Sterilucent VH2O2 Chemical Indicators

K191999 · Sterilucent, Inc. · JOJ · Oct 21, 2019 · General Hospital

Device Facts

Record IDK191999
Device NameSterilucent VH2O2 Chemical Indicators
ApplicantSterilucent, Inc.
Product CodeJOJ · General Hospital
Decision DateOct 21, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Intended Use

A chemical indicator for monitoring all cycles within the Sterilucent™ HC 80TT Hydrogen Peroxide Sterilizer (Lumen & Flexible), STERRAD® 100S (Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard & Advanced), STERIS® VPRO™ 1, V-PRO™ 1 Plus (Lumen & Non-Lumen). VPRO® maX (Flexible. Lumen & Non-Lumen) and Sterilucent PSD-85 (Lumen & Non-Lumen). The VH2O2 Indicators are intended to be used by health care providers with sterilization wraps, containers, cassettes, or pouches to distinguish between processed units. Colors other than blue such as yellow/green should be treated as a process failure.

Device Story

Chemical process indicator (Class 1/Type 1) for vaporized hydrogen peroxide (VHP) sterilization; monitors exposure to sterilization cycles. Provided as strips or labels with pressure-sensitive adhesive. Operates via chemical reaction; indicator ink (tetraethylammonium bromide) changes color from red/pink to blue upon VHP exposure. Used by healthcare providers in clinical settings; placed with sterilization wraps, containers, or pouches. Visual color change confirms load exposure to sterilization process; blue indicates successful exposure, while yellow/green indicates process failure. Benefits patient by providing visual verification that medical devices have undergone sterilization processing.

Clinical Evidence

No clinical data. Bench testing only. Performance demonstrated via compliance with ANSI/AAMI/ISO 11140-1:2014, including shelf-life testing, indicator functionality in specified sterilizers, endpoint color stability, and biocompatibility.

Technological Characteristics

Class 1/Type 1 chemical indicator. Materials: polymeric substrate (Tyvek, polypropylene, polystyrene). Indicator agent: tetraethylammonium bromide. Ink components: water-based carrier, activator, indicator agent, modulator. Form factor: strip or label. Conforms to ANSI/AAMI/ISO 11140-1:2014. Shelf life: 2 years. Storage: 15-30°C, dry, light-protected.

Indications for Use

Indicated for health care providers to monitor vaporized hydrogen peroxide sterilization cycles in specified sterilizers (Sterilucent HC 80TT, STERRAD, STERIS V-PRO, Sterilucent PSD-85) by distinguishing between processed and unprocessed units using sterilization wraps, containers, cassettes, or pouches.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 21, 2019 Sterilucent, Inc. Peter Kalkbrenner Director of Engineering 1400 Marshall St. NE Minneapolis, Minnesota 55413 Re: K191999 Trade/Device Name: Sterilucent VH2O2 Chemical Indicators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: July 25, 2019 Received: July 26, 2019 Dear Peter Kalkbrenner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K191999 Device Name Sterilucent VH2O2 Chemical Indicator #### Indications for Use (Describe) A chemical indicator for monitoring all cycles within the Sterilucent™ HC 80TT Hydrogen Peroxide Sterilizer (Lumen & Flexible), STERRAD® 100S (Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard & Advanced), STERIS® VPRO™ 1, V-PRO™ 1 Plus (Lumen & Non-Lumen). VPRO® maX (Flexible. Lumen & Non-Lumen) and Sterilucent PSD-85 (Lumen & Non-Lumen). The VH2O2 Indicators are intended to be used by health care providers with sterilization wraps, containers, cassettes, or pouches to distinguish between processed units. Colors other than blue such as yellow/green should be treated as a process failure. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) × | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary for the Sterilucent VH2O2 Chemical Indicator | Submitted by: | Sterilucent, Inc. | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Peter R. Kalkbrenner | | | Director of Engineering | | | peter.kalkbrenner@sterilucent.com | | | Phone: 612-767-3253 | | | Fax: 612-767-3261 | | Date of Summary: | 21 October 2019 | | Device Trade Name: | Sterilucent VH2O2 Chemical Indicator | | Common or Usual Name: | Chemical sterilization process indicator | | Classification: | 21 CFR 880.2800(b) | | Class: | II | | Product Code: | JOJ | | Predicate Device(s): | SPSmedical VH2O2 Indicators (K140566/ K110152) | | Device Description: | The Sterilucent VH2O2 Chemical Indicator is a Class 1/Type 1 process<br>indicator that conforms to ANSI/AAMI/ISO 11140-1:2014, and is intended<br>to be used with currently cleared hydrogen peroxide sterilizers (STERRAD®<br>100S, 200, 100NX, NX, STERIS® VPRO™ 1, V-PRO™ 1 Plus, VPRO®<br>maX and the Sterilucent PSD-85) and the Sterilucent HC 80TT Hydrogen<br>Peroxide Sterilizer. The HC 80TT Hydrogen Peroxide Sterilizer has two pre-<br>programmed sterilization cycles ("Lumen" and "Flexible") which both<br>utilize vaporized hydrogen peroxide (VHP) to rapidly sterilize a variety of<br>reusable medical devices. | | | The Sterilucent VH2O2 Chemical Indicator is provided in two different<br>formats: Strip and Label. Both formats consist of a polymeric material on<br>which indicator ink is deposited. A pressure-sensitive adhesive is provided<br>on the back of the Label, which allows the process indicator to be adhered to<br>various substrates including containers, pouches, tamper-proof arrows or<br>locks, etc. | | | The Sterilucent VH2O2 Chemical Indicator provides a visual indication that a<br>sterilization load has been exposed to the VHP sterilization process. The<br>indicator works by means of a chemical reaction, which results in a<br>recognizable color change from red to blue. | | Indication for Use: | A chemical indicator for monitoring all cycles within the Sterilucent™ HC<br>80TT Hydrogen Peroxide Sterilizer (Lumen & Flexible), STERRAD® 100S<br>(Standard & Long), 200, 100NX (Standard, Flex & Express), NX (Standard<br>& Advanced), STERIS® VPRO™ 1, V-PRO™ 1 Plus (Lumen & Non-<br>Lumen), VPRO® maX (Flexible, Lumen & Non-Lumen) and Sterilucent<br>PSD-85 (Lumen & Non-Lumen). The VH2O2 Indicators are intended to be<br>used by health care providers with sterilization wraps, containers, cassettes,<br>or pouches to distinguish between processed and unprocessed units. Colors<br>other than blue such as yellow/green should be treated as a process failure. | ## Technological Characteristics Table: | | Proposed<br>Sterilucent VH2O2 Chemical<br>Indicators | Predicate<br>SPSmedical VH2O2 Indicators<br>(K140566/ K110152) | Comparison | |--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Manufacturer | Crosstex-SPSmedical | Crosstex-SPSmedical | Same | | Proposed<br>Sterilucent VH2O2 Chemical<br>Indicators | | Predicate<br>SPSmedical VH2O2 Indicators<br>(K140566/ K110152) | Comparison | | Intended Use | Single use sterilization process indicator | Single use sterilization process indicator | Same | | Indications for Use | A chemical indicator for monitoring all cycles within the Sterilucent™ HC<br>80TT Hydrogen Peroxide Sterilizer<br>(Lumen & Flexible), STERRAD®<br>100S (Standard & Long), 200, 100NX<br>(Standard, Flex & Express), NX<br>(Standard & Advanced), STERIS®<br>VPRO™ 1, V-PRO™ 1 Plus (Lumen<br>& Non-Lumen), VPRO® maX<br>(Flexible, Lumen & Non-Lumen) and<br>Sterilucent PSD-85 (Lumen & Non-<br>Lumen). The VH2O2 Indicators are<br>intended to be used by health care<br>providers with sterilization wraps,<br>containers, cassettes, or pouches to<br>distinguish between processed and<br>unprocessed units. Colors other than<br>blue such as yellow/green should be<br>treated as a process failure. | A chemical indicator for monitoring<br>all cycles within the STERRAD®<br>100S (Standard & Long), 200, 100NX<br>(Standard, Flex & Express), NX<br>(Standard & Advanced), STERIS®<br>VPRO™ 1, V-PRO™ 1 Plus (Lumen<br>& Non-Lumen), VPRO® maX<br>(Flexible, Lumen & Non-Lumen) and<br>Sterilucent™ PSD-85 (Lumen & Non-<br>Lumen). The VH2O2 Indicators are<br>intended to be used by health care<br>providers with sterilization wraps,<br>containers, cassettes, or pouches to<br>distinguish between processed and<br>unprocessed units. Colors other than<br>blue such as yellow/green should be<br>treated as a process failure. | Similar | | Device Design | Strip, Label | Strip, Label, Tape | Similar | | Endpoint Specifications -<br>Color Change Upon<br>Exposure to H2O2 | Red/Pink changes to Blue | Red/Pink changes to Blue | Same | | Indicator Agent | Tetraethylammonium<br>Bromide(ammonium salt) | Tetraethylammonium<br>Bromide(ammonium salt) | Same | | Indicator Ink<br>Components | Water-based Carrier; Activator,<br>Indicator Agent, & Modulator | Water-based Carrier; Activator,<br>Indicator Agent, & Modulator | Same | | Substrate | Polymer (Tyvek, Polypropylene,<br>polystyrene) | Polymer (Tyvek, Polypropylene,<br>polystyrene) | Same | | Conformance Testing<br>under ISO 11140-1 for<br>VH2O2 Indicators | Comply with Standard | Comply with Standard | Same | | Full Cycle In-Use<br>Validation Testing of CI<br>with listed sterilizers | Yes | Yes | Same | | Shelf Life | Up to 2 years from date of manufacture | Up to 2 years from date of<br>manufacture | Same | | Distributor | Sterilucent & SPSmedical | Sterilucent & SPSmedical | Same | | Storage Conditions | Dry area, ambient temperature of 15° to<br>30° C, away from alkaline chemicals,<br>acids, and sources of light | Dry area, ambient temperature of 15°<br>to 30° C, away from alkaline<br>chemicals, acids, and sources of light | Same | {4}------------------------------------------------ The Sterilucent VH2O2 Chemical Indicator and predicate device are both single-use process indicators intended to monitor exposure to sterilization cycles utilizing vaporized hydrogen peroxide as a sterilant. Both devices have the same intended use, design, materials, specifications, and modeof-action, and are manufactured using the exact same production methods. {5}------------------------------------------------ ### Summary of Non-Clinical Testing: Performance testing was conducted to demonstrate the functionality of the Sterilucent VH2O2 Chemical Indicator and general conformance with the requirements for Class 1/Type 1 vaporized hydrogen peroxide sterilization indicators as detailed in ANSI/AAMI/ISO 11140-1:2014. A summary of nonclinical tests performed follows: | Test | Standard Used | Goal | Result | |-------------------------------------------------------|---------------|----------------------------------------------------------------------------------------------------|--------| | ISO 11140-1 compliance | ISO 11140-1 | Demonstrate compliance to ISO<br>11140-1 pass/fail criteria | Passed | | Shelf-life | ISO 11140-1 | Demonstrate compliance to ISO<br>11140-1 pass/fail criteria on devices<br>at the stated shelf life | Passed | | Sterilucent VH2O2 Chemical<br>Indicator Functionality | NA | Demonstrate appropriate color<br>change in applicable sterilizer | Passed | | Endpoint Color Stability | NA | Demonstrate processed indicator<br>endpoint color stability | Passed | | Biocompatibility | NA | Demonstrate CI does not release<br>any know toxic substances | Passed | Summary of Clinical Clinical evaluations were not required and therefore are not submitted with this Testing: 510(k). Overall Performance Based on the nonclinical tests performed, the subject device, the Sterilucent Conclusion: VH2O2 Chemical Indicator, is as safe, as effective and performs as well as or better than the legally marketed predicate device, the SPSmedical VH2O2 Indicators.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...