The Hintermann Series H2 Total Ankle Replacement (Hintermann Series H2) system is designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, replacing flexion and extension movement in the ankle joint, and allowing for movement at the replaced joint. The Hintermann Series H2 is indicated as a total ankle replacement in primary or revision surgery of ankle joints damaged by: · Systemic arthritis of the ankle (e.g., rheumatoid arthritis, hemochromatosis) · Primary arthritis (e.g., degenerative disease) · Secondary arthritis (e.g., post-traumatic, avascular necrosis, provided enough of the talus is preserved to support the implant) The Hintermann Series H2 is also indicated for patients with a failed previous ankle surgery and revision surgeries following failed total ankle replacement or non-union of the ankle arthrodesis, provided sufficient bone stock is present. Note: In the United States, this device is intended for cemented use only.
Device Story
Semi-constrained total ankle prosthesis; replaces ankle joint to reduce pain and restore motion. Components: metal tibial component, polyethylene inlay, metal talar component. Bone-contacting surfaces feature plasma-sprayed titanium coating. Used in orthopedic surgery; implanted by surgeons. Intended for cemented fixation in the US. Benefits: pain reduction, restored joint alignment, and flexion/extension movement. Device design and performance validated against ASTM F2665 standards.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including coating characterization, range of motion analysis, fatigue testing, contact area/stress analysis, constraint testing, tibial component locking mechanism testing, wear testing, and endotoxin testing, all performed in accordance with ASTM F2665.
Indicated for patients requiring primary or revision total ankle replacement due to systemic, primary, or secondary arthritis, or failed previous ankle surgery/arthrodesis, provided sufficient bone stock and talar support exist.
Regulatory Classification
Identification
An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Integra (Cadence) Total Ankle Replacement System (K151459)
{0}------------------------------------------------
November 7, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
DT MedTech LLC % Marcos Velez-Duran President M Squared Associates, Inc 575 8th Avenue, Suite 1212 New York, New York 10018
Re: K171004
Trade/Device Name: Hintermann Series H2 Total Ankle System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSN Dated: October 11, 2017 Received: October 12, 2017
Dear Marcos Velez-Duran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known)
## K171004
Device Name Hintermann Series H2 Total Ankle System
#### Indications for Use (Describe)
The Hintermann Series H2 Total Ankle Replacement (Hintermann Series H2) system is designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, replacing flexion and extension movement in the ankle joint, and allowing for movement at the replaced joint.
The Hintermann Series H2 is indicated as a total ankle replacement in primary or revision surgery of ankle joints damaged by:
· Systemic arthritis of the ankle (e.g., rheumatoid arthritis, hemochromatosis)
· Primary arthritis (e.g., degenerative disease)
· Secondary arthritis (e.g., post-traumatic, avascular necrosis, provided enough of the talus is preserved to support the implant)
The Hintermann Series H2 is also indicated for patients with a failed previous ankle surgery and revision surgeries following failed total ankle replacement or non-union of the ankle arthrodesis, provided sufficient bone stock is present.
Note: In the United States, this device is intended for cemented use only.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Hintermann Series H2 Total Ankle System K171004
| Sponsor: | DT MedTech, LLC.<br>110 West Road<br>Suite 227<br>Towson, MD 21204 |
|--------------------|-------------------------------------------------------------------------------------------------------------|
| Contact: | Shawn T. Huxel<br>Executive Vice President<br>Phone: 410-427-0003<br>Fax: 410-494-0515 |
| Date Prepared: | November 1, 2017 |
| Trade Name: | Hintermann Series H2™ Total Ankle System |
| Common Name: | Total ankle prosthesis |
| Classification: | 888.3110 Ankle joint metal/polymer semi-constrained cemented<br>prosthesis. |
| Product Code: | HSN |
| Predicate Devices: | Integra (Cadence) Total Ankle Replacement System, Integra<br>LifeSciences, Inc., K151459 |
| | INBONE I and II Total Ankle Replacement, Wright Medical<br>Technologies, K051023, K100886, K123059, K133585 |
| | Salto Talaris Total Ankle Prosthesis, Integra LifeSciences, Inc.,<br>K060544, K090076, K130533, K153452 |
| | Exactech Vantage Total Ankle System, Exactech Inc., K152217 |
## Description of Device:
The Hintermann Series H2 Total Ankle System includes a metal tibial component, polyethylene inlay, and metal talar component. The tibial and talar components are coated with plasma spray titanium on the bone contacting surfaces. The system is a semiconstrained device used for the indications as follows.
{4}------------------------------------------------
#### Indications for Use:
The Hintermann Series H2 Total Ankle Replacement (Hintermann Series H2) system is designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, replacing flexion and extension movement in the ankle joint, and allowing for movement at the replaced joint.
The Hintermann Series H2 is indicated as a total ankle replacement in primary or revision surgery of ankle joints damaged by:
- . Systemic arthritis of the ankle (e.g., rheumatoid arthritis, hemochromatosis)
- Primary arthritis (e.g., degenerative disease)
- Secondary arthritis (e.g., post-traumatic, avascular necrosis, provided enough of the talus is preserved to support the implant)
The Hintermann Series H2 is also indicated for patients with a failed previous ankle surgery and revision surgeries following failed total ankle replacement or non-union/mal-union of the ankle arthrodesis, provided sufficient bone stock is present. Note: In the United States, this device is intended for cemented use only.
#### Comparison of Technological Characteristics with the Predicate Device:
The technological characteristics for the Hintermann Series H2 Total Ankle System are the similar to, or the same as, the characteristics of the predicate devices as outlined in the table that follows.
| | Hintermann<br>Series H2 | Vantage | Cadence | Inbone II | Salto Talaris |
|-------------------------|----------------------------|----------------------------|----------------------------|------------------------------------------|------------------------------------------|
| Manufacturer | DT MedTech | Exactech | Integra Life<br>Sciences | Wright Medical<br>Technology, Inc | Tornier |
| Submission<br>number | K171004 | K152217 | K151459 | K051023<br>K100886<br>K123059<br>K133585 | K060544<br>K090076<br>K130533<br>K153452 |
| Indications for<br>use | Total ankle<br>replacement | Total ankle<br>replacement | Total ankle<br>replacement | Total ankle<br>replacement | Total ankle<br>replacement |
| Method of<br>fixation | Cemented | Cemented | Cemented | Cemented | Cemented |
| Sterilization<br>method | Gamma | Gamma | Gamma | Gamma | Gamma |
{5}------------------------------------------------
# Hintermann Series H2 Total Ankle System K171004 510(k) Summary
| | Hintermann<br>Series H2 | Vantage | Cadence | Inbone II | Salto Talaris |
|---------------------|--------------------------------------------------------------------------------|----------------------------|-----------------------------------------------|----------------------------------------------------|----------------------------------|
| Materials | | | | | |
| Tibial<br>component | Titanium Alloy | Titanium Alloy | Titanium Alloy | Titanium Alloy | CoCr alloy |
| Poly insert | UHMWPE | UHMWPE | UHMWPE | UHMWPE | UHMWPE |
| Talar component | CoCr alloy | CoCr alloy | CoCr alloy | CoCr alloy | CoCr alloy |
| Coating | Titanium<br>plasma spray | Porous titanium<br>coating | Porous titanium<br>coating | Titanium<br>plasma spray | Titanium<br>plasma spray |
| Sizes | | | | | |
| Tibial<br>component | 5 sizes (2, 3, 4,<br>5, 6) | Not reported | 9 sizes (1, 1X, 2,<br>2X, 3, 3X, 4, 4X,<br>5) | 9 sizes (size 2,<br>2L, 3, 3L, 4, 4L,<br>5, 5L, 6) | 4 sizes (0, 1, 2,<br>3) |
| Poly insert | 6 sizes (1, 2, 3,<br>4, 5, 6) | Not reported | 5 sizes (1, 2, 3,<br>4, 5) | 5 sizes (1+, 2, 3,<br>4, 5) | 5 sizes (size 00,<br>0, 1, 2, 3) |
| Talar component | Std: 6 sizes (1,<br>2, 3, 4, 5, 6)<br><br>Flat cut: 5 sizes<br>(1, 2, 3, 4, 5) | Not reported | 5 sizes (1, 2, 3,<br>4, 5) | 5 sizes (1, 2, 3,<br>4, 5) | 4 sizes (0, 1, 2,<br>3) |
Nonclinical Testing Summary:
Following is a summary of the non-clinical testing that was conducted:
- . Coating characterization per FDA Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements dated February 2, 2000
- . Range of motion analysis per ASTM F2665
- Fatigue testing per ASTM F2665
- Contact area/contact stress analysis per ASTM F2665
- Constraint testing per ASTM F2665
- Tibial Component locking mechanism testing
- Wear testing per ASTM F2665
- Endotoxin testing per FDA guidance
#### Substantial Equivalence:
The design features of the Hintermann Series H2 Total Ankle System are substantially equivalent to the design features of other predicate devices previously cleared for market. The methods used to establish equivalence are based on the comparison of the intended use, product technical characteristics and performance characteristics as defined by ASTM F2665 Standard Specification for Total Ankle Replacement Prosthesis. All comparisons
{6}------------------------------------------------
# Hintermann Series H2 Total Ankle System K171004 510(k) Summary
confirmed that the Hintermann Series H2 Total Ankle System is substantially equivalent to the predicate devices. The safety, effectiveness, and performance of the Hintermann Series H2 Total Ankle System are adequately supported by the substantial equivalence information, material information and analysis data provided within this Premarket Notification. Therefore, it is concluded that the Hintermann Series H2 Total Ankle System is substantially equivalent to the noted predicate devices.
### Conclusions:
While the Hintermann Series H2 Total Ankle System is not identical to the predicate devices, any differences that may exist do not significantly affect device safety and effectiveness. In addition, the differences do not add new or increased risks and complications. Therefore, it is concluded that the Hintermann Series H2 Total Ankle System is substantially equivalent to the predicate devices as outlined previously.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.