NuVasive Reline Fenestrated Screws, High V+ Bone Cement
K180498 · Nu Vasive, Incorporated · PML · Nov 28, 2018 · Orthopedic
Device Facts
Record ID
K180498
Device Name
NuVasive Reline Fenestrated Screws, High V+ Bone Cement
Applicant
Nu Vasive, Incorporated
Product Code
PML · Orthopedic
Decision Date
Nov 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
High V+ Bone Cement: When used in conjunction with the Nu Vasive Reline Fenestrated Screws, High V+ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+ Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised. NuVasive Reline Fenestrated Screws: When used in conjunction with High V+ Bone Cement, the Nu Vasive Reline Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NuVasive Reline Fenestrated Screws augmented with High V+ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System consists of fenestrated pedicle screws and high-viscosity PMMA bone cement; used for spinal stabilization in patients with advanced stage tumors. Screws feature distal fenestrations allowing controlled injection of bone cement into vertebral bodies to restore spinal integrity without fusion. Cement is a self-curing, radiopaque PMMA paste (powder/liquid components). Used in clinical settings by surgeons; cement is mixed and injected through screws. Output is a stabilized spinal construct. Benefits include temporary restoration of spinal integrity in palliative oncology cases where fusion is not feasible.
Clinical Evidence
No clinical data. Bench testing only. Mechanical performance evaluated via axial pull-out and removal torque testing per ASTM F2193. Cement flow and bolus formation compared against predicate devices.
Technological Characteristics
PMMA-based bone cement (powder: PMMA, barium sulfate, hydroxyapatite, benzoyl peroxide; liquid: MMA monomer, N,N-dimethyl-p-toluidine, hydroquinone). Fenestrated titanium-alloy screws compatible with 5.0mm-6.0mm rods. Self-curing polymerization. Sterile. No software/electronics.
Indications for Use
Indicated for patients with advanced stage tumors of the thoracic and lumbar spine with insufficient life expectancy to achieve fusion, where spinal structural integrity is not severely compromised.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue box with the letters FDA in white. To the right of the blue box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 28, 2018
NuVasive, Incorporated Ms. Olga Lewis Senior Manager, Medical Affairs 7475 Lusk Boulevard San Diego, California 92121
Re: K180498
Trade/Device Name: High V+ Bone Cement, NuVasive® Reline® Fenestrated Screws Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: PML, NKB, KWP Dated: October 25, 2018 Received: October 26, 2018
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ronald P. Jean -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180498
### Device Name
NuVasive® Reline® Fenestrated Screws with High V+ Bone Cement
Indications for Use (Describe) NuVasive® Reline® Fenestrated Screws
When used in conjunction with High V+ Bone Cement, the Nu Vasive Reline Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NuVasive Reline Fenestrated Screws augmented with High V+ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
High V+ Bone Cement
When used in conjunction with the Nu Vasive Reline Fenestrated Screws, High V+ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+ Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray, and the tagline "Speed of Innovation" in a smaller font below. The logo is clean and modern, with a focus on innovation and technology.
A.
### 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
Submitted by: Olga Lewis Senior Manager, Medical Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121
Date Prepared: October 25, 2018
#### B. Device Name
| Trade or Proprietary Name: | NuVasive® Reline® Fenestrated Screws |
|----------------------------|--------------------------------------------------------|
| Common or Usual Name: | Intervertebral Body Fusion Device |
| Classification Name: | Pedicle Screw Spinal System, Spinal Interlaminal |
| | Fixation Orthosis, Spinal Intervertebral Body Fixation |
| Device Class: | Class II |
| Classification: | 21 CFR §888.3070 |
| Product Code: | NKB, KWP |
| Trade or Proprietary Name: | High V+® Bone Cement |
| Common or Usual Name: | Bone cement |
| Classification Name: | Polymethylmethacrylate (PMMA) bone cement |
| Device Class: | Class II |
| Classification: | 21 CFR § 888.3027 |
| Product Code: | PMI |
#### C. Predicate Devices
The subject NuVasive Reline Fenestrated Screws are substantially equivalent to the primary predicate device, the NuVasive Reline System (K160989) and the additional predicate device, Medtronic Sofamor Danek USA, Inc., CD HORIZON Fenestrated Screw Set (K152604).
The subject High V+ Bone Cement is substantially equivalent to the predicate device, the Teknimed High V+ (K161114) and the additional predicate device, Medtronic Sofamor Danek USA, Inc., KYPHON HV-R Fenestrated Screw Cement (K152604).
#### D. Device Description
The NuVasive Reline Fenestrated Screws are available in a variety sizes and are designed to connect to 5.0mm, 5.5mm, and 6.0mm diameter rods and associated connecting components contained within the NuVasive Reline System. The screws contain fenestrations near the distal tip of the screw which provides a controlled means to deliver a small amount of polymethylmethacrylate (PMMA) bone cement into a targeted vertebral body. These screws are provided sterile or non-sterile.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font.
The High V+ Bone Cement is a self-curing PMMA based (high viscosity, radiopaque) bone cement. High V+ Bone Cement is provided sterile in two components: 20 grams of powder and 8.6 grams of liquid. The powder contains polymethylmethacrylate, barium sulfate and hydroxyapatite as radiopacifier, and benzoyl peroxide as an initiator. The liquid contains methylmethacrylate monomer, N, N dimethyl-p-toluidine as a promoter and hydroquinone as a stabilizer. The powder and liquid components are mixed, in the provided liquid-to-powder proportions, into a homogenous paste, to initiate the polymerization reaction of monomer into polymer.
The purpose of this 510(k) is to expand the indications for use for the NuVasive Reline Fenestrated Screws to include its use with High V+ Bone Cement in the treatment of advanced stage tumors of the thoracic and lumbar spine.
#### E. Intended Use
### High V+ Bone Cement:
When used in conjunction with the NuVasive Reline Fenestrated Screws, High V+ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+ Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
### NuVasive Reline Fenestrated Screws:
When used in conjunction with High V+Bone Cement, the NuVasive Reline Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NuVasive Reline Fenestrated Screws augmented with High V+Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
#### F. Technological Characteristics
As was established in this submission, the subject High V+ Bone Cement is identical in composition, method of manufacture and sterilization to the primary predicate Teknimed High V+ cleared by FDA in K161114. In addition, the subject High V+ Bone Cement is substantially equivalent to the predicate Medtronic Sofamor Danek USA, Inc., KYPHON HV-R Fenestrated Screw Cement cleared by the FDA in K152604 for use with CD HORIZON Fenestrated Screws in the treatment of advanced stage tumors involving the thoracic and lumbar spine.
As was established in this subject NuVasive Reline Fenestrated Screws are identical in materials and substantially equivalent in design to the primary predicate NuVasive Reline System cleared by FDA in K160989 and are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function.
{5}------------------------------------------------
#### G. Performance Data
To compare the mechanical properties of the NuVasive Reline Fenestrated Screws with High V+ Bone Cement and Medtronic CD Horizon Fenestrated Screw with Kyphon HV-R Bone Cement, axial pull-out and removal torque testing was performed according to ASTM F2193. Additionally, a comparison study of the cement flow and bolus formation of the the subject and predicate device was performed.
The subject devices were found to be substantially equivalent to the predicates. No clinical studies were conducted.
#### H. Conclusions
Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject NuVasive Reline Fenestrated Screws and High V+ Bone Cement have been shown to be substantially equivalent to legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.