CONFIDENCE™ High Viscosity Spinal Cement, VIPER® and EXPEDIUM® Fenestrated Screw Systems
K160879 · Medos International Sàrl · PML · Dec 20, 2016 · Orthopedic
Device Facts
Record ID
K160879
Device Name
CONFIDENCE™ High Viscosity Spinal Cement, VIPER® and EXPEDIUM® Fenestrated Screw Systems
Applicant
Medos International Sàrl
Product Code
PML · Orthopedic
Decision Date
Dec 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When the CONFIDENCE 114 High Viscosity Spinal Cement is used in conjunction with the VIPER® and EXPEDIUM® Fenestrated Screw Systems, the cement is integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER® and EXPEDIUM® Fenestrated Screw Systems augmented with the CONFIDENCE™ High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System comprises CONFIDENCE™ High Viscosity Spinal Cement (PMMA) and VIPER®/EXPEDIUM® Fenestrated Screw Systems. Screws feature fully cannulated shanks with lateral distal fenestrations. Physician places screws posteriorly into prepared bone canal; cement delivered through cannulated screw and out through fenestrations into vertebral body. Cement hardens in situ to augment screw fixation and restore spinal column integrity. Used in clinical/surgical settings for patients with advanced spinal tumors. Provides stabilization and immobilization of spinal segments; benefits patient by restoring structural integrity without requiring fusion in terminal cases.
Clinical Evidence
Non-clinical performance testing and clinical literature were submitted. Testing characterizes the cement and screw components. Data supports substantial equivalence to predicate devices regarding safety and effectiveness.
Indicated for patients with advanced stage tumors involving the thoracic and lumbar spine with limited life expectancy where fusion is not achievable. Used for spinal levels where structural integrity is not severely compromised.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2016
Medos International Sárl % Jeffrey Shiffman, J.D., M.S. Senior Regulatory Affairs Specialist DePuy Spine, Inc. 325 Paramount Drive Raynham, Massachusetts 02767
Re: K160879
Trade/Device Name: CONFIDENCE™ High Viscosity Spinal Cement, VIPER® and EXPEDIUM® Fenestrated Screw Systems Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: PML, MNI Dated: November 17, 2016 Received: November 18, 2016
Dear Mr. Shiffman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160879
Device Name CONFIDENCE™ High Viscosity Spinal Cement
## Indications for Use (Describe)
When the CONFIDENCE 114 High Viscosity Spinal Cement is used in conjunction with the VIPER® and EXPEDIUM® Fenestrated Screw Systems, the cement is integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER® and EXPEDIUM® Fenestrated Screw Systems augmented with the CONFIDENCE™ High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
D Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160879
Device Name
VIPER® and EXPEDIUM® Fenestrated Screw Systems
### Indications for Use (Describe)
When used in conjunction with CONFIDENCE™ High Viscosity Spinal Cement, the VIPER® and EXPEDUM® Fenestrated Screw Systems are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER® and EXPEDIUM® Fenestrated Screw Systems augmented with the CONFIDENCE™ High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
D Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| A. | Submitter Information | | |
|----|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Manufacturer: | Medos International Sárl<br>Chemin-Blanc 38<br>2400 Le Locle, Switzerland | |
| | Submitter: | DePuy Spine, Inc.<br>325 Paramount Drive<br>Raynham, MA 02767 | |
| | Contact Person: | Jeffrey Shiffman, J.D., M.S.<br>325 Paramount Drive<br>Raynham, MA 02767 | |
| | Telephone number: | 508-977-3857 | |
| | Fax number: | 508-828-3797 | |
| | Email: | jshiffma@its.jnj.com | |
| B. | Date Prepared | December 16, 2016 | |
| C. | Device Name | | |
| | Trade/Proprietary Name: | 1) CONFIDENCE™ High Viscosity Spinal<br>Cement<br>2) VIPER® and EXPEDIUM® Fenestrated Screw<br>Systems | |
| | Common/Usual Name: | 1) Polymethylmethacrylate (PMMA) bone cement;<br>2) Pedicle screw spinal system | |
| | Classification Name and<br>Product Classification Code: | 1) Bone cement, posterior screw augmentation per<br>21 CFR §888.3027<br>PML<br>2) Orthosis, Pedicle screw spinal fixation<br>per 21 CFR §888.3070:<br>MNI | |
### D. Predicate Device Name
Primary: Medtronic KYPHON HV-R Fenestrated Screw Cement, CD HORIZON Fenestrated Screw Set, K152604 Reference: EXPEDIUM Verse Spine System, K142185 EXPEDIUM and VIPER Cortical Fix Screws, K110216 CONFIDENCE High Viscosity Spinal Cement, K112907, K060300
{5}------------------------------------------------
#### E. Device Description
- 1) The CONFIDENCE™ High Viscosity Spinal Cement is a self-curing, high viscosity, radiopaque acrylic bone cement (PMMA), previously cleared by FDA in premarket notifications K060300 and K112907. The formulation of the spinal cement has not changed from the original clearance.
- 2) The VIPER® and EXPEDIUM® Fenestrated Screw Systems are designed with a cortical fix screw shank that is fully cannulated with lateral fenestrations at the distal end which allow for the controlled delivery of CONFIDENCE™ High Viscosity Spinal Cement. The design of the distally fenestrated screws provides a controlled means to deliver PMMA cement into the vertebral body. The screws are available in diameters of 4.35, 5.0, 6.0, 7.0, 8.0, 9.0 and 10.0mm and lengths of 30-80mm (in 5mm increments). Screws with lengths of 30mm have three fenestrations and screw lengths of 35mm and longer have six fenestrations. The VIPER® fenestrated screws are compatible with the VIPER System, VIPER2 and EXPEDIUM 5.5 Spine Systems (K111136) and the EXPEDIUM fenestrated screws are compatible with the EXPEDIUM VERSE® System (K142185).
#### F. Intended Use
- 1) When CONFIDENCE 100 High Viscosity Spinal Cement is used in conjunction with the VIPER® and EXPEDIUM® Fenestrated Screw Systems, the cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER® and EXPEDIUM® Fenestrated Screw Systems augmented with the CONFIDENCE 110 High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
- 2) When used in conjunction with CONFIDENCE™ High Viscosity Spinal Cement, the VIPER® and EXPEDIUM® Fenestrated Screw Systems are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The VIPER® and EXPEDIUM® Fenestrated Screw Systems augmented with the CONFIDENCE ! 10 High Viscosity Spinal Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
{6}------------------------------------------------
# G. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use
- 1) CONFIDENCE™ High Viscosity Spinal Cement is cleared via K060300 and K112907. The sterilization method for the subject cement has not changed from K060300 and K112907, and incorporates bacterial endotoxin testing applying the 20 EU/mL pyrogen limit specification utilizing the Limulus amebocyte lysate (LAL) test.
The difference between the predicate and subject cement is the delivery method of the cement. Although the cement delivery system is not changing, the additional delivery method and expanded indications enables the subject cement to be delivered through the cannulated fenestrated screw for a controlled flow of cement into the pedicle.
CONFIDENCE™ High Viscosity Spinal Cement is similar to the Medtronic KYPHON HV-R Fenestrated Screw Cement (K152604) as both devices are viscous, polymethylmethacrylate material that hardens over time. After a screw is placed posteriorly within a prepared bone canal, the cement is delivered through the cannulated screw and out through the distal fenestrations. Both the predicate and proposed devices are intended to augment the fixation of screws in a posterior spinal system construct.
- 2) The VIPER® and EXPEDIUM® Fenestrated Screw Systems are similar to Medtronic CD Horizon Fenestrated screws (K152604). The proposed screw designs are similar to the predicate with fully cannulated screw shanks with side fenestrations at the distal tip to allow cement delivery. The intended use is similar as both the predicate and proposed devices are intended to provide immobilization and stabilization of spinal segments.
#### H. Materials
- 1) The powder component of the CONFIDENCE™ High Viscosity Spinal Cement is composed of acrylic polymer, benzoyl peroxide and barium sulphate. The liquid component of the CONFIDENCE™ High Viscosity Spinal Cement contains both methyl methacrylate monomer, hydroquinone (HQ) and N. N-Dimethyl-p-Toluidine (DMpT). The complete material composition of CONFIDENCE™ High Viscosity Spinal Cement can be found in K060300 and K112907.
- 2) The VIPER® and EXPEDIUM® Fenestrated Screw Systems are manufactured from ASTM F 136 implant grade titanium alloy and ASTM F1537 cobalt chromium alloy.
#### I. Performance Data
Non-clinical testing and clinical data were submitted to characterize the subject components in this notification. The data provided supports the substantial equivalence of the subject devices.
{7}------------------------------------------------
#### J. Conclusion
The performance testing, clinical literature, and substantial equivalence justification demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.