The DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is intended for intra- and extra-articular fractures, osteotomies, nonunions of the distal radius, with or without extension into the radial diaphysis in adults and adolescents (12-21) where the growth plates have fused or will not be crossed, and the following adolescent distal radius fractures: - Intra-articular fractures exiting the epiphysis - Intra-articular fractures exiting the metaphysis - Physeal crush injuries - Any injuries which cause growth arrest to the distal radius.
Device Story
Metallic bone fixation plate system for distal radius fractures; extra-long design allows treatment of fractures extending into radial diaphysis. Device consists of stainless steel or titanium plates with variable angle locking holes; accepts various 2.4mm/2.7mm locking/cortex screws and 1.8mm buttress pins. Used by orthopedic surgeons in clinical/surgical settings to stabilize bone fragments via internal fixation. Plate head design maintains compatibility with existing guide blocks. Provides mechanical stability to bone fractures, facilitating healing and anatomical alignment. Available in multiple lengths and head widths to accommodate patient anatomy; supplied sterile.
Clinical Evidence
Bench testing only. No clinical data. Mechanical performance validated via four-point bending tests (static and fatigue strength), construct axial bending tests (plate head/strength), and torque-through testing. Biocompatibility testing performed. LAL testing confirmed endotoxin levels meet the 20EU/device limit.
Technological Characteristics
Materials: 316L stainless steel or commercially pure titanium (CP4). Design: Two-column volar distal radius plate, extra-long, with variable angle locking holes. Dimensions: 7, 10, or 13 shaft holes; variable head widths. Energy source: None (mechanical). Sterilization: Sterile packed.
Indications for Use
Indicated for adults and adolescents (12-21) with fused growth plates (or where plates are not crossed) for intra- and extra-articular fractures, osteotomies, nonunions, and malunions of the distal radius, including extension into the radial diaphysis. Specific adolescent indications include fractures exiting epiphysis/metaphysis, physeal crush injuries, and injuries causing growth arrest.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
SYNTHES (USA) MODULAR MINI FRAGMENT LCP SYSTEM (K063049)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, with three profiles overlapping to suggest a sense of community and support. The figure is positioned within a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2017
DePuy Synthes Joanna Rieder Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380
Re: K163046
Trade/Device Name: DePuy Synthes 2.4/2.7mm Va Lcp Two-column Volar Distal Radius Plate, Extra-long Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: January 10, 2017 Received: January 11, 2017
Dear Joanna Rieder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K163046
Device Name
DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long
#### Indications for Use (Describe)
#### INDICATIONS
The DePuy Synthes 2.4/2.7 mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is intended for intra- and extra-articular fractures, osteotomies, nonunions of the distal radius, with or without extension into the radial diaphysis in adults and adolescents (12-21) where the growth plates have fused or will not be crossed, and the following adolescent distal radius fractures:
- Intra-articular fractures exiting the epiphysis
- Intra-articular fractures exiting the metaphysis
- Physeal crush injuries
- Any injuries which cause growth arrest to the distal radius.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a symbol of two overlapping circles in gold, followed by the company name "DePuy Synthes" in blue. Underneath the company name, it says "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" in smaller letters. To the right of the company name, the phrase "People inspired" is written in a cursive font.
# 510(k) Summary
## Date Prepared: February 9, 2017
| Sponsor: | DePuy Synthes<br>Joanna Rieder<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Office: +41 32 720 41 17<br>Fax +41 32 720 71 73 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary<br>Name: | DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-<br>Long |
| Classification: | Classification: Class II<br>Regulation Number: 21 CRF §888.3030<br>Product Code: HRS |
| Predicate<br>Device | 2.4MM VA-LCP DORSAL DISTAL RADIUS PLATES (K102694) |
| Reference<br>Predicate | SYNTHES (USA) MODULAR MINI FRAGMENT LCP SYSTEM (K063049) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo features a gold-colored abstract symbol on the left, followed by the company name in blue. To the right of the company name, the phrase "People inspired" is written in a cursive font.
| Device<br>Description: | The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius<br>Plate, Extra-Long will be available in stainless steel and titanium. The head of<br>the plate will remain the same as the existing 2.4mm VA-LCP Two-Column<br>Volar Distal Radius Plate (K102694) and will therefore allow the use of the<br>existing guide block for these plates. The DePuy Synthes 2.4/2.7mm VA-LCP<br>Two-Column Volar Distal Radius Plate, Extra-Long will come in 3 different<br>lengths (7 shaft holes, 10 shaft holes and 13 shaft holes). To accommodate<br>different patient anatomy, the plate with 7 shaft holes will be available with<br>three different plate head widths (Standard, Narrow and Wide). The 10 shaft<br>hole and the 13 shaft hole plate will only be released with the Standard plate<br>head width and additionally they feature a curvature in the shaft. All plates will<br>feature left- and right-specific designs and will be offered in sterile packed only.<br>The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius<br>Plates, Extra-Long are designed to accept existing 2.4mm Variable Angle<br>Locking Screws, 2.4mm Cortex Screws, 2.4mm Locking Screws, 1.8mm VA<br>Buttress Pins and 1.8mm Locking Buttress Pins in the plate head holes. In the<br>plate shaft, existing 2.4mm Cortex Screws, 2.7mm Cortex Screws, 2.4mm<br>Variable Angle Locking Screws, 2.7mm Variable Angle Locking Screws,<br>2.4mm Locking Screws, and 2.7mm Locking Screws can be inserted. |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use: | The DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius<br>Plate, Extra-Long is intended for intra- and extra-articular fractures,<br>osteotomies, nonunions and malunions of the distal radius, with or without<br>extension into the radial diaphysis in adults and adolescents (12-21) where the<br>growth plates have fused or will not be crossed, such as:<br>- Intra-articular fractures exiting the epiphysis<br>- Intra-articular fractures exiting the metaphysis<br>- Physeal crush injuries<br>- Any injuries which cause growth arrest to the distal radius |
| Comparison to<br>Predicate: | The proposed DePuy Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal<br>Radius Plate, Extra-Long has similar indications, in comparison to the predicate<br>devices.<br>The features of the subject device are substantially equivalent to the predicate<br>devices based on the similarities in intended use and design, where the new<br>device can additionally treat fractures with extension to the radial shaft.<br>Mechanical testing demonstrates substantial equivalence of the subject device<br>to the predicates devices in regards to mechanical strength. In addition the<br>intended use, manufacturing methods, packaging, and sterilization of the<br>predicates compared to the subject devices are identical. The subject device and<br>predicate devices are made from stainless steel (316L) or commercially pure<br>titanium (CP4) |
| Substantial<br>Equivalence: | Functional and mechanical testing demonstrate that the proposed DePuy<br>Synthes 2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-<br>Long is comparable in mechanical and functional properties to the predicate<br>devices. |
| | Worst case plates were tested to assess the mechanical performance of the<br>newly developed implants. The defined worst cases were tested under static and<br>dynamic loading conditions and were then compared to the predicate devices: |
| | Four Point Bending Test was performed to access the mean static and median<br>fatigue strength for non-contoured and contoured subject plates and evaluate<br>the plate shaft performance. The new plates performed better than the predicate<br>device. |
| | Construct (plate with screws) axial bending was performed to access the plate<br>construct strength and evaluate the plate head performance. The new plates<br>performed better than the predicate device. |
| | Torque Through Testing was performed to assess the mechanical properties of<br>the plate to screw interface. |
| | Biocompatibility testing was performed to demonstrate that the DePuy Synthes<br>2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is<br>biocompatible. |
| | LAL testing has been performed to establish that the subject device meets the<br>specified 20EU/device limit. |
| | Based on the aforementioned mechanical testing, the subject DePuy Synthes<br>2.4/2.7mm VA-LCP Two-Column Volar Distal Radius Plate, Extra-Long is<br>substantially equivalent to the predicate device. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DePuy Synthes, a company that is part of the Johnson & Johnson family of companies. The logo consists of a circular design, the company name in blue, and the tagline "People inspired" in a handwritten-style font. The text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES" is located underneath the company name in a smaller font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.