Unity Network ID

K170199 · GE Healthcare · MWI · May 12, 2017 · Cardiovascular

Device Facts

Record IDK170199
Device NameUnity Network ID
ApplicantGE Healthcare
Product CodeMWI · Cardiovascular
Decision DateMay 12, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The Unity Network ID is indicated for use in data collection and clinical information management through networks with independent bedside devices. The Unity Network ID is not intended for monitoring purposes, nor is the Unity Network ID intended to control any of the clinical devices (independent bedside devices/ information systems) it is connected to.

Device Story

Unity Network ID v8 acts as a data gateway/bridge; acquires digital data from eight RS-232 serial ports connected to independent bedside devices (e.g., ventilators, infusion pumps, patient monitors); converts proprietary RS-232 protocols into GEMS IT Unity Network protocol; transmits data to central stations, clinical information systems, or GE patient monitors. Operates in stand-alone or peripheral mode. Bidirectional communication is limited to information requests or patient demographic data transfer; no control commands sent to bedside devices. Used in clinical environments to aggregate real-time waveforms, numeric parameters, and alarm data for clinician review. Benefits include centralized data management and improved clinical information flow.

Clinical Evidence

No clinical studies performed. Evidence consists of bench testing, including software verification and validation (unit, integration, and regression testing) per FDA guidance, and electrical safety/EMC testing per IEC 60601-1 and IEC 60601-1-2 standards.

Technological Characteristics

Gateway device with eight RS-232 serial ports. Connectivity: Networked (Unity Network protocol). No patient contact materials. Software-based data conversion. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 62336 (usability).

Indications for Use

Indicated for data collection and clinical information management in clinical settings via network connection to independent bedside devices. Not for patient monitoring or control of connected devices.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a family of three people. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 12, 2017 GE Healthcare Robert Casarsa Regulatory Affairs Leader 8200 West Tower Ave. Milwaukee, Wisconsin 53223 Re: K170199 Trade/Device Name: Unity Network ID v8 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: April 13, 2017 Received: April 14, 2017 Dear Robert Casarsa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. M.A. Hillemann for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ### K170199 Device Name Unity Network ID v8 ### Indications for Use (Describe) The Unity Network ID is indicated for use in data collection and clinical information management through networks with independent bedside devices. The Unity Network ID is not intended for monitoring purposes, nor is the Unity Network ID intended to control any of the clinical devices (independent bedside devices/ information systems) it is connected to. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <span style="font-size:12pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510k Summary ### l. SUBMITTER GE Healthcare GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223 Contact Person: Robert Casarsa Phone: 414-362-3062 Date Prepared: April 7, 2017 #### II. DEVICE Name of Device: Unity Network ID v8 Common or Usual Name: Physiological Patient Monitor Classification Name: 21 CFR 870.2300 Monitor, Physiological, Patient (without arrhythmia detection or alarms Regulatory Class: II Product Code: MWI ### III. PREDICATE DEVICE Unity Network ID v7 K142840 This predicate has not been subject to a design-related recall. No reference devices were used in this submission. ### IV. DEVICE DESCRIPTION The Unity Network ID system communicates patient data from sources other than GE Medical Systems Information Technologies, Inc. equipment to a clinical information system, central station, and/or GE Medical Systems Information Technologies Inc. patient monitors. The Unity Network ID acquires digital data from eight serial ports, converts the data to Unity Network protocols, and transmits the data over the monitoring network to a Unity Network device such as a patient monitor, clinical information system or central station. Accessories include device specific interface cables and mounting hardware. {4}------------------------------------------------ ### V. INDICATONS FOR USE The Unity Network ID is indicated for use in data collection and clinical information management through networks with independent bedside devices. The Unity Network ID is not intended for monitoring purposes, nor is the Unity Network ID intended to control any of the clinical devices (independent bedside devices/ information systems) it is connected to. The indication for use is identical to the predicate device. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The Technological Characteristics of the Unity Network ID are exactly the same as all previous versions of the Unity Network ID. The device collects data from equipment connected to any of eight RS-232 serial communications ports. The data is converted from the independent bedside manufacturer's RS-232 protocol into the GEMS IT Unity Network protocol. Data collected consists of real-time waveforms, parameter numeric data, device confiquration data and parameter alarm data. The communication between the Unity Network ID and an interfaced device is bidirectional. However, communication to an interfaced device is limited to information requests or to transfer of patient demographic data. No commands are sent which could affect how the interfaced device operates. The device operates in either a stand-alone mode or a peripheral mode of operation. Hardware: - No change to the Unity Network ID hardware 1) - Create interface cables for the newly supported devices 2) - 3) Create new interface cable for existing device: - a) Draeger Fabius GS (K011404) ## Software: This version of Unity Network ID adds support for the following devices: - Capnostream 20p Monitor (K123690) (a) - (b) SenTec Digital Monitor (K151329) - (c) Hamilton G5 Ventilator (K131774) - Vista BIS (K072286) (d) - Edwards EV1000 (K160552) (e) - (f) Draeger Perseus A500 (K133886) {5}------------------------------------------------ - Draeger Babylog VN500 (K093632) (g) - (h) Draeger Savina 300 (K121886) - (i) Draeger Evita V300 (Not for sale in US) - (j) Draeger Oxylog 3000 Plus (K103625) - (k) Puritan Bennett 980 Ventilator (K131252) The following categories of devices are supported by the Unity Network ID: - Anesthesia Machines ● - . Continuous Cardiac Output Monitors - Gas Analyzers - Infusion Pumps ● - . Patient Monitors - . Pulse Oximeters - Transcutaneous Monitors - Urometers - . Ventilators - Blood Analyzer - . Bispectral Index Monitor #### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility testing No biocompatibility testing was performed as the Unity Network ID has no patient contact. ## Electrical safety and electromagnetic compatibility (EMC) The Unity Network ID was designed and tested for compliance with the following standards: - 1. IEC 60601-1:2005. +Corr 1:2006. +Corr 2:2007 +A1:2012 Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance - 2. IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility - Requirements and tests - 3. IEC 60601-1-6:2010 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability - 4. IEC 62336: 2007 + A1:2014 Medical devices -- Part 1: Application of usability engineering to medical devices {6}------------------------------------------------ ## Software Verification and Validation Testing Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, where a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information of through the action of a care provider. The following bench tests were performed: - Unit Testing: Code reviews and/or static code analysis are the methods are . adopted for the software unit verification to ensure the unit acceptance criteria, as defined in the SDLC procedure, are met. - Integration Testing: Functional procedures were executed to test the interface requirements. - Regression Testing: Regression testing was performed to ensure additional confidence of the build or to test the regression of the software build. ## Mechanical and acoustic testing No mechanical or acoustic testing was performed on the Unity Network ID as the only changes were software based. ## Animal Study No animal studies have been performed on the Unity Network ID. ## Clinical Studies No clinical studies have been performed on the Unity Network ID. ### VIII. CONCLUSIONS This version of the Unity Network ID (v8) is substantially equivalent to the prior versions (v1-v8). The submitted device uses the same technology for the interface, has no device hardware changes, and adds new interface cables for the newly interfaced devices. Verification testing has assured that the design outputs meets the requirements of the design inputs. Testing to internationally recognized standards demonstrate the safety of the device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...