Zygomatic Implant System

K173343 · Southern Implants (Pty), Ltd. · DZE · Feb 27, 2018 · Dental

Device Facts

Record IDK173343
Device NameZygomatic Implant System
ApplicantSouthern Implants (Pty), Ltd.
Product CodeDZE · Dental
Decision DateFeb 27, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

Southern Implants Zygomatic System Standard implants, Zygan (narrow apex) implants, and Oncology implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Story

Zygomatic Implant System consists of root-form dental implants and mating abutments for placement into zygomatic bone. Implants include Standard, Zygan (narrow apex), and Oncology designs; all feature 55° angled external hex abutment interface. Abutments include Compact Conical and Titanium Cylinder designs for screw-retained overdenture support. Used by dental surgeons in clinical settings to rehabilitate edentulous maxillae. Implants provide structural support for dental prostheses; immediate loading possible with adequate primary stability. Benefits include restoration of oral function and aesthetics for patients with severe maxillary bone loss.

Clinical Evidence

Bench testing only. Includes engineering analysis, dimensional analysis, and dynamic compression-bending fatigue testing per ISO 14801. Biocompatibility referenced from predicates. Bacterial endotoxin testing per USP <85>. Clinical data for oncology implant design referenced from retrospective study (Boyes-Varley et al., 2007) of 40 implants in 20 subjects, showing 100% success at 96 months.

Technological Characteristics

Materials: Unalloyed titanium (ASTM F67) for implants/cylinder abutments; titanium alloy (ASTM F136) for conical abutments. Surface: Aluminum oxide grit-blasted. Design: Root-form, threaded, 55° head angulation. Dimensions: 3.4mm-4.3mm diameters, 27.5mm-57.5mm lengths. Sterilization: Validated per ISO 17665-1/2. Connectivity: None (mechanical device).

Indications for Use

Indicated for patients with partially or fully edentulous maxillae requiring support for fixed or removable dental prostheses via implantation in the upper jaw arch (zygomatic bone). Appropriate for immediate loading given sufficient primary stability and occlusal loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". February 27, 2018 Southern Implants (Pty) Ltd % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real Suite 400 San Diego, California 92130 Re: K173343 Trade/Device Name: Zygomatic Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: January 25, 2018 Received: January 26, 2018 ## Dear Kevin Thomas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Andrew I. Steen -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use K173343 Device Name Zygomatic Implant System Indications for Use (Describe) Southern Implants Zygomatic System Standard implants, Zygan (narrow apex) implants, and Oncology implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ ## 510(k) Summary - K173343 Zygomatic Implant System Southern Implants (Pty) Ltd February 27, 2018 #### ADMINISTRATIVE INFORMATION | Manufacturer Name | Southern Implants (Pty) Ltd<br>1 Albert Road<br>Irene, Gauteng, 0062 South Africa<br>Telephone +27 12 667 1046<br>Fax +27 12 667 1029 | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Contact | Lauranda G. Breytenbach<br>Head of Regulatory Affairs and Quality | | Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: kthomas@paxmed.com<br>flarson@paxmed.com | #### DEVICE NAME AND CLASSIFICATION | Trade/Proprietary Name | Zygomatic Implant System | |---------------------------|---------------------------| | Common Name | Dental implant | | Classification Name | Endosseous dental implant | | Classification Regulation | 21 CFR 872.3640, Class II | | Primary Product Code | DZE | | Secondary Product Code | NHA | | Classification Panel | Dental Products Panel | | Reviewing Branch | Dental Devices Branch | #### PREDICATE DEVICE INFORMATION The primary predicate device is K093562. The reference devices are K151909, K070841, K053478, and K141777. #### INDICATIONS FOR USE STATEMENT Southern Implants Zygomatic System Standard implants, Zygan (narrow apex) implants, and Oncology implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. {4}------------------------------------------------ ## SUBJECT DEVICE DESCRIPTION This submission includes fully threaded and partially threaded root-form dental implants and mating abutments designed for placement into the zygomatic bone. All implants are provided with an external hexagon abutment interface angled 55° at the head of the implants are provided in three designs: Standard implant, fully threaded, diameter 4.3 mm (coronal) tapering to 3.8 mm (apical), in lengths of 30 mm and 57.5 mm; Zygan implant, partially threaded, diameter 4.3 mm (coronal) tapering to 3.4 mm (apical), in lengths from 30 mm to 57.5 mm; and Oncology implant, partially threaded diameter 4.3 mm (coronal) tapering to 3.8 mm (apical), in lengths from 30 mm to 47. 5 mm. This submission includes additional designs of Compact Conical Abutments (gingival heights 2, 3, 4, and 5.5 mm) for use with any of the implants. This submission also includes a Titanium Cylinder Abutment, for use with the Compact Conical Abutments, with a collar (gingival) height of 5 mm, and a prosthetic platform diameter of 3.4 mm. All subject device abutments are for support of screw-retained overdenture prosthetic restorations. All subject device implants are manufactured from unalloyed titanium conforming to ASTM F67. The threaded portions of the implants have the identical aluminum oxide grit-blasted surface as the implants cleared in K093562. The subject device Compact Conical Abutments are manufactured from titanium alloy conforming to ASTM F136. The subject device Titanium Cylinder Abutment is manufactured from unalloyed titanium conforming to ASTM F67. All of the subject device components are manufactured in the same facilities using the same manufacturing processes as used for the previously cleared predicate devices in K093562 and K070841. ## PERFORMANCE DATA Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K093562 and K070841), engineering analysis, dimensional analysis, sterilization validation testing according to ISO 17665-1, ISO TS 17665-2, and dynamic compressionbending testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants. The Limulus Amebocyte Lysate (LAL) test for detection and quantitation of bacterial endotoxin was conducted in accordance with USP 39-NF 34 <85> Bacterial Endotoxin Test, using the kinetic chromogenic test method. Clinical data referenced in this submission to support the oncology implant design included the retrospective results from 40 implants (in 20 subjects) demonstrating 100% implant success and no soft tissue complications at up to 96 months [Boyes-Varley JG, et al., Int J Prosthodont 2007; 20:521-531]. #### EQUIVALENCE TO MARKETED DEVICE Southern Implants (Ptv) Ltd submits the information in this Premarket Notification to demonstrate that. for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed devices: K093562, Zygomatic Implant System, Southern Implants, Inc.; K151909, Noris Medical Zygomatic Dental Implant System, Noris Medical Ltd .; K070841, Endosseous Dental Implant System, Southern Implants, Inc .; K053478, Endosseous Dental Implant System, Northern Implants, LLC; and K141777. Neodent Implant System. JJGC Indústria e Comércio de Materiais Dentários SA. {5}------------------------------------------------ A comparison of the technological characteristics of the subject device and the primary predicate device K093562 is provided in the following table. | | Subject Device | Primary Predicate Device | |-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison | Zygomatic Implant System<br>Southern Implants (Pty) Ltd | K093562<br>Zygomatic Implant System<br>Southern Implants, Inc. | | Indications for Use<br>Statement | Southern Implants Zygomatic System Standard<br>implants, Zygan (narrow apex) implants, and Oncology<br>implants are intended to be implanted in the upper jaw<br>arch to provide support for fixed or removable dental<br>prostheses in patients with partially or fully edentulous<br>maxillae. All implants are appropriate for immediate<br>loading when good primary stability is achieved and<br>with appropriate occlusal loading. | The Zygomatic implant is intended to be implanted in<br>the upper jaw arch to provide support for fixed or<br>removable dental prostheses in patients with partially<br>or fully edentulous maxillae. | | Implants | | | | Design | Fully and partially threaded root-form implants for<br>placement into the zygoma | Fully threaded root-form implants for placement into<br>the zygoma | | Platform Ø | 4.05 mm | 4.05 mm | | Implant Ø | Standard implant:<br>4.3 mm (coronal) taper to 3.8 mm (apical)<br>Zygan implant:<br>4.3 mm (coronal) taper to 3.4 mm (apical)<br>Oncology implant:<br>4.3 mm (coronal) taper to 3.8 mm (apical) | Standard implant:<br>4.3 mm (coronal) taper to 3.8 mm (apical) | | Implant Lengths | Standard implant: 30 mm, 57.5 mm<br>Zygan implant: 30 mm - 57.5 mm<br>Oncology Implant: 27.5 mm - 47.5 mm | Standard implant: 35 mm - 55 mm | | Threaded Lengths | Standard implant: Fully threaded<br>Zygan implant: 6 mm coronal + 15 mm apical<br>Oncology implant: 20 mm (apical only) | Standard implant: Fully threaded | | Implant body-abutment<br>connection angle | 55° | 55° | | Abutments | | | | Design | One-piece, compact conical design | One-piece, compact conical design | | Implant Interface | External hex; 55° angulation at head of implant | External hex; 55° angulation at head of implant | | Gingival Height | 2.0 mm - 5.5 mm | 1 mm | | Abutment Angle | 0° (straight) | 0° (straight) | | Prosthesis Attachment | Screw-retained, multi-unit | Screw-retained, multi-unit | | Materials | | | | Implants | Unalloyed titanium, ASTM F67 | Unalloyed titanium, ASTM F67 | | Abutments | Unalloyed titanium, ASTM F67;<br>Titanium alloy, ASTM E136 | Unalloyed titanium, ASTM F67:<br>Titanium alloy, ASTM E136 | The Indications for Use Statements for the subject device and the primary predicate device K093562 are similar with the subject device statement including the additional Zygan and Oncology implants. The slight differences in wording between the Indications for Use Statements for the subject device and the primary predicate device do not affect the intended use as dental implants placed into the zygoma for rehabilitation of the edentulous maxilla. The primary predicate device K093562 is for substantial equivalence of the subject device implant designs. The subject device Standard implants have the identical design as implants in K093562 except {6}------------------------------------------------ for the additional lengths. The subject device Zygan implants have a design that is substantially equivalent design to implants in K093562, with the differences being the non-threaded region (starting 6 mm below the head of the implant), the diameter tapering to 3.4 mm at the apex (versus 3.8 mm for Standard implants), and the range of implant lengths. Similarly, the subject device Oncology implants have a design that is substantially equivalent to implants in K093562, with the differences being the nonthreaded region starting below the head of the implant, and the range of implant lengths. The reference device K151909 is for substantial equivalence of the subject device Oncology implant design. The implants cleared in K151909 and the subject device Oncology implants both have a nonthreaded region starting at the coronal end (K151909) or just below the implant head (subject device), and a threaded region that tapers to the apex. The reference device K151909 also is for the range of implant lengths. These differences in design between the subject devices and primary predicate devices were accounted for by including additional instructions in the labeling; therefore, the differences do not impact cafety and effectiveness. The reference device K070841 is for substantial equivalence of the subject device Compact Conical Abutment design. The subject device Compact Conical Abutments have the identical design as components cleared in K070841, with the only difference being additional sizes of gingival height. The reference device K053478 is for substantial equivalence of the subject device Titanium Abutment design. The subject device Titanium Abutment has the identical design as components cleared in K053478, with the only difference being a change in gingival height. The reference device K141777 is for substantial equivalence of the range of implant lengths, and for substantial equivalence of mechanical performance discussed below. The subject device and the predicate devices all incorporate the same materials and encompass similar ranges of dimensions. The surface treatment applied to the subject device implants is identical to that cleared in K093562. The subject device and all predicate devices are provided sterile for single-patient, single-use. The subject device implants are packaged preassembled with a titanium fixture mount. The fixture mount is made from the same unalloyed titanium as the implant-fixture mount assembly is attached to a stainless steel clip within a polyethylene terephthalate (PET) thermoformed tray, and sealed with a Tyvek® lid. The actual implant does not have any contact with the packaging material. The thermoformed tray with Tyvek® lid forms the primary sterile barrier and is placed in a PET box. The product label is placed on the Tyvek® lid. The outer box is sealed with an address label and a product description label. For the predicate device K093562, the sterile barrier is provided by the same packaging. a PET tray (blister) sealed with a Tyvek® Lid. The difference between the subject device packaging and the predicate device K093562 packaging is in the additional packaging of the implant and mount within a rigid plastic cylinder which is placed in the PET tray and sealed with the Tyvek® lid. For both the subject device and the predicate device, the primary sterile barrier is placed in a clear rigid PET box, which does not provide an additional sterile barrier but functions to protect the tray/Tyvek® lid package. The change in packaging has been addressed by specific labeling precautions, with images, for packaging handling and removal of the device into the surgical field, in order to maintain sterility of the device {7}------------------------------------------------ Substantial equivalence of the subject device components in terms of biocompatibility is supported by the materials being identical in formulation, processing, component interactions, and storage conditions to the predicate devices in K093562 and K070841. In support of substantial equivalence in terms of mechanical performance, dynamic compression-bending testing according to ISO 14801 was performed. Dynamic testing was performed on the worst-case subject device constructs. The results from the testing demonstrated fatigue performance substantially equivalent to that of the reference device K141777. #### CONCLUSION The subject device and the predicate devices have the same intended use, have similar technological characteristics, and are made of similar materials. The subject device and the predicate devices encompass the same range of physical dimensions, including diameter and length of the implants, and the diameter and angulation of the abutments. The subject device and the predicate devices are packaged in similar materials and sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...