Bowel Management Device, Bowel Management Irrigation Set
K180026 · Applied Medical Technology, Inc. · KNT · Jun 1, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K180026
Device Name
Bowel Management Device, Bowel Management Irrigation Set
Applicant
Applied Medical Technology, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Jun 1, 2018
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Bowel Management Device is intended to instill fluids through a stoma into the colon to promote evacuation of the contents of the lower bowel through the anus and is intended to be an aid in the management of fecal incontinence. The catheter is placed and maintained in a percutaneously prepared opening, such as a cecostomy or appendicostomy. The Bowel Management Device is intended to be used in children and adults. The Bowel Management Irrigation Set is intended to allow connection between a compatible bowel management device and the delivery tubing of an irrigation/enema bag system.
Device Story
Indwelling low-profile catheter; facilitates antegrade colonic irrigation to manage fecal incontinence. Device inserted through stoma into colon; secured by internal retention balloon. External bolster includes balloon fill-valve, interlock irrigation port, and safety plug. Irrigation set connects catheter to external irrigation/enema bag system via rotating lock or ENFit connector. Fluids instilled directly into colon to promote lower bowel evacuation. Used in clinical or home settings; operated by patients or caregivers. Benefits include non-invasive bowel evacuation management via stoma. Device features anti-reflux valve and MR conditional status.
Clinical Evidence
Bench testing only. Performance testing included strap/tube tensile testing, overmold bond strength, stoma pullout, leak testing, and flow rate testing. Balloon integrity and performance evaluated per ASTM F2528-06. Connector performance evaluated per ISO 80369-3 (fluid leakage, stress cracking, separation resistance). All components met or exceeded design specifications.
Indicated for children and adults requiring antegrade colonic irrigation for fecal incontinence management via a percutaneously prepared stoma (e.g., cecostomy or appendicostomy).
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 1, 2018
Applied Medical Technology, Inc. Joshua Meinke Quality / Regulatory Affairs Supervisor 8006 Katherine Boulevard Brecksville, OH 44141
Re: K180026
> Trade/Device Name: Bowel Management Device and Irrigation Set Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT, EXD, PIF Dated: April 23, 2018 Received: April 26, 2018
Dear Joshua Meinke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180026
Device Name Bowel Management Device and Irrigation Set
#### Indications for Use (Describe)
The Bowel Management Device is intended to instill fluids through a stoma into the colon to promote evacuation of the contents of the lower bowel through the anus and is intended to be an aid in the management of fecal incontinence. The catheter is placed and maintained in a percutaneously prepared opening, such as a cecostomy or appendicostomy. The Bowel Management Device is intended to be used in children and adults.
The Bowel Management Irrigation Set is intended to allow connection between a compatible bowel management device and the delivery tubing of an irrigation/enema bag system.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## SECTION - 5
### 510(k) Summary
#### Bowel Management Device and Irrigation Set
| Date Prepared: | May 31, 2018 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Applied Medical Technology, Inc.<br>8006 Katherine Boulevard<br>Brecksville, OH 44141<br>Phone: 440-717-4000<br>Fax: 440-717-4200<br>Contact Person: Joshua Meinke – Quality / Regulatory Affairs Supervisor<br>Email: Joshua.meinke@appliedmedical.net |
| Device<br>Information: | Trade/Device Name: Bowel Management Device and Irrigation Set<br>Common Name: Gastrointestinal Tube<br>Classification Name: 21 CFR 876.5980: Gastrointestinal tube and<br>accessories;<br>21 CFR 876.5895: Ostomy irrigator.<br>Product Code: KNT, EXD, PIF<br>Regulatory Class: II<br>Review Panel: Gastroenterology and Urology |
| Predicate<br>Device: | Primary Predicate: Cook Chait Cecostomy Catheter (K982500)<br>Reference Device: AMT Low Profile Balloon Feeding Device (K161413) |
| Intended Use: | The Bowel Management Device is intended to instill fluids through a stoma<br>into the colon to promote evacuation of the contents of the lower bowel<br>through the anus and is intended to be an aid in the management of fecal<br>incontinence. The catheter is placed and maintained in a percutaneously<br>prepared opening, such as a cecostomy or appendicostomy. The Bowel<br>Management Device is intended to be used in children and adults.<br>The Bowel Management Irrigation Set is intended to allow connection<br>between a compatible bowel management device and the delivery tubing of an<br>irrigation/enema bag system. |
| Device<br>Description: | The Bowel Management Device is an indwelling low profile catheter used to<br>promote evacuation of the lower bowel through antegrade colonic flushing. It<br>consists of an internal retention balloon and a flexible external bolster with a<br>balloon fill-valve, interlock irrigation port, and safety plug. The catheter is<br>inserted through the stoma and into the colon and is secured by the internal<br>retention balloon when inflated. The Bowel Management Irrigation Set has a<br>connector on one end that connects to the interlock on the catheter and an<br>adapter on the other end to connect to the irrigation/enema delivery system.<br>Fluids can then be instilled directly into the colon through the catheter. |
| Technological<br>Characteristics: | Antegrade colonic irrigation is the technological principle for both the subject<br>and predicate device. It is based on the use of a stoma to instill fluids into the<br>colon to evacuate the contents of the lower bowel through the anus. At a high<br>level, the subject and predicate devices are based on the following same<br>technological elements:<br>• Catheter tubing – to instill fluids into the colon<br>• External and internal bolster – to secure the device to the abdominal<br>wall and prevent inward or outward migration of the device<br>• Plug on external bolster – to prevent leakage<br>• Introducer/stiffener – to facilitate placement of the device<br>• Irrigation set/access adapter - connects the catheter to the delivery system<br>The following technological differences exist between the subject and<br>predicate device:<br>• The subject device is made of a soft, flexible silicone, while the<br>predicate is made of urethane/polyurethane<br>• The subject device uses a balloon as the internal bolster instead of the<br>catheter coiling like the predicate<br>• The subject device features one central outlet instead of having three<br>side ports like the predicate device<br>• The subject device comes in various sizes to accommodate different<br>stoma diameters and lengths and the predicate device is available in<br>three adjustable sizes<br>• The subject device has an anti-reflux valve<br>• The subject device has an irrigation set that rotates to lock into place<br>• The subject device has an ENFit connector option for the irrigation<br>set<br>• The subject device is MR conditional |
| Biocompatibility<br>Testing: | The Bowel Management Tube in its final finished form is identical to the Low<br>Profile Balloon Feeding Device (K161413) in formulation, processing,<br>sterilization, and geometry, and no other chemicals have been added (eg.<br>plasticizers, fillers, additives, cleaning agents, and mold release agents), so no<br>additional testing was required for this submission. |
| Performance<br>Testing: | AMT conducted various performance tests on the components contained<br>within the Bowel Management Device. Testing found that all components<br>and materials met or exceeded design specifications established by AMT.<br><br>In addition to testing included in K161413, bench tests have been carried out<br>on to compare performance to the primary predicate. The additional tests<br>carried out included:<br>Strap Tensile Testing Tube Tensile Testing Minimum Overmold Bond Strength Stoma Pullout 2 Minute Leak Test Flow Rate Test Testing per ASTM F2528-06: Balloon Integrity in Simulated Gastric Fluid Balloon volume maintenance Balloon size and shaft size Balloon concentricity Balloon integrity Testing per 80369-3: Fluid leakage Stress cracking Resistance to separation from axial load Resistance to separation from unscrewing Resistance to overriding Disconnection by unscrewing The Bowel Management Devices meet all the acceptance criteria and<br>performed comparable to or better than the primary predicate. |
| Conclusion: | The Bowel Management Device can be found substantially equivalent to the<br>predicate device cleared under K982500 in intended use, performance, and<br>principles of operation. The AMT Bowel Management Device is also<br>identical to the LPBFD cleared in K161413, so the technological<br>characteristics do not raise different questions of safety and effectiveness. |
{5}------------------------------------------------
## SECTION - 5
{6}------------------------------------------------
## SECTION - 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.