3M Comply Hydrogen Peroxide Chemical Indicator 1248
K150622 · 3M Company · JOJ · Sep 29, 2015 · General Hospital
Device Facts
Record ID
K150622
Device Name
3M Comply Hydrogen Peroxide Chemical Indicator 1248
Applicant
3M Company
Product Code
JOJ · General Hospital
Decision Date
Sep 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Intended Use
Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100. 100S. NX® (Standard and Advanced cycles) and 100NX® (Standard, Flex, Express and Duo cycles) sterilization processes. The chemical indicator bar turns from blue to pink after exposure to vaporized hydrogen peroxide.
Device Story
Device is a chemical indicator strip used in healthcare facilities to verify sterilization of medical instruments. It consists of a non-cellulosic plastic substrate printed with a chemical indicator bar (Alkali blue 6B dye) and a reference color match. During sterilization, the indicator bar undergoes a color change from blue to pink upon exposure to vaporized hydrogen peroxide. The device is placed inside instrument packs; healthcare personnel visually inspect the indicator after the sterilization cycle to confirm exposure. This provides a qualitative verification of the sterilization process, helping ensure instruments are safe for clinical use. The device is a passive, non-electronic tool requiring no external power or software.
Clinical Evidence
Bench testing only. Performance verified across six lots in STERRAD® NX® (Standard/Advanced) and 100NX® (Standard/Flex/Express/Duo) cycles. Testing confirmed detectable color change (blue to pink), minimum exposure parameters for color change, and endpoint color stability for at least one month post-exposure under typical office lighting.
Technological Characteristics
Non-cellulosic plastic substrate with printed chemical indicator bar (Alkali blue 6B dye). Passive chemical reaction; no energy source. Dimensions/form factor: strip. Shelf life: 2 years. No software or connectivity.
Indications for Use
Indicated for use as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in STERRAD® 100, 100S, NX® (Standard/Advanced), and 100NX® (Standard/Flex/Express/Duo) sterilization processes.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M™ Comply™ 1248 Gas Plasma Chemical Indicator (K013228)
Related Devices
K203458 — 3M Comply Hydrogen Peroxide Chemical Indicator 1248 · 3M Company · Dec 11, 2020
K192937 — 3M Comply Hydrogen Peroxide Chemical Indicator · 3M Company · Nov 16, 2019
K090650 — SPSMEDICAL VAPORIZED HYDROGEN PEROXIDE CHEMICAL INDICATORS · Spsmedical Supply Corp. · Oct 9, 2009
K183211 — 3M Comply Hydrogen Peroxide Chemical Indicator · 3M Company · Feb 26, 2019
K222508 — 3M Comply Hydrogen Peroxide Chemical Indicator 1248 · 3M Company · Nov 1, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
September 29, 2015
3M Health Care, Ms. Hilary B. Hovde Regulatory Affairs 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144
Re: K150622
Trade/Device Name: 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: August 24, 2015 Received: August 27, 2015
Dear Ms. Hovde,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K150622
Device Name
3MTM ComplyTM Hydrogen Peroxide Chemical Indicator 1248
Indications for Use (Describe)
Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100. 100S. NX® (Standard and Advanced cycles) and 100NX® (Standard, Flex, Express and Duo cycles) sterilization processes. The chemical indicator bar turns from blue to pink after exposure to vaporized hydrogen peroxide.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
### Premarket Notification [510(k)] Summary
Image /page/3/Picture/1 description: The image shows the 3M logo. The logo consists of the number "3" followed by the letter "M", both in a bold, sans-serif font. The color of the text is red, and the background is white.
### Sponsor Information:
3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact Person: Hilary B. Hovde Regulatory Affairs Phone Number: (651) 736-0364 FAX Number: (651) 737-5320
#### Date of Summary: September 29, 2015
## Device Name and Classification:
| Common or Usual Name: | Chemical Indicator |
|------------------------|---------------------------------------------------------------------------|
| Proprietary Name: | 3M™ Comply™ Hydrogen Peroxide Chemical Indicator<br>1248 |
| Classification Name: | Indicator, physical/chemical sterilization process (21 CFR §<br>880.2800) |
| Device Classification: | Class II |
| Product Code: | JOJ |
### Predicate Device:
- 3M™ Comply™ 1248 Gas Plasma Chemical Indicator (K013228) .
### Description of Device:
The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 consists of a non-cellulosic plastic material onto which a chemical indicator bar is printed. A comparison color match is also printed on the product to aid in color interpretation.
{4}------------------------------------------------
### Nonclinical Comparison to the Predicate Device
The 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is the same design as the previously cleared device of the same model number (the predicate) which is sold under the tradename 3M™ Comply™ 1248 Gas Plasma Chemical Indicator (K013228). There has been no change to the device's materials, performance specifications, or fundamental scientific technology. The intent of this submission is to expand the indications for use to include use in the STERRAD® NX® and STERRAD® 100NX® sterilizers.
## Summary of Clinical Testing
No clinical data was included in this premarket application submission.
### Indications for Use
Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as an internal pack process indicator to verify exposure to vaporized hydrogen peroxide in the STERRAD® 100, 100S, NX® (Standard and Advanced cycles) and 100NX® (Standard, Flex, Express and Duo cycles) sterilization processes. The chemical indicator bar turns from blue to pink after exposure to vaporized hydrogen peroxide.
{5}------------------------------------------------
# Comparison to Predicate Device
| Feature | Submission Device:<br>3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 | Predicate Device (K013228):<br>3M™ Comply™ 1248 Gas<br>Plasma Chemical Indicator |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | Use the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 as<br>an internal pack process indicator to verify exposure to vaporized<br>hydrogen peroxide in the STERRAD® 100, 100S, NX® (Standard and<br>Advanced cycles) and 100NX® (Standard, Flex, Express and Duo<br>cycles) sterilization processes. The chemical indicator bar turns from<br>blue to pink after exposure to vaporized hydrogen peroxide. | The 3M™ Comply™ 1248 Gas<br>Plasma Chemical Indicator is<br>indicated for use as an internal<br>pack process indicator to verify<br>exposure to vapor hydrogen<br>peroxide in the STERRAD® 100,<br>STERRAD® 100S, and<br>STERRAD® 50 Sterilization<br>processes. |
| Sterilizers | STERRAD® 100<br>STERRAD® 100S<br>STERRAD® NX® (Standard and Advanced cycles)<br>STERRAD® 100NX® (Standard, Flex, Express, and Duo cycles) | STERRAD® 100<br>STERRAD® 100S<br>STERRAD® 50 |
| Indicator Agent | Alkali blue 6B dye | Identical |
| Cycle Conditions<br>for Color Change<br>in STERRAD®<br>100 | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical Indicators 1248 turned from blue to<br>pink when exposed to the STERRAD® 100 Sterilization cycle and the minimum time required for all indicators<br>to indicate a "pass" in relation to the color match was found to be between 10 and 15 minutes.<br>Demonstrated per original submission K013228. | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical Indicators 1248 turned from blue to pink when exposed to the STERRAD® 100 Sterilization cycle and the minimum time required for all indicators to indicate a "pass" in relation to the color match was found to be between 10 and 15 minutes.<br>Demonstrated per original submission K013228. |
| for Color Change<br>in STERRAD®<br>100S | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical Indicators 1248 turned from blue to<br>pink when exposed to the STERRAD® 100S Sterilization cycle and the minimum time required for all<br>indicators to indicate a "pass" in relation to the color match was found to be between 4 and 8 minutes.<br>Demonstrated per original submission K013228. | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical Indicators 1248 turned from blue to pink when exposed to the STERRAD® 100S Sterilization cycle and the minimum time required for all indicators to indicate a "pass" in relation to the color match was found to be between 4 and 8 minutes.<br>Demonstrated per original submission K013228. |
| Cycle Conditions<br>for Color Change<br>in STERRAD®<br>50 | Not Applicable. | Testing verified that the 3M™<br>Comply™ Hydrogen Peroxide<br>Chemical Indicators 1248 turned<br>from blue to pink when exposed to<br>the STERRAD® 50 Sterilization<br>cycle and the minimum time<br>required for all indicators to<br>indicate a "pass" in relation to the<br>color match was found to be<br>between 0.25 and 2 minutes.<br>Demonstrated per original<br>submission K013228. |
| Cycle Conditions<br>for Color Change<br>in STERRAD®<br>NX® | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical<br>Indicators 1248 turned from blue to pink and the minimum time required<br>for all indicators to indicate a "pass" in relation to the color specification<br>was found when exposed to the following STERRAD® NX®<br>Sterilization Cycles:<br>Standard<br>Advanced | |
| Cycle Conditions<br>for Color Change<br>in STERRAD®<br>100NX® | Testing verified that the 3M™ Comply™ Hydrogen Peroxide Chemical<br>Indicators 1248 turned from blue to pink and the minimum time required<br>for all indicators to indicate a "pass" in relation to the color specification<br>was found when exposed to the following STERRAD® 100NX®<br>Sterilization Cycles:<br>Standard<br>Flex<br>Express<br>Duo | |
| Stability of the<br>endpoint reaction | At least one month (4 weeks) | Identical |
| Shelf life | Two (2) years | Identical |
{6}------------------------------------------------
### Biocompatibility
The components of the indicator ink used on the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 are all of low general toxicity. The exposure to the health care professional is minimal and well below any identified toxic thresholds for the compounds. There is no anticipated direct exposure to the patient. Indirect exposure to the patient via potential exposure to medical instruments with the indicator ink is minimal and well below any identified toxic thresholds for the compounds.
### Effectiveness
The effectiveness of the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 is demonstrated by the following tests:
- Color Change in Health Care Facility Cycle Samples from six different lots were verified to meet the requirements for detectable color change in the STERRAD® NX® (Standard and Advanced cycles) and the STERRAD® 100NX® (Standard, Flex, Express, and Duo cycles) sterilizers.
- Minimum Exposure Parameters to Affect the Change of the Indicator in Health Care Facility Cycle - Samples from six different lots were evaluated to determine the minimum time required for the color change of the indicator when used in the STERRAD® NX® (Standard and Advanced cycles) and the STERRAD® 100NX® (Standard, Flex, Express, and Duo cycles) sterilizers.
- End Point Color Stability Samples from four different lots were verified to be stable to storage under typical office lighting conditions for at least one month post-exposure in a Hydrogen Peroxide Health Care Facility Cycle.
## Conclusion
The composition, physical properties, specifications and technological characteristics of the 3M™ Comply" Hydrogen Peroxide Chemical Indicator 1248 remain the same as the predicate device.
Testing was conducted to confirm performance of the 3M™ Comply™ Hydrogen Peroxide Chemical Indicator 1248 in the STERRAD® NX® and the STERRAD® 100NX® per FDA's Premarket Notification [510(k)] Submissions for Chemical Indicators; Guidance for Industry and FDA Staff, December 19, 2003.
The 3M™ Comply " Hydrogen Peroxide Chemical Indicator 1248 is substantially equivalent to the predicate device. There are no new questions of safety or effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.