3M Comply Hydrogen Peroxide Chemical Indicator 1248
K170321 · 3M Company · JOJ · Aug 14, 2017 · General Hospital
Device Facts
Record ID
K170321
Device Name
3M Comply Hydrogen Peroxide Chemical Indicator 1248
Applicant
3M Company
Product Code
JOJ · General Hospital
Decision Date
Aug 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100, 100S, NX® (Standard and Advanced cycles), 100NX® (Standard, Flex, Express and Duo cycles), AMSCO® V-PRO™ 1 (Lumen cycle), AMSCO® V-PRO™ 1 Plus (Lumen and Non Lumen cycles), and AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles) sterilizers. The chemical indicator bar turns from blue to pink after exposure to vaporized hydrogen peroxide.
Device Story
Chemical indicator for vaporized hydrogen peroxide sterilization; consists of noncellulosic plastic strip with printed chemical indicator bar and reference color match; used as internal pack process indicator in healthcare facilities; indicator bar changes from blue to pink upon exposure to hydrogen peroxide vapor; provides visual confirmation of sterilization cycle exposure; aids healthcare personnel in verifying sterilization process; no direct patient contact; indirect patient exposure via sterilized instruments deemed minimal.
Clinical Evidence
Bench testing only. No clinical data included. Performance verified through color change testing across six lots in AMSCO V-PRO maX cycles (Lumen, Non Lumen, Flexible) and stability testing under office lighting for 4 weeks.
Technological Characteristics
Noncellulosic plastic substrate; alkali blue 6B dye indicator; visual color change (blue to pink); 2-year shelf life; 4-week post-exposure endpoint stability; passive chemical indicator; no energy source or software.
Indications for Use
Indicated for use as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in specified STERRAD and AMSCO V-PRO low-temperature sterilization systems.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 (K150622)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The image is a standard black and white image with the address information.
August 14, 2017
3M Company Nadia Battah Regulatory Affairs Associate 3M Health Center 3M Center, Building 275-5W-06 St Paul. Minnesota 55144-1000
Re: K170321
Trade/Device Name: 3M Comply Hydrogen Peroxide Chemical Indicator 1248 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: July 12, 2017 Received: July 13, 2017
Dear Nadia Battah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K170321
Device Name
3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248
Indications for Use (Describe)
Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100, 100S, NX® (Standard and Advanced cycles), 100NX® (Standard, Flex, Express and Duo cycles), AMSCO® V-PRO™ 1 (Lumen cycle), AMSCO® V-PRO™ 1 Plus (Lumen and Non Lumen cycles), and AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles) sterilizers. The chemical indicator bar turns from blue to pink after exposure to vaporized hydrogen peroxide.
Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Premarket Notification [510(k)] Summary
#### Sponsor Information:
3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact Person: Nadia Battah Regulatory Affairs Associate Phone Number: (651) 733-0929 FAX Number: (651) 737-5320
#### Date of Summary: July 31, 2017
#### Device Name and Classification:
| Common or Usual Name: | Chemical Indicator |
|------------------------|--------------------------------------------------------------------------|
| Proprietary Name: | 3M™ Comply™ Hydrogen Peroxide Chemical Indicator<br>1248 |
| Classification Name: | Indicator, physical/chemical sterilization process (21<br>CFR §880.2800) |
| Device Classification: | Class II |
| Product Code: | JOJ |
#### Predicate Device:
3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 (K150622)
#### Description of Device:
The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 consists of a noncellulosic plastic material onto which a chemical indicator bar is printed. A comparison color match is also printed on the product to aid in color interpretation.
{4}------------------------------------------------
#### Nonclinical Comparison to the Predicate Device
The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is the same design as the previously cleared device of the same model number (the predicate) which is sold under the tradename 3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248 (K150622). There has been no change to the device's materials, performance specifications, or fundamental scientific technology. The intent of this submission is to expand the indications for use to include use in the AMSCO® V-PRO™ 1 (Lumen cycle), AMSCO® V-PRO™ 1 Plus (Lumen and Non Lumen cycles), and AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles) sterilizers.
#### Summary of Clinical Testing
No clinical data was included in this premarket application submission.
#### Indications for Use
Use the 3M™ ComplyTM Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100, 100S, NX® (Standard and Advanced cycles), 100NX® (Standard, Flex, Express and Duo cycles), AMSCO® V-PRO™ 1 (Lumen cycle), AMSCO® V-PRO™ 1 Plus (Lumen and Non Lumen cycles), and AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles) sterilizers. The chemical indicator bar turns from blue to pink after exposure to vaporized hydrogen peroxide.
{5}------------------------------------------------
# Comparison to Predicate Device
| Feature | Submission Device:<br>3M™ Comply™ Hydrogen Peroxide<br>Chemical Indicator 1248 | Predicate Device (K150622):<br>3M™ Comply™ Hydrogen Peroxide<br>Chemical Indicator 1248 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Use the 3MTM ComplyTM Hydrogen Peroxide<br>Chemical Indicator 1248 as an internal pack<br>process indicator to verify exposure to<br>vaporized hydrogen peroxide in the<br>STERRAD® 100, 100S, NX® (Standard and<br>Advanced cycles), 100NX® (Standard, Flex,<br>Express and Duo cycles), AMSCO® V-<br>PROTM 1 (Lumen cycle), AMSCO® V-<br>PROTM 1 Plus (Lumen and Non Lumen<br>cycles), and AMSCO® V-PRO™ maX Low<br>Temperature Sterilization System (Lumen,<br>Non Lumen, and Flexible cycles) sterilizers.<br>The chemical indicator bar turns from blue to<br>pink after exposure to vaporized hydrogen<br>peroxide. | Use the 3MTM ComplyTM Hydrogen Peroxide<br>Chemical Indicator 1248 as an internal pack<br>process indicator to verify exposure to<br>vaporized hydrogen peroxide in the<br>STERRAD® 100, 100S, NX® (Standard and<br>Advanced cycles) and 100NX® (Standard,<br>Flex, Express and Duo cycles) sterilization<br>processes. The chemical indicator bar turns<br>from blue to pink after exposure to vaporized<br>hydrogen peroxide. |
| Sterilizers | STERRAD® 100<br>STERRAD® 100S<br>STERRAD® NX (Standard and Advanced<br>cycles)<br>STERRAD® 100NX (Standard, Flex, Express,<br>and Duo cycles)<br>AMSCO® V-PRO™ 1 (Lumen cycle)<br>AMSCO® V-PRO™ 1 Plus (Lumen and Non<br>Lumen cycles)<br>AMSCO® V-PRO™ maX Low Temperature<br>Sterilization System (Lumen, Non Lumen and<br>Flexible cycles) | STERRAD® 100<br>STERRAD® 100S<br>STERRAD® NX (Standard and Advanced<br>cycles)<br>STERRAD® 100NX (Standard, Flex,<br>Express, and Duo cycles) |
| Indicator Agent | Alkali blue 6B dye | Identical |
| Cycle Conditions<br>for Color Change in<br>STERRAD® 100 | Testing verified that the 3MTM Comply™ Hydrogen Peroxide Chemical Indicators 1248 turned<br>from blue to pink when exposed to the STERRAD® 100 Sterilization cycle and the minimum<br>time required for all indicators to indicate a "pass" in relation to the color match was found to<br>be between 10 and 15 minutes.<br>Demonstrated per submission K013228. | |
| Cycle Conditions<br>for Color Change in<br>STERRAD® 100S | Testing verified that the 3MTM Comply™ Hydrogen Peroxide Chemical Indicators 1248 turned<br>from blue to pink when exposed to the STERRAD® 100S Sterilization cycle and the minimum<br>time required for all indicators to indicate a "pass" in relation to the color match was found to<br>be between 4 and 8 minutes.<br>Demonstrated per submission K013228. | |
| Cycle Conditions<br>for Color Change in<br>STERRAD® NX<br>(Standard and<br>Advanced cycles) | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 turns<br>from blue to pink to indicate a "pass" in relation to the device color match printed on the<br>indicator when exposed to complete Standard and Advanced cycles in the STERRAD® NX®<br>sterilizer. The minimum time required for all samples to indicate a "pass" in these cycles is<br>listed below:<br>Standard cycle: Between 5.25 minutes and 7 minutes<br>Advanced cycle: Between 11.25 minutes and 15 minutes<br>Demonstrated per submission K150622. | |
| Cycle Conditions<br>for Color Change in<br>STERRAD®<br>100NX (Standard,<br>Flex Express and<br>Duo cycles) | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 turns<br>from blue to pink to indicate a "pass" in relation to the device color match printed on the<br>indicator when exposed to complete Standard, Flex, Express, and Duo cycles in the<br>STERRAD® 100NX® sterilizer. The minimum time required for all samples to indicate a<br>"pass" in these cycles is listed below: | |
| | Flex cycle: Between 12.75 minutes and 17 minutes | |
| | Express cycle: Between 4.5 and 6 minutes | |
| | Duo cycle. Between 10 seconds and 8 minutes | |
| | Demonstrated per submission K150622. | |
| Cycle Conditions<br>for Color Change in<br>AMSCO® V-<br>PROTM 1 (Lumen<br>cycle), AMSCO®<br>V-PROTM 1 Plus<br>(Lumen and Non<br>Lumen cycles), and<br>AMSCO® V-<br>PROTM maX Low<br>Temperature<br>Sterilization System<br>(Lumen, Non<br>Lumen, and<br>Flexible cycles)<br>sterilizers | Testing verified that the 3MTM ComplyTM<br>Hydrogen Peroxide Chemical Indicator 1248<br>turns from blue to pink to indicate a “pass” in<br>relation to the color match when exposed to a<br>complete Lumen, Non Lumen or Flexible<br>cycle in an AMSCO® V-PROTM maX Low<br>Temperature Sterilization System. The<br>minimum time required for all samples to<br>indicate a “pass” in these cycles is listed<br>below: | |
| | Lumen cycle: Between 7 seconds and 8<br>minutes | |
| | Non Lumen cycle: Between 6 minutes and 9<br>minutes. | |
| | Flexible cycle: Between 9 minutes and 12<br>minutes. | |
| | | |
| | | |
| | | |
| | | |
| | | |
| Stability of the<br>endpoint reaction | At least four (4) weeks (28 days) | Identical |
| Shelf life | Two (2) years | Identical |
{6}------------------------------------------------
## Biocompatibility
The components of the indicator ink used on the 3MTM Comply™ Hydrogen Peroxide Chemical Indicator 1248 are all of low general toxicity. The exposure to the health care professional is minimal and well below any identified toxic thresholds for the compounds. There is no anticipated direct exposure to the patient. Indirect exposure to the patient via potential exposure to medical instruments with the indicator ink is minimal and well below any identified toxic thresholds for the compounds.
#### Summary of Nonclinical Testing
The effectiveness of the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is demonstrated by the following tests:
- Color Change in Health Care Facility Cycle Samples from six different lots were verified to meet the requirements for detectable color change in the AMSCO® V-PRO™ maX Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles) sterilizer.
- Minimum Exposure Parameters to Affect the Change of the Indicator in Health Care Facility Cycle - Samples from six different lots were evaluated to determine the minimum time required for the color change of the indicator when used in the AMSCO® V-PROTM maX Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles).
{7}------------------------------------------------
- End Point Color Stability Samples from six different lots were verified to be stable to storage under typical office lighting conditions for at least one month (4 weeks) post-exposure in a Hydrogen Peroxide Health Care Facility Cycle.
### Conclusion
The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is as safe, as effective, and performs as well as the predicate, K150622, 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.