K170372 · Spiraltech Superior Dental Implants, Inc. · DZE · Jun 30, 2017 · Dental
Device Facts
Record ID
K170372
Device Name
SpiralTech Dental Implant System
Applicant
Spiraltech Superior Dental Implants, Inc.
Product Code
DZE · Dental
Decision Date
Jun 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Spiraltech Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients esthetics and chewing function. Spiraltech implants are intended for single or multiple unit restorations on splinted or non-splinted applications. The implants ESi Dynamic and Ultimate are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants [along with Premium and One Piece] can also be used for loading after a conventional healing period. Solo One Piece 3.0 and 3.3 implants. Ultimate (conical) 3.0 implants, and ESi (conical) 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.0 and/or 3.3 implants adjacent to one another.
Device Story
SpiralTech Dental Implant System comprises endosseous titanium implants, abutments, and closure screws; used by dental surgeons to support prosthetic teeth. Implants feature various thread designs (ESi, Ultimate, Dynamic, Premium, One Piece) and surface treatments (SLA, RBM). Input is surgical site in jaw; device provides mechanical foundation for prosthetic restoration. Abutments (straight, shoulder, angulated, ball, multiunit, temporary, healing) connect to implants via hex or conical interfaces. Temporary abutments utilize PEEK or zirconia; healing abutments utilize titanium alloy or zirconia. System supports immediate or conventional loading protocols. Clinical benefit is restoration of chewing function and esthetics. Device is used in clinical dental settings.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation (ISO 11737, ISO 11137, ISO 17665), fatigue testing per ISO 14801, surface analysis (SEM/EDS), and biocompatibility testing (cytotoxicity/MEM elution). Shelf life validated via accelerated aging (ASTM F1980).
Technological Characteristics
Materials: Grade 5 titanium (Ti 6Al-4V ELI, ASTM F136), PEEK, zirconia. Surface treatments: SLA, RBM. Connection types: Hex and conical. Dimensions: 3.0-6.0mm diameter; 8-15mm length. Sterilization: Gamma radiation and moist heat. No software or electronic components.
Indications for Use
Indicated for patients requiring endosseous dental implants in upper or lower jaw arches to support prosthetic devices (artificial teeth) for restoration of esthetics and chewing function. Applicable for single or multiple unit restorations (splinted or non-splinted). Specific 3.0/3.3mm implants indicated for lateral incisor (maxilla) and central/lateral incisor (mandible) replacement; mandibular incisors must be splinted if adjacent. Immediate loading permitted for ESi, Dynamic, and Ultimate lines given good primary stability and appropriate occlusive loading; otherwise, conventional healing period required.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2017
SpiralTech Superior Dental Implants Daniel Rosenthal Director of Sales and Marketing 875 N. Michigan Ave Suite 3106 Chicago, Illinois 60611
Re: K170372
Trade/Device Name: SpiralTech Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 29, 2017 Received: June 5, 2017
Dear Daniel Rosenthal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Andrew I. Steen -S
for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170372
Device Name SpiralTech Dental Implant System
#### Indications for Use (Describe)
The Spiraltech Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients esthetics and chewing function. Spiraltech implants are intended for single or multiple unit restorations on splinted applications. The implants ESi Dynamic and Ultimate are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants [along with Premium and One Piece] can also be used for loading after a conventional healing period.
Solo One Piece 3.0 and 3.3 implants. Ultimate (conical) 3.0 implants, and ESi (conical) 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.0 and/or 3.3 implants adjacent to one another.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div> <span> <span style="font-size:16px"> <b> </b> </span> </span> <span> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary June 17, 2017
#### SpiralTech Dental Implants
Company: SpiralTech Superior Dental Implants
Address: 875 N. Michigan Ave Suite 3106 Chicago, IL 60611
Phone: 312-440-7777
Contact Person: Daniel Rosenthal
Contact email: info@spiraltech.com
Trade Name: SpiralTech Dental Implant System
Common Name: dental implant and abutment system
Classification Name: endosseous dental implant
Regulation Number: 21 CFR 872.3640
Product Code: DZE
Secondary Product Code: NHA
Predicate Device: Nobel Biocare K071370
Reference Devices: Implants – Nobel Biocare K133731 K020646 K102436 Biohorizons K121787 Zimmer K082639 MIS K080162 K103089 Abutments - Biohorizons K071638 Alpha Bio Tec K063364 MSI K112162
#### Indications for Use
The Spiraltech Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients esthetics and chewing function. Spiraltech implants are intended for single or multiple unit restorations on splinted or non-splinted applications. The implants ESi Dynamic and Ultimate are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants [along with Premium and One Piece] can also be used for loading after a conventional healing period.
Solo One Piece 3.0 and 3.3 implants, Ultimate (conical) 3.0 implants, and ESi (conical) 3.0 implants are intended to replace a lateral incisor in the maxilla and/or a central incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.0 and/or 3.3 implants adjacent to one another.
{5}------------------------------------------------
#### Device Description
The SpiralTech implant system is a comprehensive product line that includes implants, corresponding abutments, and closure screws. SpiralTech dental implants are grade 5 titanium (Ti 6Al-4V ELI, conforms to ASTM F136) implants that come in 2 different surface treatments - SLA, and RBM.
SpiralTech dental implants come in five product lines with four based on their thread designs. The ESi has sharp, square, and rounded threads. The Ultimate and Dynamic lines also have sharp threads with the Dynamic having a reverse buttress thread design. The Premium line features square and sharper threads in a more conventional design. The implants have diameters ranging from 3.0 mm, and the lengths from 8mm to 15 mm. ESi, Ultimate, and Dynamic are intended to be used for immediate loading. The fifth implant type is the One Piece which comes with an abutment which cannot be used with low mechanical stability cases.
Abutments are available in various types including straight, shoulder, angulated, ball attachments, multiunit, temporary and healing. All abutments come in both hex and conical connections. Temporary abutments come in PEEK and zirconia. Healing abutments come in titanium alloy and zirconia. Abutments come in titanium alloy and as a titanium base with zirconia abutments and multi-units are titanium alloy. No SpiralTech abutments are intended to be modified.
#### Sizes
The Hex versions of the designs are available in 3.3, 3.5, 4.3, 5.0, and 6.0 with lengths 8, 10, 11, 13, and 15. The conical versions start at a diameter of 3.5 but are otherwise available in the same sizes and lengths with one exception, the Ultimate and ESi designs also come in 3.0 diameter with lengths of 10, 11, 13, and 15.
| Diameter (mm) | Implant Lengths (mm) |
|---------------|-----------------------|
| 3.3mm | 8, 10, 11.5, 13, 15mm |
| 3.5mm | 8, 10, 11.5, 13, 15mm |
| 4.3mm | 8, 10, 11.5, 13, 15mm |
| 5.0mm | 8, 10, 11.5, 13, 15mm |
| 6.0mm | 8, 10, 11.5, 13, 15mm |
Hex version implant sizes:
Conical version implant sizes:
| Diameter (mm) | Implant Lengths (mm) |
|---------------|---------------------------------------------------------|
| 3.0mm | 10, 11.5, 13, 15mm – only for Ultimate and ESi implants |
| 3.5mm | 8, 10, 11.5, 13, 15mm |
| 4.3mm | 8, 10, 11.5, 13, 15mm |
| 5.0mm | 8, 10, 11.5, 13, 15mm |
| 6.0mm | 8, 10, 11.5, 13, 15mm |
{6}------------------------------------------------
The Solo one piece implant comes in diameters 3.0, 3.3, 3.75, and 4.2. It has lengths of 10, 11, 13 and 15.
The smallest diameter conical versions for the ESi, Ultimate, Dynamic and Premium implants are only compatible with the 0° and 15° titanium conical abutment models.
The smallest diameter hex versions for the ESi, Ultimate, Dynamic and Premium implants are only compatible with the 0°, 15° and 25° titanium hex abutment models.
# Non-Clinical Testing
Sterilization validation was conducted for the implants according to the respective standards, ISO 11737-1, ISO 11737-2, ISO 11137-1, ISO 11137-2, and ISO 11137-3 (for gamma radiation validation) and ISO 17665-1 and ISO 17665-2 (for moist heat validation). This testing included USP 85, USP 161 and LAL method testing on the implants. The shelf life for the implants supplied sterile is 5 years based on accelerated testing (ASTM F1980) with one year of real time data (ASTM F1929-12, ASTM F88/F88M).
Several designs of SpiralTech Dental Implants were fatigue tested according to ISO 14801. Justification was provided to show the design tested was the worst case for that connection type. The implants with SLA and RBM surface treatments were evaluated by SEM and EDS for surface analysis.
The temporary abutments are made of PEEK. The implants are Ti6Al4V with either SLA or RBM surface treatment. Abutments are either Ti6AL4V or ZrO2. Chemical characterization of the ZrQ2 was provided. Cytotoxicity (MEM elution) has been provided to demonstrate biocompatibility for all subject devices composed of these identified materials.
## Substantial Equivalence
The widths and lengths of the implants are in the same range as the predicate and reference devices. The materials used for all the devices are commonly used in the predicate and reference devices. As can be seen in the table, the thread designs incorporate types of threads used in the predicate and reference devices. The implant connections types are ones used in several predicate devices as can be seen in the table. The indications for use are the predicate device, except for adding more restrictive placement for the Solo One Piece 3.0 and 3.3 implants, Ultimate (conical) 3.0 implants, and ESi (conical) 3.0 implants. The surface treatments available are the same types as those used on predicate devices.
The testing done for sterilization validation is the same type of testing done on the predicate and reference devices. Also the predicate and reference devices underwent the same ISO 14801 testing.
See the Substantial Equivalence Tables.
{7}------------------------------------------------
## Conclusion
The SpiralTech Dental Implants are substantially equivalent to the predicate device based on design including implant connection, materials, surface treatments, dimensions, and testing completed.
{8}------------------------------------------------
# Substantial Equivalence Table
# Implant Table
| | SpiralTech | NobelActive<br>K071370 and<br>K133731 | NobelReplace<br>K020646 | NobelActive<br>3.0<br>K102436 | Internal<br>Tapered<br>Plus from<br>Biohorizons<br>K121787 | Tapered<br>SwissPlus<br>from<br>Zimmer<br>K082639 | MIS Uno<br>K080162 | MIS<br>Short<br>implants<br>K103089 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|-------------------------------|------------------------------------------------------------|---------------------------------------------------|--------------------|-------------------------------------|
| Material | Ti6Al4V | CP Titanium | Ti6Al4V | CP Titanium | Ti6Al4V | Ti6Al4V | Ti6Al4V | Ti6Al4V |
| Diameters | 3.0 – 6.0<br>with the<br>exact sizes<br>varying<br>among the<br>designs. | 3.5, 4.3, 5.0,<br>5.5 | 3.5, 4.3, 5.0,<br>6.0 | 3.0 | 3.8, 4.6, 5.8 | 3.7, 4.8 | 3.0 and<br>3.5 | 4.2, 5.0,<br>6.0 |
| Lengths | 8, 10, 11, 13,<br>15 with not<br>all lengths<br>available for<br>all sizes and<br>designs | 8.5, 10, 11.5 | 10, 13, 16 | 10, 11.5, 13,<br>15 | 7.5, 9, 10.5,<br>12, 15 | 8, 10, 12,<br>14 | 10, 13, 16 | 6.0 |
| Indications<br>for Use | The<br>Spiraltech<br>Dental<br>Implants are<br>endosseous<br>implants<br>intended to<br>be surgically<br>placed in the<br>upper or<br>lower jaw<br>arches to | Nobel<br>Biocare's<br>NobelActive<br>implants are<br>endosseous<br>implants<br>intended to<br>be surgically<br>placed in the<br>bone of the<br>upper or<br>lower jaw | | | | | | |
{9}------------------------------------------------
| provide<br>support for<br>prosthetic<br>devices,<br>such as an<br>artificial<br>tooth, in<br>order to<br>restore<br>patients<br>esthetics<br>and chewing<br>function.<br>Spiraltech<br>implants are<br>intended for<br>single or<br>multiple unit<br>restorations<br>on splinted<br>or non-<br>splinted<br>applications.<br>The implants<br>ESi Dynamic<br>and Ultimate<br>are intended<br>for<br>immediate<br>loading<br>when good<br>primary<br>stability is<br>achieved | arches to<br>provide<br>support for<br>prosthetic<br>devices, such<br>as an<br>artificial<br>tooth, in<br>order to<br>restore<br>patient<br>esthetics and<br>chewing<br>function.<br>Nobel<br>Biocare's<br>Nobelactive<br>implants are<br>indicated for<br>single or<br>multiple unit<br>restorations<br>in splinted<br>and non-<br>splinted<br>applications.<br>Nobel<br>Biocare's<br>NobelActive<br>implants are<br>intended for<br>immediate<br>loading when<br>good primary | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| and with | stability is | | | | |
| appropriate | achieved and | | | | |
| occlusive | with | | | | |
| loading. | appropriate | | | | |
| These | occlusal | | | | |
| implants | loading. | | | | |
| [along with | | | | | |
| Premium | | | | | |
| and One | | | | | |
| Piece] can | | | | | |
| also be used | | | | | |
| for loading | | | | | |
| after a | | | | | |
| conventional | | | | | |
| healing | | | | | |
| period. | | | | | |
| Solo One | | | | | |
| Piece 3.0 | | | | | |
| and 3.3 | | | | | |
| implants, | | | | | |
| Ultimate | | | | | |
| (conical) 3.0 | | | | | |
| implants, | | | | | |
| and ESi | | | | | |
| (conical) 3.0 | | | | | |
| implants are | | | | | |
| intended to | | | | | |
| replace a | | | | | |
| lateral | | | | | |
| incisor in the | | | | | |
| maxilla | | | | | |
| and/or a | | | | | |
| central or | | | | | |
| lateral | | | | | |
{10}------------------------------------------------
{11}------------------------------------------------
| | incisor in the<br>mandible.<br>Mandibular<br>central and<br>lateral<br>incisors<br>must be<br>splinted if<br>using two or<br>more 3.0<br>and/or 3.3<br>implants<br>adjacent to<br>one another. | | | | | |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|---|---|--|
| ESI thread<br>design<br>reference<br>device | | | | | X | |
| Dynamic<br>thread<br>design<br>reference<br>device | | | | X | | |
| Premium<br>thread<br>design<br>reference<br>device | | | X | | | |
| Ultimate<br>thread<br>design<br>reference<br>and<br>predicate | | X | | | | |
{12}------------------------------------------------
| device | | | | | | | | | |
|-------------------------------------------------------------------------------------|--------------|-------------|--------------------|---------|-----|---------|--------------|-----|--|
| One Piece<br>thread<br>design<br>reference<br>and<br>predicate<br>device | | | | | | X | | | |
| Radiation<br>Sterilization<br>of the<br>Implants | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
| Validated<br>Steam<br>Sterilization<br>Instructions<br>supplied<br>for<br>abutments | Yes | Yes | Yes | Yes | Yes | Yes | No abutments | Yes | |
| Subject of<br>Fatigue<br>testing<br>meeting<br>ISO 14801 | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
| Connection<br>interface | hex, conical | hex | conical | hex | hex | octagon | N/A | hex | |
| Surface<br>Treatments<br>Available | SLA, RBM | Acid etched | HA, acid<br>etched | TiUnite | RBM | HA, RBM | SLA | SLA | |
{13}------------------------------------------------
# Abutment Table
| Abutments | | Hex | Conical<br>NP | Conical RP | Slim<br>Conical |
|----------------------------------------------------------------|-------------------|----------------------------------|----------------------------------|----------------------------------|-----------------|
| Flat Titanium<br>Straight | 3.8<br>(Narrow) | Facial<br>Heights -<br>5,7,9,11 | Facial<br>Heights -<br>5,7,9,11 | Facial<br>Heights –<br>5,7,9,11 | |
| Flat Titanium<br>Straight | 4.5<br>(Standard) | Facial<br>Heights –<br>5,7,9,11 | Facial<br>Heights -<br>5,7,9,11 | Facial<br>Heights -<br>5,7,9,11 | |
| Flat Titanium<br>Straight | 5.5 (Wide) | Facial<br>Heights -<br>5,7,9,11 | | Facial<br>Heights -<br>5,7,9,11 | |
| Flat Titanium<br>Shoulder | 3.8 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3 | |
| Flat Titanium<br>Shoulder | 4.5 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3 | |
| Flat Titanium<br>Shoulder | 5.5 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3 | |
| Anatomic<br>Titanium | 4.5 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3,4 | |
| Anatomic<br>Titanium | 5.5 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights –<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | |
| 15 Degree<br>Angulated - | 4.7 | Facial<br>Heights - | Facial<br>Heights – | Facial<br>Heights - | |
| Titanium | | | | | |
| 25 Degree<br>Angulated -<br>Titanium | 5.2 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | |
| 30 Degree<br>Angulated -<br>Titanium | 5.3 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | |
| Flat Titanium<br>Slim | 3.7 | | | Facial<br>Heights -<br>1,2,3 | |
| 15 degree<br>Angulated<br>Titanium slim | 3.7 | | | Facial<br>Heights -<br>1,2,3 | |
| Titanium<br>Multiunit<br>Abutments -<br>Straight | - | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3,4 | |
| Titanium<br>Multiunit<br>Abutments -<br>17 Degree<br>angulated | - | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3,4 | Facial<br>Heights -<br>1,2,3,4 | |
| Titanium<br>Multiunit<br>Abutments -<br>30 degree<br>angulated | - | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | Facial<br>Heights -<br>1,2,3 | |
| Titanium<br>Healing<br>Abutment | 3.8 | Facial<br>Heights -<br>2,3,4,5,6 | Facial<br>Heights -<br>2,3,4,5,6 | Facial<br>Heights -<br>2,3,4,5,6 | |
| Titanium<br>Healing | 4.5 | Facial<br>Heights - | Facial<br>Heights - | Facial<br>Heights - | |
| Abutment | | | | | |
| Titanium<br>Healing<br>Abutment | 5.5 | Facial Heights -<br>2,3,4,5,6 | Facial Heights -<br>2,3,4,5,6 | Facial Heights -<br>2,3,4,5,6 | |
| Zirconia<br>Anatomic<br>Abutments | 4.5 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | | |
| Zirconia<br>Healing<br>Abutment | 3.8 | Facial Heights -<br>2,3,4,5,6 | Facial Heights -<br>2,3,4,5,6 | Facial Heights -<br>2,3,4,5,6 | |
| Zirconia<br>Healing<br>Abutment | 5.5 | Facial Heights -<br>3,4 | | Facial Heights -<br>3,4 | |
| PEEK<br>Temporary Flat<br>Abutments | Length - 10 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | |
| PEEK<br>Temporary Flat<br>Abutments | Length - 11 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | |
| PEEK<br>Temporary Flat<br>Abutments | Length - 12 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | |
| PEEK<br>Temporary<br>anatomic<br>Abutments | Length - 10 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | |
| PEEK<br>Temporary<br>Anatomic<br>Abutments | Length - 11 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | |
| PEEK<br>Temporary<br>Anatomic<br>Abutments | Length - 12 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | Facial Heights -<br>1,2,3 | |
| Abutments | | | | | |
| Implant<br>Prosthetic<br>Indicator | - | Facial Heights –<br>1,2,3,4,5,6 | Facial Heights –<br>1,2,3,4,5,6 | Facial Heights –<br>1,2,3,4,5,6 | |
| Ball<br>Attachments | - | Facial Heights –<br>1,2,3,4,5,6 | Facial Heights –<br>1,2,3,4,5,6 | Facial Heights –<br>1,2,3,4,5,6 | |
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
| | SpiralTech (Internal Hex<br>and Conical<br>Connection) | Biohorizon (Conical<br>Internal Hex) K071638 | Alpha Biotec (Internal<br>Hex Connection)<br>K063364 | MIS (Internal Hex)<br>K112162 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Straight Titanium<br>Abutments | Diameters – 3.8,4.5,5.5<br>Facial Heights - 5,7,9,11 | Laserlok Simple<br>abutments<br>Diameters - 3.8,4.5,5.5<br>Facial Heights - 5,7,9,11 | Titanium Abutments<br>Diameters -3.85,4.5,5.3<br>Facial Heights - 6,7,8,9 | Straight Titanium<br>Abutments<br>Diameters - 3.8,4.5,5.5<br>Facial Heights - 5,7,9,11 |
| Anatomic Titanium<br>abutments | Diameter - 4.5<br>Facial Heights - 1,2,3,4<br>Diameter - 5.5<br>Facial Heights - 1,2,3 | Straight Esthetic<br>connections<br>Diameters –<br>3.0,3.5,4.5,5.7<br>Facial Heights - 1,2,3 | Anatomic Abutments<br>with Emergence Profile<br>Diameter - 4.3<br>Facial Heights - 1.5,2<br>Diameter - 5.5<br>Facial Heights - 2,3 | |
| Shoulder Titanium<br>abutments | Diameter - 3.8<br>Facial Heights - 1,2,3<br>Diameter - 4.5<br>Facial Heights - 1,2,3,4<br>Diameter - 5.5<br>Facial Heights - 1,2,3 | | Straight abutments with<br>various cuff heights<br>Diameter - 3.85<br>Facial Heights -<br>0.5,1.5,2.5<br>Diameter - 4.8<br>Facial Heights - 1,2,3,4<br>Diameter - 5.6<br>Facial Heights - 2,4 | |
| Angulated titanium<br>abutments | 15 Degree - Diameter -<br>4.7 - Heights - 1,2,3<br>25 Degree - Diameter -<br>5.2 - Heights - 1,2,3<br>30 Degree - Diameter -<br>5.3 - Heights - 2,3,4,5 | 15 Degree - Diameter -<br>4.5,5.7 - Heights - 2,3<br>25 Degree - Diameter -<br>5.2 - Heights - 1,2,3<br>30 Degree - Diameter -<br>5.3 - Heights - 2,3,4,5 | Angled Abutments<br>15 Degree - Diameter -<br>4.5,4.8,5.1 - Heights -<br>1,2,3<br>25 Degree - Diameter -<br>4.7, 5.3 - Heights - 1,2,3 | |
| | | | 35 Degree - Diameter -<br>4.65, 5.3 - Heights -<br>1,2,3 | |
| Titanium Healing<br>abutments | Diameter - 3.8<br>Facial Heights -<br>2,3,4,5,6<br>Diameter - 4.5<br>Facial Heights -<br>2,3,4,5,6<br>Diameter - 5.5<br>Facial Heights -<br>2,3,4,5,6 | Titanium Healing<br>abutments<br>Diameter - 3.8<br>Facial Heights - 1,3,5<br>Diameter - 4.7<br>Facial Heights - 1,3,5<br>Diameter - 5.7<br>Facial Heights - 1,3,5 | Titanium Healing<br>abutments<br>Diameter - 3.8<br>Facial Heights - 3,4,5<br>Diameter - 4.5<br>Facial Heights -<br>2,3,4,5,6,7<br>Diameter - 5,<br>5.5,6,7,7.8<br>Facial Heights - 3,5 | Titanium Healing<br>abutments<br>Diameter - 3.8<br>Facial Heights -<br>2,3,4,5,6,7<br>Diameter - 5.5<br>Facial Heights - 3,4,5,6 |
| Zirconia Abutments -<br>Anatomic | Diameter - 4.5<br>Facial Heights - 1,2,3 | | | Straight Anatomic<br>Zirconia<br>Diameter - 4.5<br>Facial Heights - 1,2,3,4 |
| Zirconia Healing<br>abutments | | | | |
| Multi unit abutments | Straight<br>Diameter - 4.7<br>Facial Heights - 1,2,3,4<br>17 Degree<br>Diameter - 5.3<br>Facial Heights -1,2,3<br>30 Degree<br>Diameter - 5.4<br>Facial Heights - 1,2,3 | Multi unit abutments<br>Straight<br>Diameter - 4.5<br>Facial Heights -<br>1,2,3,4,5<br>17 Degree<br>Diameter - 4.5<br>Facial Heights -2.25,3,4<br>30 Degree<br>Diameter - 3.5,4.5 | Straight<br>Diameter - 4.7<br>Facial Heights -<br>1.5,2.5,3,4,5<br>17 Degree<br>Diameter - 5.3<br>Facial Heights -2,3<br>30 Degree<br>Diameter - 5.4<br>Facial Heights - 2,3 | Straight<br>Diameter - 4.7<br>Facial Heights -<br>1,2,3,4,5,6<br>17 Degree<br>Diameter - 5.3<br>Facial Heights -2,3<br>30 Degree<br>Diameter - 5.4<br>Facial Heights - 2,3 |
| | | Facial Heights - 3,4,5 | | |
| PEEK temporary<br>abutments Straight | Diameter - 4.5<br>Facial Heights - 1,2,3 | Temporary PEEK<br>abutments<br>Diameter -<br>3.0,3.5,4.5,5.7<br>Facial Heights - 3 | | |
| PEEK temporary<br>abutments Anatomic | Diameter - 4.5<br>Facial Heights - 1,2,3 | | Temporary PEEK<br>abutments<br>Diameter - 5.4<br>Facial Heights - 1,2,3 | |
| Ball attachments | Heights - 1,2,3,4,5,6 | Ball attachments<br>Diameter - 4.5<br>Facial Heights - 1,3,5 | Titanium Ball<br>attachments<br>Heights - 0.5,2,3,4,5,6 | Titanium Ball<br>attachments<br>Heights - 1,2,3,4,5,6 |
{18}------------------------------------------------
{19}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.