Published retrospective clinical data (previously cleared in K151128)
The sponsor utilized previously cleared retrospective clinical data to demonstrate that the use of allogenic bone graft in combination with interbody fusion devices is common clinical practice and poses no new risks.
Retrospective clinical data; Allogenic bone graft; Lumbar interbody fusion
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective clinical data for lumbar interbody fusion devices (referenced from K151128); Retrospective clinical study
Patients undergoing lumbar interbody fusion
Not applicable for this study
Safety and effectiveness of allogenic bone graft use in interbody fusion
Indications for Use
The ELEVATE™ Spinal System Expandable Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine. The CAPSTONE PTC™ Spinal System is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Additionally, the CAPSTONE PTC™ Spinal System is indicated to assist in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scoliosis. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous bone. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the Food and Drug Administration (FDA) for use in the lumbar spine. The CRESCENT™ Spinal System is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants are to be used autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplementation which has been cleared by the FDA for use in the lumbar spine. The CRESCENT™ Spinal System Titanium is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with Medtronic supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
Device Story
Spinal interbody fusion cages; inserted into lumbar/lumbosacral disc space to provide structural support/correction during fusion; used with autogenous or allogenic bone graft; requires supplemental fixation instrumentation. Implants feature hollow geometry for graft packing; bullet-nose/convex/banana shapes; teeth for fixation. Materials include PEEK, Titanium Alloy (Ti-6Al-4V), or PEEK coated with Commercially Pure Titanium (CP Ti); Tantalum markers for radiographic visualization. Used in OR by surgeons via open, minimally invasive, anterior, or transforaminal approaches. Benefit: stabilization of spinal segments in patients with DDD or deformity.
Clinical Evidence
No new clinical testing performed. Relies on retrospective clinical data previously cleared in K151128 to support the safety and efficacy of using allogenic bone graft in lumbar interbody fusion.
Technological Characteristics
Materials: PEEK, Ti-6Al-4V, CP Ti, Unalloy Tantalum markers. Principle: Intervertebral body fusion cage. Energy: None (mechanical). Sterilization: Gamma irradiation. Endotoxin limit: 20 EU/device per ANSI/AAMI ST72 and USP <161>.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at 1-2 contiguous levels (L2-S1), with or without Grade 1 spondylolisthesis/retrolisthesis, who failed 6 months of non-operative treatment. CAPSTONE PTC also indicated for degenerative scoliosis as supplement to pedicle screw fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
CAPSTONE® and CLYDESDALE® Spinal Systems (K151128)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2017
Medtronic Sofamor Danek Mr. Justin O'Connor Regulatory Affairs Specialist 1800 Pyramid Place Memphis. Tennessee 38132
Re: K172199
Trade/Device Name: ELEVATE™ Spinal System, CAPSTONE PTC™ Spinal System, CRESCENT™ Spinal System, CRESCENT™ Spinal System Titanium Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: July 20, 2017 Received: July 21, 2017
Dear Mr. O'Connor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K172199
Device Name ELEVATE™ Spinal System
#### Indications for Use (Describe)
The ELEVATE™ Spinal System Expandable Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known)
K172199
Device Name CAPSTONE PTC™ Spinal System
#### Indications for Use (Describe)
The CAPSTONE PTC™ Spinal System is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
Additionally, the CAPSTONE PTC™ Spinal System is indicated to assist in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scoliosis.
These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous bone. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the Food and Drug Administration (FDA) for use in the lumbar spine.
Type of Use (*Select one or both, as applicable*)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
K172199 Device Name CRESCENTTM Spinal System
#### Indications for Use (Describe)
The CRESCENT™ Spinal System is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants are to be used autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplementation which has been cleared by the FDA for use in the lumbar spine.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known)
#### K172199
Device Name CRESCENT™ Spinal System Titanium
#### Indications for Use (Describe)
The CRESCENT™ Spinal System Titanium is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with Medtronic supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
## 510(k) SUMMARY
## MEDTRONIC Sofamor Danek
## ELEVATE™ Spinal System, CAPSTONE PTC™ Spinal System, CRESCENT™ Spinal System, and CRESCENT™ Spinal System Titanium
## July 2017
| I. | Submitter | Medtronic Sofamor Danek, USA Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133 |
|-----|----------------------|--------------------------------------------------------------------------------------------------------------------------|
| | Contact: | Justin E. O'Connor<br>Regulatory Affairs Specialist |
| | Date Prepared: | July 20, 2017 |
| II. | Subject Device | |
| | Name of Device: | ELEVATE™ Spinal System, CAPSTONE PTC™<br>Spinal System, CRESCENT™ Spinal System,<br>and CRESCENT™ Spinal System Titanium |
| | Regulation Number: | 888.3080 |
| | Code: | MAX |
| | Classification Name: | Intervertebral Fusion Device with Bone Graft,<br>Lumbar |
| | Classification: | Class II |
| | Common Name: | Intervertebral Body Fusion Device |
{7}------------------------------------------------
#### III. Predicate Device:
K151128 (S.E. 08/06/2015) CAPSTONE® and CLYDESDALE® Spinal Systems - Predicate 1, Primary Predicate
K142559 (S.E. 06/09/2015) ELEVATE™ Spinal System - Predicate 2, Additional Predicate
K133205 (S.E. 03/13/2014) CAPSTONE PTCTM and CLYDESDALE PTCTM Spinal Systems -Predicate 3, Additional Predicate
K094025 (S.E. 04/26/2010) CRESCENTTM Spinal System – Predicate 4, Additional Predicate
K110543 (S.E. 08/09/2011) CRESCENTTM Spinal System Titanium - Predicate 5, Additional Predicate
The predicates have not been subject to a design related recall.
#### IV. Device Description:
#### ELEVATE™ Spinal System
The subject ELEVATE™ Spinal System consists of bullet nose expandable cages with teeth. The top of the cage is made of Polyetheretherketone (PEEK) and the base is made of Titanium Alloy (Ti-6Al-4V). The cages also contain Unalloy Tantalum markers, which allow for radiographic visualization during the surgical procedure. The ELEVATE™ implants are provided in both standard and an extra-lordotic design. The devices can be surgically implanted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar intervertebral body fusion. The implants also contain hollow geometry which allows them to be packed with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
{8}------------------------------------------------
### CAPSTONE PTC™ Spinal System
The subject CAPSTONE PTC™ Spinal System consists of convex cages with a bullet nose and angular teeth. The implants are made of Polyetheretherketone (PEEK) and coated with Commercially Pure Titanium (CP Ti). The cages also contain Unalloy Tantalum markers, which allow for radiographic visualization during the surgical procedure. The CAPSTONE PTC™ implants are provided in various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
## CRESCENT™ Spinal System
The subject CRESCENT™ Spinal System consists of banana shaped cages with a bullet nose and angular teeth. The implants are made of Polyetheretherketone (PEEK) and contain Unalloy Tantalum markers, which allow for radiographic visualization during the surgical procedure. The CRESCENT™ implants are provided in various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
## CRESCENT® Spinal System Titanium
The subject CRESCENT™ Spinal System Titanium consists of banana shaped cages with a bullet nose and angular teeth. The CRESCENT™ Titanium implants are made of Titanium Alloy (Ti-6Al-4V) and are provided in various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.
{9}------------------------------------------------
#### V. Indications:
### ELEVATE™ Spinal System
The ELEVATE™ Spinal System Expandable Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
## CAPSTONE PTC™ Spinal System
The CAPSTONE PTC™ Spinal System is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Additionally, the CAPSTONE PTC™ Spinal System is indicated to assist in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scoliosis. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the Food and Drug Administration (FDA) for use in the lumbar spine.
{10}------------------------------------------------
### CRESCENT™ Spinal System
The CRESCENT™ Spinal System is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants are to be used autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
## CRESCENT™ Spinal System Titanium
The CRESCENT™ Spinal System Titanium is indicated for interbody fusion with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone. These devices are intended to be used with Medtronic supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
#### VI. Comparison of Technological Characteristics with the Predicate Devices:
The technological principle for both the subject and predicate devices is lumbar interbody fusion to provide correction and stabilization during intervertebral body fusion procedures. The subject interbody cages are all manufactured from the same materials as their system predicates and are inserted into the disc space along with graft material to
{11}------------------------------------------------
facilitate fusion, using supplemental fixation, at one or two contiguous levels from L2-S1.
The subject and predicate interbody cages are sterilized via gamma irradiation and provided sterile to the end user. In order to satisfy the FDA Endotoxin requirements for this submission, a bacterial endotoxin test. also known as Limulus amebocyte lysate (LAL) test, was performed utilizing worst case subject implants to verify that the subject implants meet the 20 endotoxin units (EU)/device pyrogen limit specification. Testing was successfully performed and it was confirmed that the subject implants meet the 20 EU/device testing limit for general medical devices that are implanted as outlined in ANSI/AAMI ST72, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing and USP <161>, Transfusion and Infusion Assemblies and Similar Medical Devices.
The purpose of this 510(k) bundled submission is to expand the indications to allow the subject interbody cages to be used with allogenic bone graft comprised of cancellous and/or corticocancellous bone as an alternative graft material option. With the exception of the inclusion of allogenic bone graft to the modified indications, there have been no changes to subject interbody cages; therefore, they are substantially equivalent to their predicates.
#### Discussion of Supporting Retrospective Clinical Data and Non-Clinical Testing: VII.
Published retrospective clinical data for the lumbar interbody fusion devices was cleared by the FDA in Medtronic's CAPSTONE® and CLYDESDALE® Spinal System (K151128, S.E. 08/06/2015). The retrospective clinical data demonstrated that the use of allogenic bone graft in combination with, or without, autogenous bone graft is commonly used in conjuntion with interbody fusion devices and poses no new risks to patients. Because the subject devices have the same intended use, indications, and fundamental technology as the primary predicate devices, the retrospective clincal data cleared in K151128 still applies, and no additional clinical data is required as a result of this submission.
{12}------------------------------------------------
#### VIII. Conclusion:
The design features, materials used, manufacturing processes, and sterilization methods are equivalent to the cleared ELEVATE™, CAPSTONE PTC™, CRESCENT™, and CRESCENT™ Titanium devices, with the exception of expanding the indications to include the use of allogenic bone graft comprised of cancellous and/or corticocancellous bone as an alternative graft material option. The expanded indications sought after in this application are identical to those cleared in K151128 (S.E. 08/06/2015).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.