The HIGH V+ is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Device Story
HIGH V+ is a self-hardening, radiopaque PMMA bone cement; used for stabilization of fractured vertebral bodies via vertebroplasty or kyphoplasty procedures. Device consists of two sterile components: powder (PMMA, hydroxyapatite, barium sulfate, benzoyl peroxide) and liquid monomer (methyl methacrylate). Components mixed intraoperatively; injected into vertebral body to provide consolidation and pain relief. Used in clinical settings by physicians; provides structural stabilization of fractures. Benefits include rapid pain relief and vertebral body consolidation.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including powder morphology, molecular weight, monomer elution, viscosity, setting time, handling times, compressive/bending strength, bending modulus, dynamic fatigue, radiopacity, glass transition, and bacterial endotoxin testing.
Technological Characteristics
PMMA bone cement; powder (PMMA, hydroxyapatite, barium sulfate, benzoyl peroxide) and liquid (methyl methacrylate, dimethyl-paratoluidine, hydroquinone). Complies with ISO 5833. Sterilization: Gamma radiation (powder), filtration (liquid), ethylene oxide (blister).
Indications for Use
Indicated for fixation of pathological vertebral body fractures in patients with painful vertebral compression fractures due to osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 12, 2017
Teknimed, SAS % Mr. J.D. Webb Authorized Contact Person The OrthoMedix Group, Inc. 1001 Oakwood Boulevard Round Rock. Texas 78681
Re: K161114
Trade/Device Name: HIGH V+ Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN Dated: April 10, 2017 Received: April 11, 2017
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Vincent J. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K161114
Device Name
HIGH V+
Indications for Use (Describe)
The HIGH V+ is used for the fixation of pathological fractures of the vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lessons (hemangiona), or malignant lesions (metastatic cancers, myeloma).
Type of Use (Select one or both, as applicable)
X | Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
।
{3}------------------------------------------------
#### 510(k) Summary for the HIGH V+ Bone cement
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the HIGH V+.
#### 1. GENERAL INFORMATION
| Date Prepared: | November 11, 2016 |
|----------------|-------------------|
|----------------|-------------------|
Trade Name: HIGH V+
Common Name: Polymethylmethacrylate (PMMA) bone cement
Classification Name: Cement, Bone, Vertebroplasty
Class: II
Product Code: NDN
CFR section: 21 CFR section 888.3027
Device panel: Orthopedic
PrimaryLegally Vertecem, K090435 Marketed Predicate Device:
Secondary Legally Spineplex K032945 Marketed Predicate Device:
- Submitter: Teknimed, SAS 11, rue d'Apollo 31240 L'Union France
- Contact: J.D. Webb 1001 Oakwood Blvd Round Rock, TX 78681 512-388-0199 Tele 512-692-3699 Fax e-mail: jdwebb@orthomedix.net
{4}------------------------------------------------
#### 2. DEVICE DESCRIPTION
HIGH V+ is a self-hardening and ready to use Poly Methylmethacrylatebone cement with 36.4% of radiopaque agentsfor the treatment of painfulvertebral fractures based on the predicate devices Spineplex. Vertecem. It can be injected directly into the fractured vertebral body by either Vertebroplasty or Kylphoplasty procedures to relieve pain. HIGH V+ allows an excellent consolidation of the vertebral body and an effective and rapid pain relief. The HIGH V+ cement is made of two sterile components: the polymer in powder and the liquid monomer. These two components are in a double sterile packaging. Each unit contains a sterile ampoule of liquid within a blister pack and a powder within a double peelable pouch, the whole being packaged in a box.
#### Material:
The liquid component is mainly composed of methyl methacrylate. The major powder components are polymethylmethacrylate (PMMA), hydroxyapatite and barium sulfate.Benzoyl peroxide which initiates the polymerization is included in the polymer powder.
#### Function:
The HIGH V+ bone cement allows stabilization of the vertebral body and a rapid pain relief.
#### 3. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES
HIGH V+ has similar technological characteristics as the predicate devices as shown in the detailed side by side comparison in the following table.
| Device<br>Items | HIGH V+ | Vertecem | Spineplex |
|-------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------|
| Sponsor | Teknimed | Teknimed | Stryker |
| 510(k) Number | -- | K090435 | K032945 |
| Product Code | NDN | NDN | NDN |
| Indications for Use | [2] below table | [1] below table | [3] below table |
| Chemical composition | | | |
| Powder (% w/w) | 20g ± 0.4g | 21g ± 0.4g | 20g |
| Polymethymethacrylate | 63.1% ± 1% | 43.4% ± 1% | 11.5% |
| Methyl methacrylate<br>styrene copolymer | -- | 16.1% ± 1% | 57% |
| Benzoyl peroxide | 0.5% ± 0.1% | 0.5% ± 0.1% | 1.5% |
| Barium sulfate | 27.3% +2.1% - 1.9% | 30% +2% -3% | 30% |
| Hydroxyapatite | 9.1% +1.9% / -2.1% | 10% ± 1% | -- |
| Liquid (% w/w) | 8.6g ± 0.2g | 9.2 ± 0.2g | 9.4g |
| Methylmethacrylate | 98.5% ± 1% | 99.3% ± 1% | 97.4% |
| Device<br>Items | HIGH V+ | Vertecem | Spineplex |
| Dimethyl-paratoluidine | 1.5% ± 0.1% | 0.7% ± 0.1% | 2.6% |
| Hydroquinone | 20ppm± 5ppm | 20ppm± 5ppm | 75ppm |
| Physical<br>characteristics | | | |
| Molecular weights:<br>Powder | 534,511 g/mole | 222,518 g/mole | -- |
| Cured cement | 646,966 g/mole | 394 552 g/mole | |
| Working phases @ 20°C | Mixing: 0.5 min<br>• Waiting: 3 min<br>• Application: 9 min<br>• Setting time: 16min | Mixing: 3 min<br>• Waiting: 11.5 min<br>• Application: 25 min<br>• Setting time: 30min | |
| Compressive strength | 81.91 MPa | 80.3 MPa | -- |
| Dynamic tension-<br>compression mean<br>fatigue | 4,627,970±847,458<br>cycles @ 7Mpa | 4,317,676±1,237,780<br>cycles @ 7Mpa | 3,799,425±1,990,989<br>cycles @ 7Mpa |
| Bending strength | 57.22 MPa | 52.6 MPa | -- |
| Bending modulus | 3338.82 | 4173 MPa | -- |
| Radiopacity | Equivalent radiopacity | Equivalent radiopacity | Equivalent radiopacity |
| Monomer elution testing | Increase release till<br>10 minutes after<br>mixing, then<br>stabilization | Increase release till<br>10 minutes after<br>mixing, then<br>stabilization | -- |
| Sterilization method | | | |
| Powder | Gamma radiation<br>(ISO 11137) | Gamma radiation<br>(ISO 11137) | Gamma radiation<br>(ISO 11137) |
| Liquid | Filtration (ISO 13408) | Filtration (ISO 13408) | Filtration (ISO 13408) |
| Inner Blister | Ethylene oxide (ISO<br>11135-1) | Ethylene oxide (ISO<br>11135-1) | Ethylene oxide (ISO<br>11135-1) |
{5}------------------------------------------------
{6}------------------------------------------------
#### 4. INTENDED USE
The HIGH V+ is used for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures of the vertebral body may result from osteoporosis, benign lesions (hemangioma), or malignant lesions (metastatic cancers, myeloma).
#### 5. NON-CLINICAL TEST SUMMARY
The following tests were conducted:
- · Powder morphology
- · Molecular weights
- . Monomer elution testing
- . Viscosity vs. temperature
- . Setting time vs. temperature
- Handling times .
- . Compressive strength
- Bending strength .
- Bending modulus .
- Dynamic tension-compression fatigue test .
- · Radiopacity
- Glass transition .
- Bacterial Endotoxins Test (BET) to establish non-pyrogenicity .
Test data indicate that the final properties of HIGH V+ are in compliance with the standard reference for bone cement: ISO 5833 "implants for surgery - acrylic resin cements" and are substantially equivalent to those of predicate devices.
#### 6. CLINICAL TEST SUMMARY
No clinical studies were performed.
#### 7. CONCLUSIONS NONCLINICAL AND CLINICAL
This summary includes the conclusions drawn from the nonclinical tests (discussed above) that demonstrate that HGH V+ bone cement is as safe, as effective, and perform as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.