CranioMaxillofacial Fixation (CMF) System - CMF Visionare
Applicant
Visionare, LLC
Product Code
JEY · Dental
Decision Date
Dec 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
Device Story
The CMF Visionare system is a fixation device consisting of various titanium plates and screws. It is used by surgeons in the OR to provide rigid fixation for bone fractures, osteotomies, and reconstructive procedures in the cranio-maxillofacial skeleton. The device functions by mechanically stabilizing bone segments using plates and screws, which are implanted to maintain positional and functional stability during the healing process. The system includes a wide range of plate geometries (straight, T, Y, L, H, X, Z, orbital, and orthognathic-specific designs) and screw types (locking, non-locking, self-drilling, self-tapping) to accommodate different anatomical sites and surgical needs. The surgeon selects the appropriate plate and screw configuration based on the specific fracture or reconstruction requirements. The device provides structural support to the bone, facilitating proper alignment and healing, which benefits the patient by restoring function and anatomy.
Clinical Evidence
Bench testing only. No clinical data were included. Evidence includes sterilization validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 14937), biocompatibility (ISO 10993-5, ISO 10993-12), dimensional analysis, engineering analysis, and mechanical testing (ASTM F382 for bending fatigue, ASTM F543 for torsional properties, driving torque, axial pullout, and self-tapping performance).
Technological Characteristics
Plates: Commercially pure (CP) titanium (ASTM F67). Screws: Ti-6Al-4V alloy (ASTM F136). System includes various plate geometries (straight, T, Y, L, H, X, Z, orbital, orthognathic) and screw diameters (1.5 mm to 2.7 mm). Mechanical fixation via screws and plates. Provided non-sterile; sterilized via moist heat (steam).
Indications for Use
Indicated for patients requiring fixation in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible, and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Mini and Micro Fragments Reconstruction System - NEOFACE, NEOORTHO Produtos Ortopédicos S/A (K102641)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December, 28, 2018
Visionare LLC % Kevin Thomas, Ph.D Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K180204
Trade/Device Name: CranioMaxillofacial Fixation (CMF) System - CMF Visionare Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY, DZL Dated: November 26, 2018 Received: November 27, 2018
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 Date: 2018.12.28
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K180204
Device Name
CranioMaxillofacial Fixation (CMF) System - CMF Visionare
#### Indications for Use (Describe)
CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:12px;"> <span style="font-family:Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size:12px;"> <span style="font-family:Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K180204 CranioMaxillofacial Fixation (CMF) System – CMF Visionare Visionare LLC
November 26, 2018
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Visionare LLC<br>12251 Towne Lake Drive<br>Fort Meyers, FL 33913<br>Telephone +1 239-244-2973 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Mariana de Oliveira Quinzani<br>Regulatory Affairs |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | CranioMaxillofacial Fixation (CMF) System - CMF Visionare |
|-------------------------------------|-----------------------------------------------------------|
| Common Names | Bone plate; Fixation screws |
| Primary Classification Name | Bone Plate |
| Primary Classification Regulation | 21 CFR 872.4760, Class II |
| Primary Product Code | JEY |
| Secondary Classification Name | Intraosseous fixation screw or wire |
| Secondary Classification Regulation | 21 CFR 872.4880, Class II |
| Secondary Product Code | DZL |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
flarson@paxmed.com
#### PREDICATE DEVICE INFORMATION
Primary Predicate K102641, Mini and Micro Fragments Reconstruction System - NEOFACE, NEOORTHO Produtos Ortopédicos S/A
Reference Devices K021642, Synthes Craniofacial Plates, Synthes (USA) K050934, MODUS® Titanium Osteosynthesis System, Medartis, Inc. K033065, 2.0 mm Craniofacial Locking Plates, Synthes (USA)
{4}------------------------------------------------
K972322, Lorenz 1.5mm Neuro Pack/Lorenz 2.0 FT Plates (sterile version), Walter Lorenz Surgical, Inc. K080694, OsteoMed Modular Locking Fixation System, OsteoMed LP
K091144, Synthes MatrixMANDIBLE Preformed Reconstruction Plates, Synthes USA
K992682. MODUS 2.5 Mandibular Reconstruction Set. Medartis
K063052, Mandibular Fracture/Reconstruction Devices and Pre-Bent Plates, Walter Lorenz Surgical, Incorporated
K980199, Synthes® (USA) 1.5mm/2.0mm Orthognathic Maxillary Plates and Screws, Synthes® (USA) K031708, Osteomed Maxillary/LeFort III Distraction System, OsteoMed L.P.
K944565, KLS-Martin Micro Osteosynthesis System (1.5 mm), KLS-Martin L.P.
K052061, MODUS IMF SCREWS 2.0, Medartis, Inc.
### INDICATIONS FOR USE STATEMENT
CranioMaxillofacial Fixation (CMF) System – CMF Visionare is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
### SUBJECT DEVICE DESCRIPTION
The subject device is a fixation system consisting of plates and screws in a variety of shapes and sizes. All plates are made of commercially pure titanium conforming to ASTM F67, and all screws are made of Ti-6Al-4V alloy conforming to ASTM F136. The subject device includes the following groups: 1.5 System Plate, 2.0 System Plates, 2.4 System Plates, Orthognathic Plates, Blocking Screws (for intermaxillary fixation, 2.0 mm diameter), and screws for use with the various plates ranging in diameter from 1.5 mm to 2.7 mm. The dimensions of the plates are provided in the comparison tables in this summary.
### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO 14937; biocompatibility testing according to ISO 10993-5 and ISO 10993-12; dimensional analysis; engineering analysis; mechanical testing to ASTM F382 (single cycle bending fatigue) and ASTM F543 (torsional properties, driving torque, axial pullout, self-tapping performance). No clinical data were included in this submission.
### EQUIVALENCE TO MARKETED DEVICE
Visionare LLC submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed devices cited above.
The subject device is substantially equivalent to the primary predicate device K102641 in intended use, Indications for Use, designs, materials, and function. The Indications for Use statement for the subject device is identical to that of the primary predicate K102641 except for the names of the devices. A comparison of the Indications for Use Statements and technological characteristics of the subject device, the primary predicate device, and the reference devices is provided in the following tables. Differences in the design features between the subject device and the primary predicate device K102641 are addressed by comparison to the reference devices as listed below.
{5}------------------------------------------------
### Table 1 – Substantial Equivalence 1.5 System Plates
| | Subject Device | | Primary Predicate Device | | Reference Device | | | | | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|------------------------------------|-------------|--------------------------------------------------------------------|--|
| Comparisons | K180204<br>Mini and Micro Fragments Reconstruction System - CMF Visionare<br>Visionare LLC | | K102641<br>Mini and Micro Fragments Reconstruction System - NEOFACE<br>NEOORTHO Produtos Ortopédicos S/A | | K021642<br>Synthes Craniofacial Plates<br>Synthes (USA) | | | | | |
| | | | | | | | | | | |
| Indications for Use<br>Statement | CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in<br>selective trauma of the midface, maxillofacial surgery, reconstructive procedures,<br>and selective orthognathic surgery of the maxilla, mandible and chin. | | Mini and Micro Fragments Reconstruction System - NEOFACE is intended<br>for use in selective trauma of the midface, maxillofacial surgery,<br>reconstructive procedures, and selective orthognathic surgery of the maxilla,<br>mandible and chin. | | The Synthes Craniofacial Plates are intended for use in select<br>the midface and craniofacial skeleton; craniofacial surgery re<br>procedures; and selective orthognathic surgery of the maxilla | | | | | |
| Plate Designs for<br>Comparison | 1.5 Straight Plates<br>1.5 Rectangular Plates<br>1.5 T Plates<br>1.5 Y Plates<br>1.5 Double Y Plates<br>1.5 L Plates<br>1.5 L 110° Plates<br>1.5 H Plates<br>1.5 X Plates<br>1.5 Z Plates<br>1.5 Orbital Plates<br>1.5 Orbital Floor Plates | | Straight Plates<br>Rectangular Plates<br>T Plates<br>Y Plates<br>Double Y Plates<br>L Plates<br>L 110° Plates<br>H Plates<br>X Plates<br>Z Plates<br>Orbital Plates | | Rectangular Plates<br>Y Plates<br>Double Y Plates<br>L Plates<br>H Plates<br>Z Plates<br>Orbital Plates | | | | | |
| Comparison of Designs | | Thickness (mm) | Overall Sizes (mm) | Thickness (mm) | Overall Sizes (mm) | Thickness (mm) | Overall Sizes (mm) | | | |
| 1.5 Straight Plates | 1.5 Straight Plates | 0.7 | 8.8 - 79.2 | Straight Plates | 0.7 | 10.3 - 79.2 | | | | |
| 1.5 Rectangular Plates | 1.5 Rectangular Plates | 0.7 | | Rectangular Plates | 0.7 | | Rectangular Plates<br>(Box Plates) | 0.7 | | |
| | 2x2 hole | | (7.2 - 10.3) x 10.3 | 2x2 hole | | 11.7 x 11.7 | 2x2 (holes) | | (5.0) | |
| | 2x3 hole | | 17.4 x 10.3 | 2x3 hole | | 17.4 x 10.3 | | | | |
| | 2x4 hole | | 24.4 x 10.3 | 2x4 hole | | 24.5 x 10.3 | | | | |
| 1.5 T Plates | 1.5 T Plates | 0.7 | | T Plates | 0.7 | | | | | |
| | 3x2 hole | | (11.2 - 15.7) x 11.2 | None | | | | | | |
| | 3x3 hole | | (15.2 - 19.7) x 11.2 | 3x3 hole | | 15.2 x 11.2 | | | | |
| | 3x4 hole | | (19.2 - 23.7) x 11.2 | 7x7 hole | | 31.2 x 27.2 | | | | |
| 1.5 Y Plates | 1.5 Y Plates | 0.7 | | Y Plates | 0.7 | | Y Plates | 0.6 | | |
| | 3x3 hole | | (18.0 - 22.5) x 8.9 | 3x2 hole | | (15.6 - 18.5) x 8.9 | 3x3 hole | 22.0 | | |
| | 5x4 hole | | (24.9 - 29.4) x 14.5 | None | | | | | | |
| 1.5 Double Y Plates | 1.5 Double Y Plates | 0.7 | | Double Y Plates | 0.7 | | Double Y Plates | | | |
| | 3x3 hole | | (13.1 - 17.4) x 8.9 | 3x3 hole | | (14.5-17.4) x 8.9 | | | | |
| | 5x5 hole | | (18.5 - 23.0) x 14.5 | None | | | 0.6 | (18.0) | | |
| 1.5 L Plates | 1.5 L Plates | 0.7 | | L Plates | 0.7 | | | | | |
| | 2x2 hole | | (11.4 - 15.7) x 7.2 | 2x2 hole | | (12.8-15.7) x 7.2 | | | | |
| 1.5 L 110° Plates | 1.5 L 110° Plates | 0.7 | | L 110° Plates | 0.7 | | L Oblique Plates | 0.6 | | |
| | 3x3 hole | | (18.1 - 22.4) x 11.2 | 3x3 hole | | (19.7-21.0) x 11.2 | 2x3 hole | 22.0 | | |
| | 3x4 hole | | (22.1 - 26.4) x 11.2 | None | | | 3x4 hole | 27.0 | | |
| 1.5 H Plates | 1.5 H Plates | 0.7 | 16.7 x 15.2 | H Plates | 0.7 | 20.4 x 19.2 | H Plates | 0.5 | 19.0 | |
| 1.5 X Plates | 1.5 X Plates | 0.7 | 11.2 x 11.2 | X Plates | 0.7 | 11.2 x 11.2 | | | | |
| 1.5 Z Plates | 1.5 Z Plates | 0.7 | (10.5 - 13.4) x 11.2 | Z Plates | 0.7 | (11.7-14.6) x 11.1 | Z Plates | 0.6 | (10.9-1 | |
| | Subject Device | | | Primary Predicate Device | | Reference Device | | | Reference Device | |
| Comparisons | K180204<br>Mini and Micro Fragments Reconstruction System - CMF Visionare<br>Visionare LLC | | | K102641<br>Mini and Micro Fragments Reconstruction System – NEOFACE<br>NEOORTHO Produtos Ortopédicos S/A | | K021642<br>Synthes Craniofacial Plates<br>Synthes (USA) | | | K050934<br>MODUS® Titanium Osteosynthesis System<br>Medartis, Inc. | |
| | | | | | | | | | | |
| 1.5 Orbital Plates | 1.5 Orbital Plates | 0.7 | | Orbital Plates | 0.7 | | Orbital Plates | 0.6 | | |
| | 6 hole | | 23.0 - 27.3 | 6 hole | | 23.0 | 6 hole | | 34.0 | |
| | 8 hole | | 30.6 - 34.8 | 8 hole | | 30.6 | 8 hole | | 41.0 | |
| | 10 hole | | 37.9 - 41.9 | 10 hole | | 37.9 | 10 hole | | 49.0 | |
| | | | | | | | | | | |
| 1.5 Orbital Floor Plates | 1.5 Orbital Floor Plates | 0.3 | | Orbital Floor Plates | 0.3 | | Orbital Floor Plates | | | |
| | Orbital Floor Plate, T 0.3 mm | | 59.7 x 36.8 x 0.30 mm | | | | Orbital Floor Plate | (0.2 - 0.4) | 62.2 x 38.5 | |
| | Regular Anatomical Orbital Floor<br>Plate, T 0.3 mm | | 29.0 x 17.0 x 0.30 mm | | | | Small | 0.3 | 21.0 x 13.3 | |
| | Medium Anatomical Orbital Floor<br>Plate, T 0.3 mm | | 47.7 x 39.6 x 0.30 mm | | | | Medium | 0.3 | 30.6 x 26.6 | |
| | Wide Medium Anatomical Orbital<br>Floor Plate, T 0.3 mm | | 53.1 x 39.6 x 0.30 mm | | | | Large | 0.3 | 41.3 x 34.6 | |
| | Large Anatomical Orbital Floor<br>Plate, T 0.3 mm | | 61.1 x 53.5 x 0.30 mm | | | | | | | |
| | Universal Orbital Floor Plate,<br>T 0.3 mm | | 58.1 x 52.1 x 0.30 mm | Orbital Floor Plate | | 56.7 x 49.9 | | | | |
| | | | | | | | | | | |
| Plate Material | CP Titanium | | | CP Titanium | | CP Titanium | | | CP Titanium | |
| Reference Device | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|----------------|----------------------------------------------|
| K050934<br>MODUS® Titanium Osteosynthesis System<br>Medartis, Inc. | | | |
| The MODUS® Titanium Osteosynthesis System is intended for osteotomies and fractures involving any part of the craniofacial skeleton and requiring positional and functional stability. Indications include fixation in the nasoethmoidal, intraorbital, and frontal sinus areas; fixation of comminuted fractures of maxillo-facial and craniofacial areas; tumor surgery for defect bridging; reconstruction of bony structures by means of mesh materials; coverings for burr holes in the skull; trauma of nasal bones; surgical correction of dentofacial deformations; and reconstruction after tumor surgery. | | | |
| | T Plates<br>Y Plates<br>L 110° Plates | | |
| Overall Sizes (mm) | | Thickness (mm) | Overall Sizes (mm) |
| (8.0 - 10.0) x 10 | | | |
| | T Plates<br>3x2 hole<br>3x3 hole<br>3x4 hole | 0.6 | 12.0 x 12.0<br>17.0 x 12.0<br>21.0 x 12.0 |
| (8.0 x 11.0 | Y Plates<br>3x2 hole<br>5x4 hole | 0.6<br>0.7 | (16.0 - 20.0) x 10.0<br>(25.0 - 30.0) x 16.0 |
| (18.0 - 21.0) x 11.4 | | | |
| (8.0 x 9.0<br>(8.0 x 13.0 | L 110° Plates<br>2x2 hole | 0.6 | (18.0 - 21.0) x 8.0 |
| (8.0 x 13.0 | | | |
| (10.9-14.7) x 15 | | | |
{6}------------------------------------------------
# Table 1 – Substantial Equivalence 1.5 System Plates
{7}------------------------------------------------
### Table 2 – Substantial Equivalence 2.0 System Plates
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | | | | | | | | |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------------------|-----------------------|---------------------------------------------|------------------------------------|-----------------------------|-----------------------|
| Comparison | K180204<br>Mini and Micro Fragments<br>Reconstruction System – CMF<br>Visionare<br>Visionare LLC | K102641<br>Mini and Micro Fragments<br>Reconstruction System – NEOFACE<br>NEOORTHO Produtos Ortopédicos S/A | K021642<br>Synthes Craniofacial Plates<br>Synthes (USA) | K033065<br>2.0 mm Craniofacial Locking Plates<br>Synthes (USA) | K972322<br>Lorenz 1.5mm Neuro Pack/Lorenz 2.0<br>FT Plates (sterile version)<br>Walter Lorenz Surgical, Inc. | K050934<br>MODUS® Titanium Osteosynthesis<br>System<br>Medartis, Inc. | K080694<br>OsteoMed Modular Locking Fixation<br>System<br>OsteoMed LP | | | | | | | |
| Indications for Use<br>Statement | CranioMaxillofacial Fixation (CMF)<br>System - CMF Visionare is intended for<br>use in selective trauma of the midface,<br>maxillofacial surgery, reconstructive<br>procedures, and selective orthognathic<br>surgery of the maxilla, mandible and chin. | Mini and Micro Fragments Reconstruction<br>System - NEOFACE is intended for use in<br>selective trauma of the midface, maxillofacial<br>surgery, reconstructive procedures, and<br>selective orthognathic surgery of the maxilla,<br>mandible and chin. | The Synthes Craniofacial Plates are<br>intended for use in selective trauma of the<br>midface and craniofacial skeleton;<br>craniofacial surgery; reconstructive<br>procedures; and selective orthognathic<br>surgery of the maxilla and chin. | Synthes 2.0 mm Craniofacial Locking<br>Plates are indicated for selective trauma of<br>the midface and craniofacial skeleton;<br>craniofacial surgery; reconstructive<br>procedures; and selective orthognathic<br>surgery of the maxilla and chin. | Rigid fixation of cranio, facial and<br>neurological bone. | The MODUS® Titanium Osteosynthesis<br>System is intended for osteotomies and<br>fractures involving any part of the<br>craniofacial skeleton and requiring positional<br>and functional stability. Indications include<br>fixation in the nasoethmoidal, intraorbital,<br>and frontal sinus areas; fixation of<br>comminuted fractures of maxillo-facial and<br>craniofacial areas; tumor surgery for defect<br>bridging; reconstruction of bony structures by<br>means of mesh materials; coverings for burr<br>holes in the skull; trauma of nasal bones;<br>surgical correction of dentofacial<br>deformations; and reconstruction after tumor<br>surgery. | The OsteoMed Modular Locking Fixation<br>System is intended for fracture fixation in<br>cranio-maxillofacial trauma reconstruction,<br>mandibular reconstruction and orthognathic<br>reconstruction.<br>The OsteoMed Modular Locking Fixation<br>System implants and drills are intended for<br>single use only. | | | | | | |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.