFX® Cross Connectors |
| | | EXPEDIUM® Spine System |
| | | EXPEDIUM VERSE® Spine System |
| | | E-Z Link Cross Connectors |
| | | ISOLA® Spine System |
| | | MONARCH® Spine System |
| | | MOSS® MIAMI Spine System |
| | | TiMX Low Back System |
| | | VIPER® Fenestrated Screw System |
| | | VIPER PRIME™ System |
| | | VIPER PRIME™ Fenestrated Screws |
| | | VIPER® SAI (Sacral-Alar-Iliac) |
| | | VIPER® System |
| | | VIPER® 2 System |
| | | VSP® Spine System |
{11}------------------------------------------------
K191212 Page 2 of 10
| Common/Usual Names: | Orthosis, Thoracolumbar/Sacral Pedicle Screw Spinal<br>Fixation Devices |
|-----------------------|----------------------------------------------------------------------------------------|
| Classification Names: | PGM - Unclassified |
| | JDQ – Class II – 21 CFR §888.30101<br>Bone fixation cerclage |
| | KWP - Class II – 21 CFR §888.3050<br>Spinal interlaminal fixation orthosis |
| | JDN, KWQ – Class II – 21 CFR §888.3060<br>Spinal intervertebral body fixation orthosis |
| | NKB - Class II – 21 CFR §888.3070<br>Thoracolumbosacral pedicle screw system |
#### Predicate Device Names D.
| Primary Predicate: | EXPEDIUM® Spine System (K160904) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Additional Predicates: | CrossOver Cross Connectors<br>(K012971, K021879)<br>DePuy PULSE Thoracolumbar Screw System<br>(K113396)<br>EXPEDIUM® Fenestrated Screw System<br>(K160879)<br>EXPEDIUM® SFX® Cross Connectors<br>(K111136, K160904)<br>EXPEDIUM VERSE® Spine System<br>(K142185, K160879, K173095)<br>E-Z Link Cross Connectors<br>(K001372)<br>ISOLA® Spine System<br>(K003822, K010972, K013441, K022285, K884163, K895441, K905184, K905826, K921090, K922504.) |
1 The devices subject to IDQ are non-spine FDA product code classifications. However, have previously been cleared by other companies for spinal use.
{12}------------------------------------------------
# K191212 Page 3 of 10
K922952, K943819, K944737, K951116, K952236, K962645, K964007, K971248, K974757, K980485, K981113, K984350)
MONARCH® Spine System (K010576, K021148, K023438, K024348, K111136)
MOSS® MIAMI Spine System (K001105, K002607, K010742, K011182, K022623, K023804, K030383, K103490, K760649, K894184, K933881, K950695, K953915, K955348, K962628, K964024, K972568, K980447, K982011, K982320, K983583, K984378, K992168)
TiMX Low Back System (K984348)
VIPER® Fenestrated Screw System (K160879)
VIPER PRIME™ System (K162912, K173095)
VIPER PRIME™ Fenestrated Screws (K170543, K171570, K173095)
VIPER® SAI (Sacral-Alar-Iliac) (K121020, K160904)
VIPER® System (K101993, K102701, K110216, K111136, K111571, K160904)
VIPER® 2 System (K11136, K160904)
VSP® Spine System (K944736, K951116, K952236, K965102, K980485, K984350)
#### E. Submission Purpose
Obtain clearance for magnetic resonance compatibility labeling of the systems listed.
{13}------------------------------------------------
#### F. Device Descriptions
# CrossOver Cross Connectors, and E-Z Link Cross Connectors
The Cross over and E-Z Link Cross Connectors are designed to transversely connect two rods used in spinal instrumentation constructs. The connector minimizes torsional forces on the construct, thus reducing the micromotion and the probability of the construct shifting after placement. The CrossOver and E-Z Link Cross Connectors are indicated as part of the ISOLA Spinal System.
# DePuy PULSE Thoracolumbar Screw System
The DePuy PULSE Thoracolumbar Screw System consists of a variety of rods, pedicle screws, connectors, setscrews and other connection components used to build a spinal construct. The DePuy PULSE Thoracolumbar Screw System is offered in titanium material in sizes ranging from 4.35mm to 12.0mm in shank diameter and 20mm to 130mm in length. Screw shanks with 7.0mm and smaller are assembled from the top of the outer head, while screws with 7.5mm and larger are assembled from the bottom using a flex ball. The implant components can be rigidly locked into a variety of configurations, with each construct being made for the individual case.
# EXPEDIUM® Fenestrated Screw System, VIPER® Fenestrated Screw System, and VIPER PRIME™ Fenestrated Screws
The EXPEDIUM, VIPER and VIPER PRIME screws with fenestrations are designed with a cortical fix screw shank that is fully cannulated with lateral fenestrations at the distal end and may be used in conjunction with, or without the CONFIDENCE High Viscosity Spinal Cement. The cannulation and fenestrations allow for the injection of bone cement through the screws into the vertebral body in patients with severe osteoporosis or pathological fractures due to osteoporosis and tumor in the thoracic, lumbar and sacral spine.
# EXPEDIUM® SFX® Cross Connectors
The EXPEDIUM SFX Cross Connector System is designed to transversely connect two rods used in constructs comprising spinal instrumentation components. It features an advanced top-loading design and unique snap-fit feature that simplify initial placement and tightening procedures which helps avoid impingement of surrounding anatomy. The EXPEDIUM SFX Cross Connector system incorporates a broad range of implant options in both and adjustable configurations for an optimal balance between dural clearance and implant profile.
# EXPEDIUM VERSE® Spine System
The EXPEDIUM VERSE Spine System is designed to provide intraoperative polyaxial to monoaxial conversion. It facilitates easier rod capture and provides a powerful and precise reduction mechanism. EXPEDIUM VERSE is a reduced profile thoracolumbar
{14}------------------------------------------------
implant for use for with wide range of patient statures. EXPEDIUM VERSE is a selfcontained, efficient, and intuitive instrument system that is compatible with EXPEDIUM 5.5 rods, hooks and mono screws to enhance versatility.
# ISOLA® Spine System
The ISOLA Spine System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The ISOLA Spine System consists of connectors, open and closed iliac screws, iliac bolts, and angled closed iliac screws. These components have been designed to allow for rigid fixation of the sacral and pelvic regions of the spine.
# MONARCH® Spine System
The MONARCH Spine System is both a rod-based and plate-based system designed to interface with various spinal anatomies. The plate-based system consists of pedicle screws, spine plates, transverse connectors. J-hooks, washers, nuts and set screws. The rod-based system consists of spinal rods, pedicle screws, set screws, various slotted connectors, and transverse connectors.
# TiMX Low Back System
The TiMX Low Back System is a construct that consists of pedicle and sacral screws, spine plates, nuts, washers, and transverse connectors. The TiMX Low Back System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of chronic instabilities and deformities of the thoracic, lumber, and sacral spine.
# MOSS® MIAMI Spine System
The MOSS MIAMI Spine System is a closure mechanism which secures the rod to the screw in one single, simple step. The unique square threads balance the forces vertically, creating a secure assembly. It consists of anatomic hooks, monoaxial screws, polyaxial screws and a dual closure mechanism.
# VIPER PRIME™ System
The VIPER PRIME System is novel technique for percutaneous pedicle screw placement and posterior stabilization. This innovative technique eliminates the need for guidewires, Jamshidi needles and pedicle preparation instruments; utilizing a stylet that is fully controlled by the screwdriver. The VIPER PRIME System enables surgeons to target pedicles and insert the screw, without the need for instrument exchanges or reconfirmation of their trajectory.
{15}------------------------------------------------
# VIPER® SAI (Sacral-Alar-Iliac)
The VIPER Sacral-Alar-Iliac Screw is an implant designed for sacropelvic fixation. The VIPER SAI Screw is optimized for Sacral-Alar-Iliac placement vis-à-vis a favored angle polyaxial head, smooth shank and robust drive feature. The in-line nature of this anchor allows not only stabilization but also correction of pelvic sagittal or coronal deformities.
# EXPEDIUM® Spine System, VIPER® System, and VIPER® 2 System
The EXPEDIUM Spine System, VIPER System, and VIPER 2 System are metallic implants intended to provide immobilization and stabilization of spinal segments. They can be used for skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine; or for posterior non-cervical pedicle screw fixation in pediatric patients as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM and VIPER/VIPER2 systems are intended to be used with autograft and/or allograft.
The EXPEDIUM Spine System consist of longitudinal rods, monoaxial screws, polyaxial screws, uni-planar screws, reduction screws, cable/wire screws, bolts, slotted connectors, wires, hooks, reduction hooks, transverse connectors, SFX Cross Connector System, dual rod connectors, sacral extenders, lateral connectors, and washers. The VIPER and VIPER 2 Systems consist of cannulated polyaxial screws, monoaxial screws, uni-planar screws, reduction screws, and rods used in a percutaneous approach.
### VSP® Spine System
The VSP Spine System is indicated for degenerative spondylolisthesis, in skeletally mature patients, with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion. Levels of fixation are for the thoracic, lumbar and sacral spine. The VSP Spine System is also indicated for pedicle screw fixation for severe spondylolisthesis (Grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion us attained. Levels of fixation are from L3-S1.
#### G. Intended Use
### DePuy PULSE Thoracolumbar Screw System
The DePuy PULSE Thoracolumbar Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The DePuy PULSE Thoracolumbar Screw System metallic components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with
{16}------------------------------------------------
degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
# ISOLA® Spine Svstem, VSP® Spine Svstem, MOSS® MIAMI Spine Svstem, TiMX Low Back System, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH® Spine System
The ISOLA, VSP, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are pedicle screw systems intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
The ISOLA, VSP, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The ISOLA, MOSS MIAMI, TiMX, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems are also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
The ISOLA, VSP, CrossOver Cross Connectors, E-Z Link Cross Connectors, and MONARCH Spine Systems when used with pedicle screws are indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine.
The ISOLA and MOSS MIAMI Spine Systems when used as anterior thoracic/lumbar screw fixation systems, are indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
The MONARCH Spine System Dual Rod Connectors can be used to connect MONARCH Spine System Rods to rods of other DePuy Spine 4.75mm, 5.5mm, and 6.35mm diameter rod systems.
{17}------------------------------------------------
# EXPEDIUM® SFX® Cross Connectors
The EXPEDIUM SFX Cross Connector System is designed to transversely connect two rods used in posterior spinal instrumentation constructs. The EXPEDIUM SFX Cross Connector System devices are intended for use with components of the commercially available EXPEDIUM, VIPER, VSP, ISOLA, MONARCH, MOSS MIAMI, and TiMX Spine Systems. Please refer to the labeling of the aforementioned spinal systems for each system's indications for use.
# EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System, VIPER® System, VIPER® 2 System, VIPER SAI, and VIPER PRIME
The EXPEDIUM Spine System. EXPEDIUM VERSE Spine System. VIPER System. VIPER 2 System, VIPER SAI, and VIPER PRIME is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine.
The EXPEDIUM Spine System. EXPEDIUM VERSE Spine System. VIPER System. VIPER 2 System, VIPER SAI, and VIPER PRIME is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used in a posterior percutaneous approach with MIS instrumentation, the EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER System, VIPER 2 System. VIPER SAI, and VIPER PRIME is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EXPEDIUM Spine System, EXPEDIUM VERSE Spine System, VIPER System and VIPER 2 System metallic implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The EXPEDIUM Spine System, EXPEDIUM VERSE Spine System. VIPER System. VIPER 2 System. and VIPER PRIME are intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The EXPEDIUM Growing Spine System is indicated for patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The EXPEDIUM Growing Spine System may be used with any cleared traditional 4.5 and 5.5
{18}------------------------------------------------
# K191212
Page 9 of 10
EXPEDIUM Spine Systems. The EXPEDIUM Growing Spine System is not intended to be used with 4.0mm diameter screws.
# EXPEDIUM® Fenestrated Screw System, VIPER® Fenestrated Screw System, and VIPER PRIME™ Fenestrated Screws
When used in conjunction with CONFIDENCE High Viscosity Spinal Cement, the EXPEDIUM Fenestrated Screw System, VIPER Fenestrated Screw System, and VIPER PRIME Fenestrated Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The EXPEDIUM Fenestrated Screw System, VIPER Fenestrated Screw System, and VIPER PRIME Fenestrated Screw System augmented with the CONFIDENCE High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
#### H. Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use
The subject devices maintain the design characteristics of the predicate devices. Intended use remains consistent with the predicate devices. The subject devices are provided with additional labeling language regarding magnetic resonance (MR) compatibility.
#### I. Materials
The subject device materials remain identical to the predicate device materials, which consist of Implant Grade Pure Titanium or Titanium Alloy conforming to ASTM F-67, ASTM F-136 or ASTM F-1472 specifications; Stainless Steel conforming to ASTM F-138, ASTM F-134, or F-2229 specifications; Cobalt-Nickel-Chromium-Molybdenum Alloy Wire conforming to ASTM F-562 specifications; and Cobalt-Chromium-Molybdenum Alloy conforming to ASTM F-1537 specifications.
#### J. Performance Data
Non-clinical testing was conducted in alignment with the following standards:
- ASTM F2213 Standard Test Method for Measurement of Magnetically Induced . Torque on Medical Devices in the Magnetic Resonance Environment
- ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment
- . ASTM F2119 Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants
{19}------------------------------------------------
- ASTM F2182 Standard Test Method for Measurement of Radio Frequency . Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging
Results demonstrated compatibility conditions of the subject devices in the MR environment.
#### K. Conclusion
Evaluation of subject device intended use and technological characteristics demonstrates substantial equivalence with the predicate devices. Performance data supports the addition of magnetic resonance compatibility information to subject device labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.