DALL-MILES HOMOGENOUS VITALLIUM CABLE

K961283 · Howmedica Corp. · JDQ · May 22, 1996 · Orthopedic

Device Facts

Record IDK961283
Device NameDALL-MILES HOMOGENOUS VITALLIUM CABLE
ApplicantHowmedica Corp.
Product CodeJDQ · Orthopedic
Decision DateMay 22, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The original Homogeneous cable is intended to be used for spinal wiring, trochanter reattachment, sternotomy applications, and trauma surgery of the shoulder, elbow, knee, hip, or ankle. Indications for the stabilization of bone graft material and for fracture fixation when used in conjunction with bone plates and screws have been added.

Device Story

Dall-Miles Homogeneous Vitallium Cable; bone fixation cerclage device. Used for orthopedic stabilization; spinal wiring; trochanter reattachment; sternotomy; trauma surgery; bone graft stabilization; fracture fixation with plates/screws. Operates as mechanical fixation implant. No changes in design, materials, or manufacturing from predicate (K945294).

Clinical Evidence

Bench testing only; no clinical data provided. Substantial equivalence based on design identity to predicate and equivalence of intended use to existing marketed devices.

Technological Characteristics

Material: Vitallium. Form factor: Cable. Function: Bone fixation cerclage. No changes in design, materials, or operational principles from predicate.

Indications for Use

Indicated for spinal wiring, trochanter reattachment, sternotomy, trauma surgery of shoulder, elbow, knee, hip, or ankle, stabilization of bone graft material, and fracture fixation in conjunction with bone plates and screws.

Regulatory Classification

Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0} MAY 22 1996 K961283 510(k) Summary Proprietary Name: Dall-Miles Homogeneous Vitallium® Cable Common Name: Bone Fixation Cerclage Classification Name and Reference: Bone Fixation Cerclage 21 CFR 888.3010 Proposed Regulatory Class: Class II Device Product Code: OR(87) JDQ For information contact: Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 Telephone: (201) 507-7875 Fax: (201) 507-6870 Date Summary Prepared: 4-2-96 This submission describes a modification in intended use of the previously released Dall-Miles Homogeneous Cable (K945294). The original Homogeneous cable is intended to be used for spinal wiring, trochanter reattachment, sternotomy applications, and trauma surgery of the shoulder, elbow, knee, hip, or ankle. Indications for the stabilization of bone graft material and for fracture fixation when used in conjunction with bone plates and screws have been added. There have been no changes in the design, materials, manufacturing, or operational principles of the Dall-Miles Cable. Substantial equivalence of the additional indications is based on equivalence in intended use to Howmedica's Dall-Miles Suture/Beaded Cable (K953818), Smith & Nephew Richards' Ries Cable System (K875156), and Depuy's Control Cable System (K934557). Each of the named devices are currently marketed for stabilization of bone graft and fracture fixation.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...