CranioMaxillofacial Fixation (CMF) System - CMF Visionare
Applicant
Visionare, LLC
Product Code
JEY · Dental
Decision Date
Dec 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Intended Use
CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
Device Story
The CMF Visionare system is a fixation device consisting of various titanium plates and screws. It is used by surgeons in the OR to provide rigid fixation for bone fractures, osteotomies, and reconstructive procedures in the cranio-maxillofacial skeleton. The device functions by mechanically stabilizing bone segments using plates and screws, which are implanted to maintain positional and functional stability during the healing process. The system includes a wide range of plate geometries (straight, T, Y, L, H, X, Z, orbital, and orthognathic-specific designs) and screw types (locking, non-locking, self-drilling, self-tapping) to accommodate different anatomical sites and surgical needs. The surgeon selects the appropriate plate and screw configuration based on the specific fracture or reconstruction requirements. The device provides structural support to the bone, facilitating proper alignment and healing, which benefits the patient by restoring function and anatomy.
Clinical Evidence
Bench testing only. No clinical data were included. Evidence includes sterilization validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 14937), biocompatibility (ISO 10993-5, ISO 10993-12), dimensional analysis, engineering analysis, and mechanical testing (ASTM F382 for bending fatigue, ASTM F543 for torsional properties, driving torque, axial pullout, and self-tapping performance).
Technological Characteristics
Plates: Commercially pure (CP) titanium (ASTM F67). Screws: Ti-6Al-4V alloy (ASTM F136). System includes various plate geometries (straight, T, Y, L, H, X, Z, orbital, orthognathic) and screw diameters (1.5 mm to 2.7 mm). Mechanical fixation via screws and plates. Provided non-sterile; sterilized via moist heat (steam).
Indications for Use
Indicated for patients requiring fixation in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible, and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Mini and Micro Fragments Reconstruction System - NEOFACE, NEOORTHO Produtos Ortopédicos S/A (K102641)
K032520 — ORTRAUTEK TRAUMA SYSTEM · Tekka · Nov 12, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December, 28, 2018
Visionare LLC % Kevin Thomas, Ph.D Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K180204
Trade/Device Name: CranioMaxillofacial Fixation (CMF) System - CMF Visionare Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY, DZL Dated: November 26, 2018 Received: November 27, 2018
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 Date: 2018.12.28
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K180204
Device Name
CranioMaxillofacial Fixation (CMF) System - CMF Visionare
#### Indications for Use (Describe)
CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:12px;"> <span style="font-family:Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size:12px;"> <span style="font-family:Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K180204 CranioMaxillofacial Fixation (CMF) System – CMF Visionare Visionare LLC
November 26, 2018
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Visionare LLC<br>12251 Towne Lake Drive<br>Fort Meyers, FL 33913<br>Telephone +1 239-244-2973 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Mariana de Oliveira Quinzani<br>Regulatory Affairs |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | CranioMaxillofacial Fixation (CMF) System - CMF Visionare |
|-------------------------------------|-----------------------------------------------------------|
| Common Names | Bone plate; Fixation screws |
| Primary Classification Name | Bone Plate |
| Primary Classification Regulation | 21 CFR 872.4760, Class II |
| Primary Product Code | JEY |
| Secondary Classification Name | Intraosseous fixation screw or wire |
| Secondary Classification Regulation | 21 CFR 872.4880, Class II |
| Secondary Product Code | DZL |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
flarson@paxmed.com
#### PREDICATE DEVICE INFORMATION
Primary Predicate K102641, Mini and Micro Fragments Reconstruction System - NEOFACE, NEOORTHO Produtos Ortopédicos S/A
Reference Devices K021642, Synthes Craniofacial Plates, Synthes (USA) K050934, MODUS® Titanium Osteosynthesis System, Medartis, Inc. K033065, 2.0 mm Craniofacial Locking Plates, Synthes (USA)
{4}------------------------------------------------
K972322, Lorenz 1.5mm Neuro Pack/Lorenz 2.0 FT Plates (sterile version), Walter Lorenz Surgical, Inc. K080694, OsteoMed Modular Locking Fixation System, OsteoMed LP
K091144, Synthes MatrixMANDIBLE Preformed Reconstruction Plates, Synthes USA
K992682. MODUS 2.5 Mandibular Reconstruction Set. Medartis
K063052, Mandibular Fracture/Reconstruction Devices and Pre-Bent Plates, Walter Lorenz Surgical, Incorporated
K980199, Synthes® (USA) 1.5mm/2.0mm Orthognathic Maxillary Plates and Screws, Synthes® (USA) K031708, Osteomed Maxillary/LeFort III Distraction System, OsteoMed L.P.
K944565, KLS-Martin Micro Osteosynthesis System (1.5 mm), KLS-Martin L.P.
K052061, MODUS IMF SCREWS 2.0, Medartis, Inc.
### INDICATIONS FOR USE STATEMENT
CranioMaxillofacial Fixation (CMF) System – CMF Visionare is intended for use in selective trauma of the midface, maxillofacial surgery, reconstructive procedures, and selective orthognathic surgery of the maxilla, mandible and chin.
### SUBJECT DEVICE DESCRIPTION
The subject device is a fixation system consisting of plates and screws in a variety of shapes and sizes. All plates are made of commercially pure titanium conforming to ASTM F67, and all screws are made of Ti-6Al-4V alloy conforming to ASTM F136. The subject device includes the following groups: 1.5 System Plate, 2.0 System Plates, 2.4 System Plates, Orthognathic Plates, Blocking Screws (for intermaxillary fixation, 2.0 mm diameter), and screws for use with the various plates ranging in diameter from 1.5 mm to 2.7 mm. The dimensions of the plates are provided in the comparison tables in this summary.
### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO 14937; biocompatibility testing according to ISO 10993-5 and ISO 10993-12; dimensional analysis; engineering analysis; mechanical testing to ASTM F382 (single cycle bending fatigue) and ASTM F543 (torsional properties, driving torque, axial pullout, self-tapping performance). No clinical data were included in this submission.
### EQUIVALENCE TO MARKETED DEVICE
Visionare LLC submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed devices cited above.
The subject device is substantially equivalent to the primary predicate device K102641 in intended use, Indications for Use, designs, materials, and function. The Indications for Use statement for the subject device is identical to that of the primary predicate K102641 except for the names of the devices. A comparison of the Indications for Use Statements and technological characteristics of the subject device, the primary predicate device, and the reference devices is provided in the following tables. Differences in the design features between the subject device and the primary predicate device K102641 are addressed by comparison to the reference devices as listed below.
{5}------------------------------------------------
### Table 1 – Substantial Equivalence 1.5 System Plates
| | Subject Device | | Primary Predicate Device | | Reference Device | | | | | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|------------------------------------|-------------|--------------------------------------------------------------------|--|
| Comparisons | K180204<br>Mini and Micro Fragments Reconstruction System - CMF Visionare<br>Visionare LLC | | K102641<br>Mini and Micro Fragments Reconstruction System - NEOFACE<br>NEOORTHO Produtos Ortopédicos S/A | | K021642<br>Synthes Craniofacial Plates<br>Synthes (USA) | | | | | |
| | | | | | | | | | | |
| Indications for Use<br>Statement | CranioMaxillofacial Fixation (CMF) System - CMF Visionare is intended for use in<br>selective trauma of the midface, maxillofacial surgery, reconstructive procedures,<br>and selective orthognathic surgery of the maxilla, mandible and chin. | | Mini and Micro Fragments Reconstruction System - NEOFACE is intended<br>for use in selective trauma of the midface, maxillofacial surgery,<br>reconstructive procedures, and selective orthognathic surgery of the maxilla,<br>mandible and chin. | | The Synthes Craniofacial Plates are intended for use in select<br>the midface and craniofacial skeleton; craniofacial surgery re<br>procedures; and selective orthognathic surgery of the maxilla | | | | | |
| Plate Designs for<br>Comparison | 1.5 Straight Plates<br>1.5 Rectangular Plates<br>1.5 T Plates<br>1.5 Y Plates<br>1.5 Double Y Plates<br>1.5 L Plates<br>1.5 L 110° Plates<br>1.5 H Plates<br>1.5 X Plates<br>1.5 Z Plates<br>1.5 Orbital Plates<br>1.5 Orbital Floor Plates | | Straight Plates<br>Rectangular Plates<br>T Plates<br>Y Plates<br>Double Y Plates<br>L Plates<br>L 110° Plates<br>H Plates<br>X Plates<br>Z Plates<br>Orbital Plates | | Rectangular Plates<br>Y Plates<br>Double Y Plates<br>L Plates<br>H Plates<br>Z Plates<br>Orbital Plates | | | | | |
| Comparison of Designs | | Thickness (mm) | Overall Sizes (mm) | Thickness (mm) | Overall Sizes (mm) | Thickness (mm) | Overall Sizes (mm) | | | |
| 1.5 Straight Plates | 1.5 Straight Plates | 0.7 | 8.8 - 79.2 | Straight Plates | 0.7 | 10.3 - 79.2 | | | | |
| 1.5 Rectangular Plates | 1.5 Rectangular Plates | 0.7 | | Rectangular Plates | 0.7 | | Rectangular Plates<br>(Box Plates) | 0.7 | | |
| | 2x2 hole | | (7.2 - 10.3) x 10.3 | 2x2 hole | | 11.7 x 11.7 | 2x2 (holes) | | (5.0) | |
| | 2x3 hole | | 17.4 x 10.3 | 2x3 hole | | 17.4 x 10.3 | | | | |
| | 2x4 hole | | 24.4 x 10.3 | 2x4 hole | | 24.5 x 10.3 | | | | |
| 1.5 T Plates | 1.5 T Plates | 0.7 | | T Plates | 0.7 | | | | | |
| | 3x2 hole | | (11.2 - 15.7) x 11.2 | None | | | | | | |
| | 3x3 hole | | (15.2 - 19.7) x 11.2 | 3x3 hole | | 15.2 x 11.2 | | | | |
| | 3x4 hole | | (19.2 - 23.7) x 11.2 | 7x7 hole | | 31.2 x 27.2 | | | | |
| 1.5 Y Plates | 1.5 Y Plates | 0.7 | | Y Plates | 0.7 | | Y Plates | 0.6 | | |
| | 3x3 hole | | (18.0 - 22.5) x 8.9 | 3x2 hole | | (15.6 - 18.5) x 8.9 | 3x3 hole | 22.0 | | |
| | 5x4 hole | | (24.9 - 29.4) x 14.5 | None | | | | | | |
| 1.5 Double Y Plates | 1.5 Double Y Plates | 0.7 | | Double Y Plates | 0.7 | | Double Y Plates | | | |
| | 3x3 hole | | (13.1 - 17.4) x 8.9 | 3x3 hole | | (14.5-17.4) x 8.9 | | | | |
| | 5x5 hole | | (18.5 - 23.0) x 14.5 | None | | | 0.6 | (18.0) | | |
| 1.5 L Plates | 1.5 L Plates | 0.7 | | L Plates | 0.7 | | | | | |
| | 2x2 hole | | (11.4 - 15.7) x 7.2 | 2x2 hole | | (12.8-15.7) x 7.2 | | | | |
| 1.5 L 110° Plates | 1.5 L 110° Plates | 0.7 | | L 110° Plates | 0.7 | | L Oblique Plates | 0.6 | | |
| | 3x3 hole | | (18.1 - 22.4) x 11.2 | 3x3 hole | | (19.7-21.0) x 11.2 | 2x3 hole | 22.0 | | |
| | 3x4 hole | | (22.1 - 26.4) x 11.2 | None | | | 3x4 hole | 27.0 | | |
| 1.5 H Plates | 1.5 H Plates | 0.7 | 16.7 x 15.2 | H Plates | 0.7 | 20.4 x 19.2 | H Plates | 0.5 | 19.0 | |
| 1.5 X Plates | 1.5 X Plates | 0.7 | 11.2 x 11.2 | X Plates | 0.7 | 11.2 x 11.2 | | | | |
| 1.5 Z Plates | 1.5 Z Plates | 0.7 | (10.5 - 13.4) x 11.2 | Z Plates | 0.7 | (11.7-14.6) x 11.1 | Z Plates | 0.6 | (10.9-1 | |
| | Subject Device | | | Primary Predicate Device | | Reference Device | | | Reference Device | |
| Comparisons | K180204<br>Mini and Micro Fragments Reconstruction System - CMF Visionare<br>Visionare LLC | | | K102641<br>Mini and Micro Fragments Reconstruction System – NEOFACE<br>NEOORTHO Produtos Ortopédicos S/A | | K021642<br>Synthes Craniofacial Plates<br>Synthes (USA) | | | K050934<br>MODUS® Titanium Osteosynthesis System<br>Medartis, Inc. | |
| | | | | | | | | | | |
| 1.5 Orbital Plates | 1.5 Orbital Plates | 0.7 | | Orbital Plates | 0.7 | | Orbital Plates | 0.6 | | |
| | 6 hole | | 23.0 - 27.3 | 6 hole | | 23.0 | 6 hole | | 34.0 | |
| | 8 hole | | 30.6 - 34.8 | 8 hole | | 30.6 | 8 hole | | 41.0 | |
| | 10 hole | | 37.9 - 41.9 | 10 hole | | 37.9 | 10 hole | | 49.0 | |
| | | | | | | | | | | |
| 1.5 Orbital Floor Plates | 1.5 Orbital Floor Plates | 0.3 | | Orbital Floor Plates | 0.3 | | Orbital Floor Plates | | | |
| | Orbital Floor Plate, T 0.3 mm | | 59.7 x 36.8 x 0.30 mm | | | | Orbital Floor Plate | (0.2 - 0.4) | 62.2 x 38.5 | |
| | Regular Anatomical Orbital Floor<br>Plate, T 0.3 mm | | 29.0 x 17.0 x 0.30 mm | | | | Small | 0.3 | 21.0 x 13.3 | |
| | Medium Anatomical Orbital Floor<br>Plate, T 0.3 mm | | 47.7 x 39.6 x 0.30 mm | | | | Medium | 0.3 | 30.6 x 26.6 | |
| | Wide Medium Anatomical Orbital<br>Floor Plate, T 0.3 mm | | 53.1 x 39.6 x 0.30 mm | | | | Large | 0.3 | 41.3 x 34.6 | |
| | Large Anatomical Orbital Floor<br>Plate, T 0.3 mm | | 61.1 x 53.5 x 0.30 mm | | | | | | | |
| | Universal Orbital Floor Plate,<br>T 0.3 mm | | 58.1 x 52.1 x 0.30 mm | Orbital Floor Plate | | 56.7 x 49.9 | | | | |
| | | | | | | | | | | |
| Plate Material | CP Titanium | | | CP Titanium | | CP Titanium | | | CP Titanium | |
| Reference Device | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|----------------|----------------------------------------------|
| K050934<br>MODUS® Titanium Osteosynthesis System<br>Medartis, Inc. | | | |
| The MODUS® Titanium Osteosynthesis System is intended for osteotomies and fractures involving any part of the craniofacial skeleton and requiring positional and functional stability. Indications include fixation in the nasoethmoidal, intraorbital, and frontal sinus areas; fixation of comminuted fractures of maxillo-facial and craniofacial areas; tumor surgery for defect bridging; reconstruction of bony structures by means of mesh materials; coverings for burr holes in the skull; trauma of nasal bones; surgical correction of dentofacial deformations; and reconstruction after tumor surgery. | | | |
| | T Plates<br>Y Plates<br>L 110° Plates | | |
| Overall Sizes (mm) | | Thickness (mm) | Overall Sizes (mm) |
| (8.0 - 10.0) x 10 | | | |
| | T Plates<br>3x2 hole<br>3x3 hole<br>3x4 hole | 0.6 | 12.0 x 12.0<br>17.0 x 12.0<br>21.0 x 12.0 |
| (8.0 x 11.0 | Y Plates<br>3x2 hole<br>5x4 hole | 0.6<br>0.7 | (16.0 - 20.0) x 10.0<br>(25.0 - 30.0) x 16.0 |
| (18.0 - 21.0) x 11.4 | | | |
| (8.0 x 9.0<br>(8.0 x 13.0 | L 110° Plates<br>2x2 hole | 0.6 | (18.0 - 21.0) x 8.0 |
| (8.0 x 13.0 | | | |
| (10.9-14.7) x 15 | | | |
{6}------------------------------------------------
# Table 1 – Substantial Equivalence 1.5 System Plates
{7}------------------------------------------------
### Table 2 – Substantial Equivalence 2.0 System Plates
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | | | | | | | | |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------------------|-----------------------|---------------------------------------------|------------------------------------|-----------------------------|-----------------------|
| Comparison | K180204<br>Mini and Micro Fragments<br>Reconstruction System – CMF<br>Visionare<br>Visionare LLC | K102641<br>Mini and Micro Fragments<br>Reconstruction System – NEOFACE<br>NEOORTHO Produtos Ortopédicos S/A | K021642<br>Synthes Craniofacial Plates<br>Synthes (USA) | K033065<br>2.0 mm Craniofacial Locking Plates<br>Synthes (USA) | K972322<br>Lorenz 1.5mm Neuro Pack/Lorenz 2.0<br>FT Plates (sterile version)<br>Walter Lorenz Surgical, Inc. | K050934<br>MODUS® Titanium Osteosynthesis<br>System<br>Medartis, Inc. | K080694<br>OsteoMed Modular Locking Fixation<br>System<br>OsteoMed LP | | | | | | | |
| Indications for Use<br>Statement | CranioMaxillofacial Fixation (CMF)<br>System - CMF Visionare is intended for<br>use in selective trauma of the midface,<br>maxillofacial surgery, reconstructive<br>procedures, and selective orthognathic<br>surgery of the maxilla, mandible and chin. | Mini and Micro Fragments Reconstruction<br>System - NEOFACE is intended for use in<br>selective trauma of the midface, maxillofacial<br>surgery, reconstructive procedures, and<br>selective orthognathic surgery of the maxilla,<br>mandible and chin. | The Synthes Craniofacial Plates are<br>intended for use in selective trauma of the<br>midface and craniofacial skeleton;<br>craniofacial surgery; reconstructive<br>procedures; and selective orthognathic<br>surgery of the maxilla and chin. | Synthes 2.0 mm Craniofacial Locking<br>Plates are indicated for selective trauma of<br>the midface and craniofacial skeleton;<br>craniofacial surgery; reconstructive<br>procedures; and selective orthognathic<br>surgery of the maxilla and chin. | Rigid fixation of cranio, facial and<br>neurological bone. | The MODUS® Titanium Osteosynthesis<br>System is intended for osteotomies and<br>fractures involving any part of the<br>craniofacial skeleton and requiring positional<br>and functional stability. Indications include<br>fixation in the nasoethmoidal, intraorbital,<br>and frontal sinus areas; fixation of<br>comminuted fractures of maxillo-facial and<br>craniofacial areas; tumor surgery for defect<br>bridging; reconstruction of bony structures by<br>means of mesh materials; coverings for burr<br>holes in the skull; trauma of nasal bones;<br>surgical correction of dentofacial<br>deformations; and reconstruction after tumor<br>surgery. | The OsteoMed Modular Locking Fixation<br>System is intended for fracture fixation in<br>cranio-maxillofacial trauma reconstruction,<br>mandibular reconstruction and orthognathic<br>reconstruction.<br>The OsteoMed Modular Locking Fixation<br>System implants and drills are intended for<br>single use only. | | | | | | |…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.