K123690 · Oridion Medical 1987 , Ltd. · CCK · Aug 5, 2014 · Anesthesiology
Device Facts
Record ID
K123690
Device Name
CAPNOSTREAM20P WITH HIFI C02 MONITORING
Applicant
Oridion Medical 1987 , Ltd.
Product Code
CCK · Anesthesiology
Decision Date
Aug 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Capnostream®20p combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoqlobin (SpO2) and pulse rate. It is also indicated for continuous noninvasive monitoring of carboxyhemoglobin saturation (measured by an SpCO/SpMet/SpHb sensor), methemoglobin saturation (measured by an SpCO/SpMet/SpHb sensor) and total hemoglobin concentration (measured by an SpCO/SpMet/SpHb sensor). It is intended for use with neonatal, pediatric, and adult patients in hospital-type facilities, intra-hospital transport and home environments. Capnostream®20p is to be operated by qualified healthcare personnel only. The Capnostream®20p monitor provides the clinician with an integrated pulmonary index (IPI). The IPI is based on four parameters provided by the monitor: end tidal carbon dioxide, respiration rate, oxygen saturation and pulse rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status. The IPI is an adjunct to, and is not intended to replace, vital sign monitoring. The Capnostream20p HiFi mode provides a measurement of expired carbon dioxide and detects spontaneous breaths for infant/neonatal patients ventilated using high frequency oscillatory ventilation (HFOV).
Device Story
Bedside monitor integrating microMediCO2 capnography and pulse oximetry modules; processes CO2, respiration rate, SpO2, and pulse rate. Features Integrated Pulmonary Index (IPI) algorithm, providing single 1-10 score of ventilatory status. HiFi mode enables CO2 measurement and spontaneous breath detection for infants/neonates on High Frequency Oscillatory Ventilation (HFOV). Operated by qualified healthcare personnel in hospitals, transport, or home. Output displayed on host device; alerts clinicians to changes in pulmonary status. IPI serves as adjunct to vital sign monitoring; assists in clinical decision-making by providing consolidated status index. Benefits include continuous monitoring of complex ventilation states.
Clinical Evidence
Bench testing only. Performance validated using lung simulator system; accuracy of CO2 measurements in HFOV mode verified up to 900 BPM (15 Hz) for both passive and spontaneous breaths. CO2 range tested: 16-76 mmHg; spontaneous respiratory rates: 30, 45, and 60 BPM.
Technological Characteristics
Two-parameter bedside monitor; microMediCO2 capnography and pulse oximetry modules. Standards: ISO 21647, ISO 9919, IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 60601-1-8, ISO 14971, EN 980. Software-based IPI algorithm and HiFi mode for HFOV signal processing.
Indications for Use
Indicated for continuous, non-invasive monitoring of CO2, respiration rate, SpO2, pulse rate, carboxyhemoglobin, methemoglobin, and total hemoglobin in neonatal, pediatric, and adult patients in hospitals, transport, and home settings. HiFi mode indicated for infant/neonatal patients on HFOV.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2014
Ms Dalia Givony Regulatory Affairs Manager Oridion Medical 1987 Limited 7 Hamarpe Street Har Hotzvim Industrial Park Jerusalem, Israel 91450
Re: K123690
Trade/Device Name: Capnostream® 20p with HiFi CO2 Monitoring Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: June 15, 2014 Received: June 17, 2014
Dear Ms. Givony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Givony
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Response to FDA RFI- Capnostream®20p with HiFi CO2 monitoring - K123690
## Statement of Indications for Use
Capnostream®20p with HiFi CO2 monitoring (This document is not confidential)
510(k) Number (if known) K123690
Device Name: Capnostream®20p with HiFi CO2 monitoring
#### Intended Use/ Indication for Use
The Capnostream®20p combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoqlobin (SpO2) and pulse rate. It is also indicated for continuous noninvasive monitoring of carboxyhemoglobin saturation (measured by an SpCO/SpMet/SpHb sensor), methemoglobin saturation (measured by an SpCO/SpMet/SpHb sensor) and total hemoglobin concentration (measured by an SpCO/SpMet/SpHb sensor). It is intended for use with neonatal, pediatric, and adult patients in hospital-type facilities, intra-hospital transport and home environments.
Capnostream®20p is to be operated by qualified healthcare personnel only. The Capnostream®20p monitor provides the clinician with an integrated pulmonary index (IPI). The IPI is based on four parameters provided by the monitor: end tidal carbon dioxide, respiration rate, oxygen saturation and pulse rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status. The IPI is an adjunct to, and is not intended to replace, vital sign monitoring.
The Capnostream20p HiFi mode provides a measurement of expired carbon dioxide and detects spontaneous breaths for infant/neonatal patients ventilated using high frequency oscillatory ventilation (HFOV).
Prescription Use
AND/OR
Over-The-Counter Use
(Per 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Oridion Medical 1987 Ltd.
Todd D. Courtney -S 2014.08.04 14:22:31 -04'00'
August 28, 2013
{4}------------------------------------------------
Page 1 of 6
## 510(k) SUMMARY
This summary is being submitted in accordance with the requirements of 21 CFR 807.92
(This section is not confidential)
#### DATE THIS SUMMARY WAS PREPARED
July 31, 2014
### SUBMITTER'S NAME AND ESTABLISHMENT ADDRESS:
Oridion Medical 1987 Ltd 7 HaMarpe Street, Har Hotzvim Industrial Park, Jerusalem, 91450, Israel
### ESTABLISHMENT REGISTRATION NUMBER
8044004
## CONTACT PERSON:
Dalia Givony, Director of Regulatory Affairs Oridion Medical 1987 Ltd Har Hotzvim Science Based Industrial Park POB 45025 91450 Jerusalem, Israel Telephone: +972-2-589 9115 FAX: +972-2-586-6680
{5}------------------------------------------------
Page 2 of 6
## DEVICE INFORMATION
Trade Name: Capnostream20p with HiFi CO2 monitoring Common Name: Two Parameter Bedside Monitor Classification Name: Capnograph/Pulse Oximeter/Ventilatory Effort Recorder Regulation Number: 21 CFR 868.1400, Carbon dioxide analyzer, (product code CCK) Secondary product codes: DQA and MNR.
#### PREDICATE DEVICE
Capnostream20p with HiFi CO2 monitoring is substantially equivalent to the following commercially available devices:
#### MANUFACTURER DEVICE 510(k) No. Clearance Date
Oridion 1987 Medical Ltd Capnostream20p with Smart A/hr &ODI K112368 Julv 19, 2012
#### DEVICE DESCRIPTION
The Capnostream20p bedside monitor is a two parameter monitor consisting of a microMediCO2 capnography module and a pulse oximetry module implemented in a host device. The host device displays parameters received from the respective modules and generates alarms when preset alarm thresholds are crossed.
The HiFi capnography software feature, presented in this submission, is intended to enable measurement of airway CO2 for infant/neonatal patients during high frequency oscillatory ventilation (HFOV) and detects spontaneous breaths for infant/neonatal patients ventilated using high frequency oscillatory ventilation (HFOV).
{6}------------------------------------------------
Page 3 of 6
#### INTENDED USE
The Capnostream®20p combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is also indicated for continuous noninvasive monitoring of carboxyhemoglobin saturation (measured by an SpCO/SpMet/SpHb sensor), methemoglobin saturation (measured by an SpCO/SpMet/SpHb sensor) and total hemoglobin concentration (measured by an SpCO/SpMet/SpHb sensor). It is intended for use with neonatal, pediatric, and adult patients in hospital-type facilities, intra-hospital transport and home environments.
Capnostream®20p is to be operated by qualified healthcare personnel only.
The Capnostream®20p monitor provides the clinician with an integrated pulmonary index (IPI). The IPI is based on four parameters provided by the monitor: end tidal carbon dioxide, respiration rate, oxygen saturation and pulse rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10. where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status. The IPI is an adjunct to, and is not intended to replace, vital sign monitoring. The Capnostream20p HiFi mode provides a measurement of expired carbon dioxide and detects spontaneous breaths for infant/neonatal patients ventilated using high frequency oscillatory ventilation (HFOV).
{7}------------------------------------------------
## COMPARISON TO PREDICATE DEVICE
The Capnostream20p with HiFi CO2 monitoring is substantially equivalent to the predicate Capnostream20p with Smart A/hr ODI™ .
| Attribute | Predicate Device: Capnostream20p with<br>Smart A/hr ODI™ (K112368) | Capnostream 20p<br>with HiFi CO2<br>monitoring |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | = | Same |
| Product<br>Code<br>& Regulation<br>number | 868.1400, Carbon Dioxide Analyzer<br>(Classification CCK)<br>870.2700 Pulse Oximeter (Classification<br>DQA)<br>868.2375 Ventilatory Effort Recorder<br>(Classification MNR) | Same |
| Intended use | The Capnostream®20p combined<br>capnograph/pulse oximeter monitor and its<br>accessories are intended to provide<br>professionally trained health care providers<br>with continuous, non-invasive measurement<br>and monitoring of carbon dioxide<br>concentration of the expired and inspired<br>breath and respiration rate, and with<br>continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate. It is also<br>indicated for continuous non-invasive<br>monitoring of carboxyhemoglobin saturation<br>(measured by an SpCO/SpMet/SpHb<br>sensor),<br>methemoglobin saturation (measured by an<br>SpCO/SpMet/SpHb sensor) and total<br>concentration (measured by an<br>SpCO/SpMet/SpHb sensor). It is intended<br>for<br>use with neonatal, pediatric, and adult<br>patients in hospitals, hospital-type facilities,<br>intra-hospital transport and home<br>environments.<br>The Capnostream®20p monitor provides the<br>clinician with an integrated pulmonary index<br>(IPI). The IPI is based on four parameters<br>provided by the monitor: end tidal carbon | Same.<br>In addition:<br>The Capnostream20p<br>HiFi mode provides a<br>measurement of<br>expired carbon<br>dioxide and detects<br>spontaneous breaths<br>for infant/neonatal<br>patients ventilated<br>using high frequency<br>oscillatory ventilation<br>(HFOV). |
| Target<br>population | It is intended for use with neonatal, pediatric,<br>and adult patients. The Smart A/hr and ODI<br>indication for use is for adult patients aged<br>22 and higher only. | Identical target<br>population for<br>monitoring. In addition<br>HiFi mode is intended<br>for infants and<br>neonates only |
| Design | Equivalent to the Capnostream20 described<br>in K094012 | same |
| Performance<br>Standards | ISO 21647<br>ISO 9919 | same |
| Safety<br>Standards | IEC/EN 60601-1<br>IEC/EN 60601-1-2<br>IEC 60601-1-8<br>ISO 14971<br>EN 980 | same |
{8}------------------------------------------------
Page 5 of 6
## PERFORMANCE TESTING
The Capnostream20p with HiFi is substantially equivalent to the predicate Capnostream20p with A/hr and ODI.
The new device meets the safety and performance standards met by the predicate device.
Software testing was performed to validate the performance of the new monitor and its substantial equivalence to the predicate device. The functional features and the intended use of Capnostream20p with HiFi are substantially equivalent to the predicate device.
{9}------------------------------------------------
Page 6 of 6
Performance testing conducted on a lung simulator system demonstrated the accuracy of the CO2 measurements in HFOV mode up to a rate of 900 BPM (breaths per minute) (15 Hz) for both passive (ventilator driven) and spontaneous breaths. The testing was conducted for a CO2 range of 16mmHg to 76 mmHg and at spontaneous respiratory rates of 30, 45, and 60 BPM.
A hazard analysis was carried out on the Capnostream host monitor displaying the HiFi values. This hazard analysis concluded that any residual risks were judged as acceptable when weighed against the intended benefits of use of the system.
## CONCLUSION
The Capnostream20p with the HiFi software feature does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices. The minor differences between the Capnostream20p with HiFi and its predicate device raise no issue of safety and effectiveness. Therefore, the device is substantially equivalent to the predicate devices with respect to safety, effectiveness, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.