AxSOS 3 Ti is intended for long bone fracture fixation. Indications include: - Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures - Non-unions and malunions - Normal and osteopenic bone - Osteotomies - Periprosthetic fractures of the femur and proximal tibia. The AxSOS 3 Ti Waisted Compression Plates are also indicated for fracture fixation of: - Periprosthetic fractures - Diaphyseal and metaphyseal areas of long bones in pediatric patients The 4mm Waisted Compression Plate indications also include fixation of the scapula and the pelvis
Device Story
AxSOS 3 Ti is an internal bone fixation system comprising anatomically contoured monoaxial locking plates, screws (locking, cortical, cancellous), and a 5.0mm Cable Plug. The system is used by surgeons in clinical settings to stabilize fractures. The Cable Plug, a new component, threads or clicks into the plate's universal locking hole to provide a positioning point for cerclage cables. The device provides mechanical support for bone healing. It is manufactured from titanium alloy (Ti6Al4V-ELI). The system is designed for compatibility with MR environments.
Clinical Evidence
No clinical data was required or provided for this submission. Substantial equivalence is supported by non-clinical bench testing.
Technological Characteristics
System components manufactured from titanium alloy (Ti6Al4V-ELI) per ASTM F136. Cable Plug is Type II anodized. Mechanical fixation via locking/non-locking screws and cerclage cable positioning. MR compatibility verified per ASTM F2052, F2213, F2182, and F2119. Mechanical performance verified per ASTM F543-13 and ASTM STP 731.
Indications for Use
Indicated for long bone fracture fixation in patients with diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures, non-unions, malunions, osteotomies, and periprosthetic fractures of the femur and proximal tibia. Includes pediatric patients for waisted compression plates and fixation of scapula/pelvis for 4mm plates. Suitable for normal and osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
December 15, 2017
Stryker GmbH Kemine Hale Sr. Regulatory Affairs Specialist Bohnackerweg 1 Selzach, 2545 Switzerland
Re: K172350
Trade/Device Name: AxSOS 3 Ti Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 10, 2017 Received: November 13, 2017
Dear Kemine Hale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K172350
Device Name AxSOS 3 Ti
Indications for Use (Describe)
AxSOS 3 Ti is intended for long bone fracture fixation
Indications include:
- * Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures
- * Non-unions and malunions
- * Normal and osteopenic bone
- * Osteotomies
- * Periprosthetic fractures of the femur and proximal tibia.
The AxSOS 3 Ti Waisted Compression Plates are also indicated for fracture fixation of:
- * Periprosthetic fractures
- * Diaphyseal and metaphyseal areas of long bones in pediatric patients
The 4mm Waisted Compression Plate indications also include fixation of the scapula and the pelvis
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary
| Proprietary Name: | AxSOS 3 Ti |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bone Plates<br>Bone Screws |
| Regulation Number &<br>Regulation Description: | 21 CFR 888.3030: Single/multiple component<br>metallic bone fixation appliances and accessories. |
| | 21 CFR 888.3040: Smooth or threaded metallic bone<br>fixation fastener |
| Product Code: | HRS (Plate, Fixation, bone)<br>HWC (Screw, Fixation, Bone/Washer) |
| Device Class: | Class II |
| Sponsor: | Stryker GmbH<br>Bohnackerweg 1<br>2545 Selzach / Switzerland |
| Contact Person: | Kemine Hale<br>Sr. Regulatory Affairs Specialist<br>Bohnackerweg 1<br>2545 Selzach, Switzerland<br>kemine.hale@stryker.com<br>Phone: +41 32 641 6935 |
| Date Prepared: | December 14, 2017 |
| Primary Predicate: | AxSOS 3 Ti (K153379 & K162439)<br>Dall-Miles Homogenous Vitallium Cable (K961283) |
| Reference Devices: | Fixos Screw System (K133451)<br>Asnis III Cannulated Screw System (K000080) |
510(k) Summary
{4}------------------------------------------------
#### Description
This traditional 510(k) submission is intended to update the labeling of the AxSOS 3 Ti system to include MR compatibility and to introduce an additional component, the AxSOS 3 Ti 5.0mm Cable Plug. The AxSOS 3 Ti system was previously cleared in K153379 and K162439.
The AxSOS 3 Ti system (also referred to as the AxSOS 3 Ti Locking Plate system) is a system used for internal fixation applications. The system includes anatomically contoured monoaxial locking plates, as well as screws and blind screws (limited to waisted compression plates) . The plates have been designed with holes that can accommodate either locking or non-locking screws, and also include holes for suture and Kirschner wires (K-wires). The screws come in 3 types (locking, cortical, and cancellous) and are available in several different diameters. All implantable components are available sterile and non-sterile and are manufactured from titanium alloy (Ti6Al4V-ELI) per ASTM F136.
The AxSOS 3 Ti 5.0mm Cable Plug is categorized as permanent, direct bone/tissue patient contacting implant with contact duration >30 days. It is manufactured from titanium alloy (ASTM F136, Ti6Al4V-ELI) and is Type II anodized. It is designed to be used in combination with the 5.0mm AxSOS 3 Ti system. It is an aid for internal fixation component used in conjunction with the AxSOS 3 Ti 5.0mm distal lateral femur, broad compression and narrow compression plates. The Cable Plug is threaded or clicked into a 5.0mm universal locking hole of the AxSOS 3 Ti plates and provides a positioning point for a cerclage cable.
#### Indications for Use
AxSOS 3 Ti is intended for long bone fracture fixation.
Indications include:
- Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures ●
- . Non-unions and malunions
- Normal and osteopenic bone ●
- Osteotomies
- Periprosthetic fractures of the femur and proximal tibia.
{5}------------------------------------------------
The AxSOS 3 Ti Waisted Compression Plates are also indicated for fracture fixation of:
- Periprosthetic fractures ●
- o Diaphyseal and metaphyseal areas of long bones in pediatric patients
The 4mm Waisted Compression Plate indications also include fixation of the scapula and the pelvis.
## Summary of Technologies
A comparison of the systems demonstrated the subject AxSOS 3 Ti system is substantially equivalent to the previously cleared AxSOS 3 Ti system (K162439) and the Dall-Miles Homogenous Vitallium Cable (K961283) with regards to intended use, material, design and operational principles.
## Non-Clinical Testing
Testing was performed to ASTM F543-13 – Pull Out Test and ASTM STP 731-Dynamic Compression Test for the Cable Plug component.
Testing was performed to determine the compatibility of the system in a MR environment. These tests included an assessment of:
- o Magnetically Induced Displacement Force per ASTM F2052
- Magnetically Induced Torque per ASTM F2213 ●
- Heating by RF Fields per ASTM F2182
- Image Artifacts per ASTM F2119
Testing demonstrated that the subject device is substantially equivalent to the marketed predicate device.
Animal testing was not required for this submission.
#### Clinical Testing
Clinical testing was not required for this submission.
#### Conclusion
510(k) Summary
{6}------------------------------------------------
The AxSOS 3 Ti system described in this submission is identical to the previously cleared predicate, K153379, K162439 and K961283, with the exception of the additional component. Non-clinical testing has been used to support substantial equivalence. The testing demonstrated that the subject system performs as intended and at least as well as the predicate. Based on these attributes the subject AxSOS 3 Ti is substantially equivalent to the Stryker GmbH AxSOS 3 Ti system and the Dall-Miles Homogenous Vitallium Cable.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.